BioSphere Flex is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. BioSphere Flex is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. BioSphere Flex must be hydrated with autogenous bone marrow prior to implantation. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
Device Story
BioSphere Flex is a sterile, resorbable bone void filler; composed of porous 45S5 bioactive glass granules in a collagen and sodium hyaluronate scaffold carrier. Device is hydrated with autogenous bone marrow by the surgeon prior to implantation; becomes flexible and moldable. Implanted into non-structural bony voids or gaps in extremities or pelvis. Provides osteoconductive scaffold; supports new bone formation on granule surfaces and throughout porosity. Resorbs over time; replaced by host bone. Used in surgical settings by orthopedic surgeons. Benefits patient by filling bone defects and facilitating natural bone healing.
Clinical Evidence
Bench and animal testing only. Bench testing evaluated physical/chemical properties (density, hydration, handling, microstructure, bioactivity via hydroxyapatite formation in simulated body fluid) and biocompatibility (ISO 10993). In vivo testing in a critical-sized sheep femoral defect used x-ray, micro-CT, histology, and histomorphometry to compare healing characteristics against the predicate. Results demonstrated equivalent healing performance.
Technological Characteristics
Composition: 45S5 bioactive glass granules, collagen, and sodium hyaluronate scaffold. Principle: Osteoconductive bone void filler. Form: Sterile, single-use, resorbable, flexible/moldable upon hydration. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for bony voids or gaps in the skeletal system (extremities and pelvis) not intrinsic to structural stability. Applicable to surgically created or traumatic osseous defects. Requires hydration with autogenous bone marrow prior to implantation.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a blue square and the full name, "U.S. Food & Drug Administration," written in blue text.
September 12, 2018
Synergy Biomedical % Randy Prebula Partner Hogan Lovells U.S. LLP 555 Thirteenth Street N.W. Washington, District of Columbia 20004
Re: K173424
Trade/Device Name: BioSphere Flex Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: August 10, 2018 Received: August 10, 2018
Dear Mr. Prebula:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurence D. Coyne -S
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: June 30, 2020 See PRA Statement on last page
510(k) Number (if known)
K173424
Device Name
BioSphere Flex
Indications for Use (Describe)
BioSphere Flex is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. BioSphere Flex is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis). These defects may be surgically created osseous defects created from traumatic injury to the bone. BioSphere Flex must be hydrated with autogenous bone marrow prior to implantation. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY BioSphere Flex
#### Submitter
Synergy Biomedical, LLC 100 Springhouse Drive, Suite 108 Collegeville, PA 19426
Phone: 484-902-8141 Facsimile: 484-902-8179
### Contact Person
Randy Prebula Hogan Lovells U.S. LLP 555 Thirteenth Street, NW Washington, DC 20004
Date Prepared: September 10, 2018
Name of Device: BioSphere Flex Common or Usual Name: Bioactive Bone Graft Classification Requlation: 21 C.F.R. § 888.3045, Bone Void Filler Device Class: Class II Product Code: MQV 510(k) Review Panel: Orthopedic
# Predicate/Reference Devices
Primary Predicate: K122868 – BioSphere Putty (Synergy Biomedical, LLC) Reference Device: K130392 - Kinex Bioactive (Globus Medical, Inc)
# Device Description
BioSphere Flex is a sterile, single-use resorbable bone void filler that is composed of porous bioactive glass granules suspended in a carrier. BioSphere Flex utilizes 45S5 bioactive glass, which has been shown to be osteoconductive and bioactive. The carrier is composed of a porous scaffold of collagen and sodium hyaluronate. BioSphere Flex has the ability to absorb fluid such as bone marrow aspirate and becomes flexible following hydration. Following implantation into the target site, BioSphere Flex supports bone formation through the porosity of the implant and on the surface of the bioactive glass granules. As the materials resorb, the device is eventually replaced by host bone. BioSphere Flex is provided in a variety of forms.
# Intended Use / Indications for Use
BioSphere Flex is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. BioSphere Flex is indicated to be gently packed into bony voids or gaps of the
{4}------------------------------------------------
skeletal system (i.e., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. BioSphere Flex must be hydrated with autogenous bone marrow prior to implantation. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
The only difference between the BioSphere Flex indications for use and those cleared for the BioSphere Putty predicate is the requirement for hydration with autogenous bone marrow, obtained by aspiration, prior to implantation. This difference has no impact on the device's therapeutic purpose or ability to achieve its function as a bone void filler; it also does not affect the device's safety or effectiveness when used as labeled, because it merely serves to ensure that BioSphere Flex, which has a slightly different composition and configuration than the predicate, will still be sufficiently moldable/flexible to fit various sizes and shapes of bony Moreover. the reference device, Kinex Bioactive (K130392), has identical defects. indications for use to BioSphere Flex.
### Summary of Technological Characteristics
BioSphere Flex is composed of the same 45S5 bioactive glass component as the predicate device, mixed with an inert carrier. In BioSphere Flex, the bioactive glass is in the form of porous granules composed of thermally bonded 45S5 bioactive glass spheres. The bioactive glass granules are suspended in a 3-dimensional, porous scaffold carrier composed of freezedried collagen and sodium hyaluronate. BioSphere Flex is designed to be mixed with autogenous bone marrow. The structure readily absorbs liquid and becomes flexible and moldable when fully hydrated.
BioSphere Flex and its predicates rely on the same technological principles for bone healing, When implanted, the device provides an osteoconductive environment for new bone formation. The porosity of BioSphere Flex allows it to support new bone formation throughout its structure and on the surface of the device. As natural healing progresses, the new bone remodels and the device is eventually replaced by host bone.
#### Performance Data
The device's physical characteristics and ability to function as intended were evaluated in bench and animal testing. In addition, BioSphere Flex was subjected to ISO 10993 biocompatibility testing and was shown to be biocompatible, including testing which demonstrated that the device meets recommended endotoxin limits.
Bench testing assessed various physical, chemical, and biological characteristics of the device, including density, hydration, handling properties, particle distribution, microstructure, particle weight, oxide content, crystallinity, and bioactivity. Bioactivity was assessed through demonstration of hydroxyapatite formation on the surface of the bioglass components following exposure to simulated body fluid. This has not been correlated to human clinical use. The results of Bioactive glass testing were compared against those from BioSphere Putty (K122868). Additionally, BioSphere Flex was determined to have the same component composition (45S5 bioactive glass, collagen, hyaluronic acid), same porous particle size, same carrier (porous collagen/hyaluronic acid), and a similar bioactive glass amount as Kinex Bioactive (K130392).
The ability of the device to support new bone formation was confirmed through in vivo testing in a critical-sized sheep femoral defect. In vivo testing, including x-ray and micro-CT
{5}------------------------------------------------
radiographic analysis, histology, and histomorphometry, demonstrated that BioSphere Flex had the same healing characteristics as the BioSphere Putty predicate (K122868).
## Conclusions
BioSphere Flex is as safe and effective as the predicate devices. BioSphere Flex has the same intended use and similar indications for use, technological characteristics, and principles of operation as its predicate devices. The minor differences in indications do not alter the intended therapeutic/surgical use of the device and do not affect its safety or effectiveness when used as labeled. Similarly, the minor technological differences between BioSphere Flex and the primary predicate, Biosphere Putty, raise no new issues of safety or effectiveness and are supported by similarities to the Kinex Bioactive reference device. Performance data further demonstrate that BioSphere Flex is as safe and effective as the identified predicates. Thus, BioSphere Flex is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.