K173548 · Merit Medical Systems, Inc. · DQY · Mar 30, 2018 · Cardiovascular
Device Facts
Record ID
K173548
Device Name
Merit Pursue Microcatheter
Applicant
Merit Medical Systems, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Mar 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Microcatheter is intended for general intravascular use, including peripheral and coronary vasculature. Once the subselective region has been accessed, the Microcatheter can be used for the controlled and selective infusion of diagnostics, embolic, or therapeutic materials into vessels. The catheter should not be used in the cerebral vessels.
Device Story
Merit Pursue Microcatheter is a percutaneous catheter for peripheral and coronary vasculature. Device features 2.9F/2.0F or 2.8F/1.7F configurations; lengths of 110cm, 130cm, 150cm; straight or pre-shaped (45 degree/swan neck) distal tips. Proximal end includes molded winged hub with strain relief; distal 80cm shaft has hydrophilic coating for insertion; radiopaque marker at tip for fluoroscopic visualization. Used by physicians in clinical settings to access subselective regions and deliver diagnostic, embolic, or therapeutic agents. Output is visual confirmation via fluoroscopy and delivery of infused materials. Benefits include precise, controlled delivery of therapeutic/diagnostic agents to target vessels.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, thrombogenicity, complement activation), ISO 10555-1 performance testing (surface, leakage, tensile force, corrosion), and ISO 594-2 luer fitting testing. Additional design verification included lubricity, dimensional testing, marker band retention, torque strength, kink resistance, and particulate analysis.
Technological Characteristics
Percutaneous catheter; hydrophilic coating on distal 80cm; radiopaque marker band; molded winged hub with tapered strain relief. Materials meet ISO 10993. Sterilization via ethylene oxide (ISO 11135). Conforms to ISO 10555-1 (general requirements) and ISO 594-2 (luer fittings). Dimensions: 2.9F/2.0F or 2.8F/1.7F; 110-150cm lengths. Standalone device.
Indications for Use
Indicated for general intravascular use in peripheral and coronary vasculature for controlled and selective infusion of diagnostic, embolic, or therapeutic materials. Contraindicated for use in cerebral vessels.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Excelsior™ 1018™ and SL-10 Pre-Shaped Microcatheter (K042568)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The Department of Health and Human Services logo features a stylized depiction of a human figure embraced by a wing-like form. The FDA part of the logo has the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 30, 2018
Merit Medical Systems, Inc. Niloufar Samimi Regulatory Affairs Specialist II 1600 West Merit Parkwav South Jordan, Utah 84095
Re: K173548
Trade/Device Name: Merit Pursue Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: March 5, 2018 Received: November 17, 2017
Dear Niloufar Samimi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
M.A. Hillemann
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173548
Device Name Merit Pursue Microcatheter
#### Indications for Use (Describe)
The Microcatheter is intended for general intravascular use, including peripheral and coronary vasculature. Once the subselective region has been accessed, the Microcatheter can be used for the controlled and selective infusion of diagnostics, embolic, or therapeutic materials into vessels. The catheter should not be used in the cerebral vessels.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K173548- 510(k) Summary
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Contact Person:<br>Date Prepared:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 208-4583<br>Niloufar Samimi<br>11/15/2017<br>1721504 |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel: | Merit Pursue Microcatheter<br>Microcatheter<br>Percutaneous catheter<br>2<br>DQY<br>870.1250<br>Cardiovascular |
| Predicate<br>Device | Premarket Notification Predicate #1:<br>Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer:<br><br>Premarket Notification Reference Predicate #2:<br>Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer:<br>These predicate devices have not been subject to a design-related<br>recall. | Excelsior™ 1018™ and SL-10 Pre-Shaped<br>Microcatheter<br>Percutaneous catheter<br>K042568<br>Stryker<br><br>Maestro Microcatheter<br>Catheter, Continuous Flush<br>K172081<br>Merit Medical System, Inc. |
| Device<br>Description | | The Merit Pursue Microcatheter is available in two French<br>configurations, 2.9F (proximal) / 2.0F (distal) and 2.8F (proximal / 1.7F<br>(distal) and 110cm, 130cm and 150cm lengths. The distal tip of the |
| | microcatheter is offered in straight or pre-shaped 45 degree and swan<br>neck configurations. The proximal end of the microcatheter consists of<br>a molded winged hub with a tapered strain relief. The outer surface of<br>the distal 80cm of the microcatheter shaft is coated with a hydrophilic<br>coating designed to facilitate the introduction of the microcatheter into<br>the vasculature. The microcatheter incorporates a radiopaque marker<br>at the distal tip to facilitate fluoroscopic visualization.<br>The Merit Pursue Microcatheter is offered with two 3mL syringes. | |
| Indications for<br>Use | The Microcatheter is intended for general intravascular use, including<br>peripheral and coronary vasculature. Once the subselective region has<br>been accessed, the Microcatheter can be used for the controlled and<br>selective infusion of diagnostic, embolic, or therapeutic materials into<br>vessels. The catheter should not be used in the cerebral vessels. | |
| Comparison to<br>Predicate<br>Devices | The design and technological characteristics of the subject Merit<br>Pursue Microcatheter are substantially equivalent to those of the<br>Predicate Device #1, Excelsior™ 1018™ and SL10- Pre-shaped<br>Microcatheter and Reference Predicate Device #2 Maestro<br>Microcatheter. The subject device has a similar basic design as the<br>predicates.<br>The comparison between the subject and the predicate devices is<br>based on the following:<br>• Similar Intended use<br>• Similar Indications for use<br>• Similar material types that meet ISO 10993 biocompatibility<br>requirements<br>• Similar design<br>• Same sterilization methods<br>• Same fundamental technology/principle of operation | |
| Performance<br>Data | No performance standards have been established under Section 514 of<br>the Food, Drug and Cosmetic Act for these devices. Performance<br>testing of the subject Merit Pursue Microcatheter was conducted based<br>on the risk analysis and based on the requirements of the following<br>international standards:<br>• ISO 10555-1:2013, Intravascular Catheters – Sterile and single-<br>use catheters – Part 1: General requirements<br>• ISO 11135:2014, Sterilization of health care products - Ethylene<br>oxide - Requirements for the development, validation and routine<br>control of a sterilization process for medical devices<br>• ISO 10993-1: 2009/Technical Corrigendum 1 2010 Biological<br>Evaluation of Medical Devices Part 1: Evaluation and Testing | |
{4}------------------------------------------------
{5}------------------------------------------------
within a risk management process, and FDA quidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (2010)
- . ISO 10993-4: 2017, Biological Evaluation of Medical Devices -Part 4: Selection of Tests for Interactions with Blood
- ISO 10993-4:2002/Amendment 1 2006, Biological evaluation of . medical devices – Part 4: Selection of tests for interaction with blood
- . ISO 10993-5:2009. Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2010. Biological evaluation of medical devices -● Part 10: Tests for irritation and skin sensitization
- ISO 10993-11:2006, Biological evaluation of medical devices -Part 11: Tests for systemic toxicity
- United States Pharmacopeia 40, National Formulary 35, 2017 <151> Pyrogen Test
- ASTM F756-17 Standard Practice for Assessment of Hemolytic . Properties of Materials
- ISO 10993-7:2008, Biological evaluation of medical devices -. Part 7: Ethylene oxide sterilization residuals
- ISO 2233:2000, Packaging complete, filled transport packages . and unit loads -conditioning for testing
- . ASTM D4169-14:2014, Standard practice for performance testing of shipping containers and systems
- . ISO 594-2: 1998 Conical fittings with 6% (Luer) taper for syringes, needles and certain other medical equipment – Part 2: Lock fittings
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility testing
The biocompatibility evaluation for the Merit Pursue Microcatheter was conducted in accordance with FDA guidance use of International standard ISO-10993, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. The battery of testing included the following tests:
- Cytotoxicity ●
- Sensitization
- Irritation .
- Acute Systemic Toxicity ●
- Pvrogenicitv ●
- Hemolysis ●
- Thrombogenicity ●
- Complement Activation ●
The Merit Pursue Microcatheter is considered to have circulating blood contact for a limited (< 24 hours) duration.
{6}------------------------------------------------
#### Performance Testing-Bench
### ISO 594-2 Performance testing/ Design verification
- Gauging ●
- Separation Force ●
- Unscrewing Torque ●
- Ease of Assembly ●
- Resistance to Overriding
- Stress Cracking .
#### ISO 10555-1 Performance testing
- Surface ●
●
- Freedom from Liquid Leakage ●
- Freedom from Air Leakage ●
#### Performance Data cont.
- Power injection ● Peak Tensile Force
- Corrosion Resistance ●
## Additional Testing (Design verification and validation)
- Lubricity .
- Dimensional testing ●
- Design Validation
- Marker Band Retention ●
- Tip Shape
- Torque Strength ●
- Radio-detectability
- . Kink Resistance
- Chemical Compatibility .
- Particulate Analysis ●
## Summary of Substantial Equivalence
Based on the indications for use, design, safety and performance testing, the subject Merit Pursue Microcatheter meets the requirements that are considered essential for its intended use and is substantially equivalent to the Predicate Device #1 Stryker, the Excelsior 1018 and SL-10 [K042568].
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.