Celerity 20 Steam Process Challenge Device for Dynamic Air Removal Cycles
K173626 · STERIS Corporation · FRC · Feb 21, 2018 · General Hospital
Device Facts
Record ID
K173626
Device Name
Celerity 20 Steam Process Challenge Device for Dynamic Air Removal Cycles
Applicant
STERIS Corporation
Product Code
FRC · General Hospital
Decision Date
Feb 21, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The Celerity™ 20 Steam Challenge Pack is used for qualification, routine microbial monitoring, and load monitoring of steam sterilizers. The validated steam sterilization cycles include: - 270°F (132°C) 4-minute dynamic air removal (Prevac, SFPP) . - 275°F (135°C) 3-minute dynamic air removal (Prevac, SFPP) .
Device Story
The Celerity 20 Steam Process Challenge Device (PCD) is a monitoring tool for steam sterilizers. It consists of a Celerity 20 Steam Biological Indicator (BI) and a VERIFY Steam Integrating Indicator sealed within a plastic tray featuring a molded channel; this channel creates a tortuous pathway that limits steam penetration and air removal, simulating the challenge of a standard 16-towel test pack. Used in clinical settings to verify sterilization efficacy, the device provides a 20-minute fluorescent readout via the BI and a visual chemical indicator result. By providing a greater challenge to the sterilization process than the indicators alone, it ensures that if the PCD passes, the load has been adequately sterilized. This aids healthcare providers in confirming the safety of sterilized instruments.
Clinical Evidence
No clinical data. Bench testing only. Simulated use testing demonstrated that the BI and chemical integrator performance within the PCD is equivalent to those in the AAMI reference pack. Testing confirmed the PCD provides an equivalent or greater challenge to the sterilization process than the AAMI standardized test pack and that the PCD provides a greater challenge than the BI or chemical integrator alone.
Technological Characteristics
The device is a plastic tray with a foil cover containing a molded channel for steam/air flow. It includes a biological indicator (20-minute fluorescent result) and a chemical integrating indicator. It is designed to be equivalent to the ANSI/AAMI ST79 16-towel test pack. It is a standalone, non-electronic device used for monitoring sterilization cycles.
Indications for Use
Indicated for qualification, routine microbial monitoring, and load monitoring of steam sterilizers using 270°F (132°C) 4-minute or 275°F (135°C) 3-minute dynamic air removal (Prevac, SFPP) cycles.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
3M Attest Super Rapid 5 Steam-Plus Challenge Pack (K121593)
Reference Devices
Verify Assert Steam Process Challenge Device (K162945)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health and Human Services logo. The logo on the right is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
February 21, 2018
STERIS Corporation Anthony Piotrkowski Senior Manager, Regulatory Affairs 5960 Heisley Rd Mentor, Ohio 44060
Re: K173626
Trade/Device Name: Celerity 20 Steam Process Challenge Device for Dynamic Air Removal Cycles Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: November 22, 2017 Received: November 24, 2017
Dear Anthony Piotrkowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K173626
### Device Name
CELERITY 20 Steam Process Challenge Device for Dynamic Air Removal Cycles
### Indications for Use (Describe)
The Celerity™ 20 Steam Challenge Pack is used for qualification, routine microbial monitoring, and load monitoring of steam sterilizers.
The validated steam sterilization cycles include:
- 270°F (132°C) 4-minute dynamic air removal (Prevac, SFPP) .
- 275°F (135°C) 3-minute dynamic air removal (Prevac, SFPP) .
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for STERIS. The logo consists of the word "STERIS" in a bold, sans-serif font, with the registered trademark symbol (®) to the right of the word. Below the word "STERIS" is a graphic of several horizontal, wavy lines in blue, resembling water waves.
# 510(k) Summary For Celerity 20 Steam Process Challenge Device for Dynamic Air Removal Cycles
### Sponsor Facility
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
## Manufacturing Facility
STERIS Corporation 9325 Pinecone Drive Mentor, OH 44060 Phone: (440) 392-7800 Fax No: (440) 392-7896
Contact: Anthony Piotrkowski Senior Manager Regulatory Affairs
Telephone: (440) 392-7437 (440) 639-6258 Fax No: e-mail: tony_piotrkowski@steris.com
Submission Date: January 23, 2018
Premarket Notification Number: K173626
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{4}------------------------------------------------
## K173626 STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION Celerity 20 Steam Process Challenge Device for Dynamic Air Removal Cycles
#### Device Name 1.
| Trade Name: | Celerity 20 Steam Process Challenge Device for<br>Dynamic Air Removal Cycles |
|------------------------|------------------------------------------------------------------------------|
| Common/usual Name: | Biological Indicator Pack (PCD) |
| Device Classification: | Class II |
| Classification Name: | Sterilization Process Indicator<br>(21 CFR 880.2800, FRC) |
#### 2. Predicate Device
3M Attest Super Rapid 5 Steam-Plus Challenge Pack K121593
Reference Device: Verify Assert Steam Process Challenge Device, K162945
#### 3. Description of Device
The Celerity 20 Steam Process Challenge Device for Dynamic Air Removal Cycles (PCD), contains a Celerity 20 Steam Biological Indicator (BI) and a VERIFY Steam Integrating Indicator - Short, sealed within a plastic tray with foil cover. The tray has a molded channel which serves as the tortuous pathway for air removal/steam penetration during steam sterilization. The PCD is designed and validated to be equivalent to the standard 16-towel test pack described in ANSI/AAMI ST79.
#### 4. Intended Use/ Indications for Use
The Celerity™ 20 Steam Challenge Pack is used for qualification, routine microbial monitoring, and load monitoring of steam sterilizers. The validated steam sterilization cycles include:
- 270°F (132°C) 4-minute dynamic air removal (Prevac, SFPP)
- 275°F (135°C) 3-minute dynamic air removal (Prevac, SFPP)
#### 5. Summary of Technical Characteristics
A comparison of technical characteristics are summarized in Table 5-1.
{5}------------------------------------------------
| Feature | Celerity PCD<br>(K173626) Proposed | Attest 41482V<br>(K121593) Predicate | Comparison |
|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use | The Celerity 20<br>STEAM Process<br>Challenge Device is<br>used for qualification,<br>routine microbial<br>monitoring, and load<br>monitoring of steam<br>sterilizers.<br>The validated steam<br>sterilization cycles<br>include:<br>• 270F (132°C), 4-<br>minute dynamic<br>air removal;<br>(Prevac, SFPP)<br>• 275F(135°C), 3-<br>minnute dynamic<br>air removal;<br>(Prevac, SFPP) | Use 3M Attest Super<br>Rapid 5 Steam-Plus<br>Challenge Pack in<br>conjunction with the<br>3M Attest Auto-<br>reader 490 to qualify<br>or monitor dynamic<br>air removal<br>(prevacuum) steam<br>sterilization cycles of<br>4 minutes at 270 F<br>(132 C) and 3 minutes<br>at 275 F (135 C).<br>The 3M Attest Super<br>Rapid Readout<br>Biological Indicator<br>contained within the<br>challenge pack<br>provides a final<br>fluorescent result in 1<br>hour. An optional<br>visual pH color<br>change result is<br>observed in 48 hours. | Both are intended for<br>monitoring 270F<br>(132C), 4-minute<br>dynamic air removal<br>275F(135C), 3-minnute<br>dynamic air removal,<br>steam sterilization<br>cycles. |
| General<br>Design | Indicators are sealed<br>in plastic tray with<br>channel to limit steam<br>penetration/air<br>removal. | Layers of medical<br>index cards, some of<br>which are die-cut to<br>contain indicators,<br>overwrapped and<br>secured with a label. | Both the devices are<br>contained in a steam<br>barrier. Simulated use<br>testing demonstrated<br>equivalence |
| Biological<br>Indicator | Celerity 20 Steam<br>Biological Indicator<br>(20-minute<br>fluorescent result) | Attest 1492V<br>Biological Indicator<br>(1-hour fluorescent<br>result) | Both contain 510(k)-<br>cleared Biological<br>Indicators with rapid<br>fluorescent results. |
| Chemical<br>Integrator | VERIFY STEAM<br>Integrating Indicator-<br>Short | SteriGage Chemical<br>Integrator | Both contains 510(k)-<br>cleared chemical<br>integrators |
| Means to<br>distinguish<br>processed<br>PCD from<br>unprocessed | Proposed device's<br>internal integrator is<br>visible through the<br>pack. | External process<br>indicator. Yellow to<br>brown color change<br>when exposed to<br>steam | Both have a means to<br>distinguish processed<br>PCD from unprocessed |
## Table 5-1. Summary of PCD Physical Description and Technological Properties
{6}------------------------------------------------
## K173626 STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION Celerity 20 Steam Process Challenge Device for Dynamic Air Removal Cycles
The Celerity 20 Steam Process Challenge Device for Dynamic Air Removal Cycles is identical to the reference device VERIFY Assert Steam Process Challenge Device except for the biological indicator within the device. The difference in the biological indicator allows the Celerity BI to have a 20-minute read (versus 40-minutes for the reference device). The different BI also requires a different reader.
#### 6. Summary of Nonclinical Tests
Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 5-2 below.
| Test | Acceptance Criteria | Conclusion |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Simulated Use | • Performance of the BI in the PCD is equivalent to<br>the performance of the BI in the AAMI reference<br>pack in their respective sterilization processes | PASS |
| | • Performance of the chemical integrator in the PCD<br>is equivalent to the performance of the chemical<br>integrator in AAMI reference pack in their<br>respective processes | PASS |
| | • PCD provides an equivalent or greater challenge<br>than the AAMI standardized test pack | PASS |
| BI in pack vs BI<br>outside of pack | PCD provides a greater challenge to the process than<br>the BI itself. | PASS |
| CI in pack vs CI<br>outside of pack | PCD provides a greater challenge to the process than<br>the integrator by itself. | PASS |
Table 5-2. Summary of Non-clinical Testing
### 7. Conclusion
The Celerity 20 Steam Process Challenge Device for Dynamic Air Removal Cycles has met the established performance criteria. The results of the studies performed demonstrate that the biological indicator performs as intended, and based on the nonclinical tests performed, the subject device is as effective, and performs at least as safely and effectively as the legally marketed predicate devices, Class II (21 CFR 880.2800, Product code FRC).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.