syngo.CT Coronary Analysis is an image analysis software package for evaluating cardiac CT angiography (CTA) volume data sets. Combining digital image processing and visualization tools (multiplanar reconstruction (MPR) thin/thick, maximum intensity projection (MIP) thin/thick, inverted MIP thin/thick, volume rendering technique (VRT), curved planar reformation (CPR)), evaluation tools (coronary vessel centerline calculation, stenosis calculation and plaque analysis) and reporting tools (lesion location, lesion characteristics and key images), the software package is designed to support the physician in confirming the presence of physician-identified coronary lesions and evaluation, documentation and follow-up of any such lesion. These visualization tools allow for characterization (geometry (length, lumen diameter, cross section area, stenosis grade) and appearance (HU values)) of coronary lesions and lesion size over time, helping the physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue. syngo.CT Vascular Analysis is an image analysis software package for evaluating enhanced CT images. Combining digital image processing and visualization tools (multiplanar reconstruction (MPR) thin/thick, maximum intensity projection (MIP) thin/thick, inverted MIP think, volume rendering technique (VRT), curved planar reformation (CPR), processing tools (bone removal (based both on single energy and Dual Energy), table removal) and evaluation tools (vessel centerline calculation, lumen calculation) and reporting tools (lesion location, lesion characteristics and key images), the software package is designed to support the physician in confirming the presence or absence of physician-identified lesions in blood vessels and evaluation and follow-up of any such lesion. These visualization/processing/evaluation tools allow for characterization (geometry (length, lumen diameter, cross section area, stenosis grade) and appearance (HU values)) of vascular lesions and lesion size over time, helping the physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue.
Device Story
Software packages for post-processing cardiac CTA and enhanced CT images; inputs include CT volume data sets. Tools include MPR, MIP, VRT, CPR, bone/table removal, and vessel centerline/lumen/stenosis calculation. Used by physicians in clinical settings to confirm, characterize, and monitor coronary or vascular lesions over time. Output includes quantitative measurements (length, diameter, area, stenosis grade, HU values) and key images for reporting. Assists physicians in clinical decision-making regarding lesion growth and tissue classification.
Clinical Evidence
Bench testing only. Performance testing evaluated coronary centerline centeredness, branch tracing, and vessel coverage, demonstrating increased sensitivity, specificity, and improved coverage compared to predicates. Vascular analysis testing evaluated detection rates, showing higher levels of successfully traced vessels. Software verification and validation confirmed all specifications met acceptance criteria.
Technological Characteristics
Software-based image analysis platform (SOMARIS/8 VB30). Features 2D/3D visualization (MPR, MIP, VRT, CPR), automated vessel tracing, bone/table removal, and quantitative measurement tools. Conforms to DICOM, IEC 62304 (software lifecycle), ISO 14971 (risk management), IEC 62366-1 (usability), and AAMI/ANSI ES60601-1 (safety).
Indications for Use
Indicated for use by physicians to evaluate cardiac CTA volume data sets (syngo.CT Coronary Analysis) or enhanced CT images (syngo.CT Vascular Analysis) for the presence, characterization, and follow-up of coronary or vascular lesions. Supports lesion documentation, measurement (geometry/HU values), and classification of tissue regions.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".
Siemens Medical Solutions USA, Inc. % Ms. Kimberly Mangum Regulatory Affairs Specialist 40 Liberty Boulevard MALVERN PA 19355
March 30, 2018
Re: K173637
Trade/Device Name: syngo.CT Coronary Analysis; syngo.CT Vascular Analysis Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: March 2, 2018 Received: March 9, 2018
Dear Ms. Mangum:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173637
Device Name
syngo.CT Coronary Analysis
#### Indications for Use (Describe)
syngo.CT Coronary Analysis is an image analysis so0ftware package for evaluating cardiac CT angiography (CTA) volume data sets. Combining digital image processing and visualization tools (multiplanar reconstruction (MPR) thin/ thick, maximum intensity projection (MP) thin/thick, inverted MP thin/thick, volume rendering technique (VRT), curved planar reformation (CPR)), evaluation tools (coronary vessel centerline calculation, stenosis calculation and plaque analysis) and reporting tools (lesion location, lesion characteristics and key images), the software package is designed to support the physician in confirming the presence of physician-identified coronary lesions and evaluation, documentation and follow-up of any such lesion. These visualization tools allow for characterization (geometry (length, lumen diameter, cross section area, stenosis grade) and appearance (HU values)) of coronary lesions and lesion size over time, helping the physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 10pt;"></span> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;"></span> <span> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173637
#### Device Name
syngo.CT Vascular Analysis
#### Indications for Use (Describe)
syngo.CT Vascular Analysis is an image analysis software package for evaluating enhanced CT images. Combining digital image processing and visualization tools (multiplanar reconstruction (MPR) thin/thick, maximum intensity projection (MIP) thin/thick, inverted MIP think, volume rendering technique (VRT), curved planar reformation (CPR), processing tools (bone removal (based both on single energy and Dual Energy), table removal) and evaluation tools (vessel centerline calculation, lumen calculation) and reporting tools (lesion location, lesion characteristics and key images), the software package is designed to support the physician in confirming the presence or absence of physician-identified lesions in blood vessels and evaluation and follow-up of any such lesion. These visualization/processing/evaluation tools allow for characterization (geometry (length, lumen diameter, cross section area, stenosis grade) and appearance (HU values)) of vascular lesions and lesion size over time, helping the physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The color of the text is a light blue or teal. The letters are bold and evenly spaced, creating a clear and recognizable logo. The background is plain and white, which makes the text stand out.
#### 510(K) Summary FOR SYNGO.CT CORONARY ANALYSIS AND SYNGO.CT VASCULAR ANALYSIS
Submitted by: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Date Prepared: March 29, 2018
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### l. Submitter
Importer/Distributor Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Establishment Registration Number 2240869
Manufacturing Site Siemens Healthcare GmbH Siemensstr. 1 D-91301 Forchheim, Germany Establishment Registration Number 3004977335
#### Contact Person
Kimberly Mangum Regulatory Affairs Specialist Siemens Medical Solutions, Inc. USA 40 Liberty Boulevard Malvern, PA 19355 Phone: (610) 448 - 6477 Fax: (610) 640 - 4481 Email: kimberly.mangum@siemens-healthineers.com
#### II. Device Name and Classification
| Product Name: | syngo.CT Coronary Analysis |
|-----------------------|----------------------------------|
| Proprietary Name: | syngo.CT Coronary Analysis |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| Product Name: | syngo.CT Vascular Analysis |
| Proprietary Name: | syngo.CT Vascular Analysis |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
#### III. Predicate Device Predicate Device for syngo.CT Coronary Analysis: Trade Name: syngo.CT Coronary Analysis 510(k) Number: K100637
{5}------------------------------------------------
# SIEMENS
Clearance Date: 05/26/2010 Classification Name: Computed tomography X-ray system Classification Panel: Radiology CFR Section: 21 CFR §892.1750 Device Class: Class II JAK Product Code: This predicate has not been subject to a design-related recall Recall Information:
# Predicate Device for syngo.CT Vascular Analysis:
Trade Name: syngo.CT Vascular Analysis 510(k) Number: K112020 Clearance Date: 08/18/2011 Classification Name: Computed tomography X-ray system Classification Panel: Radiology 21 CFR §892.1750 CFR Section: Device Class: Class II Product Code: JAK Recall Information: This predicate has not been subject to a design-related recall
# IV. Device Description
# syngo.CT Coronary Analysis
syngo.CT Coronary Analysis is an image analysis post-processing software application designed to support evaluation of the vessels of the heart. This submission describes the following modifications that were made to predicate device syngo.CT Coronary Analysis (K100637, clearance date 5/26/2010):
- 1) Support of software version SOMARIS/8 VB30 which supports the following functionality:
- a. Coronary Tracing Quality Improvement
- Rapid Results Technology Support for Curved Vessel Ranges ڪ
- 2) Update 510(k) Information
Combining digital image processing and visualization tools (multiplanar reconstruction (MPR) thin/thick, minimum intensity projection (MIP) thin/thick, inverted MIP thin/thick, volume rendering technique (VRT), curved planar reformation (CPR)), evaluation tools (coronary vessel centerline calculation, stencils calculation and plaque analysis) and reporting tools (lesion location, lesion characteristics and kev images), the software package is designed to support the physician in confirming the presence of physician-identified coronary lesions and evaluation, documentation and follow-up of any such lesion. These visualization/evaluation tools allow for characterization of coronary lesions and lesion size over time, helping the physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue.
# syngo.CT Vascular Analysis
syngo.CT Vascular Analysis is a post-processing software application to support evaluation of the large vessels of the body. This submission describes the following modifications that were made to predicate device syngo.CT Vascular Analysis (K112020, clearance date 08/18/2011):
- Modified Indications for Use Statement 1)
- Support of software version SOMARIS/8 VB30 which supports the following 2) functionality:
- a. Rapid Results Technology Support for curved vessel ranges
- Bone Removal Quality Improvement b.
- Vessel Tracing Quality Improvement C.
- 3) Update 510(k) Information
Combining digital image processing and visualization tools (multiplanar reconstruction (MPR) thin/thick, maximum intensity projection (MIP) thin/thick, inverted MIP thin/thick, volume rendering technique (VRT), curved planar reformation (CPR), processing tools (bone removal (based both on single energy and Dual Energy), table removal) and evaluation tools (vessel centerline calculation, lumen calculation, stenosis calculation) and reporting tools (lesion location, lesion characteristics and key images), the software package is designed to support
{6}------------------------------------------------
# SIEMEN
the physician in confirming the presence or absence of physician-identified lesions in blood vessels and evaluation, documentation and follow-up of any such lesion. These visualization/processing/evaluation tools allow for characterization of vascular lesions and lesion size over time, helping the physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue.
## V. Indications for Use
## Indications for Use for syngo.CT Coronary Analysis
syngo.CT Coronary Analysis is an imaqe analysis soOftware package for evaluating cardiac CT angiography (CTA) volume data sets. Combining digital image processing and visualization tools (multiplanar reconstruction (MPR) thin/thick, maximum intensity projection (MIP) thin/thick, inverted MIP thin/thick, volume rendering technique (VRT), curved planar reformation (CPR)), evaluation tools (coronary vessel centerline calculation, stenosis calculation and plaque analysis) and reporting tools (lesion characteristics and key images), the software package is designed to support the physician in confirming the presence or absence of physician-identified coronary lesions and evaluation, documentation and follow-up of any such lesion. These visualization/evaluation tools allow for characterization (geometry (length, lumen diameter, cross section area, stenosis grade) and appearance (HU values)) of coronary lesions and lesion size over time, helping the physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue.
### Indications for Use for syngo.CT Vascular Analysis
syngo.CT Vascular Analysis is an image analysis software package for evaluating enhanced CT images. Combining digital image processing and visualization tools (multiplanar reconstruction (MPR) thin/thick, maximum intensity projection (MIP) thin/thick, inverted MIP thin/thick, volume rendering technique (VRT), curved planar reformation (CPR), processing tools (bone removal (based both on single energy and Dual Energy), table removal) and evaluation tools (vessel centerline calculation, lumen calculation, stenosis calculation) and reporting tools (lesion location, lesion characteristics and kev images), the software package is designed to support the physician in confirming the presence or absence of physicianidentified lesions in blood vessels and evaluation, documentation and follow-up of any such lesion. These visualization/processing/evaluation tools allow for characterization (geometry (length, lumen diameter, cross section area, stenosis grade) and appearance (HU values)) of vascular lesions and lesion size over time, helping the physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue.
#### VI. Comparison of Technological Characteristics with the Predicate Device
At a very high-level overview, the subject and predicate devices are based on the following same/modified technological characteristics.
- Software operating platform SOMARIS8/8 .
- Support of basic reading Functionality ●
- 2D/3D vessel visualization tools ●
- Automated segmentation and removal of obstructive bones and/or vessels ●
- . Support of automated pre-processing and layouts
- Measurement tools that support quantitative assessment and documentation ●
The following technological differences exist between the subject devices and the predicate devices:
- Software version SOMARIS/8 VB30 ●
- Support of Rapid Results Technology feature ●
- lmproved bone removal functionality ●
- Improved vessel tracing functionality ●
{7}------------------------------------------------
# SIEMENS
A tabular summary of the subject and predicate device technological differences is provided as Table 1 and Table 2 below:
| | | | | | Table 1: syngo.CT Coronary Analysis Predicate and Subject Device Technological |
|--|----------------------------|--|--|--|---------------------------------------------------------------------------------|
| | Characteristics Comparison | | | | |
| Property | Subject Device<br>syngo.CT Coronary<br>Analysis | Predicate Device<br>syngo.CT Coronary<br>Analysis<br>(K100637) | Comparison Results |
|-----------------------------------------|-------------------------------------------------|----------------------------------------------------------------|--------------------|
| Software Operating<br>Platform | SOMARIS/8 VB30 | SOMARIS/8 VA10 | Modified |
| | Basic Visualization and<br>Navigation Tools | Basic Visualization and<br>Navigation Tools | Same |
| | Automatic Organ<br>Segmentation | Automatic Organ<br>Segmentation | Same |
| | Automatic Vessel<br>Tracing | Automatic Vessel<br>Tracing | Modified. |
| | 3D Vessel Visualization<br>Tools | 3D Vessel Visualization<br>Tools | Same |
| Visualization and<br>Segmentation Tools | 2D Vessel Visualization<br>Tools | 2D Vessel Visualization<br>Tools | Same |
| | Vessel Navigation<br>Tools | Vessel Navigation<br>Tools | Same |
| | Vessel Definition Tools | Vessel Definition Tools | Same |
| | Vessel Evaluation<br>Tools | Vessel Evaluation<br>Tools | Same |
| | Result Image Creation | Result Image Creation | Same |
| | Integrated Reporting | Integrated Reporting | Same |
| | Workflow Automation<br>Support | Workflow Automation<br>Support | Modified |
| | User Interface | User Interface | Same |
| Archiving and<br>Reporting | Archiving/Storing | Archiving/Storing | Same |
| | Communication | Communication | Same |
Table 2: syngo.CT Vascular Analysis Predicate and Subject Device Technological Characteristics Comparison
| Property | Subject Device<br>syngo.CT Coronary<br>Analysis | Predicate Device<br>syngo.CT Vascular<br>Analysis<br>(K112020) | Comparison Results |
|-----------------------------------------|-------------------------------------------------|----------------------------------------------------------------|--------------------|
| Software Operating<br>Platform | SOMARIS/8 VB30 | SOMARIS/8 VA10 | Modified |
| Visualization and<br>Segmentation Tools | Basic Visualization and<br>Navigation Tools | Basic Visualization and<br>Navigation Tools | Same |
| | Automatic Organ<br>Segmentation | Automatic Organ<br>Segmentation | Modified |
| | Automatic Vessel<br>Tracing | Automatic Vessel<br>Tracing | Modified. |
| | 3D Vessel Visualization<br>Tools | 3D Vessel Visualization<br>Tools | Same |
| | 2D Vessel Visualization<br>Tools | 2D Vessel Visualization<br>Tools | Same |
{8}------------------------------------------------
| Property | Subject Device<br>syngo.CT Coronary<br>Analysis | Predicate Device<br>syngo.CT Vascular<br>Analysis<br>(K112020) | Comparison Results |
|-----------------------------------------|-------------------------------------------------|----------------------------------------------------------------|--------------------|
| Visualization and<br>Segmentation Tools | Vessel Navigation<br>Tools | Vessel Navigation<br>Tools | Same |
| | Vessel Definition Tools | Vessel Definition Tools | Same |
| | Vessel Evaluation<br>Tools | Vessel Evaluation<br>Tools | Same |
| | Result Image Creation | Result Image Creation | Same |
| | Integrated Reporting | Integrated Reporting | Same |
| | Workflow Automation<br>Support | Workflow Automation<br>Support | Modified |
| Archiving and<br>Reporting | User Interface | User Interface | Same |
| | Archiving/Storing | Archiving/Storing | Same |
| | Communication | Communication | Same |
The subject device modifications do not alter the fundamental scientific technology of the 510(k) cleared predicate devices syngo.CT Coronary Analysis and syngo.CT Vascular Analysis. The software version has been modified to support improved clinical workflows.
# VII. Performance Data
#### Non-Clinical Testing Summary
Non-clinical tests (integration and functional) were conducted for syngo.CT Coronary Analysis and syngo.CT Vascular Analysis during product development. Performance tests were conducted to test the functionality of syngo.CT Coronary Analysis and syngo.CT Vascular Analysis. The modifications described in this Premarket Notification were supported with verification/validation testing. These tests have been performed to test the ability of the included features of the subject device. The results of these tests demonstrate that the subject device performs as intended. The result of all conducted testing was found acceptable to support the claim of substantial equivalence. Siemens claims conformance to the following performance standards:
| Product<br>Area | Title of Standard | Publicatio<br>n Date | Standards<br>Development<br>Organization |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|------------------------------------------|
| Radiology | Digital Imaging and Communications in<br>Medicine (DICOM) Set; PS 3.1 – 3.20 | 06/27/2016 | NEMA |
| Software | Medical Device Software -Software Life<br>Cycle Processes; 62304:2006 (1st<br>Edition) | 08/20/2012 | AAMI, ANSI, IEC |
| Software/<br>Informatics | Medical devices - Application of risk<br>management to medical devices; 14971<br>Second Edition 2007-03-01 | 08/20/2012 | ISO |
| General I<br>(QS/RM) | Medical devices - Part 1: Application of<br>usability engineering to medical devices<br>IEC 62366-1:2015 | 06/27/2016 | IEC |
| General II<br>(ES/EMC) | AAMI / ANSI ES60601-1:2005/(R)2012<br>and A1:2012, Medical electrical<br>equipment - part 1: general requirements<br>for basic safety and essential<br>performance (IEC 60601-1:2005, mod) | 07/09/2014 | AAMI, ANSI |
{9}------------------------------------------------
# SIEMENS
## Non-Clinical Performance (Bench) Testing
Additional non-clinical quantitative performance testing was conducted to demonstrate the functionality and applicability of the subject devices modified features. To demonstrate the coronary tracing quality improvement for subject device syngo.CT Coronary Analysis, quantitative performance testing was conducted to test the following functions:
- Centeredness of the Centerlines ●
- Tracing of branches ●
- Coverage of Centerlines .
As part of this testing, a sampling of centerlines from a predetermined number of datasets were evaluated. To test for accuracy of centeredness, the derived centerlines were assessed in comparison to the predicate device to demonstrate an improvement in the ability of the subject device application to more accurately follow the vessels in the lumen. Testing for the tracing of branches and coverage of centerlines was conducted on branches that were considered both clinically relevant and clinically irrelevant to assess sensitivity and specificity of branch tracing and coverage. The results of this test demonstrated an increased sensitivity and specificity and improved vessel coverage for the subject device tracing feature in comparison to the predicate device.
To demonstrate the functionality of the vessel tracing quality improvement for subject device syngo.CT Vascular Analysis, the detection rate of a predetermined set of vessels was evaluated to determine whether a vessel was successfully traced if a vessel was expected based on activation of the algorithm. The result of this test demonstrated that the subject device had a higher level of traced vessels in comparison to the predicate device, which supports an improvement of vessel tracing quality in comparison to the predicate device.
#### Software Verification and Validation
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. The Risk Analysis was completed and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claim of substantial equivalence.
Siemens Healthcare conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Cybersecurity information in accordance with quidance document "Content of Premarket Submissions for Management of Cybersecurity Medical Devices issues on October 2, 2014" is included within this submission.
#### Summary
This subject devices provide tools designed for the evaluation of CT images. The fundamental software technology which is provided within the scope of the subject devices is already cleared and remains unchanged in comparison to the predicate devices. The Indications for Use for the subject devices have been adapted to provide a more specific description of the subject device functionality, but does not represent a new intended use. The modifications described in this Premarket Notification were supported with non-clinical performance testing as well as verification and validation testing. The Risk analysis was completed and risk control implemented to mitigate identified hazards.
#### General Safety and Effectiveness Concerns
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. syngo.CT Coronary Analysis and syngo.CT Vascular Analysis is designed to fulfill the requirements of the applicable safety and performance standards as listed above.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is written in all capital letters. The color of the text is a light blue or turquoise.
# VIII. Conclusions
The subject devices have the same intended use as the predicate devices. The technological characteristics such as image visualization, operating platform, and image manipulation remain unchanged from the predicate device. The predicate devices were cleared based on non-clinical testing including verification and validation testing. The results of these tests demonstrate that the predicate devices are adequate for the intended use. The subject devices are also tested using the same methods and workflows as used for the predicate devices. The comparison of technological characteristics, non-clinical performance data, and software validation included in this submission demonstrates that the subject device is as safe and effective when compared to the predicate devices that are currently marketed for the same intended use. Since both devices were tested using the same methods, Siemens believes that the data generated from the subject devices syngo.CT Coronary Analysis and syngo.CT Vascular Analysis testing supports a finding of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.