K173420 · Microsoft Corp. · LLZ · Dec 27, 2017 · Radiology
Device Facts
Record ID
K173420
Device Name
Radiomics App v1.0
Applicant
Microsoft Corp.
Product Code
LLZ · Radiology
Decision Date
Dec 27, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Organ and tumor contouring
—
—
—
—
—
—
—
Indications for Use
Microsoft Radiomics App v1.0 is a software-only medical device intended for use by trained radiation oncologists, dosimetrists and physicists to derive optimal organ and tumor contours for input to radiation treatment planning. Supported image modalities are Computed Tomography and Magnetic Resonance. Radiomics App assists in the following scenarios: Load, save and display of medical images and contours for treatment evaluation and treatment planning. Creation, transformation, and modification of contours for applications including, but not limited to: transferring contours to radiotherapy treatment planning systems, aiding adaptive therapy, and archiving contours for patient follow-up. Localization and definition of both solid tumors and healthy anatomical structures. Fusion display of compatible images for treatment planning. Three-dimensional rendering of medical images and the segmented contours. Images reviewed using the Radiomics App software should not be used for primary image interpretations. Radiomics App is not for use with digital mammography.
Device Story
Software-only device for radiation treatment planning; integrates into clinical networks to receive CT and MR scans. Assists clinicians in segmenting solid tumors and healthy anatomical structures. Features include manual assisted contouring, machine-learning-based automatic contouring, and contour refinement tools. Outputs include 3D contours and rendered images for export to radiotherapy treatment planning systems. Used by radiation oncologists, dosimetrists, and physicists in clinical settings to aid adaptive therapy and patient follow-up. Benefits include optimized contouring workflows for treatment planning.
Clinical Evidence
No clinical data. Substantial equivalence supported by software verification and validation testing, including manual and automated verification, and validation of measurements, volumetric rendering, multi-planar reconstruction, and assisted/automatic contouring functions.
Technological Characteristics
Software-only device; operates on Microsoft Windows. Functions include assisted and automatic contouring, 2D distance measurement, tissue density calculation, 3D rendering, and CT/MR image fusion. Software level of concern: moderate.
Indications for Use
Indicated for trained radiation oncologists, dosimetrists, and physicists to derive organ and tumor contours for radiation treatment planning using CT and MR images. Not for primary image interpretation or digital mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
December 27, 2017
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Microsoft Corp. % Donna-Bea Tillman, Ph.D. Senior Consultant Biologics Consulting Group, Inc. 1555 King Street. Suite 300 ALEXANDRIA VA 22314
Re: K173420
Trade/Device Name: Radiomics App v1.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 31, 2017 Received: November 1, 2017
Dear Dr. Tillman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and
Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173420
Device Name Radiomics App v1.0
### Indications for Use (Describe)
Microsoft Radiomics App v1.0 is a software-only medical device intended for use by trained radiation oncologists, dosimetrists and physicists to derive optimal organ and tumor contours for input to radiation treatment planning. Supported image modalities are Computed Tomography and Magnetic Resonance. Radiomics App assists in the following scenarios:
· Load, save and display of medical images and contours for treatment evaluation and treatment planning.
· Creation, transformation, and modification of contours for applications including, but not limited to: transferring contours to radiotherapy treatment planning systems, aiding adaptive therapy, and archiving contours for patient followup.
· Localization and definition of both solid tumors and healthy anatomical structures.
- · Fusion display of compatible images for treatment planning.
· Three-dimensional rendering of medical images and the segmented contours.
Images reviewed using the Radiomics App software should not be used for primary image interpretations. Radiomics App is not for use with digital mammography.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
#### 1. SUBMITTER
| Submitter: | Microsoft Corp<br>1 Microsoft Way<br>Redmond, WA 98052<br>Phone: 425-538-9419 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Ivan Tarapov<br>Ivan.Tarapov@microsoft.com |
| Submission Correspondent: | Donna-Bea Tillman, Ph.D, MPA<br>Biologics Consulting Group, Inc.<br>1555 King Street, Suite 300<br>Alexandria, VA 22314<br>Phone: 410-531-6542<br>dtillman@biologicsconsulting.com |
| Date Prepared: | October 31, 2017 |
#### 2. DEVICE
| Name of Device: | Radiomics App v1.0 |
|-----------------------|----------------------------------------------------------------|
| Common or Usual Name: | Radiological Image Processing System |
| Classification Name: | Picture Archiving and Communications System<br>21 CFR 892.2050 |
| Regulatory Class: | Class II |
| Product Code: | LLZ |
#### 3. PREDICATE DEVICE
| Predicate Device Name: | MIM 5.2 (BRACHY) |
|------------------------|----------------------------------------------------|
| Manufacturer: | MIM Software Inc. |
| 510(k) Number: | K103576 |
| Reference Devices: | No reference devices were used in this submission. |
### DEVICE DESCRIPTION 4.
Microsoft Radiomics App v1.0 is a software-only medical device intended for use by trained radiation oncologists, dosimetrists and medical physicists for radiation treatment planning.
Radiomics App stems from more than eight years of research in computerized medical image analysis, computer vision and machine learning. It applies well tested, state-of-the-art algorithms for the assisted delineation of anatomical structures of interest in three-dimensional, clinical radiological scans.
Radiomics App works on computed tomography (CT) and magnetic resonance (MR) scans, and is designed to contour/delineate both healthy anatomical structures as well as lesions such as solid tumors.
{4}------------------------------------------------
Radiomics App integrates into the clinical data network of radiation therapy treatment centers, receiving data from imaging devices such as CT and MR scanners. The purpose of the tool is to assist the expert user in producing segmentations (three-dimensional contours) of anatomical structures, for both solid tumors and healthy tissue structures. The following segmentation tools are provided:
- Assisted Contouring. This module allows for the manual, user-guided segmentation of . structures of interest in both CT and MR images.
- . Machine-learning based contouring. This module uses machine learning algorithms (ML) to provide an initial segmentation of certain structures of interest automatically. The user has the option to accept this initial segmentation or edit and refine it.
- . Contour refinement. This module allows the user to edit and improve segmentations created by either the machine learning or the assisted contouring algorithms.
These segmentations are then exported back into the clinical data network, and subsequently utilized in a radiotherapy treatment planning system to generate a treatment plan for a patient.
### ട്. INDICATION FOR USE
Microsoft Radiomics App v1.0 is a software-only medical device intended for use by trained radiation oncologists, dosimetrists and physicists to derive optimal organ and tumor contours for input to radiation treatment planning. Supported image modalities are Computed Tomography and Magnetic Resonance. Radiomics App assists in the following scenarios:
- Load, save and display of medical images and contours for treatment evaluation and treatment planning.
- . Creation, transformation, and modification of contours for applications including, but not limited to: transferring contours to radiotherapy treatment planning systems, aiding adaptive therapy, and archiving contours for patient follow-up.
- Localization and definition of both solid tumors and healthy anatomical structures.
- . Fusion display of compatible images for treatment planning.
- Three-dimensional rendering of medical images and the segmented contours.
Images reviewed using the Radiomics App software should not be used for primary image interpretations.
Radiomics App is not for use with digital mammography.
## Comparison of Indications for Use
Both devices are only intended for use by trained medical professionals. As noted below, both devices are intended for use during radiation therapy planning. The subject device supports a subset of the image modalities supported by the predicate device. Both devices accept DICOM images generated elsewhere as inputs. Both devices provide the user with tools to contour medical images for use in radiation therapy planning and patient follow-up. Both devices can be used to localize and define normal tissues as well as solid tumors. Both devices support the fusion of different image modalities. Both devices provide three-dimensional image rendering. The subject device is not intended for primary image interpretation or for use with digital mammography. This is a subset of the indications of the predicate device.
{5}------------------------------------------------
The minor differences in indications of use between the subject Radiomics App and the predicate MIM software do not constitute a different intended use.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS 6. WITH THE PREDICATE DEVICE
| | Proposed Device | Predicate Device |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | TBD | K103576 |
| Applicant | Microsoft | MIM Software |
| Device Name | Radiomics App | MIM 5.2 |
| Classification<br>Regulation | 892.2050 - Picture Archiving and<br>Communication System | 892.2050 – Picture Archiving and<br>Communication System |
| Product Code | LLZ | LLZ |
| Indications for Use | Microsoft Radiomics Advanced<br>Image Contouring v1.0 (Radiomics<br>App) is a software-only medical<br>device intended for use by trained<br>radiation oncologists, dosimetrists<br>and physicists to derive optimal organ<br>and tumor contours for input to<br>radiation treatment planning.<br>Supported image modalities are<br>Computed Tomography and Magnetic<br>Resonance. Radiomics App assists in<br>the following scenarios:<br>Load, save and display of medical<br>images and contours for treatment<br>evaluation and treatment planning.<br>Creation, transformation, and<br>modification of contours for<br>applications including, but not limited<br>to: transferring contours to<br>radiotherapy treatment planning<br>systems, aiding adaptive therapy, and<br>archiving contours for patient follow-<br>up.<br>Localization and definition of both<br>solid tumors and healthy anatomical<br>structures.<br>Fusion display of compatible images<br>for treatment planning.<br>Three-dimensional rendering of<br>medical images and the segmented<br>contours.<br>Images reviewed using the Radiomics | MIM 5.2 software is used by trained<br>medical professionals as a tool to aid in<br>evaluation and information management<br>of digital medical images. The medical<br>image modalities include, but are not<br>limited to, CT, MRI, CR, DX, MG, US,<br>SPECT, PET and XA as supported by<br>ACRINEMA DICOM 3.0. MIM 5.2<br>assists in the following indications<br>Receive, transmit, store, retrieve, display,<br>print, and process medical images and<br>DICOM objects.<br>Create, display and print reports from<br>medical images.<br>Registration, fusion display, and review<br>of medical images for diagnosis,<br>treatment evaluation, and treatment<br>planning.<br>Evaluation of cardiac left ventricular<br>function and perfusion, including left<br>ventricular end-diastolic volume, end-<br>systolic volume, and ejection fraction.<br>Localization and definition of objects<br>such as tumors and normal tissues in<br>medical images.<br>Creation, transformation, and<br>modification of contours for applications<br>including, but not limited to, quantitative<br>analysis, aiding adaptive therapy,<br>transferring contours to radiation therapy<br>treatment planning systems, and<br>archiving contours for patient follow-up. |
| | Proposed Device | Predicate Device |
| | App software should not be used for<br>primary image interpretations.<br>Radiomics App is not for use with<br>digital mammography. | and management.<br>Quantitative and statistical analysis of<br>PET/SPECT brain scans by comparing to<br>other registered PET/SPECT brain scans. |
| | | Planning and evaluation of permanent<br>implant brachytherapy procedures.<br>Lossy compressed mammographic<br>images and digitized film screen images<br>must not be reviewed for primary image<br>interpretations. Images that are printed to<br>film must be printed using a FDA-<br>approved printer for the diagnosis of<br>digital mammography images.<br>Mammographic images must be viewed<br>on a display system that has been cleared<br>by the FDA for the diagnosis of digital<br>mammography images. The software is<br>not to be used for mammography CAD. |
| Intended users | Healthcare providers | Healthcare providers |
| Contouring modes | Assisted and automatic | Assisted and automatic |
| Types of tissue<br>contoured | Tumors and normal tissues | Tumors and normal tissues |
| Measurements | 2D distance measurement, average<br>tissue density within a region (for<br>CT), segmentation volume | 2D distance measurement, segmentation<br>volume, average tissue density within a<br>region (for CT) |
| Image Fusion | Fuse only two 3D images, CT and<br>MR | Fuse medical images from multiple<br>modalities |
| 3D image<br>rendering | Yes | Yes |
| Cardiac<br>applications | No | Yes |
| Dose visualization<br>and manipulation | No | Yes |
| Image modalities | CT and MR | CT, MRI, CR, DX, MG, US, SPECT,<br>PET and XA |
| Platform | Stand-alone package which operates<br>on Microsoft Windows operating<br>systems only. | Stand-alone package which operates on<br>both Windows and Mac computer<br>systems. |
The table below provides a side-by-same comparison of the technological characteristics of the proposed Radiomics App and the predicate MIM Software.
{6}------------------------------------------------
### 7. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility
{7}------------------------------------------------
The subject device is a software-only device. There are no direct patient-contacting components of the subject device. Therefore, patient contact information is not needed for this device.
## Electrical Safety and electromagnetic compatibility (EMC)
Not applicable. The subject device is a software-only device. It contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a moderate level of concern.
## Verification Testing
Two different types of verification testing were conducted to verify the software requirements: Manual and Automated. All tests passed, demonstrating that the software performance is in accordance with the stated software requirements.
## Validation Testing
Validation testing of the following functions of the Radiomics App demonstrated that the software meets user needs and intended uses and to support substantial equivalence:
- Measurements - Validation Test
- Volumetric Rendering - Validation Test
- Multi-Planar Reconstruction Validation Test ●
- Assisted Contouring Validation Test ●
- Automatic Contouring - Validation Test
## Animal Study
Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.
## Clinical Study
Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.
#### CONCLUSIONS 8.
The minor differences in indications of use between the subject Radiomics App and the predicate MIM software do not constitute a different intended use. The technological characteristics of the Radiomics App are similar to those of the MIM Software, with the biggest differences being that the Radiomics App offers a subset of the features provided by the MIM software. Results of software verification and validation testing demonstrate that the Radiomics App performs in accordance with specifications and that the performance is comparable to that of the predicate device. Therefore, the Radiomics App can be found to be substantially equivalent to the predicate MIM Software.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.