← Product Code [KPS](/productcode/KPS) · K173547

# ImagenUniversal (K173547)

_Cardiovascular Imaging Technologies · KPS · Jan 12, 2018 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173547

## Device Facts

- **Applicant:** Cardiovascular Imaging Technologies
- **Product Code:** [KPS](/productcode/KPS.md)
- **Decision Date:** Jan 12, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1200
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The ImagenUniversaJTM system is a software. application that allows the user to visualize raw PET and/or PET/CT data, to correct for transmission and emission, and to correct PET 3D data for photon scatter and prompt gamma photons. The resulting corrected tomograms can be exported to cardiac interpretive software for interpretation. The use of this system is limited to qualified, licensed healthcare providers (radiologists or nuclear med icine physicians) trained in the use of nuclear medicine imaging devices.

## Device Story

ImagenUniversal™ is a Windows-based software application for nuclear medicine imaging. It inputs raw PET and/or PET/CT emission and transmission data. The device performs image processing to correct for transmission/emission misregistration, photon scatter, prompt gamma events, and image noise. Users (radiologists, nuclear cardiologists, or nuclear medicine physicians) select correction parameters, filter settings, reconstruction ranges, and registration settings. The system outputs corrected 3D tomographic data, which is exported to external cardiac interpretive software. The device assists clinicians in preparing PET images for diagnostic interpretation, potentially improving image quality and diagnostic accuracy for cardiac patients.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Windows-based software application. Processes PET/CT emission and transmission data. Performs mathematical corrections for photon scatter, prompt gamma, and misregistration. Standalone software for radiological image processing.

## Regulatory Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- Siemens Biograph 64 and 40 ([K060631](/device/K060631.md))
- Siemens Biograph mCT ([K151486](/device/K151486.md))
- ImagenPRO ([K050366](/device/K050366.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 12, 2018

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Cardiovascular Imaging Technologies % Ms. Melanie Hasek Associate Director, Regulatory Publishing PRA Health Sciences 9755 Ridge Drive LENEXA KS 66219

Re: K173547

Trade/Device Name: ImagenUniversal™ Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS, LLZ Dated: November 10, 2017 Received: November 16, 2017

Dear Ms. Hasek:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn

(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K173547

Device Name ImagenUniversalTM

Indications for Use (Describe)

The ImagenUniversaJTM system is a software. application that allows the user to visualize raw PET and/or PET/CT data, to correct for transmission and emission, and to correct PET 3D data for photon scatter and prompt gamma photons. The resulting corrected tomograms can be exported to cardiac interpretive software for interpretation. The use of this system is limited to qualified, licensed healthcare providers (radiologists or nuclear med icine physicians) trained in the use of nuclear medicine imaging devices.

Type of Use (Select one or both, as applicable)

{ X } Prescription Use (Part 21 CFR 801 Subpart D)

{ } Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE F NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review.instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An aqency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with 21 CFR 807.92.

## General Information:

| A. | Submitted By:        | Cardiovascular Imaging Technologies<br>4320 Wornall Road, Suite 114<br>Kansas City, MO 64111<br>Tel: 816-531-2842<br>Fax: 816-531-0643 |
|----|----------------------|----------------------------------------------------------------------------------------------------------------------------------------|
|    | Contact Person:      | James A. Case                                                                                                                          |
|    | Date Prepared:       | November 13, 2017                                                                                                                      |
| B. | Device Trade Name:   | ImagenUniversal™                                                                                                                       |
|    | Classification Name: | System, Emission Computed Tomography<br>21 CFR 892.1200 (KPS)<br>System, Image Processing, Radiological<br>21 CFR 892.2050 (LLZ)       |
| C. | Predicate Devices:   | Siemens Biograph 64 and 40 (K060631)<br>Siemens Biograph mCT (K151486)<br>ImagenPRO (K050366)                                          |

#### D. Device Description:

ImagenUniversal™ is a Windows application which allows physicians and healthcare professionals to correct myocardial perfusion PET images for transmission and emission misregistration, prompt gamma events, photon scatter and image noise. The system processes perfusion and gated PET reconstructed emission and transmission data to create corrected 3D tomographic data. The user can select correction parameters, filter settings, range of reconstruction, and registration setting. The use of this system is limited to qualified, licensed healthcare providers (radiologists, nuclear cardiologists or nuclear medicine physicians) trained in the use of nuclear medicine imaging devices.

- E. Indications for Use:
The ImagenUniversal™ system is a software application that allows the user to visualize raw PET and/or PET/CT data, to correct for transmission and emission misregistration, and to correct PET 3D data for photon scatter and prompt gamma photons. The resulting corrected tomograms can be exported to cardiac interpretive 

{4}------------------------------------------------

software for interpretation. The use of this system is limited to qualified, licensed healthcare providers (radiologists, nuclear cardiologists or nuclear medicine physicians) trained in the use of nuclear medicine imaging devices.

- F. Comparison of Technical Characteristics to Predicate Device:
The ImagenUniversal™ system and its predicates, the Biograph™ systems, and the ImagenPRO™ utilize the same type of data sets for analysis and calculation of data.

- H. Summary:
Testing and comparison of technological characteristics and intended uses found that all components of the ImagenUniversal™ system are equivalent to the predicates.

---

**Source:** [https://fda-staging.innolitics.com/device/K173547](https://fda-staging.innolitics.com/device/K173547)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com