K173664 · Medtronic, Inc. · OWB · Dec 29, 2017 · Radiology
Device Facts
Record ID
K173664
Device Name
Medtronic O-arm O2 Imaging System
Applicant
Medtronic, Inc.
Product Code
OWB · Radiology
Decision Date
Dec 29, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60lbs or greater and having an abdominal thickness greater than 16cm and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System is compatible with certain image guided surgery systems.
Device Story
Mobile cone-beam x-ray system; provides 2D fluoroscopic and 3D imaging. Consists of Image Acquisition System (IAS) and Mobile View Station (MVS). IAS performs 360-degree gantry rotation for 3D reconstruction; MVS provides visualization, window leveling, and oblique view rendering. Used in clinical settings by physicians for intraoperative guidance. Software revision 4.1 adds angular annotation (closed, open, Cobb angles), automated image transfer to navigation systems, and enhanced dynamic range for metal implant visualization. Cybersecurity updates include user authentication and software integrity checks. Output assists surgeons in visualizing anatomy and metallic implants during procedures, potentially improving surgical accuracy and workflow efficiency.
Clinical Evidence
Bench testing only. Verification and validation activities performed per risk/hazard analysis. Compliance with AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, and IEC 60601-2-43 standards confirmed.
Technological Characteristics
Mobile cone-beam CT; 40x30cm CsI flat-panel detector; 32kW generator. Isocentric motion; 360-degree gantry rotation. Connectivity via cable between IAS and MVS. Software revision 4.1.0. Cybersecurity: user authentication, integrity checks. Standards: AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-43.
Indications for Use
Indicated for adult and pediatric patients weighing 60 lbs or greater with abdominal thickness >16cm requiring 2D fluoroscopic or 3D imaging of anatomic structures and high-attenuation objects (e.g., bone, metal). Compatible with image-guided surgery systems.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG' in bold blue letters, and the word 'ADMINISTRATION' in smaller blue letters below.
Medtronic Inc. % Mr. Dean Honkonen Regulatory Affairs Manager 300 Foster Street LITTLETON MA 01460
December 29, 2017
Re: K173664
Trade/Device Name: Medtronic O-arm O2 Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, OXO, JAA, Dated: November 28, 2017 Received: November 29, 2017
Dear Mr. Honkonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
## Page 2 - Mr. Dean Honkonen
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn
(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
#### 510(k) Number (if known)
K173664
Device Name Medtronic O-arm O2 Imaging System
Indications for Use (Describe)
The O-arm O2 Imaging System is a mobile x-ray system, designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60 lbs or greater and having an abdominal thickness greater than 16cm, and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.
The O-arm O2 Imaging System is compatible with certain image guided surgery systems.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
#### 510(k) Summary
| Submitter: | Medtronic Navigation, Inc. (Littleton)<br>300 Foster Street<br>Littleton, MA 01460 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dean Honkonen<br>Regulatory Affairs Manager<br>Phone: (978)698-6065<br>(978)698-6090<br>Fax:<br>E-mail: dean.honkonen@medtronic.com |
| Date Summary Prepared: | October 11, 2017 |
| Device Trade Name: | Medtronic O-arm O2 Imaging System |
| Common Name: | Interventional Fluoroscopic X-ray System |
| Device Classification: | Class II |
| Product Code: | Primary OWB<br>Secondary OXO, JAA |
| Classification Name: | 892.1650 - Image Intensified Fluoroscopic X-ray System |
| Predicate Device: | K151000 – Medtronic O-arm O2 Imaging System |
### Device Description:
The O-arm O2 Imaging System is a mobile x-ray system that provides 3D imaging as well as 2D fluoroscopic imaging. It was originally cleared for market under 510(k) K151000. The device is classified under primary product code OWB (secondary product codes OXO, JAA) ref 21 CFR 892.1650.
### Modified Device:
This is a special submission for the addition of the following software features:
- Angular Annotation
- Easy Image Transfer ●
- Enhance Dynamic Range ●
- Cybersecurity Enhancements ●
These changes are only changes in software. There are no changes related to radiation performance. There are no changes to the device hardware required for these software changes. These features are described in more detail below and in the Device Description. This software also includes defect corrections.
{4}------------------------------------------------
The O-arm O2 Imaging System consists of two main assemblies that are used together:
- The Image Acquisition System (IAS) ●
- . The Mobile View Station (MVS)
The two units are interconnected by a single cable that provides power and signal data. The IAS has an internal battery pack that provides power for motorized transportation and gantry positioning. In addition the battery pack is used to power the X-ray tank. The MVS has an internal UPS to support its function when mains power is disconnected.
The O-arm O2 operates off standard line voltage within the following voltages:
- VAC 100, 120 or 240
- Frequency 60Hz or 50Hz
- . Power Requirements 1440 VA
# Indications for Use:
The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60lbs or greater and having an abdominal thickness greater than 16cm and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.
The O-arm O2 Imaging System is compatible with certain image guided surgery systems.
# Substantial Equivalence:
O-arm O2 with revision 4.1.0 software is viewed as substantially equivalent to the unmodified device as there are no changes that impact indications for use or fundamental technology. We have compared the unmodified device (predicate) and the modified device in Table 5-1 below:
{5}------------------------------------------------
| Table 5-1: Device Comparison Table |
|------------------------------------|
|------------------------------------|
| | Predicate | Modified Device | Discussion |
|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | O-arm O2 Imaging System | O-arm O2 Imaging System | |
| | | With revision 4.1 Software | |
| Classification | Class 2 | Class 2 | Identical |
| Product Code | OWB; 892.1650 | OWB; 892.1650 | Identical |
| Indications for<br>Use | The O-arm O2 Imaging System is a<br>mobile x-ray system designed for<br>2D fluoroscopic and 3D imaging for<br>adult and pediatric patients weighing<br>60lbs or greater and having an<br>abdominal thickness greater than<br>16cm and is intended to be used<br>where a physician benefits from 2D<br>and 3D information of anatomic<br>structures and objects with high x-<br>ray attenuation such as bony<br>anatomy and metallic objects.<br>The O-arm O2 Imaging System is<br>compatible with certain image<br>guided surgery systems. | The O-arm O2 Imaging System is a<br>mobile x-ray system designed for<br>2D fluoroscopic and 3D imaging for<br>adult and pediatric patients weighing<br>60lbs or greater and having an<br>abdominal thickness greater than<br>16cm and is intended to be used<br>where a physician benefits from 2D<br>and 3D information of anatomic<br>structures and objects with high x-<br>ray attenuation such as bony<br>anatomy and metallic objects.<br>The O-arm O2 Imaging System is<br>compatible with certain image<br>guided surgery systems. | Identical |
| | Predicate | Modified Device | Discussion |
| | O-arm O2 Imaging System | O-arm O2 Imaging System<br>With revision 4.1 Software | |
| Cone Beam CT | The O-arm O2 Imaging System is a mobile cone-beam x-ray system with isocentric motion options. It allows 3D image reconstruction using a 360 degree rotation of the x-ray source and detector within closed gantry. | The O-arm O2 Imaging System is a mobile cone-beam x-ray system with isocentric motion options. It allows 3D image reconstruction using a 360 degree rotation of the x-ray source and detector within closed gantry. | Identical |
| Detector Technology | 40 x 30 cm (RoHS compliant, Flat-Panel Detector using a CsI scintillation) | 40 x 30 cm (RoHS compliant, Flat-Panel Detector using a CsI scintillation) | Identical |
| Generator Technology | 32 kW, RoHS compliant generator with improved electrical interface. | 32 kW, RoHS compliant generator with improved electrical interface. | Identical |
| 2D Imaging | 2D Fluoroscopic | 2D Fluoroscopic | Identical |
| 3D Imaging<br>(20 cm FOV) | Full Fan (20cm FOV) scan acquisition | Full Fan (20cm FOV) scan acquisition | Identical |
| 3D Imaging<br>Protocols (20 cm FOV) | Available presets:<br>1. Standard 3D<br>2. HD3D (High Definition)<br>3. Enhanced Cranial<br>4. Low Dose 3D | Available presets:<br>5. Standard 3D<br>6. HD3D (High Definition)<br>7. Enhanced Cranial<br>8. Low Dose 3D | Identical |
| 3D Imaging<br>(40 cm FOV) | Half-fan single scan acquisition | Half-fan single scan acquisition | Identical |
| | Predicate | Modified Device | Discussion |
| | O-arm O2 Imaging System | O-arm O2 Imaging System | |
| | | With revision 4.1 Software | |
| 3D Imaging<br>Protocols (40<br>cm FOV) | Available presets:<br>HD3D (high definition)<br>equivalent to 750 projections | Available presets:<br>HD3D (high definition)<br>equivalent to 750 projections | Identical |
| Annotation | Stereotaxy protocols<br>Allows for adding arrows, lines and<br>text to 2D images | Stereotaxy protocols<br>Allows for adding arrows lines and<br>text to 2D images<br><br><b>Added:</b><br>Additional annotation capability to<br>perform angular measurements onto<br>a 2D images. These measurements<br>include closed, open and Cobb<br>angles. It also provides the ability to<br>place a right angle on the image. | Additional functionality |
| Image<br>Transfer | Allows for transfer of data to<br>external devices. | Automatically transfers auto-<br>registered navigation scans.<br><br><b>Added:</b><br>Depending upon the clinical<br>application and workflow within the<br>procedure, this will automatically<br>transfer non-auto-registered (non-<br>navigated) images to the navigation<br>system | Improves workflow<br>The capability to transfer<br>non-auto-registered (non-<br>navigated) images to the<br>navigation system existed in<br>510(k) K15000 but required<br>several manual steps |
| | Predicate | Modified Device | Discussion |
| | O-arm O2 Imaging System | O-arm O2 Imaging System | |
| | | With revision 4.1 Software | |
| 3D<br>Visualization<br>(Enhanced<br>Dynamic<br>Range) | 3D visualization of CBCT image on<br>the MVS. It allows the user to<br>window level the images as well as<br>render oblique views | 3D visualization of CBCT image on<br>the MVS. It allows the user to<br>window level the images as well as<br>render oblique views<br><br><b>Added:</b><br>Improved visualization of images<br>that contain objects of high-x-ray<br>attenuation such as metal implants<br>on the Mobile View Station. | Improved visualization<br>capability |
| Cybersecurity | Hardcoded passcodes for system<br>access<br><br>Industry standard protocols with<br>error detection for data transmission<br>and storage. | Industry standard protocols with<br>error detection for data<br>transmission and storage.<br><br><b>Removed</b> hardcoded passwords<br><br><b>Added:</b><br>Authentication that includes user<br>names and passcodes<br><br>Software integrity check | Increased cybersecurity<br>protection. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# Performance Testing:
Testing conducted demonstrates the product will perform as intended according to the outlined design requirements. The following testing was conducted in accordance to the following FDA recognized standards:
- AAMI/ANSI ES 60601-1:2012 Medical Electrical Equipment Part 1: General . Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
- . IEC 60601-1-2:2007 - Medical Electrical Equipment - Part 1-2: General requirements for safety; Electromagnetic Compatibility - Requirements and Tests (2/2014)
- IEC 60601-1-3:2008 Medical Electrical Equipment Part 1-3:General Requirements for ● Basic Safety and Essential Performance - Collateral Standard: Radiation Protection in Diagnostic X-ray Equipment (1/2014)
- IEC 60601-2-28:2010 Medical electrical equipment part 2: Particular requirements for . the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis (8/2013)
- IEC 60601-2-43:2010 Medical electrical equipment Part 2-43: Particular requirements ● for the basic safety and essential performance of X-ray equipment for interventional procedures (3/2010)
As required in Special 510(k) submissions Medtronic Navigation Inc. (Littleton) performed a Risk / Hazard Analysis in accordance to company procedures which was used to assess the impact of the modifications listed above.
Based on the Risk / Hazard Analysis an identification of the verification and / or, validation activities required was made and appropriate methods or tests and the applicable pass / fail criteria applied were performed.
The following guidance documents used in the development of 0-arm O2 Imaging System:
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices .
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
- . Off-The-Shelf Software Use in Medical Devices
# Conclusion:
The O-arm O2 Imaging system is similar in technological characteristics, imaging performance and indications for use as the predicate devices listed. These aspects, along with the functional testing conducted to the FDA recognized standards, demonstrate that 0-arm 02 Imaging System with 4.1 software does not raise new risks of safety and effectiveness when compared to the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.