ChemoCLAVE Cytotoxic Medication Preparation and Delivery System
K173477 · Icu Medical, Inc. · ONB · Aug 31, 2018 · General Hospital
Device Facts
Record ID
K173477
Device Name
ChemoCLAVE Cytotoxic Medication Preparation and Delivery System
Applicant
Icu Medical, Inc.
Product Code
ONB · General Hospital
Decision Date
Aug 31, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The ChemoCLAVE is a needle-free Closed System Transfer Device (CSTD) that mechanically prohibits the transfer of environmental contaminants, including bacterial and airborne contaminants into the system, and the escape of drug or vapor concentrations outside the system during drug preparation and administration, thereby minimizing exposure of individuals, healthcare personnel and the environment to hazardous drugs.
Device Story
ChemoCLAVE is a needle-free Closed System Transfer Device (CSTD) used during hazardous drug preparation and administration. System comprises two primary components: Spiros and ChemoCLAVE. Components connect to form a closed fluid path; both independently self-seal upon disconnection. Device includes vial access devices, bag access devices, adapters, and administration sets. Used in clinical settings by healthcare personnel to prevent transfer of environmental contaminants into the system and escape of hazardous drug/vapor concentrations outside the system. Minimizes exposure to hazardous drugs for personnel and environment. Device is sterile, single-use, and provides a mechanical barrier to leakage.
Clinical Evidence
Bench testing only. Testing included microbial ingress, vapor leakage, hazardous drug exposure, and urinary catheter compatibility. Sterilization validated per ISO 11137-2:2013; packaging validated per ISO 11607:2017. Risk management conducted per ISO 14971:2007. All pre-determined acceptance criteria met.
Technological Characteristics
Needle-free CSTD; components include Spiros and ChemoCLAVE. Materials consistent with predicate K081361. Sterilization via E-beam or Gamma radiation (SAL 10^-6) per ISO 11137-2:2013. Packaging per ISO 11607:2017. Mechanical self-sealing interface.
Indications for Use
Indicated for healthcare personnel and individuals involved in the preparation and administration of hazardous drugs to minimize exposure to contaminants and drug vapors.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
ChemoCLAVE Cytotoxic Medication Preparation and Delivery System (K081361)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 31, 2018
ICU Medical, Inc. Catherine Kang Manager, Global Regulatory Affairs 600 N. Field Drive Lake Forest, Illinois 60045
Re: K173477
Trade/Device Name: ChemoCLAVE Cytotoxic Medication Preparation and Delivery System Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Dated: July 20, 2018 Received: July 24, 2018
#### Dear Catherine Kang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan M.
Stevens -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173477
#### Device Name
ChemoCLAVE Cytotoxic Medication Preparation and Delivery System
Indications for Use (Describe)
The ChemoClave is a needle-free Closed System Transfer Device (CSTD) that mechanically prohibits the transfer of environmental contaminants, including bacterial and airborne contaminants into the system, and the escape of drug or vapor concentrations outside the system during drug preparation and administration, thereby minimizing exposure of individuals, healthcare personnel and the environment to hazardous drugs.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K173477 510(k) Summary
| Submitter: | ICU Medical, Inc.<br>600 North Field Drive<br>Lake Forest, IL 60045<br>Establishment Registration: 3013319212<br>(Owner/Operator #2025816) |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Application Correspondent: | Catherine Kang, Ph.D<br>Manager, Global Regulatory Affairs<br>Phone: (224) 706-2372<br>Fax: (224) 706-2917<br>E-mail: catherine.kang@icumed.com |
| Preparation Date: | August 29, 2018 |
| Trade Name: | ChemoCLAVE® Cytotoxic Medication<br>Preparation and Delivery System |
| Common or Usual Name: | Closed Antineoplastic and Hazardous Drug<br>Reconstitution and Transfer System |
| Regulation Name: | Intravascular Administration Set |
| Regulation Number: | 21 CFR 880.5440 |
| Product Code: | ONB |
| Device Class: | Class II |
| Primary Predicate Device: | K081361 - ChemoCLAVE Cytotoxic<br>Medication Preparation and Delivery<br>System |
### Submission Purpose:
The purpose of this submission is to provide information demonstrating that the ChemoClave devices meet the design input requirements of a Closed System Transfer Device (CSTD) and to demonstrate substantial equivalence to K081361.
### Device Indications For Use:
The ChemoClave is a needle-free Closed System Transfer Device (CSTD) that mechanically prohibits the transfer of environmental contaminants, including bacterial and airborne contaminants into the system, and the escape of drug or vapor concentrations outside the system
{4}------------------------------------------------
during drug preparation and administration, thereby minimizing exposure of individuals, healthcare personnel and the environment to hazardous drugs.
## Device Description:
The ChemoCLAVE® Cytotoxic Medication Preparation and Delivery System devices are sterile, single use CSTDs, which prohibit the transfer of environmental contaminants into the system and the escape of hazardous drug or vapor concentrations outside the system. The Spiros and ChemoClave components comprise the primary components of the ChemoClave CSTD System, which connect to form a closed fluid path for transfer of hazardous medications. The devices which include the ChemoClave component are vial access devices, bag access devices, adapters, and administration sets. The devices which include the mating Spiros component are syringes, adapters, and administration sets. The Spiros and the ChemoClave will each independently self-seal when they are disconnected from one another.
## Technological Characteristics and Substantial Equivalence:
A technological comparison table is provided below that compares the subject device and predicate device:
| Characteristic | Proposed Device | Predicate 510(k) K081361 | Assessment of Differences |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The ChemoCLAVE is a needle-free<br>Closed System Transfer Device<br>(CSTD) that mechanically prohibits<br>the transfer of environmental<br>contaminants, including bacterial<br>and airborne contaminants into the<br>system, and the escape of drug or<br>vapor concentrations outside the<br>system during drug preparation and<br>administration, thereby minimizing<br>exposure of individuals, healthcare<br>personnel and the environment to<br>hazardous drugs. | The ChemoCLAVE Cytotoxic<br>Medication Preparation and<br>Delivery System consists of 6<br>previously cleared components<br>(CLAVE®, Spikes, SPIROSTM,<br>GENIETM, Vial Access, and Admin<br>Sets) that can be combined into<br>various configurations intended for<br>use in the preparation and patient<br>administration of cytotoxic<br>medications. | The Indications for Use has<br>been updated to provide<br>information consistent with a<br>Closed System Transfer<br>Device (CSTD) and devices<br>under the ONB product code.<br><br>The differences do not<br>impact the safety or<br>effectiveness of device. |
| Product Code | ONB | LHI | Updating devices as ONB. |
| Common<br>name | Closed Antineoplastic And<br>Hazardous Drug Reconstitution<br>And Transfer System | Set, I.V. Fluid Transfer | Renamed to provide clarity<br>consistent with product code. |
| Design and<br>Materials of<br>Construction | ChemoClave Adapter | ChemoClave Adapter | Same. |
| | Spiros Adapter | Spiros Adapter | Addition of non-fluid path<br>cap to the proposed device. |
| | Genie Vial Access Device | Genie Vial Access Device | Minor change in silicone<br>formulation in the proposed<br>device. |
| | Vial Spikes | Vial Spikes | Same. |
{5}------------------------------------------------
| | Bag Spikes | Bag Spikes | Minor dimensional changes<br>to improve manufacturing of<br>the proposed device. |
|--|----------------------|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Administration sets | Administration sets | Minor material modifications<br>to connecting components in<br>the proposed device; material<br>is identical to materials used<br>in other components on the<br>set. |
| | Syringes with Spiros | Syringes with Spiros | Same. |
## Substantial Equivalence Discussion
The differences between predicate and proposed devices are identified in the table above. The proposed devices have equivalent technological characteristics as ICU Medical's current legally marketed devices cleared under 510(k) premarket notification K081316.
These minor differences, between subject device and predicate device, do not raise different questions of safety and effectiveness.
## Non-Clinical Testing Summary:
Non-clinical tests were conducted to verify that the proposed device met all design specifications and is substantially equivalent (SE) to the ChemoCLAVE Cytotoxic Medication Preparation and Delivery System. The device was subjected to the following functional and performance tests to demonstrate that it performs as intended:
- Microbial Ingress Testing ●
- Emission Testing- Assess Vapor Leakage
- Hazardous Drug Exposure Testing - Assess Liquid/Vapor Leakage
- Urinary Catheter Compatibility Testing
- ANSI/AAMI/ISO 11137-2 2013 "Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose".
- ANSI/AAMI ST72:2011/(R)2016 Bacterial endotoxins -- Test methods, routine . monitoring, and alternatives to batch testing
- ISO 11607:2017 Packaging For Terminally Sterilized Medical Devices ●
The materials of construction are the same as those used in the predicate device cleared under K081361. Biocompatibility testing provided in K081361 demonstrated the biocompatibility of ChemoCLAVE products.
The ChemoCLAVE products are supplied sterile for single use. The subject device is sterilized by E-beam or Gamma Radiation to achieve a sterility assurance level (SAL) of 10th. Sterility was validated in accordance with ISO 11137-2: 2013 - Dose setting using Bioburden. Packaging testing was performed per ISO 11607:2017.
{6}------------------------------------------------
A risk analysis was conducted in accordance with ISO 14971: 2007 - Medical devices -Application of risk management to medical devices.
In all testing, the pre-determined acceptance criteria were met.
## Substantially Equivalence Conclusion
Differences between the intended use and technological characteristics of the subject device compared to the predicate do not raise different questions of safety and effectiveness. The performance of the device is supported by non-clinical testing and risk management activities. The ChemoCLAVE products are Substantially Equivalent (SE) to the ChemoCLAVE Cytotoxic Medication Preparation and Delivery System, cleared under K081361.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.