The NeoCoil Wireless Audio System is intended to provide entertainment and facilitate patient communication in MRI environments. The product is not intended for medical diagnosis or treatment. The product is intended for "MR Conditional" use in MRI environments at 3 Tesla and below. Technologist control units are intended to be used outside the MRI scan room. Wireless receivers are intended for use outside the imaging field of view.
Device Story
Modular system providing entertainment and patient-operator communication in MRI environments (up to 3T). Components include wireless patient headphones, remote audio data source, and communication infrastructure (transmitters, routers, interface modules). Operated by healthcare professionals (technologists) outside the scan room. System transmits audio signals wirelessly to headphones worn by the patient; facilitates two-way communication and entertainment delivery. Benefits include improved patient comfort and reduced anxiety during scanning. Does not affect MR image quality (SNR/uniformity) or introduce artifacts. Designed for MR conditional use.
Clinical Evidence
Clinical performance testing conducted using a mix of pulse sequences and imaging options (axial, sagittal, coronal) per FDA guidance for MR diagnostic devices. Results showed no adverse events and no adverse impact on MR image production in worst-case environments. Bench testing confirmed NRR ≥29 dBA (ANSI S3.19-1974), RF heating ≤41°C, and compliance with NEMA MS 6-2008 for SNR and uniformity.
Technological Characteristics
Modular wireless audio system for MRI environments. Components: headphones, transmitters, routers, interface modules. MR Conditional (≤3T). Connectivity: wireless, wired, or pneumatic. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-6/62366 (usability), IEC 60601-2-33 (MR safety), ISO 10993-1 (biocompatibility), NEMA MS6 (coil characterization).
Indications for Use
Indicated for patients undergoing MRI procedures at 3 Tesla field strength or below to provide entertainment and facilitate communication with the operator.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Patient Communication and Entertainment System (K133670)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
February 16, 2018
NeoCoil, LLC Michael Leigh Director. Regulatory Affairs N27 W23910A Paul Rd. Pewaukee, Wisconsin 53072
Re: K173409
Trade/Device Name: Wireless Audio System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: January 22, 2018 Received: January 24, 2018
Dear Michael Leigh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
# Page 2 - Michael Leigh
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K173409
Device Name NeoCoil Wireless Audio System
Indications for Use (Describe)
The NeoCoil Wireless Audio System is intended to provide entertainment and facilitate patient communication in MRI environments at 3 Tesla field strength and below.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. Traditional 510(k) Summary
# 5.1. Applicant
NeoCoil, LLC N27 W23910A Paul Rd Pewaukee, WI 53072 USA
### 5.2. Contact
Michael Leigh Director, Regulatory 262-347-1250 (office) 261-347-1251 (fax) mike.leigh@neocoil.com
### 5.3. Preparation Date
### 5.4. Name of Device
●
●
- Proprietary Name: ● Common Name:
Classification:
NeoCoil Wireless Audio System
- Nuclear Magnetic Resonance System 21 CFR 892.1000, Product Code LNH
- 5.5. Model Numbers
| NeoCoil Model Number | NeoCoil Model Name |
|----------------------|--------------------------------|
| NC069201 | Technologist Console Interface |
| NC079202 | Penetration Panel Transmitter |
| NC071200 | Wireless Patient Headphones |
| NC075200 | Wireless Audio Interface |
| NC107200 | S1 Audio Interface Module |
| NC130200 | S2 Audio Interface Module |
| NC130202 | P2 Audio Interface Module |
| NC126200 | Scan Room Module |
| NC126201 | Wireless Router |
| NC126202 | Technologist Console Interface |
### 5.6. Device Description
The NeoCoil Wireless Audio System is intended to provide entertainment and facilitate communications between the patient and the operator in a Magnetic Resonance Imaging (MRI) scanner environment. The Wireless Audio System is intended to be used by healthcare professionals.
The NeoCoil Wireless Audio System is a modular system comprised of wireless patient headphones, a remote audio data source unit, and communications infrastructure.
### 5.7. Predicate Device
- . Patient Communication and Entertainment System , K133670, as cleared on 07/24/2014
### 5.8. Comparison to Predicate
The NeoCoil Wireless Audio System is similar in physical, performance, design and material characteristics to the legally marketed device Patient Communication and Entertainment System, K133670, as cleared on 07/24/2014.
{4}------------------------------------------------
Use of the device in conjunction with an MRI scanner is similar.
The NeoCoil Wireless Audio System includes the following differences to the predicate.
- Improved Noise Reduction Rating to 29dBA ●
- Optional Wireless Audio Interface for use with a head coil
- Additional installation options for wired, pneumatic, or wireless communication with ● the control room
Clinical testing demonstrates that use of the NeoCoil Wireless Audio System does not affect the safety and/or the effectiveness of the device when used as labeled.
#### 5.9. Indications for Use
The NeoCoil Wireless Audio System is intended to provide entertainment and facilitate patient communication in MRI environments at 3 Tesla field strength and below.
#### 5.10. Intended Use
The NeoCoil Wireless Audio System is intended to provide entertainment and facilitate patient communication in MRI environments. The product is not intended for medical diagnosis or treatment.
The product is intended for "MR Conditional" use in MRI environments at 3 Tesla and below. Technologist control units are intended to be used outside the MRI scan room.
Wireless receivers are intended for use outside the imaging field of view.
#### 5.11. Testing
The following data has been submitted, referenced or relied on to demonstrate that the NeoCoil Wireless Audio System is safe and effective. The device's performance meets the requirements of pre-defined acceptance criteria and intended uses.
| Test | Pass/Fail Criteria | Result |
|----------------------------------------------------------------|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Max B1 in first fault<br>conditions | Pre-defined<br>performance standards | PASS: Wireless Audio System does not arc or<br>show any signs of voltage breakdown. |
| Surface Temperature<br>in normal and first<br>fault conditions | Pre-defined<br>performance standards | PASS: RF heating is not greater than 41° C. |
| NEMA MS 6-2008 | Pre-defined<br>performance standards | PASS: Wireless Audio System does not<br>adversely impact MR image SNR and<br>Uniformity |
| Coherent Noise Test | Pre-defined<br>performance standards | PASS: Wireless Audio System does not<br>introduce image artifacts noise in center<br>frequency range of compatible field strengths. |
| Noise Reduction | Pre-defined<br>performance standards | PASS: ≥29 dBA NRR (ANSI S3.19-1974) |
| Quality of Service<br>and Coexistence<br>Test | Pre-defined<br>performance standards | PASS: Wireless Audio System provides<br>adequate quality of service for patient<br>communication during MRI scanning. |
### Performance Testing - Bench:
### Published Standards Testing:
{5}------------------------------------------------
The NeoCoil Wireless Audio System has been evaluated to the following standards:
| Standard | Purpose |
|----------------|---------------------------------------------------------------------------------------|
| IEC 60601-1 | Electromechanical safety |
| IEC 60601-1-2 | Electromagnetic Compatibility |
| IEC 60601-1-6 | Usability |
| IEC 60601-2-33 | Electromechanical safety for magnetic resonance equipment |
| IEC 62366 | Guidance on the application of usability engineering to medical devices |
| ISO 10993-1 | Biocompatibility |
| NEMA MS6 | Characterization of Special Purpose Coils for Diagnostic Magnetic Resonance<br>Images |
### Performance Testing - Clinical:
Clinical data submitted exhibits a mix of pulse sequences and imaging options in the axial, sagittal and coronal planes as recommended in the FDA guidance, Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, issued November 18, 2016.
No adverse events were reported during clinical performance testing; the NeoCoil Wireless Audio System does not adversely affect MR image production in the worst-case environment.
#### 5.12. Conclusion
This submission demonstrates that the Indications for Use associated with the NeoCoil Wireless Audio System are as safe and effective as the predicate device, Patient Communication and Entertainment System, K133670, as cleared on 07/24/2014. As such, the NeoCoil Wireless Audio System is equivalent to its predicate, Patient Communication and Entertainment System, K133670, as cleared on 07/24/2014.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.