K173573 · Ivoclar Vivadent, AG · EBF · Jun 7, 2018 · Dental
Device Facts
Record ID
K173573
Device Name
Tetric CAD
Applicant
Ivoclar Vivadent, AG
Product Code
EBF · Dental
Decision Date
Jun 7, 2018
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Tetric® CAD is intended for: - Veneers - Inalys - Onlays (e.g. occlusal veneers, partial crowns) - Crowns in the anterior and posterior region
Device Story
Tetric® CAD is a pre-cured composite block used for chairside fabrication of permanent dental restorations (veneers, inlays, onlays, crowns). The device consists of cross-linked dimethacrylate and inorganic fillers attached to a metal mandrel compatible with CAD/CAM milling systems (e.g., Cerec/InLab, PlanMill, Programill One). The dentist performs intraoral imaging, designs the restoration, and mills the block in a compatible CAD/CAM system. Post-milling, the restoration is polished, characterized, and cemented. The device enables efficient, esthetic, permanent tooth restoration in a single clinical visit. It is intended for professional use by dentists.
Clinical Evidence
Bench testing only. Evaluated physical properties including water solubility (ISO 10477:2004) and flexural strength (ISO 6872:2015) in direct comparison to the predicate. Biocompatibility evaluated per ISO 10993-1:2009, including cytotoxicity (EN ISO 7405:2008, ISO 10993-5:2009) and genotoxicity (ISO 10993-3:2014). Results confirmed the device is non-cytotoxic and non-genotoxic, with physical properties substantially equivalent to the predicate.
Technological Characteristics
Composite block composed of cross-linked dimethacrylate and inorganic fillers. Form factor: blocks with metal mandrel for CAD/CAM milling. Standards: EN 1641:2009 (Dentistry - Medical devices for dentistry - Materials). Flexural strength: 272 MPa. Storage: 2-28°C. Industrially cured; no sterilization required.
Indications for Use
Indicated for patients requiring dental restorations including veneers, inlays, onlays (e.g., occlusal veneers, partial crowns), and crowns in the anterior and posterior regions. Contraindicated for bridge constructions, patients with severely reduced residual dentition, and patients with known allergies to ingredients.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Ivoclar Vivadent AG % Donna Hartnett Director QA/Regulatory Affairs Ivoclar Vivadent, Inc. 175 Pineview Drive Amherst, New York 14228
Re: K173573
Trade/Device Name: Tetric® CAD Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: May 3, 2018 Received: May 8, 2018
Dear Donna Hartnett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
June7, 2018
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Mary S. Runner -S
forTina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K173573
Device Name Tetric® CAD
Indications for Use (Describe)
Tetric® CAD is intended for:
- Veneers
- Inalys
- Onlays (e.g. occlusal veneers, partial crowns)
- Crowns in the anterior and posterior region
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) SUMMARY
Image /page/3/Picture/2 description: The image shows the logo for Ivoclar Vivadent, a dental company. The logo features the company name in a stylized font, with "Ivoclar" in green and "Vivadent" in blue. The tagline "passion vision innovation" is written in a smaller font below the company name. The logo also includes a series of green and blue dots and squares arranged in an arc above and to the right of the company name.
| Contact: | Donna Hartnett, Director QA/Regulatory<br>Ivoclar Vivadent, Inc.<br>175 Pineview Drive<br>Amherst, New York 14228<br>716-691-2260<br>Donna.hartnett@ivoclarvivadent.com |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company: | Ivoclar Vivadent, AG<br>Bendererstrasse 2, Schaan, FL-9494, Liechtenstein<br>+423-235-3535 |
| Date Prepared: | June 5, 2018 |
| Proprietary Name: | Tetric® CAD |
| Classification Name: | Material, Tooth Shade, Resin (872.3690)<br>(Product Code: EBF) |
| Predicate Device: | Cerasmart (K133824) by GC America, Inc. |
Tetric® CAD is a composite block for the fabrication of esthetic Device Description: permanent inlays, onlays, veneers and crowns in the anterior and posterior region. Fabrication of the dental restoration is performed at the dentist's office in a compatible milling machine. The blocks are composed of cross-linked dimethacrylate and inorganic fillers. In the processing technique, the restoration is polished and seated after milling in the CAD/CAM system. The blocks are attached to a metal mandrel, the geometry of such is compatible with a milling machine, e.g. Cerec/InLab (Sirona), PlanMill (PlanMeca) or Programill One (Ivoclar Vivadent).
Tetric® CAD composite blocks are available in the translucency levels MT (5 shades - A1, A2, A3, A3.5, BL) and HT (4 shades 0 A1, A2, A3, A3.5), and in the sizes I12 and C14.
Indications for Use:
Tetric® CAD is intended for:
- Veneers "
- -Inalys
- -Onlays (e.g. occlusal veneers, partial crowns)
- Crowns in the anterior and posterior region -
{4}------------------------------------------------
K173573
Image /page/4/Picture/1 description: The image shows the logo for Ivoclar Vivadent. The logo is in blue and green, with the words "Ivoclar" and "Vivadent" stacked on top of each other. Above the words are a series of green dots and squares. The logo is simple and modern, and it is likely used to represent the company's brand.
510(K) SUMMARY
Comparison to Predicate: The primary predicate devices to which Tetric® CAD has been compared is GC Americ a's Cerasmart (K133824).
| Device | GC America: Cerasmart (K133824) | Ivoclar Vivadent: Tetric® CAD (K173573) |
|----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The product is indicated for inlays, onlays,<br>veneers and full crown restorations,<br>including crowns on implants. | Tetric® CAD is intended for:<br>- Veneers<br>- Inlays<br>- Onlays (e.g. occlusal<br>veneers, partial crowns)<br>- Crowns in the anterior and<br>posterior region |
| Precaution Measures/<br>Contraindications/<br>Processing restrictions/<br>Side effects | In rare cases this product may cause<br>sensitivity in<br>some people. If such reactions are<br>experienced, discontinue<br>the use of product and consult a physician.<br>Personal protective equipment such as<br>gloves, face<br>masks and safety eyewear should always be<br>worn. | Contraindications:<br>- Bridge constructions<br>- Conventional and self-adhesive<br>cementation<br>- Temporary cementation<br>- Patients with severely reduced residual<br>dentition<br>- Any other use not listed in the indications<br>Failure to observe the following restrictions<br>may compromise<br>the results achieved with Tetric® CAD:<br>- Falling short of the required minimum<br>layer<br>thicknesses<br>- Milling the blocks in a non-compatible<br>CAD/CAM system<br>- Deviations from the recommended luting<br>protocol<br>If a patient is known to be allergic to any of<br>the ingredients of Tetric® CAD, the material<br>should not be used.<br>Do not inhale the composite resin dust<br>during finishing.<br>Use suction equipment and wear a dust<br>mask |
| Summary of<br>Indications, Precaution<br>Measures/<br>Contraindications/<br>Processing restrictions/<br>Side effects | The precautions, processing restrictions and contraindications are described in more detail<br>for Tetric® CAD, but in total the indications and contraindications are substantially<br>equivalent. | |
| Technology | Cerasmart is a pre-cured composite block<br>for milling CAD/CAM indirect restorations.<br>The milled device is used for the<br>restorations of both anterior and posterior<br>teeth. | Tetric® CAD is a composite block for the<br>fabrication of permanent inlays, onlays,<br>veneers and crowns in the anterior and<br>posterior region.<br>In this processing technique, the<br>restoration is polished and incorporated<br>immediately after milling in the CAD/CAM<br>system. |
| Summary of Technology<br>Delivery forms/dosage | The devices are similar, there is a minor difference in the wording.<br>The blocks are available in 3 sizes (12, 14,<br>14L) in 33 shades (Translucency HT, LT and<br>one Bleach shade). | The blocks are available in the<br>translucency levels MT and HT, in 5 and 4<br>shades respectively, and in the sizes I12<br>and C14. |
| Summary of Delivery<br>forms/dosage | The predicate device is available in 3 sizes, one more than Tetric® CAD (14L is not<br>available for Tetric® CAD). Additionally the predicate device offers 33 shades<br>(Translucency HT, LT and a Bleach shade), Tetric® CAD offers 5 shades for MT and 4<br>shades for HT. Even though less shades are available the dentist can still meet the<br>aesthetic demands of the patient. | |
| Storage Conditions | Storage at room temperature (4 - 25 $^{\circ}$ C/39.2 -<br>77.0 $^{\circ}$ F), away from direct sunlight and<br>high humidity;<br>Shelf life: 5 years from date of manufacture | Storage conditions: 2 - 28 $^{\circ}$ C/36 - 82 $^{\circ}$ F,<br>avoid sunlight<br>Shelf life: 36 months |
| Summary of Storage<br>Conditions | The recommended shelf life is shorter for the new product, as less data is currently<br>available. | |
| Principles of Operation | Step by Step:<br>-Preparation Design, shade selection<br>-Scanning and Milling<br>-Finishing and polishing<br>-Characterization<br>-Pre-Treatment<br>-Cementation<br>-Maintenance and repair | Step by Step:<br>- Preparation, shade determination<br>- Intraoral imaging, CAD/CAM process<br>- Polishing<br>- Characterization<br>- Preparing for cementation<br>- Cementation<br>- Checking the articulation/occlusion,<br>polishing<br>- Aftercare |
| Summary Principles of<br>operation | There are minor differences in the wording between the predicate and the new product<br>Tetric® CAD, but these are not significant and the principle of the operation is substantially<br>equivalent. | |
| Summary Chemical<br>Composition | Both products are pre-cured composite blocks with differences in composition. Tetric® CAD<br>has evaluated biocompatibility according to ISO 10993- 1:2009. | |
| Finished Device<br>Specification | Applicable standard: No information<br>Applicable FDA Guidance: No information | EN 1641:2009 – Dentistry - Medical<br>devices for dentistry -<br>Materials<br><br>Dental Composite Resin Devices –<br>Premarket Notification<br>[510(k)] Submission |
| Sensitivity to light | Not relevant - industrially cured | Not relevant - industrially cured |
| Color stability | Not relevant - industrially cured | Not relevant - industrially cured |
| Flexural<br>Strength | Brochure: 238 MPa | Typical mean value (IFU): 272 MPa |
| Sterilization | Not applicable. No sterilization<br>recommendation. | Not applicable. No sterilization<br>recommendation. |
| Single use | Consumable material | Consumable material |
| Summary of Finished<br>Device Specification | Tetric® CAD fulfils the relevant standard and follows the relevant Guidance. | |
| Summary of<br>Performance<br>Specification | The value for flexural strength of Cerasmart (GC`s brochure) correlates with Ivoclar<br>Vivadent's Tetric® CAD. | |
{5}------------------------------------------------
## 510(K) SUMMARY
Image /page/5/Picture/2 description: The image shows the logo for Ivoclar Vivadent. The logo is in two lines, with "ivoclar" on the first line and "vivadent" on the second line. The word "ivoclar" is in green, and the word "vivadent" is in blue. There is a series of green and blue dots above the word "ivoclar".
3
#### Substantial equivalence to the predicate:
The subject and the predicate device are pre-cured composite blocks for milling CAD/CAM restorations. Thus, the devices have the same function, have similar principles of operation and a similar chemical composition. The forgoing comparison supports that the subject device is substantially equivalent to the predicate device.
#### Differences :
The recommended storage temperature is marginally different (2-28 °C) compared to 4-25 °C for the predicate device. This is not substantial and is due to company policy. The shelf life is shorter (36 months) compared to 5 years for the predicate device. This is not substantial and is due to ongoing real-time testing.
{6}------------------------------------------------
### 510(K) SUMMARY
Image /page/6/Picture/2 description: The image shows the logo for Ivoclar Vivadent, a dental company. The logo features the company name in a stylized font, with "Ivoclar" in a darker blue and "Vivadent" in a lighter blue. Above the company name is a series of green and blue dots arranged in an arc. Below the company name is the slogan "passion vision innovation" in a smaller font.
The chemical composition is not fully published for the predicate device and therefore it is assumed there is some difference in ingredients, although the published data shows both products to be substantially equivalent.
#### Non-clinical performance testing:
Bench testing was performed to test the physical properties included in the Finished Device Specification for the subject device including: water solubility according to ISO 10477:2004 and flexural strength according to ISO 6872:2015. The subject device was tested in direct comparison to the predicate device and the results of the bench testing show the products to be substantially equivalent.
#### Biocompatibility:
The subject device was also evaluated for Biocompatibility according to ISO 10993-1:2009 +AC:2010. The following testing was conducted on the subject device: Cytotoxicity according to EN ISO 7405:2008 + A1:2013; ISO 10993-5:2009 and Genotoxicity according to 10993-3:2014 and the device was found to be non-cytotoxic and non-genotoxic.
Conclusion: The data show that Tetric® CAD is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.