← Product Code [FRC](/productcode/FRC) · K173634

# CELERITY 20 Steam Biological Indicator (K173634)

_STERIS Corporation · FRC · Feb 12, 2018 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173634

## Device Facts

- **Applicant:** STERIS Corporation
- **Product Code:** [FRC](/productcode/FRC.md)
- **Decision Date:** Feb 12, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Celerity 20 Steam Biological Indicator is used for routine monitoring, qualification testing, load monitoring and product testing of the following steam sterilization cycles: - Dynamic Air Removal 270°F (132°C) 4 minutes - Dynamic Air Removal 275°F (135°C) 3 minutes - Gravity 250°F (121°C) 30 minutes - Gravity 270°F (132°C) 15 minutes. When used in conjunction with the Celerity™ Steam Incubator, the Incubator provides a fluorescent result within 20 minutes.

## Device Story

Celerity 20 Steam Biological Indicator (BI) monitors steam sterilization efficacy. Device contains Geobacillus stearothermophilus spores and nutrient media with a fluorogenic substrate. Following sterilization, BI is placed in Celerity 20 Steam Incubator. If spores survive, a reporter enzyme reacts with the substrate to produce fluorescence, detected by the incubator within 20 minutes. Used by healthcare personnel in clinical settings to confirm sterilization cycle success. Output provides binary pass/fail result for sterilization monitoring. Benefits include rapid verification of sterilization process, ensuring instrument safety for patient use.

## Clinical Evidence

Bench testing only. Performance validated via RIT testing (>97% alignment with 7-day conventional incubation), spore population counts (2.3-2.5 x 10^6), resistance (D121 ≥ 2.53 min, D132 ≥ 46 s, D135 ≥ 37 s), survival/kill time testing, and simulated use testing in sterilizers.

## Technological Characteristics

Biological indicator containing Geobacillus stearothermophilus (ATCC 7953) and fluorogenic nutrient media. Plastic vial packaging. Sensing principle: enzymatic fluorescence detection. Connectivity: requires Celerity 20 Steam Incubator. Complies with ISO 11138-1 and ISO 11138-3 standards.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- 3M Attest Super Rapid Readout Biological Indicator ([K121484](/device/K121484.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 12, 2018

STERIS Corporation Anthony Piotrkowski Senior Manager, Regulatory Affairs 5976 Heisley Rd Mentor, Ohio 44060

Re: K173634

Trade/Device Name: CELERITY 20 Steam Biological Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: November 22, 2017 Received: November 24, 2017

Dear Anthony Piotrkowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known)

K173634

Device Name

CELERITY 20 Steam Biological Indicator

### Indications for Use (Describe)

The Celerity 20 Steam Biological Indicator is used for routine monitoring, qualification testing, load monitoring and product testing of the following steam sterilization cycles:

- · Dynamic Air Removal 270°F (132°C) 4 minutes
- · Dynamic Air Removal 275°F (135°C) 3 minutes
- · Gravity 250°F (121°C) 30 minutes
- · Gravity 270°F (132°C) 15 minutes.

When used in conjunction with the Celerity™ Steam Incubator, the Incubator provides a fluorescent result within 20 minutes.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

[X] Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for STERIS. The word "STERIS" is written in a bold, sans-serif font at the top of the image. Below the word, there is a graphic of several horizontal, wavy lines in blue. The STERIS logo is simple and modern.

## 510(k) Summary For K173634 Celerity 20 Steam Biological Indicator

### Sponsor Facility

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198

### Manufacturing Facility

STERIS Corporation 9325 Pinecone Drive Mentor, OH 44060 Phone: (440) 392-7800 Fax No: (440) 392-7896

Contact: Tony Piotrkowski Senior Manager, Regulatory Affairs

Telephone: (440) 392-7437 Fax No: (440) 357-9198 e-mail: tpiotrko@steris.com

Submission Date: January 17, 2018

Premarket Notification Number: K173634

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{4}------------------------------------------------

### STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION K173634 Celerity 20 Steam Biological Indicator

#### Device Name 1.

| Trade Name:            | Celerity 20 Steam Biological Indicator                                |
|------------------------|-----------------------------------------------------------------------|
| Common/usual Name:     | Biological Indicator (BI, SCBI)                                       |
| Device Classification: | Class II                                                              |
| Classification Name:   | Indicator, Biological Sterilization Process<br>(21 CFR 880.2800, FRC) |

#### Predicate Device 2.

3M Attest Super Rapid Readout Biological Indicator, K121484

#### 3. Description of Device

The product is intended to monitor the critical parameters of steam sterilization cycles described in the indications for use by producing an optical change (signal) that is detected by the STERIS proprietary reader. Celerity 20 Steam Incubator in 20 minutes to confirm the viability of the biological indicator at the end of a steam sterilization process. The product consists of a biological organism known to be resistant to steam (Geobacillus stearothermophilus) and a defined nutrient media. A reporter enzyme, which is produced by the native organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.

#### 4. Intended Use/ Indications for Use

The Celerity 20 Steam Biological Indicator is used for routine monitoring, qualification testing, load monitoring and product testing of the following steam sterilization cycles:

- · Dynamic Air Removal 270°F (132°C) 4 minutes
- · Dynamic Air Removal 275°F (135°C) 3 minutes
- · Gravity 250°F (121°C) 30 minutes
- · Gravity 270°F (132°C) 15 minutes.

When used in conjunction with the Celerity™ Steam Incubator, the Incubator provides a fluorescent result within 20 minutes.

#### ട. Summary of Technical Characteristics

A comparison of technical characteristics are summarized in Table 5-1.

{5}------------------------------------------------

| Feature                    | Celerity SCBI<br>(proposed)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Attest 1492V<br>(K121484)                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use            | The Celerity 20 Steam<br>Biological Indicator (BI) is<br>for routine monitoring,<br>qualification testing and<br>product testing of the<br>following steam<br>sterilization processes:<br>270F, 4-minute dynamic<br>air removal; 275F, 3-<br>minute dynamic air<br>removal; 250 F, 30-minute<br>gravity; 270, 15-minutes<br>gravity.<br>When used in conjunction<br>with the Celerity 20 Steam<br>Incubator, the Celerity 20<br>Steam Biological Indicator<br>provides a fluorescent result<br>within 20 minutes. | Use 3M Attest Super Rapid<br>Readout Biological<br>Indicator 1492V in<br>conjunction with the 3M<br>Attest Auto-reader 490 to<br>qualify or monitor dynamic<br>air removal (prevacuum)<br>steam sterilization cycles<br>of 4 minutes at 270 F (132<br>C) and 3 minutes at 275 F<br>(135 C).<br>The 3M Attest Super Rapid<br>Readout Biological<br>Indicator 1492V provides a<br>final fluorescent result in 1<br>hour. An optional visual<br>pH color change result is<br>observed in 48 hours. | Both are intended for<br>monitoring steam<br>sterilization cycles.<br><br>The Celerity BI is claiming<br>additional cycles compared<br>to the predicate. Simulated<br>use testing in sterilizers with<br>a load under pass and fail<br>conditions demonstrate<br>performance of the Celerity<br>BI in these additional cycles.<br><br>The Celerity BI has a shorter<br>claimed readout time.<br>Reduced Incubation Time<br>(RIT) testing per the FDA<br>guidance demonstrates the<br>read time is appropriate. |
| Indicator<br>organism      | > 90% similarity to ATCC<br>7953 Geobacillus<br>stearothermophilus                                                                                                                                                                                                                                                                                                                                                                                                                                                | > 90% similarity to ATCC<br>7953 Geobacillus<br>stearothermophilus                                                                                                                                                                                                                                                                                                                                                                                                                              | Same criteria                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Mechanism<br>of action     | An enzyme, which is<br>produced by the organism,<br>reacts with a fluorogenic<br>substrate within the defined<br>nutrient media to produce a<br>fluorescent moiety                                                                                                                                                                                                                                                                                                                                                | An enzyme, which is<br>produced by the organism,<br>reacts with a fluorogenic<br>substrate within the defined<br>nutrient media to produce a<br>fluorescent moiety                                                                                                                                                                                                                                                                                                                              | Same mechanism                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Accessories                | Automated incubator /<br>reader                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Automated incubator /<br>reader                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | RIT testing performed with<br>the proposed<br>incubator/reader.                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Viable spore<br>population | 1.0 - 4.0 x 106 spore/SCBI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | > 1.0 x 106 spore/SCBI                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Both proposed and predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Resistance                 | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | D132 ≥ 10 s<br>D135 ≥ 8 s                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | meet criteria of ISO 11138-3<br>and FDA guidance                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Culture<br>Conditions      | 55- 59 °C, media included<br>in SCBI, 20-minute<br>incubation time.                                                                                                                                                                                                                                                                                                                                                                                                                                               | 55- 59 °C, media included<br>in SCBI, 60 minute<br>incubation time.                                                                                                                                                                                                                                                                                                                                                                                                                             | RIT Testing and ISO 11138<br>media testing verifies<br>performance                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Primary<br>Packaging       | Direct inoculum on plastic<br>vial, cap with recovery<br>media.                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Inoculated paper in plastic<br>vial with cap and glass<br>ampoule with recovery<br>media in capped vial.                                                                                                                                                                                                                                                                                                                                                                                        | Similar configuration.<br>Component testing per ISO<br>11138-1 Annex B<br>demonstrates packaging is<br>compatible with indicator<br>and sterilization process.                                                                                                                                                                                                                                                                                                                                                 |

### Table 5-1. Summary of SCBI Physical Description and Technological Properties

{6}------------------------------------------------

### STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION K173634 Celerity 20 Steam Biological Indicator

#### Summary of Nonclinical Tests 6.

Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 5-2 below.

| Test                                                     | Acceptance Criteria                                                                                                       | Conclusion                                                |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Reduced<br>Incubation Time<br>(RIT) Testing              | Meets FDA's requirement of > 97%<br>alignment of the 40-minute results with the<br>conventional incubation time of 7 days | PASS                                                      |
| Viable spore<br>population                               | 1.0 - 4.0 x 106 spore/SCBI                                                                                                | 2.3 - 2.5 x 106<br>spore/SCBI                             |
| Resistance                                               | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s                                                                               | D121 ≥ 2.53 min<br>D132 ≥ 46 s<br>D135 ≥ 37 s             |
| Survival Time                                            | Meets the longer of FDA and ISO 11138-3<br>requirements                                                                   | 121 C ≥ 11.11 min<br>132 C ≥ 3.42 min<br>135 C ≥ 2.77 min |
| Kill Time                                                | Meets the shorter of FDA and ISO 11138-3<br>requirements                                                                  | 121 C ≤ 31.81 min<br>132 C ≤ 8.62 min<br>135 C ≤ 7.11 min |
| Carrier growth<br>inhibition / media<br>growth promotion | Positive growth of less than 100 spores after<br>primary packaging and media are subject to<br>worst case steam exposure  | PASS                                                      |
| Hold Time                                                | Performance not affected if incubated within<br>72 hours of exposure to steam sterilization                               | PASS                                                      |
| Simulated Use                                            | Demonstrate growth when exposed to<br>abbreviated cycle and all kill in a full cycle                                      | Abbreviated cycle - growth<br>Full cycle - no growth      |

Table 5-2. Summary of Non-clinical Testing

#### 7. Conclusion

The Celerity 20 Steam Biological Indicator has met the established performance criteria. The conclusions drawn from the nonclinical tests performed demonstrate the subject device is as safe, as effective, and performs as well or better than the legally marketed predicate device, K121424 Class II (21 CFR 880.2800, Product code FRC).

---

**Source:** [https://fda-staging.innolitics.com/device/K173634](https://fda-staging.innolitics.com/device/K173634)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).