← Product Code [DRY](/productcode/DRY) · K173591

# Quantum Diagnostic Module, Quantum Diagnostic Module - No Gas (K173591)

_Spectrum Medical , Ltd. · DRY · Mar 28, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173591

## Device Facts

- **Applicant:** Spectrum Medical , Ltd.
- **Product Code:** [DRY](/productcode/DRY.md)
- **Decision Date:** Mar 28, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Quantum Diagnostic Module is intended for the continuous monitoring of critical clinical parameters during procedures that require extracorporeal circulation. The Quantum Diagnostic Module is an accessory that only works with the Quantum Workstation. Parameters provided by the Quantum Diagnostic Module include: - Measurement of up to three blood flow channels and arterial and venous flow differential - Indication of gas bubbles - Extracorporeal gas flow measurements (O2, CO2, gas flow, and CO2 removal) - Predicted PO2 and PCO2 - Temperature - Up to three circuit pressure channels - Reservoir level indication The Quantum Diagnostic Module is to only be used by an experienced and trained clinician. The device is not intended to be used by the patient or other untrained personnel.

## Device Story

Quantum Diagnostic Module (QDM) is an accessory to the Quantum Workstation for extracorporeal circulation monitoring. QDM connects to the workstation via cable; positioned between gas blender and oxygenator. Inputs: data from embedded/attached sensors measuring blood flow (up to 3 channels), arterial/venous flow differential, gas bubbles, extracorporeal gas flow (O2, CO2, gas flow, CO2 removal), predicted PO2/PCO2, temperature, circuit pressure (up to 3 channels), and reservoir level. QDM processes sensor signals and transmits physiological measurements to the Quantum Workstation for display. Used in clinical settings (e.g., OR) by trained clinicians. Output allows real-time visualization of circuit parameters, aiding clinical decision-making during cardiopulmonary bypass. Benefits include continuous, non-invasive monitoring of critical circuit status.

## Clinical Evidence

No clinical data submitted. Substantial equivalence supported by non-clinical testing including electrical safety, EMC, electrosurgery interference, hardware testing, software verification/validation, in vitro comparison testing, and usability validation.

## Technological Characteristics

Accessory module for Quantum Workstation. Measures flow, pressure, level, temperature, and gas diagnostics via attached sensors. Connectivity: cable-linked to Quantum Workstation. Software-based data processing. Technical characteristics include sensor-based monitoring of extracorporeal circuits.

## Regulatory Identification

A cardiopulmonary bypass on-line blood gas monitor is a device used in conjunction with a blood gas sensor to measure the level of gases in the blood.

## Predicate Devices

- M4 Monitor ([K110957](/device/K110957.md))

## Reference Devices

- Quantum Workstation ([K163657](/device/K163657.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized depiction of a human figure, while the FDA part includes the acronym "FDA" in blue, followed by "U.S. FOOD & DRUG ADMINISTRATION" in a larger, bolder font, also in blue.

March 28, 2018

Spectrum Medical Ltd Mark Drain Chief Financial Officer Harrier 4, Meteor Business Park, Cheltenham Road East Gloucester, GL2 9OL Gb

Re: K173591

Trade/Device Name: Quantum Diagnostic Module, Quantum Diagnostic Module - No Gas Regulation Number: 21 CFR 870.4330 Regulation Name: Cardiopulmonary Bypass On-Line Blood Gas Monitor Regulatory Class: Class II Product Code: DRY Dated: February 23, 2018 Received: February 26, 2018

Dear Mark Drain:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Nicole G. Ibrahim -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

#### K173591

Device Name

Quantum Diagnostic Module

#### Indications for Use (Describe)

The Quantum Diagnostic Module is intended for the continuous monitoring of critical clinical parameters during procedures that require extracorporeal circulation. The Quantum Diagnostic Module is an accessory that only works with the Quantum Workstation. Parameters provided by the Quantum Diagnostic Module include:

- · Measurement of up to three blood flow channels and arterial and venous flow differential
- · Indication of gas bubbles
- · Extracorporeal gas flow measurements (02, CO2, gas flow, and CO2 removal)
- · Predicted PO2 and PCO2
- · Temperature
- · Up to three circuit pressure channels
- · Reservoir level indication

The Quantum Diagnostic Module is to only be used by an experienced and trained clinician. The device is not intended to be used by the patient or other untrained personnel.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

EF

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

SpectrumMedical

# I. SUBMITTER

Spectrum Medical Ltd Harrier 4, Meteor Business Park Cheltenham Road East Gloucester GL2 9QL United Kingdom

Phone: +44 (0) 1242 650 120 Fax: +44 (0) 8452 808 127

Contact Person: Mr. Mark Drain, Chief Financial Officer Date Summary Prepared: November 17, 2017

### II. DEVICE

| Proprietary Name:    | Quantum Diagnostic Module                                                                                                    |
|----------------------|------------------------------------------------------------------------------------------------------------------------------|
| Common Name:         | Blood Gas Monitor                                                                                                            |
| Classification Name: | Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass<br>(21 CFR 870.4330)                                                     |
| Regulatory Class:    | II                                                                                                                           |
| Product Code:        | DRY                                                                                                                          |
| Panel:               | Office of Device Evaluation (ODE) /<br>Division of Cardiovascular Devices (DCD)<br>Circulatory Support Devices Branch (CSDB) |

### III. PREDICATE DEVICE

Spectrum Medical Ltd's M4 Monitor (K110957)

This predicate has not been subject to a design-related recall.

Spectrum Medical Ltd's Quantum Workstation (K163657) was used as a reference device.

### IV. DEVICE DESCRIPTION

The Quantum Diagnostic Module is an accessory to the Quantum Workstation that provides continuous non-invasive monitoring of critical clinical parameters during procedures requiring extracorporeal circulation. The Quantum Diagnostic Module is connected to the Quantum Workstation via a cable and positioned between the gas blender and the oxygenator devices in

{4}------------------------------------------------

SpectrumMedical

the extracorporeal circuit. When paired with the Quantum Workstation, the combination of the Quantum Workstation and Quantum Diagnostic Module is known as the Quantum Diagnostic System.

The Quantum Diagnostic Module performs three functions:

- 1. Provides measurements from embedded and attached sensors to monitor gases into and out of a blood oxygenator.
- 2. Provides measurements from attached sensors for blood flow, bubble detection, pressure, level and temperature to monitor an extracorporeal blood loop.
- 3. Sends these physiological measurements to the Quantum Workstation for display to the user.

The Quantum Diagnostic Module, with its attached sensors, can measure flow, pressure, reservoir level, temperature and gas diagnostics. The Quantum Diagnostic Module only works with the Quantum Workstation.

# V. INTENDED USE / INDICATIONS FOR USE

The Quantum Diagnostic Module is intended for the continuous monitoring of critical clinical parameters during procedures that require extracorporeal circulation. The Quantum Diagnostic Module is an accessory that only works with the Quantum Workstation. Parameters provided by the Quantum Diagnostic Module include:

- . Measurement of up to three blood flow channels and arterial and venous flow differential
- Indication of gas bubbles
- . Extracorporeal gas flow measurements (O2. CO2 gas flow, and CO2 removal)
- . Predicted PO2 and PCO2
- . Temperature
- . Up to three circuit pressure channels
- . Reservoir level indication

The Quantum Diagnostic Module is to only be used by an experienced and trained clinician. The device is not intended to be used by the patient or other untrained personnel.

# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Spectrum Medical Ltd's M4 Monitor and Quantum Diagnostic Module have the same manufacturer, clinical application, clinical setting, target patient population, and equivalent sensor performance. The Quantum Diagnostic Module has the same intended use

{5}------------------------------------------------

SpectrumMedical

when connected to the Quantum Workstation and the principles of operation of its sensors are identical or equivalent to those of the M4 Monitor.

Whereas the user interface is integrated into the M4 Monitor, the Quantum Diagnostic Module uses the Quantum Workstation to display the information and to power the module. The Quantum Diagnostic Module also has the capability to measure circuit pressures and blood reservoir levels.

These differences in technological characteristics do not raise new issues of safety or effectiveness.

## VII. PERFORMANCE DATA - NON-CLINICAL TESTING

No animal testing was submitted to support the substantial equivalence of the Quantum Diagnostic Module to the M4 Monitor. The following non-clinical testing was performed to support the substantial equivalence of the Quantum Diagnostic Module to the M4 Monitor:

- Electrical safety .
- . Electromagnetic compatibility (EMC)
- Electrosurgery interference ●
- Hardware testing of printed circuit boards
- Software verification and validation ●
- In vitro evaluation testing (comparison testing)
- Usability validation ●

# VIII. PERFORMANCE DATA - CLINICAL TESTING

No clinical data were submitted to support the substantial equivalence of the Quantum Diagnostic Module to the M4 Monitor.

### IX. CONCLUSIONS

Based on the indications for use, technological characteristics, results of non-clinical testing, and comparison to predicate devices, the Quantum Diagnostic Module has been shown to be substantially equivalent to legally marketed predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K173591](https://fda-staging.innolitics.com/device/K173591)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com