The Confidence Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip. The Confidence Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The Confidence Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Confidence Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. The Confidence Blood Glucose Test Strips are for use with the Confidence Blood Glucose Meter to quantitatively measure glucose in capillary whole blood samples drawn from the fingertip
Device Story
Confidence Blood Glucose Monitoring System; electrochemical biosensor; measures glucose in capillary whole blood via fingertip. User inserts test strip; manually codes meter; applies 1.2 uL blood sample. Device uses FAD-GDH enzyme and redox mediator; amperometric detection; converts current to plasma-equivalent glucose concentration (mg/dL). Results displayed within 10 seconds. Intended for home self-testing by patients with diabetes. Not for clinical/assisted-use settings. Benefits include convenient, rapid glycemic monitoring for diabetes management. System includes meter, test strips, and control solutions. Single-patient use only; requires disinfection with EPA-certified germicidal cloths.
Clinical Evidence
User performance study with 187 lay users comparing self-tested capillary blood glucose against YSI 2300 laboratory comparator. Results: For glucose <75 mg/dL, 78% within 5 mg/dL and 100% within 10 mg/dL. For glucose >75 mg/dL, 90% within 10% and 100% within 15% of comparator. Linear regression: y = 1.01x + 0.29, R^2 = 0.99. Analytical performance supported by precision, linearity, and interference studies.
Technological Characteristics
Electrochemical biosensor; FAD-GDH enzyme; amperometric detection. Manual calibration coding. Measuring range 50-600 mg/dL. Sample volume 1.2 uL. Operating conditions: 5-40°C, 10-85% RH. Compliant with IEC 61010-1, CLSI EP06-A, and CLSI EP07-A2. Traceable to NIST SRM #917c. Single-patient use; validated for 200 disinfection cycles.
Indications for Use
Indicated for people with diabetes for self-testing of capillary whole blood glucose from the fingertip to monitor diabetes control. Contraindicated for neonatal use, diagnosis or screening of diabetes, and use in critically ill patients. Not for use in patients who are dehydrated, hypotensive, in shock, or in hyperglycemic/hyperosmolar state with or without ketosis.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
OneTouch Verio Flex Blood Glucose Monitoring System (k150214)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k173658
B. Purpose for Submission:
New Device
C. Measurand:
Glucose in capillary whole blood from the fingertip
D. Type of Test:
Quantitative, amperometric detection (FAD-GDH)
E. Applicant:
Applied Biomedical LLC
F. Proprietary and Established Names:
Confidence Blood Glucose Monitoring System
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1345, Glucose test system
2. Classification:
Class II
3. Product code:
NBW, System Test, Blood Glucose, Over the Counter
{1}
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indication(s) for use below.
2. Indication(s) for use:
The Confidence Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip. The Confidence Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
The Confidence Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.
The Confidence Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. The Confidence Blood Glucose Test Strips are for use with the Confidence Blood Glucose Meter to quantitatively measure glucose in capillary whole blood samples drawn from the fingertip
3. Special conditions for use statement(s):
- For over-the-counter use
- For single patient use only
- Not for neonatal use
- Should not be used for diagnosis of or screening for diabetes mellitus.
- Not for use in critically ill patients
- The system should not be used on people who are dehydrated, hypotensive, in shock, or in hyperglycemic/hyperosmolar state, with or without ketosis.
- A hematocrit that is either very high (above 55%) or very low (below 30%) can cause false results.
- Only for use with fresh capillary whole blood
- You may receive wrong results at altitudes above 10,000 ft
- This device is not intended for use in healthcare or assisted-use settings such as hospitals, physician offices, or long-term care facilities because it has not been part of glycemic control procedures. Use of this device on multiple patients may lead to transmission of Human Immunodeficiency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or other bloodborne pathogens.
2
{2}
4. Special instrument requirements:
Confidence Meter
I. Device Description:
The Confidence Blood Glucose Monitoring System consists of Confidence Blood Glucose Test Strips, the Confidence Meter, and three levels of Confidence Glucose Control Solutions (Levels 1, 2 and 3). To perform a blood glucose measurement, a test strip is inserted in the meter. On inserting the strip, the meter turns on and displays a code, and the user must select a code corresponding to their strip vial using meter's interface. The meter will be ready to apply sample by showing a droplet on the LCD display. On applying blood on the end of the test strip the meter displays the plasma equivalent value (mg/dL) in the blood sample within 10 seconds.
J. Substantial Equivalence Information:
1. Predicate device name(s):
OneTouch Verio Flex Blood Glucose Monitoring System
2. Predicate 510(k) number(s):
k150214
3. Comparison with predicate:
| Item | Candidate Device Confidence Blood Glucose Monitoring System | Predicate Device OneTouch Verio Flex Blood Glucose Monitoring System k150214 |
| --- | --- | --- |
| Indications for Use | Quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip as an aid to monitor the effectiveness of diabetes control | Same |
| Specimen type | Capillary whole blood | Same |
| Sample site | Fingertip | Same |
| Test Principle | Amperometry | Same |
| Enzyme | FAD-GDH | Same |
{3}
| Item | Candidate Device Confidence Blood Glucose Monitoring System | Predicate Device OneTouch Verio Flex Blood Glucose Monitoring System k150214 |
| --- | --- | --- |
| Altitude | Up to 10,000 | Same |
| Sample Volume | 1.2 uL | 0.4 uL |
| Measuring Range | 50-600 mg/dL | 20-600 mg/dL |
| Measuring Time | 10 seconds | 5 seconds |
| Hematocrit | 30%-55% | 20%-60% |
| Calibration Coding | Manual code input | Non-coding |
K. Standard/Guidance Document Referenced (if applicable):
FDA Guidance “Self-Monitoring Blood Glucose Test System for Over-the-Counter Use” released on 11 October, 2016.
CLSI EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition.
CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
ISO 14971 - Medical devices - Application of risk management to medical devices (2007).
ISO 10993-1 Biological Evaluation Of Medical Devices -- Part 1: Evaluation And Testing Within A Risk Management Process Fourth Edition (2009).
IEC 61010-1 (2nd Edition) - Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 1: General Requirements.
L. Test Principle:
The Confidence Blood Glucose Monitoring System is an electrochemical biosensor using a glucose-oxidizing enzyme, Flavin Adenine Dinucleotide-Glucose Dehydrogenase (FAD-GDH), along with a redox mediator on a disposable test strip (the electrochemical embedded sensor), and a hand-held device to measure current generated by test strip. The software in the hand-held device converts the measured current into glucose concentration, where
{4}
generated current is directly proportional to glucose concentration. The resulting glucose value is displayed on the meter in $\mathrm{mg / dL}$ for the user.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Within-Run Precision
The sponsor performed repeatability (within-day) precision studies using venous whole blood samples adjusted to four different glucose concentration levels (51 to 110, 111 to 150, 151 to 250, and 251 to $350\mathrm{mg / dL}$ ). Each glucose concentration level was analyzed in replicates of 10, with 3 test strip lots and 10 meters, for a total of 100 replicates per glucose level per test strip lot (for a total of 300 tests per each glucose level). Results are summarized below:
| Glucose Level (mg/dL) | Strip lot | Mean (mg/dL) | SD (mg/dL) | % CV |
| --- | --- | --- | --- | --- |
| 1 (51-110) | 1 | 59 | 1.7 | 2.9 |
| | 2 | 61 | 1.7 | 2.8 |
| | 3 | 63 | 2.2 | 3.5 |
| | Combined | 61 | 1.9 | 3.0 |
| 2 (111-150) | 1 | 136 | 2.7 | 2.0 |
| | 2 | 135 | 3.6 | 2.7 |
| | 3 | 132 | 3.4 | 2.6 |
| | Combined | 134 | 3.1 | 2.4 |
| 3 (151-250) | 1 | 214 | 4.9 | 2.3 |
| | 2 | 220 | 6.2 | 2.8 |
| | 3 | 221 | 4.1 | 1.9 |
| | Combined | 218 | 5.4 | 2.3 |
| 4 (251-350) | 1 | 320 | 6.6 | 2.1 |
| | 2 | 327 | 6.6 | 2.0 |
| | 3 | 328 | 8.5 | 2.6 |
| | Combined | 325 | 7.2 | 2.2 |
# Intermediate Precision
Intermediate (day-to-day) precision was evaluated using four levels of glucose control solution $(50 - 100\mathrm{mg / dL},111 - 150\mathrm{mg / dL},151 - 250\mathrm{mg / dL},$ and $251 - 400\mathrm{mg / dL})$ . Each control level was analyzed on 10 meters using 3 lots of test strips each day for a total of 10 days, giving a total of 100 measurements per lot for a total of 300 measurements for each glucose level. Results are summarized below:
{5}
| Glucose Level (mg/dL) | Strip lot | Mean (mg/dL) | SD (mg/dL) | % CV |
| --- | --- | --- | --- | --- |
| 1 (50-110) | 1 | 63 | 2.2 | 3.4 |
| | 2 | 59 | 2.3 | 4.0 |
| | 3 | 63 | 2.5 | 3.9 |
| | Combined | 62 | 2.4 | 3.8 |
| 2 (111-150) | 1 | 139 | 5.6 | 4.0 |
| | 2 | 121 | 4.8 | 3.9 |
| | 3 | 125 | 4.6 | 3.7 |
| | Combined | 128 | 5.0 | 3.9 |
| 3 (151-250) | 1 | 215 | 8.0 | 3.7 |
| | 2 | 201 | 5.7 | 2.8 |
| | 3 | 215 | 5.9 | 2.8 |
| | Combined | 210 | 6.5 | 3.1 |
| 4 (251-400) | 1 | 420 | 9.0 | 2.2 |
| | 2 | 402 | 5.1 | 1.3 |
| | 3 | 400 | 7.5 | 1.9 |
| | Combined | 407 | 7.2 | 1.8 |
# b. Linearity/assay reportable range:
Linearity testing was performed using venous whole blood samples. Twelve samples with the following glucose concentrations were prepared: 40, 48, 74, 151, 204, 261, 361, 419, 473, 522, 553, and $630\mathrm{md / dL}$ (as established using a laboratory comparator method YSI 2300). Testing was conducted in quadruplicate on three meters using three test strip lots $(n = 12$ tests per glucose concentration). The evaluation yielded the following regression results:
| Strip Lot | Slope | Y-Intercept | R2 |
| --- | --- | --- | --- |
| 1 | 1.05 | 4.03 | 0.999 |
| 2 | 1.01 | 0.13 | 0.999 |
| 3 | 1.04 | 4.1 | 0.999 |
The results of the study support the sponsor's claimed glucose measurement range of 50 to $600\mathrm{mg / dL}$ . If a sample is less than $50\mathrm{mg / dL}$ glucose, the result is flagged by the meter as LO. If a sample result exceeds $600\mathrm{mg / dL}$ glucose, the result is flagged by the meter as HI. The LO and HI functions were validated and demonstrated to function as intended.
{6}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
The system is traceable to the NIST SRM #917c glucose reference material.
Open and Closed Vial Stability:
Test strip stability was assessed using accelerated and real-time stability studies. Testing protocols and acceptance criteria were reviewed and found to be acceptable. The labeling includes claims that the Confidence Test Strips are stable for 6 months after opening and 12 months unopened when stored between 41-104°F (5-40°C) and 10-85% relative humidity.
d. Detection limit:
The reportable range for the Confidence Blood Glucose Monitoring System is 50-600 mg/dL. This range was verified by the linearity study (M.1.b).
e. Analytical specificity:
Interference studies were performed by spiking endogenous and exogenous substances into venous whole blood. Twenty-eight substances were tested at three glucose levels (50-70, 110-130, and 225-270 mg/dL) in replicates of 10 on each of 3 strip lots per glucose levels yielding thirty replicates per glucose level. Results of spiked samples with potential interferent were compared to samples with zero potential interferent by measurement on the Confidence meter. The compounds at the concentrations listed below did not have significant interference (defined by the sponsor as average bias between spiked and control sample glucose results within ± 10% with glucose >100 mg/dL and within 10 mg/dL with glucose <100 mg/dL). The highest concentraitons at which no significant interference was observed are presented in the following table:
| Interfering Substance | Maximum tested concentration with no significant interference (mg/dL) |
| --- | --- |
| Acetaminophen | 9.0 |
| Ascorbic acid | 5.0 |
| Bilirubin | 35.0 |
| Cholesterol (total) | 325 |
| Creatinine | 20.0 |
| Dopamine HCl | 1.5 |
| Ephedrine | 0.51 |
| Galactose | 15.1 |
| Gentisic Acid | 2.5 |
| Glutathione | 90.0 |
{7}
| Interfering Substance | Maximum tested concentration with no significant interference (mg/dL) |
| --- | --- |
| Hemoglobin | 210 |
| Ibuprofen | 55.2 |
| Isomalt | 1.3 |
| L-dopa | 3.0 |
| Maltitol | 1.1 |
| Maltose | 300 |
| Methyl-dopa | 1.54 |
| Pralidoxime Iodide (PAM) | 20 |
| Salicylate | 58 |
| Salicylic acid | 60.55 |
| Sodium Chloride (HCl) | 457 |
| Tetracycline | 1.7 |
| Tolazamide | 8.6 |
| Tolbutamide | 78.0 |
| Triglycerides | 3,000 |
| Uric Acid | 10.0 |
| Xylose | 50.0 |
Based on the test results, the sponsor included the following statements in the labeling:
- PAM (Pralidoxime Iodide): Do not test blood glucose during or soon after PAM treatment (>20 mg/dL), it may affect the glucose results.
- Uric acid may cause inaccurate results in certain condition, such as gout or kidney disease, if concentration of the substance is greater than physiological concentration (>10 mg/dL), if you are not sure consult your doctor before using it.
- You should not use this meter if you are taking acetaminophen or acetaminophen containing drugs (e.g. Tylenol™). Acetaminophen in your blood (>9 mg/dL) may cause an inaccurate result.
- Do not test blood glucose during or soon after a xylose absorption test, within 24 hours, it may give falsely elevated results.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device
Not applicable. See User Performance Studies below (section M.3.d) for demonstration of system accuracy in the hands of the intended user.
{8}
b. Matrix comparison:
Not applicable
# 3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
# User Performance Study
To assess the performance of the Confidence Glucose Monitoring System in the hands of the intended users, the sponsor performed a study with 187 lay user participants who collected and tested samples from their own fingertip. Results were analyzed by comparing blood glucose results obtained by the lay users with the Confidence Glucose Monitoring System against results obtained using a laboratory comparator method (YSI 2300 analyzer). Glucose concentrations in the samples ranged from approximately 60 to $480\mathrm{mg / dL}$ as measured by the laboratory comparator method. Results are summarized in the tables below:
| For glucose concentrations <75 mg/dL | | | |
| --- | --- | --- | --- |
| Within 5 mg/dL | Within 10 mg/dL | Within 15 mg/dL | |
| 78% (14/18) | 100% (18/18) | 100% (18/18) | |
| For glucose concentrations >75 mg/dL | | | |
| Within 5 % | Within 10 % | Within 15 % | Within 20 % |
| 59% (100/169) | 90% (152/169) | 100% (169/169) | 100% (169/169) |
Results of linear regression analysis:
$$
y = 1.01x + 0.29, R^2 = 0.99
$$
A readability assessment indicated a Flesch-Kincaid Score of 8 and lower for all instructional materials included with this device.
# 4. Clinical cut-off:
Not applicable
# 5. Expected values/Reference range:
{9}
The sponsor states that American Diabetes Association (ADA) suggests the following ranges of glucose levels for nonpregnant adults without diabetes:
Before a meal: <100 mg/dL
After a meal: < 140 mg/dL
Reference: American Diabetes Association (ADA) Clinical Practice Recommendations: Classification and Diagnosis of Diabetes. Diabetes Care 2018; 41 (Supplement 1):S13-S22.
## N. Instrument Name:
Confidence Blood Glucose Monitoring System
## O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X ☐ or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☐ X ☐
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
The system is intended to be used with capillary whole blood from the finger. The whole blood sample is applied directly to the test strip by capillary action.
10
{10}
5. Calibration:
The meter is manually coded by the user.
6. Quality Control:
Confidence Control solutions are aqueous solutions containing glucose and are available at three levels (level 1, level 2 and level 3). Instructions on how to order the control solutions are included in the user manual. The control solution readings are not included in the average of the patient results.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. Sample Volume Study:
To verify the test strip minimum sample volume requirement, fresh venous blood with five volumes (0.6, 0.7, 0.9, 1.2, and 1.5 uL) were tested. Sample concentrations included four glucose concentrations (56, 121, 194, and 302 mg/dL), established using a laboratory comparator method YSI 2300 analyzer, were tested at each volume. Values obtained with the candidate device were compared to values obtained using a laboratory-based comparator method. Results support the claimed minimum sample volume of 1.2 μL. The sponsor provided validation studies demonstrating that with blood volumes below 1.2 μL, the insufficient sample volume error message functioned as intended.
2. Hematocrit Study:
The effect of different hematocrit levels was evaluated using venous whole blood samples with hematocrit levels of 25-60% (25, 30, 35, 40, 45, 50, 55 and 60%) at four levels of glucose (50 - 100, 111-150, 151-250, and 251-400 md/dL). Each sample was tested in replicates of ten using two meters and three test strip lots. Results from the meter were compared to results obtained using a laboratory-based comparator measurement (YSI 2300 analyzer). The evaluation of bias and percent bias relative to values obtained on the YSI 2300 analyzer demonstrate acceptable performance across the claimed hematocrit range of 30-55%.
3. Altitude Study:
The effect of altitude was evaluated using blood samples spiked to four glucose levels (50-110, 111-150, 151-250, and 251-400 mg/dL). The samples were tested at 340 - 10,010 feet above sea level. Results obtained were compared with those obtained with the comparator method (YSI 2300). The results demonstrate acceptable bias to the comparator to support the claims in the labeling at altitudes up to 10,000 feet have no significant effect on blood glucose measurements from the Confidence Blood Glucose Monitoring System.
4. Operating Conditions:
The sponsor performed temperature and humidity studies using venous blood samples with three glucose levels (50-110, 111-150, 151-300). Temperatures ranging from 41°F - 104°F (5-40°C) and relative humidity from 10% to 85% were tested. Meter results were compared
11
{11}
to the YSI 2300 comparator analyzer. Four temperature and humidity combinations were tested including low temperature/low humidity, low temperature/high humidity, high temperature/low humidity and high temperature/high humidity. The results support the claims in the labeling that the system can be used in conditions of 43°F - 104°F (5-40°C) with relative humidity of 10 to 85%.
5. Infection Control and Robustness Studies:
The device system is intended for single-patient use only. Disinfection efficacy studies were performed on the materials comprising the meter by an outside commercial testing laboratory demonstrating complete inactivation of hepatitis B virus (HBV) with the chosen disinfectant, EPA Certified Germicidal Disposable Cloth (PDI® SUPER SANI-CLOTH®), EPA Reg. No.: 9480-4 (PN1000617). Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or external materials of the meter after 200 cleaning and disinfection cycles. The robustness studies support 3 years of single-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
6. EMC Testing:
The sponsor provided appropriate documentation certifying that electromagnetic (EMC) testing was performed and the Confidence Blood Glucose Monitoring System was found to be compliant.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.