Dental implant system for the functional and esthetic replacement of missing teeth.
Device Story
Threaded root-form dental implant and abutment system; Ø3.0 mm diameter implants with lengths 8.5–15 mm; includes standard and machined surface coronal (MSC) versions; external hexagonal implant-abutment interface. Used by dental clinicians in surgical procedures for tooth replacement. Implants placed into alveolar bone; abutments (cover screws, healing, gold, passive, compact conical) attached to support prosthetic restorations. Provides functional and esthetic tooth replacement. Biocompatible materials (CPTi, titanium alloy, gold alloy) ensure osseointegration. Performance validated via engineering analysis and dynamic fatigue testing (ISO 14801).
Clinical Evidence
Bench testing only. No clinical data included. Performance demonstrated via engineering analysis, dimensional analysis, and dynamic compression-bending fatigue testing per ISO 14801. Biocompatibility confirmed per ISO 10993-1 and ISO 10993-5. Endotoxin testing performed per USP 39-NF 34 <85>.
Technological Characteristics
Threaded root-form dental implants; Ø3.0 mm diameter; CPTi (cold worked) material. Abutments: CPTi, titanium alloy, gold alloy. External hexagonal interface. Sterilization: Gamma radiation (ISO 11137-1/2) or moist heat (ISO 17665-1/2). Packaging validated per ISO 11607 and ASTM F-1980. Mechanical testing per ISO 14801.
Indications for Use
Indicated for one- and two-stage surgical procedures for replacing maxillary lateral incisors and mandibular incisors; immediate placement in extraction sites and partially or completely healed alveolar ridges; immediate loading when good primary stability with appropriate occlusal loading is achieved, except in soft bone (type IV).
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Deep Conical (DC) Implants and Accessories (K163060)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".
Southern Implants (Pty) Ltd. Lauranda Breytenbach Regulatory Affairs and Quality 1 Albert Road Irene, Gauteng 0062 SOUTH AFRICA
Re: K173706
Trade/Device Name: Piccolo Implants and Accessories Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: March 19, 2018 Received: March 23, 2018
Dear Lauranda Breytenbach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
June 8, 2018
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K173706
Device Name Piccolo Implants and Accessories
Indications for Use (Describe)
The Piccolo Implants and Accessories are intended for both one- and two-stage surgical procedures in the following situations and with the following clinical protocols:
· replacing maxillary lateral incisors and mandibular incisors,
· immediate placement in extraction situations with a partially or completely healed alveolar ridge,
· immediate loading when good primary stability with appropriate occlusal loading is achieved, except in soft bone (type IV) where implant stability may be difficult to obtain.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Piccolo Implants and Accessories
# Southern Implants (Pty)
## Ltd. June 8, 2018
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Southern Implants (Pty) Ltd.<br>1 Albert Road<br>Irene, Gauteng, 0062 South Africa |
|--------------------------------|------------------------------------------------------------------------------------|
| | Telephone + 27 12 667 1046<br>Fax + 27 12 667 1029 |
| Official Contact | Lauranda Breytenbach<br>Regulatory Affairs and Quality |
| DEVICE NAME AND CLASSIFICATION | |
| Trade/Proprietary Name | Piccolo Implants and Accessories |
| Common Name | Dental implant<br>Dental implant abutment |
| Classification Name | Endosseous dental implant<br>Endosseous dental implant abutment |
| Classification Regulations | 21 CFR 872.3640, Class II |
| Product Code | DZE (Primary product code)<br>NHA (Secondary product code) |
Classification Panel Dental Products Panel Reviewing Branch Dental Devices Branch
### PREDICATE DEVICE INFORMATION
| | Manufacturer | Device Name | K number |
|------------|------------------------------|--------------------------------------------|----------|
| Primary: | Southern Implants (Pty) Ltd. | Deep Conical (DC) Implants and Accessories | K163060 |
| Reference: | Southern Implants (Pty) Ltd. | External Hex Implants and Accessories | K163634 |
| Reference: | Astra Tech AB | OsseoSpeed™ Plus | K120414 |
#### INDICATIONS FOR USE
The Piccolo Implants and Accessories are intended for both one- and two-stage surgical procedures in the following situations and with the following clinical protocols:
- replacing maxillary lateral incisors and mandibular incisors,
- immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge,
- immediate loading when good primary stability with appropriate occlusal loading is achieved, except ● in soft bone (type IV) where implant stability may be difficult to obtain.
{4}------------------------------------------------
The Piccolo Implants and Accessories is a threaded root-form dental implant and abutment system.
The implants are Ø3.0 mm in diameter with lengths of 8.5, 10, 11.5, 13 and 15 mm. There is a standard version and a MSC version with a 1 and 3 mm machined surface coronal region respectively. The rest of the body has a roughened surface. The system has an external hexagonal implant-abutment interface.
The following abutments are included in the range: Cover Screws, Healing Abutments (all made from commercially pure titanium (CPTi)), Gold Abutments (made from gold alloy), Passive Abutments and Compact Conical Abutments (both made from titanium alloy). All abutments (except the onepiece abutments) are used with a titanium alloy retaining screw. An overview is given in Table 1.
| Abutment Model | Type | Body/gingival<br>height | Platform Ø | Body Ø |
|---------------------------|---------------------|-------------------------|------------|--------|
| Cover Screws | One-piece | 1.2 | Ø3.0 | Ø3.0 |
| Healing Abutments | One-piece Narrow | 3, 4, 6 | Ø3.0 | Ø3.0 |
| | One-piece Wide | 3, 4, 6 | Ø3.0 | Ø4.0 |
| Titanium Abutments | Engaging | 1 | | Ø3.5 |
| | Non-engaging | 1 | Ø3.0 | |
| Gold Abutments | Engaging | 0.75 | | Ø3.5 |
| | Non-engaging | 0.75 | Ø3.0 | |
| Passive Abutments | Engaging | 0.6 | | Ø3.5 |
| | Non-engaging | 0.6 | Ø3.0 | |
| | Engaging (Vertical) | 1.0 | | Ø3.0 |
| Compact Conical Abutments | One-piece | 1.7, 3, 4, 5.5 | Ø3.0 | Ø4.8 |
Table 1: List of Abutments Included in the Subject Device
#### PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence include:engineering analysis, dimensional analysis, and dynamic compression-bending to ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants. No clinical data were included in this submission.
The sterilization method for the subject device is the same as the primary predicate (K163060) and reference device (K163634). The sterilization method is Gamma radiation and has been validated in accordance with ISO 11137-1 Sterilization of health care products - radiation - part 2: Requirements for development, validation, and routine control of a sterilization process for medical devices and ISO 11137-2 Sterilization of health care products - radiation - part 1: establishing the sterilization dose. The sterilization method is moist heat for the devices to be end user sterilized and has been validated in accordance with ISO 17665-1 Sterilization of health care products -- Moist heat -- Part 1: Reguirements for the development, validation and routine control of a sterilization process for medical devices and ISO 17665-2 Sterilization of health care products -- Moist heat -- Part 2: Guidance on the application of ISO 17665-1 . The Limulus Amebocyte Lysate (LAL) test for detection and quantitation of bacterial endotoxin was conducted in accordance with USP 39-NF 34 <85> Bacterial Endotoxin Test, using the kinetic chromogenic test method.
The packaging for the subject device is the same as the primary predicate (K163060) and reference device (K163634). Packaging was validated in accordance with ISO 11607 Packaging for terminally sterilized medical devices -- Part 1: Requirements for materials, sterile barrier systems. Accelerated aging per ASTM-F-1980 Standard guide for accelerated aging of sterile medical device packages has been applied on the final packaging followed by validating durability to peel and dye tests conditions, in order to substantiate 5 years shelf life.
The subject device is manufactured from the same material using the same manufacturing method as the applicant's own primary predicate (K163060) and reference device (K163634), has the same intended use, and the same patient contact type and duration.
{5}------------------------------------------------
The subject device is biocompatible in accordance with ISO 10993-1. Cytotoxicity testing in accordance with ISO 10993-5 was also performed.
### EQUIVALENCE TO MARKETED DEVICE
Southern Implants (Pty) Ltd submits the information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
K163060, Southern Implants (Pty) Ltd., Deep Conical (DC) Implants and Accessories
K163634, Southern Implants (Pty) Ltd., External Hex Implants and Accessories
K120414, Astra Tech AB, OsseoSpeed™ Plus
A table comparing the subject device and primary predicate devices is provided in Table 2.
| | Subject Device | Primary Predicate |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission name | Piccolo Implant and Accessories | Deep Conical (DC) Implants and Accessories |
| Manufacturer | Southern Implants (Pty) Ltd. | Southern Implants (Pty) Ltd. |
| 510(k) Number | | K163060 |
| Indications for Use Statement | The Piccolo Implants and Accessories are<br>intended for both one- and two-stage<br>surgical procedures in the following<br>situations and with the following clinical<br>protocols:<br>• replacing maxillary lateral incisors and<br>mandibular incisors,<br>• immediate placement in extraction sites<br>and in situations with a partially or<br>completely healed alveolar ridge,<br>• immediate loading when good<br>primary stability with appropriate<br>occlusal loading is achieved, except in<br>soft bone (type IV) where implant<br>stability may be difficult to obtain. | Southern Implants Dental Implants are intended for both<br>one- and two-stage surgical procedures in the following<br>situations and with the following clinical protocols:<br>• replacing single and multiple missing teeth in the<br>mandible and maxilla,<br>• immediate placement in extraction sites and in<br>situations with a partially or completely healed alveolar<br>ridge,<br>• immediate loading in all indications, except in single<br>tooth situations on implants shorter than 8mm or in<br>soft bone (type IV) where implant stability may be<br>difficult to obtain and immediate loading may not be<br>appropriate.<br>The intended use for 3.0 Deep Conical implants is limited<br>to replacement of maxillary lateral incisors and<br>mandibular incisors |
| Product Code | DZE, NHA | DZE, NHA |
| Intended Use | Dental implant system for the functional<br>and esthetic replacement of missing teeth. | Dental implant system for the functional and esthetic<br>replacement of missing teeth. |
| Implant Body Design | Threaded root-form implants:<br>Tapered | Threaded root-form implants:<br>Tapered<br>Cylindrical<br>Co-axis tapered<br>Co-axis cylindrical |
| Implant body diameters and<br>lengths (mm) | $Ø$ 3.0: length 8.5, 10, 11.5, 13, 15 | $Ø$ 3.0: length 9, 11, 13, 15<br>$Ø$ 3.5 – $Ø$ 5.0: length 6 – 15 |
| Implant Material | CPTi - (cold worked) | CPTi – (cold worked) |
| Implant Surface | Southern Implants' roughened surface<br>Machined collar versions available | Southern Implants' roughened surface |
| Implant Sterilization | Sterile — irradiation | Sterile — irradiation |
| Implant / Abutment Interface | External Hex | Deep Conical |
| Abutment Designs | Cover screw<br>Healing abutment | Cover Screw<br>Healing Abutment, Healing Caps |
| | Subject Device | Primary Predicate |
| | Titanium Abutment<br>Gold Abutment<br>Passive Abutments<br>Compact Conical Abutment | Titanium Abutment, Angled Abutments<br>Overdenture Abutment<br>Gold Abutment<br>Passive Abutments<br>Compact Conical Abutment |
| Abutment platform diameters | Ø3.0 | Ø3.0, Ø3.5, Ø4.0, Ø5.0 |
| Abutment angle | 0° only | 0°, 12°, 24° |
| Abutment Material | CPTi<br>Titanium alloy<br>Gold alloy | CPTi<br>Titanium alloy<br>Gold alloy |
| Abutment Surface | Machined<br>Knurled (grooved)<br>Titanium nitride coated | Machined<br>Knurled (grooved)<br>Titanium nitride coated |
| Restoration | Single-unit or multi-unit | Single-unit or multi-unit |
#### Table 2: Predicate 1 and Subject Device Comparison Table
{6}------------------------------------------------
Primary predicate (K163060) is for substantial equivalence of the diameter, materials and surface of the implants and abutments. The subject device implants have the same outer diameter as the primary predicate Ø3.0 implants. The subject device is made from the same manufacturing facilities, with the same surface treatment as implants in primary predicate (K163060). The subject device has Ø3.0 abutments of the same type as the primary predicate (K163060) with substantially equivalent design and materials. Primary predicate (K163060) has implants with a range of sizes including diameter 3.0 mm implants. The subject device only has diameter 3.0 mm implants. The differences in wording of the indications for use statement reflect this difference, but with regards to the Ø3.0 implant and accessories the indications for use statements are substantially equivalent.
Reference device (K163634) is for substantial equivalence of the subject device implants and abutments. The subject device Piccolo implants are Ø3.0 tapered implants with a similar shape. lengths, thread, flutes, taper, apex, collar and connection interface to the corresponding Ø3.25 tapered implant in reference device (K163634).
The subject device and all primary predicate (K163060) and reference devices (K163634, K120414) are provided sterile for single-patient, single-use.
Substantial equivalence of the subject device components in terms of biocompatibility is supported by the fact that materials are identical in formulation, processing, component interactions, and storage conditions to the devices in primary predicate (K163060) and reference device (K163634).
### CONCLUSION
The subject device and the primary predicate (K163060) and reference devices (K163634, K120414) have the same intended use, have similar technological characteristics, and are made of the same materials. The subject device and the primary predicate (K163060) and reference devices (K163634, K 120414) are packaged in similar materials and sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence of the subject device to the primary predicate and reference devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.