K173646 · Palodex Group OY · MUH · Dec 19, 2017 · Dental
Device Facts
Record ID
K173646
Device Name
FS Ergo
Applicant
Palodex Group OY
Product Code
MUH · Dental
Decision Date
Dec 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
The FS Ergo sensor is a digital sensor which is indicated for acquiring dental intra-oral radiography images. The FS Ergo sensor shall be operated by healthcare professionals, who are educated and competent to perform the acquisition of dental intraoral radiographs.
Device Story
FS Ergo is a USB-connected digital intraoral dental X-ray sensor; captures X-ray photons and converts them into electronic images. Operated by dental healthcare professionals in clinical settings. Device features a bendable, soft silicone housing with urethane coating, small diameter flexible cable, and 32-degree cable exit for ergonomic patient comfort. Sensor automatically detects X-ray exposure (60-70 kV range; min 40 µGy dose) when connected to a workstation running compatible image-capturing software. Images are transmitted to the workstation for viewing and diagnostic use. Benefits include improved patient comfort via ergonomic design and efficient digital image acquisition for dental diagnostics.
Clinical Evidence
No clinical data was required. Substantial equivalence was supported by bench testing, including electrical, mechanical, and safety testing per IEC 60601-1, IEC 60601-1-6, IEC 62366, IEC 62304, and EMC testing per IEC 60601-1-2. Biocompatibility was confirmed per ISO 10993-1. Image quality was validated via bench testing using skull phantoms, reviewed by a qualified clinician to be of acceptable quality.
Technological Characteristics
Digital intraoral X-ray sensor; soft silicone housing with urethane coating. Dimensions: 40 x 30 mm (Size 1.5). Pixel size: 17 x 17 µm. Connectivity: USB 2.0. Power: 5 Vdc / 500 mA. IP68 rating. Standards: IEC 60601-1:2015, IEC 60601-1-2:2014, ISO 10993-1:2009. Continuous mode of operation.
Indications for Use
Indicated for acquiring dental intra-oral radiography images. For use by healthcare professionals educated and competent in dental intraoral radiography acquisition.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
December 19, 2017
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
PaloDEx Group Oy % Anni Lundholm Regulatory Manager Nahkelantie 160 Tuusula 04300 FINLAND
Re: K173646
Trade/Device Name: FS Ergo Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: November 10, 2017 Received: November 27, 2017
Dear Anni Lundholm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn
(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert Oolo
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
### 510(k) Number (if known)
### K173646
Device Name FS Ergo
### Indications for Use (Describe)
The FS Ergo sensor is a digital sensor which is indicated for acquiring dental intra-oral radiography images. The FS Ergo sensor shall be operated by healthcare professionals, who are educated and competent to perform the acquisition of dental intraoral radiographs.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## December 19, 2017
## 510(k) SUMMARY for FS Ergo
## Submitter Information
| Name: | Palodex Group Oy |
|-------------------|------------------------------------------------|
| Address: | Nahkelantie 160<br>FI-04300 Tuusula<br>Finland |
| Contact Person: | Anni Lundholm |
| Email: | regulatory@kavokerr.com |
| Telephone Number: | +358 10 270 2000 |
| Fax Number: | +358 10 270 2230 |
## Device Name
Proprietary Name: FS Ergo Common Name: Digital intraoral sensor system Classification Name: System, X-Ray, Extraoral source, Digital FDA CDRH Panel: Radiology Product Code: MUH Requlation Number: 21 CFR 872.1800 Device Class: Class 11
# Primary Predicate Device
| Proprietary Name: | Snapshot |
|--------------------|------------------------------------------|
| 510(k) Number: | K081925 |
| Manufacturer: | Instrumentarium Dental, PaloDEX Group Oy |
| Regulation Number: | 21 CFR 872.1800 |
| Regulation Name: | Extraoral source x-ray system |
| Product Code: | MUH |
| Device Class: | II |
## Reference Predicate Device
| Proprietary Name: | EzSensor Soft (EzSensor Soft, EzSensor Soft i, EzSensor Bio and EzSensor Bio I) | | |
|--------------------|---------------------------------------------------------------------------------|--|--|
| 510(k) Number: | K151707 | | |
| Manufacturer: | Rayence Co., Ltd. | | |
| Regulation Number: | 21 CFR 872.1800 | | |
| Regulation Name: | Extraoral source x-ray system | | |
| Product Code: | MUH | | |
| Device Class: | II | | |
# Description of the Device
The FS Ergo is a USB-driven sensor designed to capture intraoral digital dental x-ray images. FS Ergo can be used with any X-ray units intended for dental intraoral purposes. When connected to a computer with an appropriate image capturing software application, images are automatically acquired when the FS Ergo sensor receives a perceptible X-ray dose. The FS Ergo sensor is designed with advanced ergonomic principles. It has a bendable sensor housing together with other physical
{4}------------------------------------------------
features supporting advanced erqonomics: small diameter flexible cable, 32 degrees' cable exit and sensor dimensions that fit all needs (size 1.5). Comfort is achieved in with large corner chamfers, rounded edges, and thicker sensor structure.
The basic system of the FS Ergo consists of senor unit, sensor holders and hygienic covers (which are medical devices separately cleared by the FDA). A connection to a computer is required but the computer is provided by the users e.q. the computer is not part of the sensor system itself.
# Indications for Use
The FS Ergo sensor is a digital sensor which is indicated for acquiring dental intra-oral radiography images. The FS Ergo sensor shall be operated by healthcare professionals, who are educated and competent to perform the acquisition of dental intraoral radiographs.
# Description of Substantial Equivalence
The substantial equivalence of the FS Ergo is demonstrated by two predicate devices: primary predicate device Snapshot cleared under K081925 and manufactured by Instrumentarium Dental, Palodex Group Oy and secondary predicate device EzSensor Soft cleared under K151707 and manufactured by Rayence Co., Ltd. Indications for use of the FS Erqo and the predicate devices Snapshot (K081925) and EzSensor Soft (K151707) are not identical but they include the same fundamental elements of producing x-ray radiography images for professional use.
Snapshot (K081925) is the primary predicate device since its technological characteristics is most similar to the FS Ergo. EzSensor Soft (K151707) is the secondary predicate device since its structure (sensor head) is soft like the FS Ergo. EzSensor Soft is not, however, available for technical testing and image quality comparison.
Technical characteristics of the FS Ergo sensor and the predicate devices Snapshot cleared under K081925 and EzSensor Soft cleared under K151707 are described in the table 7-1 below.
| Characteristics | Primary Predicate Device<br>Snapshot<br>(K081925) | Secondary Predicate Device<br>EzSensor Soft<br>(K151707) | Subject Device<br>FS Ergo |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Snapshot is<br>intended to be used<br>by dentists and<br>other qualified<br>professional for<br>producing diagnostic<br>x-ray radiographs of<br>definition, jaws and<br>other oral<br>structures. | EzSensor Soft,<br>EzSensor Soft i,<br>EzSensor Bio and<br>EzSensor Bio i<br>Digital Dental Intra<br>Oral Sensors are<br>intended to collect<br>dental x-ray<br>photons and convert<br>them into electronic | The FS Ergo sensor<br>is a digital sensor<br>which is indicated<br>for acquiring dental<br>intra-oral<br>radiography images.<br>The FS Ergo sensor<br>shall be operated by<br>healthcare<br>professionals, who |
Table 7-1. Significant characteristics comparison table.
{5}------------------------------------------------
| Characteristics | Primary Predicate<br>Device<br>Snapshot<br>(K081925) | Secondary<br>Predicate Device<br>EzSensor Soft<br>(K151707) | Subject Device<br>FS Ergo |
|------------------------------------------|------------------------------------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| | | impulses that may<br>be stored, viewed<br>and manipulated for<br>diagnostic use by<br>dentists. | are educated and<br>competent to<br>perform the<br>acquisition of dental<br>intraoral<br>radiographs. |
| Sensor type | Intra-oral | Intra-oral | Intra-oral |
| Sensor<br>Dimensions<br>(mm) | Size 1: 37 x 26<br>Size 2: 42 x 31 | Size 1.0: 38 x 27<br>Size 1.5: 41 x 31<br>Size 2.0: 44 x 33 | Size 1.5: 40 x 30 |
| Sensor<br>Thickness (mm) | 12 | 5.5 | 8.5 |
| Sensor Shape | Rectangular, four<br>rounded corners | Rectangular, four<br>rounded corners | Rectangular, four<br>rounded corners |
| Resolution –<br>Pixel Size ( μm) | 19 x 19 | 14.8 x 14.8 | 17 x 17 |
| Theoretical<br>Resolution | 26.3 lp/mm | 33.78 lp/mm | 29.4 lp/mm |
| Sensor Active<br>Area (mm) | Size 1: 30 x 20<br>Size 2: 36 x 26 | Size 1.0: 30 x 20<br>Size 1.5: 33 x 24<br>Size 2.0: 36 x 26 | Size 1.5: 32 x 25 |
| Sensor Housing | Rigid, polyamide<br>(PA) | Soft, silicone with<br>urethane coating | Soft, silicone with<br>urethane coating |
| Connection to<br>PC | USB 2.0 | USB 2.0 | USB 2.0 |
| Electrical Rating | 5 Vdc / 500 mA | 5 Vdc / 500 mA | 5 Vdc / 500 mA |
| IP Classification | Sensor: IP67 | Sensor: IP68 | Sensor: IP68 |
| Mode of<br>Operation | Continuous mode | Continuous mode | Continuous mode |
| Electrical Safety | IEC 60601-1:1998 | IEC 60601-1:2015 | IEC 60601-1:2015 |
# Principle of Operation
The FS Ergo sensor is automatically powered when connected to the USB port on a workstation. No power switch is needed. User activates image capturing from an image capturing software installed on the connected workstation. The sensor detects x-ray exposure automatically and transfers the image to the software without additional user interaction. The FS Ergo can be used with any x-ray units intended for dental intraoral purposes in the range of 60 to 70 kV and at minimal 40 µGy incident dose.
After transmitting the image, the sensor is ready for the next exposure. A full mouth s series of image can be easily captured by sequentially activating the sensor from the software application. Image capture session can be terminated from the software application.
{6}------------------------------------------------
# Non-Clinical Performance Data
The FS Ergo sensor has successfully passed internal design verification and validation.
Electrical, mechanical, safety and performance testing according to standards IEC 60601-1:2005, IEC 60601-1-6:2013, IEC 62366:2014 and IEC 62304:2006 were performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2:2014. Testing was completed by 3rd party test house, and the FS Ergo sensor has passed all tests.
Biocompatibility evaluation was conducted on patient contacting accessory parts and the materials. The parts are found to be in conformance with ISO 10993-1:2009.
Bench test images of skull phantoms were acquired using FS Ergo sensor, the images were reviewed by qualified external clinician to be of acceptable quality for the proposed intended use. Bench testing for the FS Ergo sensor was performed as recommended by the FDA's Guidance for the Submission of 510(k)s for Solid State X-Ray Imaging Devices. Bench testing results indicate substantial equivalence to the predicate device Snapshot.
# Clinical Performance Data
Clinical data was not needed to support substantial equivalence.
# Conclusion as to Substantial Equivalence
Based on the substantial equivalence comparison of intended use, design, technological characteristics, performance, labeling, biocompatibility, standards, and other characteristics, there are no significant differences found between the FS Ergo and the predicate devices Snapshot (cleared under K081925) and EzSensor Soft (cleared under K151707). Minor differences have not been shown to raise new concerns for the proposed indications for use.
The device does not introduce a fundamentally new scientific technology and the nonclinical tests demonstrate that the device is substantially equivalent to the predicate device. All internal verification and validation has been completed successfully.
In summary, FS Ergo sensor described in this submission is substantially equivalent to the Snapshot (cleared under K081925) and the EzSensor Soft (cleared under K151707) and satisfies all criteria of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.