K173734 · Promepla Sam · FAD · Mar 12, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K173734
Device Name
RocaJJ Soft Stents
Applicant
Promepla Sam
Product Code
FAD · Gastroenterology, Urology
Decision Date
Mar 12, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4620
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RocaJJ Soft Stents are used for temporary internal drainage from the ureteropelvic junction to the bladder. Ureteral stents have been used to relieve obstruction in a variety of benign, malignant, and post-traumatic conditions. The stents may be placed using endoscopic techniques. The stents are not intended as a permanent indwelling devices, it is recommended that: - The indwelling time not exceed 6 days when the device is used with the removal string; - The indwelling time not exceed 365 days when the device is used without the removal string.
Device Story
Tubular double pigtail ureteral stent; made of flexible radiopaque polyurethane; includes polypropylene-monofilament suture loop for extraction; optional pusher (with radiopaque tip) and guidewire. Used for temporary internal drainage from ureteropelvic junction to bladder. Placed by physicians using endoscopic techniques. Stent relieves ureteral obstruction; improves patient outcomes by facilitating drainage. Indwelling time limited to 6 days with removal string or 365 days without. Device provides mechanical drainage path; radiopacity allows visualization during placement.
Clinical Evidence
Bench testing only. Evaluations included tensile strength, ultimate elongation, opening force of retaining loops, drainage capacity, biocompatibility (ISO 10993), sterilization, shelf life, and transportation validation. All results met specifications.
Technological Characteristics
Flexible radiopaque polyurethane double pigtail stent; 4.8-8.0 Fr diameter; 24-30 cm length. Includes polypropylene-monofilament suture. Pusher made of polyamide. Guidewire interface 0.035". Biocompatibility per ISO 10993. Sterilization validated.
Indications for Use
Indicated for temporary internal drainage from ureteropelvic junction to bladder in adults (at least 22 years old) to relieve obstruction caused by benign, malignant, or post-traumatic conditions.
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
Predicate Devices
Bard InLayOptima Ureteral Stent With Suture (K043193)
Universa Soft Ureteral Stents and Stent Sets (K151051)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.
March 12, 2018
Promepla SAM Alexandre Bareille Regulatory Affairs Manager 9 Avenue Albert II Monaco, 98000 Monaco
Re: K173734
Trade/Device Name: RocaJJ Soft Stents Regulation Number: 21 CFR§ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD Dated: February 15, 2018 Received: February 20, 2018
Dear Alexandre Bareille:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173734
Device Name RocaJJ Soft Stents
| Indications for Use (Describe) | |
|--------------------------------|--|
|--------------------------------|--|
The Roca.IJ Soft Stents are used for temporary internal drainage from the ureteropelvic junction to the bladder. Ureteral stents have been used to relieve obstruction in a variety of benign, malignant, and post-traumatic conditions. The stents may be placed using endoscopic techniques.
The stents are not intended as a permanent indwelling devices, it is recommended that:
- The indwelling time not exceed 6 days when the device is used with the removal string;
- The indwelling time not exceed 365 days when the device is used without the removal string.
Target population: adults only (at least 22 years old)
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (k) Summary
# A. Submitter Information
| Submitter's Name: | PROMEPLA SAM |
|-------------------------|----------------------------------------------------------------|
| Address | 9 Avenue Albert II<br>98000 Monaco<br>MONACO (Principality of) |
| Contact Person | Alexandre Bareille |
| Contact Person's email: | alb@promepla.com |
| Contact Person's Number | (377) 979-842-44 |
| Contact Person's Fax | (377) 920-561-50 |
| Submission date | December 01, 2017 |
## B. Device Name
| Trade Name of Device: | RocaJJ Soft Stents |
|--------------------------|--------------------------|
| Common Name: | Stent, Ureteral |
| Regulation Number: | 21 CFR 876.4620 |
| Regulation Name : | Ureteral stent |
| Regulation Class : | II |
| Product Code : | FAD |
| Panel : | Gastroenterology/Urology |
| Official Contact Person: | Alexandre Bareille |
## C. Predicates Devices
| Nº | Product name | Manufacturer | 510(k) number |
|----|-------------------------------------------------|--------------|---------------|
| 1 | Bard InLayOptima Ureteral Stent<br>With Suture | Bard | K043193 |
| 2 | Universa Soft Ureteral Stents and<br>Stent Sets | Cook | K151051 |
{4}------------------------------------------------
### D. Device Description:
The RocaJJ Soft Stents system is a set consisting of a tubular double pigtail stent made up of flexible radiopaque polyurethane, with a polypropylene-monofilament suture loop in order to facilitate the extraction of the ureteral stent.
The devices may include a pusher with a radiopaque tip, and a guidewire. Otherwise, these sets may include only a pusher.
The RocaJJ Soft Stents are from 4.8 to 8.0 Fr in diameter and from 24 to 30 cm in specified length.
### E. Intended Use:
The RocaJJ Soft Stents are used for temporary internal drainage from the ureteropelvic junction to the bladder. Ureteral stents have been used to relieve obstruction in a variety of benign, malignant, and post-traumatic conditions. The stents may be placed using endoscopic techniques.
- The stents are not intended as a permanent indwelling devices, it is recommended that:
- The indwelling time not exceed 6 days when the device is used with the removal string; -
- The indwelling time not exceed 365 days when the device is used without the removal string.
Target population: adults only (at least 22 years old)
### F. Technological Characteristics and Performance Data:
In support of this 510(k) premarket notification. Promepla SAM has conducted bench testing to demonstrate that RocaJJ Soft Stents provide adequate mechanical strength for their intended use. All bench testing results confirmed that the products described in this submission met the necessary specification.
In addition, the biocompatibility of the devices has been confirmed in accordance with ISO 10993, and the company has conducted sterilization, shelf life and transportation validation in accordance with recognized industry standards.
A list of the tests performed to support substantial equivalence is provided below:
- Sterilization Validation; -
- -Transportation Validation;
- -Tensile strength:
- -Ultimate elongation;
- -Opening force of the retaining loops;
- -Drainage capacity;
- Biocompatibility; -
- Shelf life Validation. -
The results of these evaluations demonstrate that the RocaJJ Ureteral Stents are safe and effective when used in accordance with their intended use and labeling.
{5}------------------------------------------------
## G. Substantial Equivalence:
The RocaJJ Soft Stents are substantially equivalent to the Bard InLayOptima Ureteral Stent and to the Cook Universa Soft Stents in regard to intended use, design, materials and function.
| Product Name | RocaJJ Soft Stents | Bard InLayOptima<br>Ureteral Stent with<br>Suture | Universa Soft Ureteral<br>Stents and Stent Sets |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | -- | K043193 | K151051 |
| Product Code | FAD | FAD | FAD |
| Regulation Name | Ureteral Stent | Ureteral Stent | Ureteral Stent |
| Manufacturer | PROMEPLA SAM | C.R. Bard. Inc. Urological<br>Division | Cook Incorporated (Cook) |
| Intended Use | The RocaJJ Soft Stents<br>are used for temporary<br>internal drainage from the<br>ureteropelvic junction to<br>the bladder. Ureteral<br>stents have been used to<br>relieve obstruction in a<br>variety of benign,<br>malignant, and post-<br>traumatic conditions. The<br>stents may be placed<br>using endoscopic<br>techniques.<br>The stents are not<br>intended as a permanent<br>indwelling devices, it is<br>recommended that:<br>- The indwelling time not<br>exceed 6 days when the<br>device is used with the<br>removal string;<br>- The indwelling time not<br>exceed 365 days when<br>the device is used<br>without the removal<br>string. | The Bard InLayOptima<br>Ureteral Stent and Multi-<br>Length Ureteral Stent is<br>indicated to relieve<br>obstruction in a variety of<br>benign, malignant and<br>post-traumatic conditions<br>in the ureter. These<br>conditions include stones<br>and/or stone fragments or<br>other ureteral obstructions<br>such as those associated<br>with ureteral stricture,<br>malignancy of abdominal<br>organs, retroperitoneal<br>fibrosis or ureteral trauma,<br>or in association with<br>ESWL. The stent may be<br>placed using endoscopic<br>surgical techniques or<br>percutaneously using<br>standard radiographic<br>technique. It is<br>recommended that the<br>indwelling time not exceed<br>365 days. The stent is not<br>intended as a permanent<br>indwelling device. | The Universa Soft<br>Ureteral Stents and Stent<br>Sets aer used for<br>temporary internal<br>drainage from the<br>ureteropelvic junction to<br>the bladder. Ureteral<br>stents have been used to<br>relieve obstruction in a<br>variety of benign,<br>malignant, and post-<br>traumatic conditions. The<br>stents may be placed<br>using endoscopic,<br>percutaneous, or open<br>surgical techniques. |
The RocaJJ Soft Stents and its predicates have a similar intended use, diameter, length, indwelling period without the withdrawal string, sterility process, radiopaque properties, materialprocessing and design.
The RocaJJ Soft Stents have a shorter indwelling time withdrawal string than the Universa Soft Stent, which is more favorable.
{6}------------------------------------------------
K173734
RocaJJ Soft Stents pusher is made of Polyamide instead of HDPE for Bard InLay Optima, which is a more resistant material. Length of pusher is equivalent for 6/7/8Fr versions while 4.8Fr pusher is longer in RocaJJ Soft Stents, which is more favorable as it allows an easier handling by physician. RocaJJ Soft Stents exist in a span of lengths from 24 to 30cm, which is narrower than predicate devices and address most of patients.
Guidewire interface is 0.035" for RocaJJ Soft Stents, while predicates allow a guidewire diameter up to 0.038". Standard endoscopic procedures use 0.035" guidewires, while 0.038" guidewires are more commonly used in percutaneous procedures which are not included in RocaJJ Soft Stents intended use.
Hence, the RocaJJ Soft Stents are substantially equivalent to the predicates devices since the basic features and intended uses are the same.
The minor differences between RocaJJ Soft Stents and the predicate devices raise no new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.