The VALEO® C+CSC with Lumen Interbody Fusion Device is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level or two contiguous levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Valeo® C+CSC with Lumen Interbody Fusion Device is used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach from the C2-C3 disc space to the C7-T1 disc space using autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. VALEO® C+CSC with Lumen Interbody Fusion Device is to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.
Device Story
Valeo C+CSC with Lumen is a cervical interbody fusion device; manufactured from silicon nitride ceramic. Implanted via anterior approach (C2-T1) by surgeons to facilitate spinal fusion in patients with degenerative disc disease. Device acts as a structural spacer to maintain disc space height while bone graft promotes fusion. Requires supplemental fixation. Provided sterile. Benefits include stabilization of spinal segment and facilitation of fusion in symptomatic patients.
Clinical Evidence
Bench testing only. Mechanical testing (static/dynamic compression, static/dynamic torsion, subsidence, expulsion) performed per ASTM F2077 and ASTM F2267. Biocompatibility testing performed per ISO 10993 standards (cytotoxicity, sensitization, intracutaneous extract, systemic toxicity, genotoxicity, muscle implantation) and ANSI/AAMI ST72:2011 (endotoxin). All tests passed.
Technological Characteristics
Material: Silicon nitride ceramic. Form factor: Cervical interbody fusion cage with lumen. Sterilization: Provided sterile. Standards: ASTM F2077 (mechanical), ASTM F2267 (subsidence), ISO 10993 (biocompatibility), ANSI/AAMI ST72 (endotoxin).
Indications for Use
Indicated for skeletally mature patients with cervical degenerative disc disease (DDD) and radicular symptoms at 1-2 contiguous levels. Requires 6 weeks of prior non-operative treatment. Used with supplemental fixation and autograft/allogenic bone graft via anterior approach (C2-T1).
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Valeo Spacer System and Valeo II Interbody Fusion Device (K142264)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 5, 2018
Amedica Corporation Shanna Ryan Regulatory Affairs Project Manager 63 East 11400 South 192 Sandy, Utah 84070
Re: K173453
Trade/Device Name: Valeo® C+CSC with Lumen Interbody Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: February 3, 2018 Received: February 5, 2018
Dear Ms. Ryan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katherine D. Kavlock -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
#### K173453
#### Device Name
Valeo® C+CSC with Lumen Interbody Fusion Device
#### Indications for Use (Describe)
The VALEO® C+CSC with Lumen Interbody Fusion Device is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level or two contiguous levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies.
Valeo® C+CSC with Lumen Interbody Fusion Device is used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach from the C2-C3 disc space using autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
VALEO® C+CSC with Lumen Interbody Fusion Device is to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
EF PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Amedica. The logo consists of an impossible triangle shape in shades of orange, above the word "AMEDICA" in a simple, sans-serif font. The logo is clean and modern, with a focus on the company name.
SECTION 2
VOLUME 2
Date of summary: 2 February 2018 (Updated)
### Applicant
AMEDICA® Corporation 1885 West 2100 South Salt Lake City, UT 84119 Phone: 801.839.3562 Fax: 801.839.3605 Contact: Shanna Ryan, Regulatory Affairs Project Manager Email: shanna.ryan@gmail.com
### Device Classification
| Trade Name: | Valeo® C+CSC with Lumen Interbody Fusion Device System |
|----------------------|--------------------------------------------------------|
| Class: | II |
| Panel: | Orthopedic |
| Product Code: | ODP |
| Regulation: | 21 CFR 888.3080 |
| Classification Name: | Intervertebral body fusion device, cervical |
# Predicate Device
| Primary | |
|---------------|----------------------------------------------------------|
| 510k: | k142264 |
| Trade Name: | Valeo Spacer System and Valeo II Interbody Fusion Device |
| Manufacturer: | Amedica® Corporation |
| Secondary | |
| 510k: | K103033 |
| Trade Name: | Zimmer Trabecular Metal Fusion Device |
| Manufacturer: | Zimmer |
# Device Description
The Valeo® C+CSC with Lumen Interbody Fusion Device consists of cervical spine interbody fusion devices, as well as instrumentation designed specifically for the implantation of these implants. The Valeo C+CSC with Lumen is manufactured from Amedica's Silicon Nitride ceramic material and is provided sterile.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for AMEDICA. The logo consists of an impossible triangle shape in yellow and orange gradient, positioned above the word "AMEDICA" in a simple, sans-serif font. The logo is clean and modern.
Valeo® C+CSC with Lumen Traditional 510(k) Summary 21 CFR 807.92(a)
VOLUME 2
SECTION 2
### Indications for Use
The VALEO® C+CSC with Lumen Interbody Fusion Device is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level or two contiguous levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies.
Valeo® C+CSC with Lumen Interbody Fusion Device is used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach from the C2-C3 disc space to the C7-T1 disc space using autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
VALEO® C+CSC with Lumen Interbody Fusion Device is to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment with an intervertebral cage.
### Technological Characteristics and Comparison to the Predicate Device
The Valeo® C+CSC with Lumen Interbody Fusion Device System is substantially equivalent to the predicates with respect to intended use, indications, design, and performance.
#### Performance Testing Summary
Testing performed on this device indicates that the Valeo C+CsC with Lumen Device is substantially equivalent to the Valeo Spacer System-C devices cleared under K142264. ASTM F2077 and ASTM F2267 performance standards were adhered to and all applicable requirements were met. Any deviations to the standards are noted at appropriate sections.
Mechanical Testing
| Test | Result |
|---------------------|--------|
| Static Compression | PASS |
| Dynamic Compression | PASS |
| Static Torsion | PASS |
| Dynamic Torsion | PASS |
| Subsidence | PASS |
| Expulsion | PASS |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Amedica. The logo consists of an impossible triangle above the word "AMEDICA" in a simple, sans-serif font. The triangle is a gradient of yellow and orange, and the text is gray.
VOLUME 2
#### Biocompatibility evaluations
| Test / Biological Endpoint | Standard | Result |
|-----------------------------------------|---------------------|--------|
| Cytotoxicity | ISO 10993-5:1999 | PASS |
| Maximization Sensitization | ISO 10993-10:2002 | PASS |
| ISO Intracutaneous Extract | ISO 10993-10:2002 | PASS |
| ISO Systemic Toxicity Study | ISO 10993-11:2006 | PASS |
| Genotoxicity | ISO 10993-3:2003 | PASS |
| ISO Muscle Implantation Study | ISO 10993-6:1994 | PASS |
| System Toxicity | ISO 10993-11:2006 | PASS |
| Bacterial Endotoxin Testing (LAL assay) | ANSI/AAMI ST72:2011 | PASS |
### Conclusions
In summary, the Valeo C+CSC with Lumen has the same indications for use, technological characteristics, principle of operation, materials, and sterilization method as the cited predicate devices. Based on this and data provided in this pre-market notification, the subject device has been shown to be substantially equivalent to the predicate devices. Performance testing demonstrate the subject device performs as well or better than the predicate, therefore, is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.