The iQMR is intended for networking, communication, processing and enhancement of MRI images in DICOM format. The device processing is not effective for lesion, mass, or abnormalities of sizes less than 1.5mm. This device is indicated for use by qualified trained medical professionals.
Device Story
iQMR is a software package for MRI image enhancement via noise reduction. It operates on an off-the-shelf PC server connected to a local area network (LAN). The device receives DICOM-formatted MRI images from workstations, processes them using proprietary algorithms, and transmits the enhanced images back to selected workstations. It is intended for use by qualified medical professionals in clinical environments. The output is used by healthcare providers to assist in viewing MRI data; the enhancement aims to improve image quality for diagnostic review. The device does not detect lesions or abnormalities smaller than 1.5mm.
Clinical Evidence
No formal clinical trials were conducted. Performance was validated using non-clinical bench testing with MRI standard phantoms and processing of clinical MRI images to ensure conformance to user needs and intended use under actual conditions.
Technological Characteristics
Software package running on off-the-shelf PC server with graphical boards. Operating system: Linux Ubuntu. Inputs/outputs: DICOM format MRI images. Connectivity: LAN. Risk management per ISO 14971.
Indications for Use
Indicated for qualified trained medical professionals requiring networking, communication, processing, and enhancement of MRI images in DICOM format. Not effective for lesions, masses, or abnormalities smaller than 1.5mm.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Medic Vision - Imaging Solutions Ltd % Mr. Dan Laor Q&R Advisor 6 Sireni St. Haifa, 3297206 ISRAEL
Re: K173605
Trade/Device Name: iOMR Regulation Number: 21 CFR 892,2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 19, 2018 Received: February 27, 2018
Dear Mr. Laor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
March 8, 2018
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert Sauer
for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173605
Device Name iQMR
Indications for Use (Describe)
The iQMR is intended for networking, communication, processing and enhancement of MRI images in DICOM format. The device processing is not effective for lesion, mass, or abnormalities of sizes less than 1.5mm. This device is indicated for use by qualified trained medical professionals.
Type of Use (Select one or both, as applicable)
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
# Submitter details
Medic Vision Imaging Solutions Ltd. 3 Yozma St, Tirat Carmel 39032, Israel
Telephone +972.73.7262226
Submission Contact: Dan Laor Israel.
#### Details of the submitted Device
| Proprietary Name: | iQMR |
|-------------------|-----------|
| Regulation Number | 892.2050 |
| Product Code: | LLZ |
| Committee/Panel: | Radiology |
| Device Class: | 2 |
6 Sireni St., Haifa, 3297206,
# Type of 510(k) Submission:
Traditional
#### Identification of the Legally Marketed Predicate Device K993802 SharpView
### Device Description
iQMR is a software package that is aimed to process MRI images. The iQMR processing enhances MRI images by reduction of the image noise. The software runs on a PC server, which is connected to the Local Area Network (LAN). The device receives MRI images from different work stations over the network in DICOM format, processes the images and transmits them in DICOM format, to selected work stations.
#### Intended use and indications for Use
The iQMR is intended for networking, communication, processing and enhancement of MRI images in DICOM format. The device processing is not effective for lesion, mass, or abnormalities of sizes less than 1.5mm. This device is indicated for use by qualified trained medical professionals.
#### Technological Characteristics
The iQMR software package is installed in an off-the-shelf PC server, which include off-the -shelf graphical boards. The Operating System is Linux Ubuntu. The iQMR software package consists of three software modules: The User Interface, Algorithm, and Interface module. Data inputs and output are MRI images in DICOM format.
#### Performance Tests
#### Non-Clincal tests
The device software has been verified by testing the software following predefined software test plan.
The iQMR system has been verified by testing that it meets its specified performance, using MRI standard phantoms.
The iQMR performance has been validated by testing it performing its intended and indications for use, in end-user environment.
Vol_003 002_510K Summary
{4}------------------------------------------------
The above testing methods adhered to state-of-art standards and procedures. The tests results demonstrate that the device output meet the design input and the intended use.
#### Clinical tests:
Clinical tests were not conducted. However, MRI clinical images were processed in order to ensure that the iQMR conforms to defined user needs and intended uses under actual conditions.
## Risk Management
The device risks were managed and controlled following the requirements of ISO 14971standard. The device hazards were identified, their risk levels were evaluated and mitigation measures were taken to reduce the risk levels. In Medic Vision opinion the benefits of providing the iQMR features, overweight the device residual risks.
### Substantial Equivalence
| Parameter | Predicate Device<br>SharpView (K993802) | Subject Device<br>iQMR |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Intended use | Intended for use by qualified<br>trained medical professionals for<br>enhancement of MRI images that<br>are transferred in the network in<br>DICOM format. | The same |
| 21CFR section | 892.2050 | The same |
| Product Code | LLZ | The same |
| <b>Technological<br/>Characteristics</b> | | |
| Device nature | SW package | The same |
| Operating System | Windows | Linux |
| Data input | MRI images in DICOM format | The same |
| Data output | MRI images in DICOM format | The same |
| Processing<br>Algorithms | GOP Enhancement Software | Medic Vision's<br>Algorithms |
| User Interface | Included | The same |
Comparison with the predicate device
# Substantial Equivalence conclusion
The iQMR has the same intended use as the legally marketed SharpView predicate device. The two devices have similar technological characteristics. Results of tests, which were adhered to state-of-art standards and procedures, demonstrate that differences in the technological characteristics do not raise new questions of safety or effectiveness. Based on this discussion, it is Medic Vision's opinion that iQMR is substantially equivalent in terms of safety and effectiveness to the SharpView (K993802) predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.