Amsco Evolution Medium Steam Sterilizer

K173481 · STERIS Corporation · FLE · Mar 9, 2018 · General Hospital

Device Facts

Record IDK173481
Device NameAmsco Evolution Medium Steam Sterilizer
ApplicantSTERIS Corporation
Product CodeFLE · General Hospital
Decision DateMar 9, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6880
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Amsco Evolution Medium Steam Sterilizer Models HC-900 and HC-1200 are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations: - Prevacuum – is equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. - Steam Flush Pressure-Pulse (SFPP) - is equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.

Device Story

Steam sterilizer for healthcare facilities; processes heat/moisture-stable materials (instrument trays, fabric packs). Operates via Prevacuum or Steam Flush Pressure-Pulse (SFPP) cycles; includes Gravity, Liquid, Leak Test, and DART (Bowie-Dick) cycles. Device uses steam to achieve sterilization; parameters (temperature, time, dry time) are factory-programmed. Operated by healthcare personnel. Output consists of sterilized items; status/cycle data provided via integrated printer. Benefits include effective sterilization of surgical instruments and facility materials.

Clinical Evidence

Bench testing only. Verification and validation activities included performance testing per ST8, electrical safety testing per IEC 61010-1, software validation, and mechanical operation checks (door function, switch actuation). All tests passed.

Technological Characteristics

Steam sterilizer; chamber sizes 26"x37.5"x42" (HC-900) and 26"x37.5"x54" (HC-1200). Energy source: steam. Connectivity: integrated printer. Software: validated for control and printing functions. Standards: ST8 (sterilization performance), IEC 61010-1 (electrical safety).

Indications for Use

Indicated for sterilization of heat and moisture-stable materials in healthcare facilities. Includes wrapped instrument trays and fabric packs. Liquid cycle is for non-patient contact use only.

Regulatory Classification

Identification

A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus symbol is composed of three human profiles facing to the right. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 9, 2018 STERIS Corporation Tony Piotrkowski Sr. Manager. Regulatory Affairs 5960 Heisley Rd. Mentor, Ohio 44060 Re: K173481 Trade/Device Name: Amsco® Evolution Medium Steam Sterilizer (Models HC-900 and HC-1200) Regulation Number: 21 CFR 880.6880 Regulation Name: Sterilizer, Ethylene Oxide Gas Regulatory Class: Class II Product Code: FLE Dated: December 12, 2017 Received: January 11, 2018 Dear Mr. Tony Piotrkowski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Michael J. Ryan -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use ### 510(k) Number (if known) K173481 ### Device Name Amsco Evolution Medium Steam Sterilizer (Models HC-900 and HC-1200) ### Indications for Use (Describe) The Amsco Evolution Medium Steam Sterilizer Models HC-900 and HC-1200 are designed for sterilization of heat and moisturestable materials used in healthcare facilities and are available in two configurations: - Prevacuum is equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. . - Steam Flush Pressure-Pulse (SFPP) is equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-• Dick) cycles. The Amsco Evolution Medium Prevacuum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 5-1): Table 5-1. Amsco Evolution Medium Steam Sterilizer factory-programmed sterilization cycles and cycle values | Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load | |-------------------------|--------------------------------|----------------------------|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------| | Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument<br>trays, maximum weight 25<br>lbs (11.3 kg) each and<br>Fabric Packs. Refer to<br>Table 5-3 for recommended<br>quantities. | | Prevac<br>Gravity | 270°F (132°C)<br>250°F (121°C) | 4 minutes<br>30 minutes | 5 minutes<br>15 minutes | Single Fabric Pack<br>Fabric Packs. Refer to<br>Table 5-3 for recommended<br>quantities. | | Liquid* | 250°F (121°C) | 45 minutes | N/A | Refer to Table 5-4 for<br>recommended quantities. | | Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument<br>trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to<br>Table 5-3 for recommended<br>quantities. | | Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument<br>trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to<br>Table 5-3 for recommended<br>quantities. | | Gravity | 250°F (121°C) | 30 minutes | 30 minutes | Double wrapped instrument<br>trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to<br>Table 5-3 for recommended<br>quantities. | | Gravity | 270°F (132°C) | 25 minutes | 15 minutes | Fabric Packs. Refer to<br>Table 5-3 for recommended<br>quantities. | | DART<br>Warm-Up<br>DART | 270°F (132°C)<br>270°F (132°C) | 3 minutes<br>3 1/2 minutes | 1 minute<br>1 minute | N/A<br>Bowie-Dick Test Pack,<br>DART Test Pack | | Leak Test | N/A | N/A | N/A | N/A | * The liquid cycle is for non-patient contact use only. {3}------------------------------------------------ The Amsco Evolution Medium Steam Fluse (SFPP) Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 5-2): Table 5-2. Amsco Evolution Medium Steam Flush Pressure-Pulse (SFPP) Sterilizer factory-programmed sterilization cycles and cycle | Cycles | Sterilize Temperature | Sterilize Time | Dry Time | Recommended Load | |--------------|-----------------------|----------------|------------|----------------------------------------------------------------------------------------------------------------------------------------| | SFPP | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 5-3 for recommended quantities. | | SFPP | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. | | Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 5-3 for recommended quantities. | | Gravity | 250°F (121°C) | 30 minutes | 15 minutes | Fabric Packs. Refer to Table 5-3 for recommended quantities. | | SFPP | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack | | Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. | | Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 5-3 for recommended quantities. | | Liquid | 250°F (121°C) | 45 minutes | N/A | Refer to Table 5-4 for recommended quantities. | | DART Warm-Up | 270°F (132°C) | 3 minutes | 1 minute | N/A | | DART | 270°F (132°C) | 3 ½ minutes | 1 minute | Bowie-Dick Test Pack, DART Test Pack | | Leak Test | N/A | N/A | N/A | N/A | *The liquid cycle is for non-patient contact use only {4}------------------------------------------------ The following table lists STERIS's recommended loads by sterilizer size: Table 5-3. Amsco Evolution Medium Steam Sterilizer recommended loads per sterilizer size | Sterilizer Size | Wrapped Instrument<br>Trays | Fabric Packs | |-------------------|-----------------------------|--------------| | 26" x 37 ½" x 42" | 12 | 18 | | 26" x 37 ½" x 54" | 16 | 30 | The following table is a guideline for liquid cycle processing: Table 5-4. Amsco Evolution Medium Steam Sterilizer Liquid Cycle* Guideline | Number of<br>Containers | Volume of Liquid In One<br>Container | Minimum Recommended<br>Sterilize Time at 250°F (121°C) | |-------------------------|--------------------------------------|--------------------------------------------------------| | 3 | 1000 ml | 45 minutes | *The liquid cycle is for non-patient contact use only. The Amsco Evolution Medium Steam Sterilizer is offered in the following chamber sizes: - 26" x 37½" x 42" (Model HC 900) . - 26" x 37½" x 54" (Model HC – 1200) {5}------------------------------------------------ Type of Use (Select one or both, as applicable) _ | Prescription Use (Part 21 CFR 801 Subpart D) X | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the STERIS logo. The word STERIS is in large, bold, sans-serif font. Below the word STERIS is an image of several horizontal wavy lines in blue. The lines are stacked on top of each other and appear to represent water. # 510(k) Summary For Amsco Evolution Medium Steam Sterilizer STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198 Contact: Tony Piotrkowski Senior Manager, Regulatory Affairs Tel: 440-392-7437 Fax: 440-357-9198 e-mail: tony_piotrkowski@steris.com Submission Date: February 12, 2018 Premarket Notification Number: K173481 STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600 {7}------------------------------------------------ #### 1. Device Name Trade Name: Amsco® Evolution Medium Steam Sterilizer (Models HC-900 and HC-1200) | Device Class: | Class II | |------------------------|-------------------| | Common/usual Name: | Steam Sterilizer | | Classification Name: | Sterilizer, Steam | | Classification Number: | 21 CFR 880.6880 | | Product Code: | FLE | #### 2. Predicate Device AMSCO Evolution Medium Steam Sterilizer (Models HC-900 and HC-1200), K091136. #### 3. Description of Device The Amsco Evolution Medium Steam Sterilizer models HC-900 and HC-1200 are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations: - Prevacuum is equipped with Prevacuum, Gravity, Liquid, Leak Test and ● DART (Bowie-Dick) cycles. - Steam Flush Pressure-Pulse (SFPP) is equipped with SFPP, Prevacuum, . Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. The chamber sizes of the proposed Amsco Evolution Medium Steam Sterilizer Models are as follows: - · 26'' x 371/2' x 42" (Model HC 900) - 26'' x 37½' x 54'' (Model HC 1200) #### 4. Intended Use The Amsco Evolution Medium Steam Sterilizer Models HC-900 and HC-1200 are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations: - Prevacuum – is equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. - . Steam Flush Pressure-Pulse (SFPP) - is equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. {8}------------------------------------------------ The Amsco Evolution Medium Prevacuum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 6-1): | Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load | |-----------------|--------------------------|-------------------|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------| | | | | | | | Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument<br>trays, maximum weight 25<br>lbs (11.3 kg) each and<br>Fabric Packs. Refer to<br>Table 6-3 for recommended<br>quantities. | | Prevac | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack | | Gravity | 250°F (121°C) | 30 minutes | 15 minutes | Fabric Packs. Refer to<br>Table 6-3 for recommended<br>quantities. | | Liquid* | 250°F (121°C) | 45 minutes | N/A | Refer to Table 6-4 for<br>recommended quantities. | | Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument<br>trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to<br>Table 6-3 for recommended<br>quantities. | | Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument<br>trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to<br>Table 6-3 for recommended<br>quantities. | | Gravity | 250°F (121°C) | 30 minutes | 30 minutes | Double wrapped instrument<br>trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to<br>Table 6-3 for recommended<br>quantities. | | Gravity | 270°F (132°C) | 25 minutes | 15 minutes | Fabric Packs. Refer to<br>Table 6-3 for recommended<br>quantities. | | DART<br>Warm-Up | 270°F (132°C) | 3 minutes | 1 minute | N/A | | DART | 270°F (132°C) | 3 ½ minutes | 1 minute | Bowie-Dick Test Pack,<br>DART Test Pack | | Leak Test | N/A | N/A | N/A | N/A | Table 6-1. Amsco Evolution Medium Prevacuum Steam Sterilizer factory-programmed sterilization cycles and cycle values *The liquid cycle is for non-patient contact use only. {9}------------------------------------------------ The Amsco Evolution Medium Steam Flush Pressure-Pulse (SFPP) Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 6-2): | Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load | |---------|--------------------------|-------------------|------------|----------------------------------------------------------------------------------------------------------------------------------------| | SFPP | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 6-3 for recommended quantities. | | SFPP | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 6-3 for recommended quantities. | | Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 6-3 for recommended quantities. | | Gravity | 250°F (121°C) | 30 minutes | 15 minutes | Fabric Packs. Refer to Table 6-3 for recommended quantities. | | SFPP | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack | | Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 6-3 for recommended quantities. | | Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 6-3 for recommended quantities. | | Liquid* | 250°F (121°C) | 45 minutes | N/A | Refer to Table 6-4 for recommended quantities. | | Table 6-2. Amsco Evolution Medium Steam Flush Pressure-Pulse (SFPP) | | | | | |---------------------------------------------------------------------|---------------------------------------------------------------------|--|--|--| | | Sterilizer factory-programmed sterilization cycles and cycle values | | | | {10}------------------------------------------------ ### K173481 STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION Amsco Evolution Medium Steam Sterilizer (Models HC-900 and HC-1200) | Cycles | Sterilize Temperature | Sterilize Time | Dry Time | Recommended Load | |-----------------|-----------------------|----------------|----------|-----------------------------------------| | DART<br>Warm-Up | 270°F (132°C) | 3 minutes | 1 minute | N/A | | DART | 270°F (132°C) | 3 ½ minutes | 1 minute | Bowie-Dick Test Pack,<br>DART Test Pack | | Leak Test | N/A | N/A | N/A | N/A | *The liquid cycle is for non-patient contact use only. The following table lists STERIS's recommended loads by sterilizer size: | Sterilizer Size | Wrapped Instrument<br>Trays | Fabric Packs | |---------------------|-----------------------------|--------------| | 26" x 37 1/2" x 42" | 12 | 18 | | 26" x 37 ½" x 54" | 16 | 30 | ## Table 6-3. Amsco Evolution Medium Steam Sterilizer recommended loads per sterilizer size The following table is a guideline for liquid cycle processing: | Table 6-4. Amsco Evolution Medium Steam | |-----------------------------------------| | Sterilizer Liquid Cycle* Guideline | | Number of<br>Containers | Volume of Liquid In One<br>Container | Minimum Recommended<br>Sterilize Time at 250°F (121°C) | |-------------------------|--------------------------------------|--------------------------------------------------------| | 3 | 1000 ml | 45 minutes | *The liquid cycle is for non-patient contact use only. The Amsco Evolution Medium Steam Sterilizer is offered in the following chamber sizes: - 26" x 37½" x 42" (Model HC - 900) - 26" x 37½" x 54" (Model HC 1200) ● #### 5. Description of Safety and Substantial Equivalence The Evolution Medium Sterilizers are the same as the predicate devices except for the specific modifications described in this submission. There are no differences in the indications for use or technology (including features, materials, and principles of operation) between the proposed and predicate devices. Therefore, the differences between the proposed and predicate devices are limited to the described modifications, and these proposed changes raise no new concerns of safety and effectiveness when compared to the predicate device. {11}------------------------------------------------ The following table summarizes the verification activities that were performed with their respective acceptance criteria to ensure that these modifications do not affect the safety or effectiveness of the Evolution Medium Steam Sterilizers. | Device<br>Modification | Verification /<br>Validation<br>Activity | Acceptance<br>Criteria | Results of<br>Verification<br>/Validation | |---------------------------------------|------------------------------------------|-----------------------------------------------------------|-------------------------------------------| | Control replacement | Performance testing<br>Electrical safety | Per ST8<br>Per IEC 61010-1 | Pass | | Software modifications | Software validation | Software shall be appropriately<br>verified and validated | Pass | | Vacuum pump<br>replacement | Performance testing | Per ST8 | Pass | | Printer replacement | Software validation | Software shall be appropriately<br>verified and validated | Pass | | Valve modifications | Performance testing | Per ST8 | Pass | | Switch modification | Actuation | Proper actuation, no alarms | Pass | | Door mounting bracket<br>modification | Door operation check | Proper function after<br>multiple cycles | Pass | #### 6. Conclusion Based on the intended uses, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and perform at least as well as the legally marketed predicate device (K091136), Class II (21 CFR 880.6880), product code FLE.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...