K173594 · Institut Straumann AG · NPM · Oct 19, 2018 · Dental
Device Facts
Record ID
K173594
Device Name
Straumann cerabone
Applicant
Institut Straumann AG
Product Code
NPM · Dental
Decision Date
Oct 19, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Straumann® cerabone® is used in the following intra-oral surgical procedures: - Alveolar ridge augmentation/ reconstruction - Filling of defects after root resection and apicoectomy - Filling alveoli after tooth extraction - Sinus elevation
Device Story
Straumann® cerabone® is a xenogeneic hydroxyapatite bone grafting material derived from bovine cancellous bone. The manufacturing process involves high-temperature sintering (up to 1250°C) to eliminate organic components, reducing rejection risks. The resulting material is a porous scaffold (>95% pentacalcium-hydroxy-tri-phosphate) with interconnected macro- and micro-pores (100–1,500 µm) that mimic human bone structure. Used by dental clinicians in intra-oral surgical procedures, the granules are implanted into a vital bone bed. The material acts as an osteoconductive scaffold, facilitating cell adhesion and stabilizing new bone matrix deposition. Over 6–8 months, the graft integrates into newly formed bone. It is provided in various particle sizes and volumes, packaged in a triple sterile barrier system, and terminally sterilized via gamma irradiation.
Clinical Evidence
Evidence includes bench testing, animal studies, and clinical literature review. Bench testing confirmed physicochemical properties per ASTM F1185-03 and F1581-99, and viral inactivation per ISO 22442-3. Biocompatibility was validated per ISO 10993 series (cytotoxicity, sensitization, irritation, subchronic toxicity, genotoxicity, implantation). An animal kinetic study in minipigs demonstrated osteoconductive capacity equivalent to the predicate. Clinical literature review supports the safety and efficacy for the proposed indications.
Technological Characteristics
Material: Xenogenic hydroxyapatite (≥95% Ca5(PO4)3(OH)) derived from bovine cancellous bone. Structure: Interconnected macro- (65–80 vol%) and micro-porous. Granule sizes: 0.5–1.0 mm and 1.0–2.0 mm. Sterilization: Gamma irradiation. Packaging: Triple sterile barrier (glass vial and two blisters). Standards: ASTM F1185-03, ASTM F1581-99, ISO 10993, ISO 11137-1, ISO 22442-3.
Indications for Use
Indicated for patients requiring intra-oral surgical procedures including alveolar ridge augmentation/reconstruction, filling of defects after root resection and apicoectomy, filling of alveoli after tooth extraction, and sinus elevation.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 19, 2018
Institut Straumann AG % Jennifer Jackson Director, Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01801
Re: K173594
Trade/Device Name: Straumann® cerabone® Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPM Dated: September 17, 2018 Received: September 20, 2018
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
for
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173594
Device Name Straumann® cerabone®
Indications for Use (Describe)
Straumann® cerabone® is used in the following intra-oral surgical procedures:
- · Alveolar ridge augmentation/ reconstruction
- · Filling of defects after root resection and apicoectomy
- · Filling alveoli after tooth extraction
- Sinus elevation
| Type of Use (Select one or both, as applicable) | <table><tr><td><div> <input checked="checked" type="checkbox"/> Research Use (Part 21 CFR 201.3 Subpart D) </div></td></tr><tr><td><div> <input type="checkbox"/> Testing Conducted (21 CFR 201.3 Subpart G) </div></td></tr></table> | <div> <input checked="checked" type="checkbox"/> Research Use (Part 21 CFR 201.3 Subpart D) </div> | <div> <input type="checkbox"/> Testing Conducted (21 CFR 201.3 Subpart G) </div> |
|---------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| <div> <input checked="checked" type="checkbox"/> Research Use (Part 21 CFR 201.3 Subpart D) </div> | | | |
| <div> <input type="checkbox"/> Testing Conducted (21 CFR 201.3 Subpart G) </div> | | | |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
| Submitter: | Straumann USA, LLC<br>(on behalf of Institut Straumann AG) |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 60 Minuteman Road<br>Andover, MA 01810 |
| | Registration No.: 1222315<br>Owner/Operator No.: 9005052 |
| Contact Person: | Jennifer M. Jackson, MS<br>Director, Regulatory Affairs<br>+1 (978) 747-2509 |
| Prepared By &<br>Secondary Contact: | Christelle Gerspach-Gasser<br>Regulatory Affairs and Compliance Manager<br>Institut Straumann AG<br>+41 61 965 1666 |
| Date Prepared: | October 19, 2018 |
| Product Code(s): | NPM 21 CFR 872.3930 |
| Device Class: | II 21 CFR 872.3930 |
| Classification Panel: | Dental Devices |
| Classification Name: | Bone Grafting Material |
| Common Name: | Bone Grafting Material |
| Proprietary Name: | Straumann® cerabone® |
| Predicate Device(s): | K122894 Geistlich Bio-Oss |
| Device Description: | Straumann® cerabone® is a xenogeneic hydroxyapatite<br>ceramic for implantation into a vital bone bed during<br>intra-oral surgical procedures. It is manufactured from<br>bovine cancellous bone in a high temperature sintering<br>process lasting several hours at temperatures of up to<br>1250°C. This high temperature process results in all<br>organic components being eliminated so that the risk of<br>occurrence of allergic side effects or the occurrence of<br>rejection reactions is reduced.<br>Straumann® cerabone® is made of over 95% of<br>Pentacalcium-hydroxy-[tri]-phosphate Ca5(PO4)3(OH) |
on
{4}------------------------------------------------
Due to the animal-derived starting material, Straumann® cerabone® possesses an interconnected macro- and micro-porous structure that strongly resembles human bone structure. The macroporosity lies within a range of 65–80 vol % and the pore size lies within a range of approximately 100 - 1,500 um. After implantation, the particles serve as an osteoconductive guide rail for new bone formation. The grafting material has high stability over several years with only minor signs of graft resorption.
Straumann® cerabone® acts as a scaffold for stabilization and the deposition of new bone matrix. Due to porosity and rough surface structure, Straumann® cerabone® facilitates cell adhesion. Subsequently, bone formation starts by deposition of primarily not yet mineralized bone matrix. Within a time period of six to eight months following implantation, the graft is integrated into the newly formed bone matrix.
Straumann® cerabone® is biocompatible and packed in a triple sterile barrier system consisting of a glass vial and two outer blisters and subject to sterilization by qammairradiation.
Straumann® cerabone® is offered in two particle size ranges (0.5 – 1.0 mm and 1.0 – 2.0 mm) and various pack sizes (0.5, 1.0, 2.0 and 5.0 cc).
| Art. No. | Article Name |
|----------|------------------------------------------------|
| BS-1510 | Straumann® cerabone®, 0.5 - 1.0 mm, 1 x 0.5 cc |
| BS-1511 | Straumann® cerabone®, 0.5 - 1.0 mm, 1 x 1.0 cc |
| BS-1512 | Straumann® cerabone®, 0.5 - 1.0 mm, 1 x 2.0 cc |
| BS-1515 | Straumann® cerabone®, 0.5 - 1.0 mm, 1 x 5.0 cc |
| BS-1520 | Straumann® cerabone®, 1.0 - 2.0 mm, 1 x 0.5 cc |
| BS-1521 | Straumann® cerabone®, 1.0 - 2.0 mm, 1 x 1.0 cc |
| BS-1522 | Straumann® cerabone®, 1.0 - 2.0 mm, 1 x 2.0 cc |
| BS-1525 | Straumann® cerabone®, 1.0 - 2.0 mm, 1 x 5.0 cc |
{5}------------------------------------------------
Straumann® cerabone® is used in the following intra-oral Indications For Use: surgical procedures:
- Alveolar ridge augmentation/ reconstruction ●
- Filling of defects after root resection and ● apicoectomy
- Filling alveoli after tooth extraction .
- Sinus elevation .
Technological The subject and the predicate device are both sterile, Characteristics: biocompatible porous bone mineral materials for use in the same intra-oral surgical procedures.
> The subject and predicate devices are based on the following same technological elements:
- Pure cancellous bovine bone mineral (free from . organic compounds)
- . Rough surface structure, interconnecting macroand micro pores
- . Scaffold provision as mode of action
- . Same sterilization method and shelf life
The comparison between the features of the subject device and the predicate are provided in the table below. The Indications for Use for the subject device are a subset of the Primary Predicate Indications for Use.
| Feature | Subject Device<br>Straumann® cerabone® | Primary Predicate Devices<br>Geistlich Bio-Oss<br>K122894 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | Straumann® cerabone® is used in the following intra-oral surgical procedures:<br>Alveolar ridge augmentation/reconstruction Filling of defects after root resection and apicoectomy Filling alveoli after tooth extraction Sinus elevation | Geistlich Bio-Oss® is indicated for:<br>Augmentation or reconstructive treatment of the alveolar ridge Filling of intrabony periodontal defects Filling of defects after root resection, apicoectomy and cystectomy Filling of extraction sockets to enhance preservation of the alveolar ridge Elevation of the maxillary sinus floor Filling of periodontal defects in conjunction with products identified for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR) |
{6}------------------------------------------------
| Feature | Subject Device<br>Straumann® cerabone® | Primary Predicate Devices<br>Geistlich Bio-Oss<br>K122894 |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | xenogenic hydroxyapatite (calcium<br>phosphate) ceramic manufactured<br>from cancellous (spongiosa)<br>bovine bone | xenogenic hydroxyapatite (calcium<br>phosphate) ceramic manufactured<br>from cancellous (spongiosa)<br>bovine bone |
| Origin | Femoral heads of cattle | Femoral heads of cattle |
| Form | Granules | Granules |
| Composition | - Hydroxyapatite (Pentacalcium-<br>hydroxy-[tri]-phosphate<br>Ca5(PO4)3(OH)) (≥95%)<br>- ≤2.5% other calcium phosphates<br>(traces of calcium carbonate<br>(CaO3) (<0.1%))<br>- No water (0%)<br>- No organic components (0%) | - Hydroxyapatite (Pentacalcium-<br>hydroxy-[tri]-phosphate<br>Ca5(PO4)3(OH)) (93.6%),<br>- Calcium carbonate (CaO3)<br>(3,4%)<br>- Water (3%)<br>- No organic components (0%) |
| Appearance<br>(dry state) | Whitish up to beige | Whitish |
| Structure /<br>Density | Matrix possessing an<br>interconnected macro- and micro-<br>porous structure. | Matrix consisting of interconnected<br>macro- and micropores, highly<br>porous with a large inner surface<br>area. |
| Specifications<br>/ Granules<br>sizes | Granule sizes<br>0.5 – 1.0 mm<br>1.0 – 2.0 mm | Granule sizes<br>0.25 – 1.0 mm<br>1.0 – 2.0 mm |
| Packaging &<br>Sterilization | Single use, triple sterile barrier<br>system, terminally gamma<br>sterilized by gamma irradiation | Single use, double sterile barrier<br>system, terminally gamma<br>sterilized by gamma irradiation |
| Sterilization | Gamma-irradiation | Gamma-irradiation |
| Shelf life | 3 years | 3 years |
Performance Data:
The following performance data were provided in support of the substantial equivalence determination
Sterilization and Shelf Life:
- Validation of sterilization process per ISO 11137-1 ●
- . Validation of shelf life per ASTM F1980
Biocompatibility Testing:
- Biocompatibility testing per ISO 10993-1 and the . following:
- Cytotoxicity per ISO 10993-5 o
- Sensitization per ISO 10993-10 o
- Irritation per ISO 10093-10 O
- Subchronic Toxicity per ISO 10993-11 o
- O Genotoxicity per ISO 10993-3
- Implantation per ISO 10993-6 o
{7}------------------------------------------------
Performance Testing - Bench:
- Physicochemical and material characterization per ● ASTM F 1185-03 and ASTM F 1581-99
- . Viral inactivation demonstrated through validation of manufacturing steps, and the assessment of the potential pathogenic burden, quality management system in place. The Zoonosis assessment demonstrated that the validated manufacturing processes are sufficient to ensure a reduction of viral burden in accordance with the recommendation of the FDA recognized standard ISO 22442-3 and are sufficient for the demonstration of substantial equivalence to the predicate device.
Performance Testing - Animal:
- The osteoconductive capacity of Straumann® . cerabone® in intra-oral defects was assessed in a kinetic study in minipigs with comparison to the primary predicate Bio-Oss®. The performance of cerabone was equivalent to the primary predicate device Bio-Oss.
Performance Testing – Clinical:
- A review of the clinical literature was performed. . The reviewed literature supports the use of Straumann® cerabone® in the proposed Indications for Use.
- Conclusions: Following assessment of the design and applicable performance data, the subject devices have been determined to be substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.