15.5F X 19 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 23 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 28 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 33 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 37 cm Symetrex Long Term Hemodialysis Catheter with Sideholes
K173667 · Medical Components, Inc. (dba MedComp) · MSD · May 17, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K173667
Device Name
15.5F X 19 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 23 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 28 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 33 cm Symetrex Long Term Hemodialysis Catheter with Sideholes, 15.5F X 37 cm Symetrex Long Term Hemodialysis Catheter with Sideholes
Applicant
Medical Components, Inc. (dba MedComp)
Product Code
MSD · Gastroenterology, Urology
Decision Date
May 17, 2018
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5540
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Long term, greater than 30 days, vascular access for Hemodialysis and Apheresis treatments.
Device Story
Symetrex™ Long Term Hemodialysis Catheter with Sideholes is a 15.5 French, dual-lumen, radiopaque polyurethane catheter; features polyester retention cuff for subcutaneous anchoring; includes two female luer adapters. Device provides vascular access for chronic hemodialysis and apheresis; inserted percutaneously or via cut down by clinicians. Symmetrical side channels and distal tip configuration separate intake and output flow. Sideholes at distal tip enhance flow dynamics. Device is used in clinical settings for long-term (>30 days) blood access. Healthcare providers use the catheter to facilitate extracorporeal blood treatment; output is the physical conduit for blood flow between patient and dialysis/apheresis machine. Benefits include reliable, long-term vascular access for patients with end-stage renal disease or those requiring apheresis.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included in vitro recirculation studies (demonstrating <1% recirculation in forward/reverse flow), air/liquid leak testing, extension/hub strength, and biocompatibility per ISO 10993-1 (hemocompatibility, genotoxicity, cytotoxicity, irritation, systemic toxicity, implantation).
Technological Characteristics
15.5 French dual-lumen catheter; materials: Carbothane (hub/lumen) with 20% BaSO4, polyester felt (cuff), Pellethane (extension), Isoplast (luer), ABS (ID rings), polypropylene (clamp). EO sterilization. ISO 10555-1, ISO 11607-1/2, ISO 594-1/2 compliance. Radiopaque.
Indications for Use
Indicated for chronic hemodialysis and apheresis in patients requiring long-term vascular access. May be inserted percutaneously or by cut down. Catheters >37cm are indicated for femoral placement.
Regulatory Classification
Identification
A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
Special Controls
*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Symetrex™ Long Term Hemodialysis Catheter (K171618)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in a blue box, and the full name of the agency, "U.S. Food & Drug Administration," is in blue text.
May 17, 2018
Medical Components, Inc. Dba Medcomp® Courtnev Nix Regulatory Affairs Manager, North America and Europe 1499 Delp Drive Harleysville, Pennsylvania 19438
Re: K173667
Trade/Device Name: Symetrex™ Long Term Hemodialysis Catheter with Sideholes Regulation Number: 21 CFR 876.5540 Regulation Name: Blood Access Device and Accessories Regulatory Class: Class II Product Code: MSD Dated: April 12, 2018 Received: April 17, 2018
Dear Courtney Nix:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173667
Device Name
SymetrexTM Long Term Hemodialysis Catheter with Sideholes
Indications for Use (Describe)
The Symetrex™ Long Term Hemodialysis Catheter with Sideholes is a symmetric tip dual lumen catheter designed for chronic hemodialysis and apheresis. It may be inserted percutaneously or by cut down. Catheters with greater than 37cm implant length are indicated for femoral placement.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains a logo for medCOMP. The logo features a red geometric shape that resembles a stylized letter "C" or an abstract building design. Below the geometric shape, the text "medCOMP" is written, with "med" in red and "COMP" in black. The overall design is clean and modern.
1499 Delp Drive Harleysville, PA 19438 Tel: 215-256-4201 Fax. 215-256-1787 www.medcompnet.com Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes
| Section 6 | 510(k) SUMMARY K173667 Traditional 510K |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| 1. Submitter Information: | |
| Submitter: | Medical Components Inc. dba Medcomp® 1499 Delp Drive Harleysville, PA 19438 Tel: (215) 256-4201, x 2285 Fax: (215) 256-9191 |
| Registration Number: | 2518902 |
| Contact: | Courtney Nix Cnix@Medcompnet.com Regulatory Affairs Manager: North America and EU |
| Date Prepared: | November 27, 2017 |
| 2. Proposed or Subject Device Information: | |
| Trade Name: | Symetrex™ Long Term Hemodialysis Catheter with Sideholes |
| Device: | Catheter, Hemodialysis, Implanted |
| Product Code: | MSD |
| Regulation Description: | Blood access device and accessories |
| C.F.R. Section: | 21 CFR 876.5540 |
| Class: | II |
| Regulation Medical Specialty and Review Panel: | Gastroenterology/Urology |
| 3. Predicate Device Information: | |
| 510(k) Number: | K171618 |
| 510(k) Holder: | Medcomp® |
| | 6-1 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo for medCOMP. The logo features a red geometric shape that resembles a stylized letter "G" or an abstract building. Below the shape, the word "med" is written in red, followed by "COMP" in black. The overall design is simple and modern.
Harlevsville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes
| Trade Name: | Symetrex™ Long Term Hemodialysis Catheter |
|------------------------------------------------------|-------------------------------------------|
| Device: | Catheter, Hemodialysis, Implanted |
| Product Code: | MSD |
| Regulation Description: | Blood access device and accessories |
| C.F.R Section: | 21 CFR 876.5540 |
| Class: | II (Special Controls) |
| Regulation Medical<br>Specialty and Review<br>Panel: | Gastroenterology/Urology |
#### 4. Device Description:
The Symetrex™ Long Term Hemodialysis Catheter with Sideholes is a chronic, 15.5 French, dual lumen, radiopaque catheter made of polyurethane. It has a polyester retention cuff and two female luer adapters. The retention cuff promotes tissue ingrowth to anchor the catheter in the subcutaneous tunnel. The luer adapters are identical in color to indicate the reversibility of this catheter. This catheter features symmetrical side channels with a distal tip configuration designed to separate the intake flow from the output flow in both directions.
#### 5. Indications for Use:
The Symetrex™ Long Term Hemodialysis Catheter with Sideholes is a symmetric tip dual lumen catheter designed for chronic hemodialysis and apheresis. It may be inserted percutaneously or by cut down. Catheters with greater than 37cm implant length are indicated for femoral placement.
#### 6. Intended Use:
Long term, greater than 30 days, vascular access for Hemodialysis and Apheresis treatments.
#### 7. Comparison to Predicate Devices:
Symetrex™ Long Term Hemodialysis Catheter with Sideholes is substantially equivalent to the predicate device, Symetrex™ Long Term Hemodialysis Catheter (K171618), in terms of indications for use, labeling, intended use. anatomical location, performance, components, and method of sterilization.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "G" or a house with a missing side. Below the geometric shape, the word "medCOMP" is written in a combination of red and black letters. The "med" portion is in red, while the "COMP" portion is in black.
- 1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes
The difference or changes between the Symetrex™ Long Term Hemodialysis Catheter Sideholes and the predicate, Symetrex™ Long Term Hemodialysis Catheter (K171618) are the sideholes at the distal tip on both the arterial and venous lumen.
Comparison table on the following page.
## Table 6.1: 510(K) Summary: Design Comparison Matrix
| Device | Proposed Device:<br>Symetrex™ Long Term<br>Hemodialysis Catheter<br>with Sideholes | Predicate Device:<br>Symetrex™ Long Term<br>Hemodialysis Catheter<br>(K171618) |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design | Dual Kidney Shaped<br>Lumen Catheter | Dual Kidney Shape<br>Lumen Catheter |
| Tip Design | Symmetric Tip, Sideholes<br>at Distal Tip | Symmetric Tip, No Side<br>Holes |
| Lengths | 19cm-42cm | 19cm-42cm |
| Indications<br>for Use | The Symetrex™ Long<br>Term Hemodialysis<br>Catheter with Sideholes<br>is a symmetric tip dual<br>lumen catheter designed<br>for chronic hemodialysis<br>and aphersis. It may be<br>inserted percutaneously<br>of by cut down. Catheters<br>with greater than 37cm<br>implant length are<br>indicated for femoral<br>placement. | The Symetrex™ Long<br>Term Hemodialysis<br>Catheter is a symmetric<br>tip dual lumen catheter<br>designed for chronic<br>hemodialysis and<br>aphersis. It may be<br>inserted percutaneously<br>of by cut down. Catheters<br>with greater than 37cm<br>implant length are<br>indicated for femoral<br>placement. |
| Catheter<br>O.D. | 15.5FR | 15.5FR |
| Sterilization<br>Method | EO Sterilization | EO Sterilization |
| Materials | Hub (Green):<br>Carbothane w/ 20%<br>BaSO4<br>Lumen (White):<br>Carbothane w/ 20%<br>BaSO4<br>Cuff:<br>Polyester Felt | Hub (Green):<br>Carbothane w/ 20%<br>BaSO4<br>Lumen (White):<br>Carbothane w/ 20%<br>BaSO4<br>Cuff:<br>Polyester Felt |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for medCOMP. The logo is red and white. The logo has a stylized letter "C" above the word "medCOMP".
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes
| Extension (Clear):<br>Pellethane<br>Barbed Luer (Green) :<br>Isoplast<br>ID Ring (White with red ink):<br>ABS<br>ID Ring (White with blue ink):<br>ABS<br>Clamp (White):<br>Polypropylene | Extension (Clear):<br>Pellethane<br>Barbed Luer (Green) :<br>Isoplast<br>ID Ring (White with red ink):<br>ABS<br>ID Ring (White with blue ink):<br>ABS<br>Clamp (White):<br>Polypropylene | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-----------------|
| Components | Dilator, Stylet, Tunneler,<br>Valved Introducer, and<br>Guidewire | Dilator, Stylet, Tunneler,<br>Valved Introducer, and<br>Guidewire | | |
| Packaging<br>Systems | Clamshell (Catheters<br>under 33cm):<br>One unit gets a snap lid<br>one unit is sealed into a<br>header bag<br>5 Header bags are<br>placed in a shipping<br>carton<br><br>Translumbar tray<br>(Catheters greater than<br>33cm):<br>One unit gets a snap lid<br>One unit is sealed into a<br>header bag.<br>5 units are placed into a<br>shipping carton. | Clamshell (Catheters<br>under 33cm):<br>One unit gets a snap lid<br>one unit is sealed into a<br>header bag<br>5 Header bags are<br>placed in a shipping<br>carton<br><br>Translumbar tray<br>(Catheters greater than<br>33cm):<br>One unit gets a snap lid<br>One unit is sealed into a<br>header bag.<br>5 units are placed into a<br>shipping carton. | | |
| Performance<br>Testing | Priming Volumes: | | | |
| | Tip to Cuff<br>Length | Lumen<br>Volume | Tip to Cuff<br>Length | Lumen<br>Volume |
| | 19cm | 2.1cc | 19cm | 2.1cc |
| | 23cm | 2.3cc | 23cm | 2.3cc |
| | 28cm | 2.5cc | 28cm | 2.5cc |
| | 33cm | 2.6cc | 33cm | 2.6cc |
| | 37cm | 2.8cc | 37cm | 2.8cc |
| | 42cm | 3.2cc | 42cm | 3.2cc |
| | Recirculation: | | Recirculation: | |
Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes Section 6: 510 Summary
6-4
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "G" or a box with an open side. Below the geometric shape, the word "medCOMP" is written in a stylized font, with "med" in red and "COMP" in black.
- 1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes
| Symetrex™ Long Term<br>Hemodialysis Catheter<br>with Sideholes has less<br>than 1% recirculation in<br>forward and reverse flow<br>when tested in vitro | Symetrex™ Long Term<br>Hemodialysis Catheter<br>has less than 1%<br>recirculation in forward<br>and reverse flow when<br>tested in vitro |
|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
#### 8. Bench / Performance Data / Non-Clinical Testing:
The results of performance testing, in conjunction with the substantial equivalence claims, effectively demonstrate the proposed device, Symetrex™ Long Term Hemodialysis Catheter with Sideholes, is equivalent to the predicate device, Symetrex™ Long Term Hemodialysis Catheter (K171618). The performance testing was performed in accordance with the following standards:
#### Table 6.2: Performance Standards
| Standard | Standard Title | Revision/ Date | Performance<br>Testing |
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--------------------------------------------------------------------------------|
| ISO 10555-1 | Intravascular<br>catheters --<br>sterile and single-<br>use<br>intravascular<br>catheters -- part<br>1: general<br>requirements. | Second Edition<br>2013-06-15 | Air Leak, Catheter<br>Leak, Extension-<br>Hub, Extrusion-<br>Hub, Gravity Flow |
| ISO 11607-1 | Packaging for<br>terminally<br>sterilized medical<br>devices - part<br>1: requirements<br>for materials,<br>sterile barrier<br>systems and<br>packaging<br>systems<br>[including:<br>amendment 1<br>(2014)]. | First Edition<br>2006-04-15 | Shipping and Shelf<br>Life testing |
| ISO 11607-2 | Packaging for<br>terminally<br>sterilized medical<br>devices - part<br>2: validation<br>requirements for | First Edition<br>2006-04-15 | Shipping and Shelf<br>Life testing |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "G" or an abstract building. Below the geometric shape, the word "medCOMP" is written in a stylized font, with "med" in red and "COMP" in black.
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes
| | forming, sealing<br>and assembly<br>processes<br>[including:<br>amendment 1<br>(2014)]. | | |
|-----------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 594-1 | conical fittings<br>with a 6% (luer)<br>taper for<br>syringes, needles<br>and<br>certain other<br>medical<br>equipment - part<br>1: general<br>requirements. | First edition<br>1986-06-15, | Gauging |
| ISO 594-2 | Conical fittings<br>with a 6% (luer)<br>taper for<br>syringes, needles<br>and<br>certain other<br>medical<br>equipment - part<br>2: lock fittings. | Second Edition<br>1998-09-01 | Liquid Leakage<br>Air Leakage<br>Separation Force<br>Unscrewing Torque<br>Ease of Assembly<br>Resistance to<br>Overriding<br>Stress Cracking |
## 9. Biocompatibility:
Biocompatibility was performed for the Symetrex™ Long Term Hemodialysis Catheter with Sideholes per ISO 10993-1 for a blood implant device with permanent exposure (i.e. > 30 days). Biocompatibility was performed on the final, finished device. The biological end points include:
- Hemocompatibility: 0
- · ISO 10993-4 Biological Evaluation Of Medical Devices Part 4: Selection Of Tests For Interaction With Blood
- · ASTM F 756-08, Standard Practice for Assessment of Hemolytic Properties of Materials, 2008
- 0 Genotoxicity:
- o ISO 10993-3, Biological Evaluation of Medical Devices Part 3: Tests for Genotoxicity, Carcinogenicity and Reproductive Toxicity.
- 0 Cytotoxicity:
- o AAMI / ANSI / ISO 10993-5: Biological evaluation of medical devices part 5: tests for in vitro cytotoxicity.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "M" or an abstract building. Below the geometric shape, the word "medCOMP" is written in a lowercase, italicized font, also in red. The overall design is simple and modern, with a focus on the geometric shape and the company name.
Harlevsville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Medcomp®: Symetrex™ Long Term Hemodialysis Catheter with Sideholes
- Irritation/Intracutaneous: 0
- o ISO 10993-10 Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization.
- Acute Systemic Toxicity: 0
- o ISO 10993-11 Biological Evaluation of Medical Devices Part 11: Tests for Systemic Toxicity
- Implantation: 0
- o ISO 10993-6 Biological Evaluation of Medical Devices Part 6: Tests for Local Effects after Implantation.
- ASTM F763-04 Standard Practice For Short-Term Screening of Implant O Materials.
- ASTM F981-04 Standard Practice For Assessment of Compatibility of o Biomaterials for Surgical Implants With Respect To Effect of Materials on Muscle and Bone.
- Additional Testing: 0
- o ISO 10993-18 Biological Evaluation of Medical Devices Part 18: Chemical Characterization of Materials
- ASTM D4128-06 Standard Guide for Identification of Quantitation of O Organic Compounds in Water by Combined Gas Chromatography and Electron Impact Mass Spectrometry
- ASTM D1971-11 Standard Practices for Digestion of Water Samples for O Determination of Metals by Flame Atomic Absorption, Graphite Furnace Atomic Absorption, Plasma Emission Spectroscopy, or Plasma Mass Spectroscopy
- ISO 10993-17 Biological Evaluation of Medical Devices Part 17: O Establishment of Allowable Limits for Leachable Substances
## 10. Summary of Substantial Equivalence:
In conclusion, the proposed device, Symetrex™ Long Term Hemodialysis Catheter with Sideholes, is considered substantially equivalent to the predicate device. Symetrex™ Long Term Hemodialysis (K171618) as demonstrated through non-clinical testing performed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.