ORISE Tissue Retractor System

K173400 · Boston Scientific · FED · Feb 9, 2018 · Gastroenterology, Urology

Device Facts

Record IDK173400
Device NameORISE Tissue Retractor System
ApplicantBoston Scientific
Product CodeFED · Gastroenterology, Urology
Decision DateFeb 9, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This device is indicated for use in conjunction with an endoscope in the gastrointestinal tract for tissue or foreign body manipulation and/or where multiple removal and insertions of an endoscope are required.

Device Story

ORISE Tissue Retractor System facilitates endoscopic procedures in the gastrointestinal tract; provides stable environment via deployment of retractor arms. System components include ORISE Tissue Retractor (OTR) with handle, flexible shaft, expandable distal chamber; and ORISE Instrument Guides (OIGs) with 45°, 60°, or 90° tip angles. Operated by gastroenterologist or surgeon; used in conjunction with endoscope. OIGs act as flexible conduits for secondary instruments (e.g., graspers) to perform tasks like grasping, retracting, dissecting, cutting, and coagulating tissue. Shaft markers aid in retractor placement. System enhances procedural efficiency by allowing instrument access without repeated endoscope removal/insertion.

Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included dimensional analysis, tensile strength, simulated use, and compatibility testing (OTR/OIG). Biocompatibility testing included cytotoxicity, sensitization, intracutaneous irritation, material-mediated pyrogen, and chemical characterization. All tests met specifications.

Technological Characteristics

System comprises OTR (handle, flexible shaft, expandable distal chamber) and OIGs (flexible conduits). OIGs available in 45°, 60°, 90° tip configurations. Mechanical device; no energy source. Biocompatibility evaluated per EN ISO 10993-1:2009. Non-software device.

Indications for Use

Indicated for patients requiring gastrointestinal tissue or foreign body manipulation or multiple endoscope insertions/removals.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 9, 2018 Boston Scientific Corporation Thomas Hirte Senior Manager, Regulatory Affairs 100 Boston Scientific Way Marlborough, MA 01752 Re: K173400 Trade/Device Name: ORISE™ Tissue Retractor System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FED. ODB. GAD Dated: January 11, 2018 Received: January 12, 2018 Dear Thomas Hirte: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Charles Viviano -S For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### SECTION 4 INDICATIONS FOR USE STATEMENT | 510(k) Number (if known): | K173400 | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | ORISE™ Tissue Retractor System | | Indications For Use: | This device is indicated for use in conjunction with an<br>endoscope in the gastrointestinal tract for tissue or foreign<br>body manipulation and/or where multiple removal and<br>insertions of an endoscope are required. | Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ ## SECTION 5 510(k) SUMMARY ### 510(k) SUMMARY #### 1. Submitter: | Emitter: | Boston Scientific Corporation<br>100 Boston Scientific Way<br>Marlborough, MA 01752<br>Telephone: 508-683-4454<br>Fax: 508-683-5939 | |----------|-------------------------------------------------------------------------------------------------------------------------------------| |----------|-------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Thomas Hirte | |----------------|------------------------------------| | | Senior Manager, Regulatory Affairs | | Date Prepared: | October 30, 2017 | #### 2. Proposed Device: | Trade Name: | ORISE™ Tissue Retractor System | |----------------------|--------------------------------| | Classification Name: | Endoscope and Accessories | | Regulation Number: | 876.1500 | | Product Code: | FED, ODB, GAD | | Classification: | Class II | ### 3. Predicate Device: | Trade Name: | LumenR Cannula Retractor System | |----------------------|---------------------------------| | 510(k) Number: | K153698 | | Classification Name: | Endoscope and Accessories | | Regulation Number: | 876.1500 | | Product Code: | FED, ODB, GAD | | Classification: | Class II | ## 4. Proposed Device Description: The ORISE Tissue Retractor System is comprised of the ORISE Tissue Retractor (OTR) consisting of the handle, the flexible shaft with expandable distal end (chamber) and the associated instrument guides called the ORISE Instrument Guides (OIGs) which are flexible conduits, guiding other instruments such as graspers. The OIG is available in three (3) tip configurations; with 45°, 60° and 90° tip bend angles. Shaft Markers are used for reference purposes for aid in placing the retractor. The proposed ORISE™ Tissue Retractor System has the same design intent and meets the same performance requirements as the predicate LumenR Cannula Retractor System. Minor design modifications and material modifications have been made to enhance the manufacturability. {4}------------------------------------------------ # 5. Indications for Use: This device is indicated for use in conjunction with an endoscope in the gastrointestinal tract for tissue or foreign body manipulation and/or where multiple removal and insertions of an endoscope are required. # 6. Technological Characteristics: The ORISE™ Tissue Retractor System is a device that can accommodate an endoscope and is used to create a stable environment in the GI tract by the deployment of retractor arms. The system allows the Gastroenterologist or surgeon to perform surgical tasks such as to grasp. retract, dissect, cut and coagulate tissue. ## 7. Performance Data: A series of performance tests were conducted which demonstrated the quantitative mechanical performance, tolerance and usability of the ORISE™ Tissue Retractor System in endoscopic procedures. ## Bench Testing: Functional and performance tests were performed on the proposed ORISE™ Tissue Retractor system to demonstrate substantial equivalence and to satisfy all design verification requirements. The ORISE™ Tissue Retractor System passed all tests. In-vitro testing that has been performed and all components, subassemblies, and/or full devices met the required specifications. The proposed ORISE™ Tissue Retractor System met the same performance requirements as the predicate LumenR Cannula Retractor System in order to establish substantial equivalence in performance, in the following tests: OTR and OIG dimensional testing, OTR and OIG performance testing including simulated use testing, Tensile strength testing, OTR scope compatibility, OIG instrument compatibility and OTR/OIG compatibility. ## Biocompatibility Testing: The proposed ORISE™ Tissue Retractor System was evaluated in accordance with EN ISO 10993-1: 2009. The following tests were performed: Cytotoxicity, Sensitization, Intracutaneous Irritation, Material Mediated Pyrogen, Chemical Characterization and Analysis. ## 8. Conclusion: All biocompatibility tests conducted on the ORISE™ Tissue Retractor System passed. Therefore, the System is considered biocompatible for its intended use. All device bench and biocompatibility test results were acceptable. The data demonstrate that the System sufficiently meets the design specifications and is suitable for the intended use. Boston Scientific Corporation has demonstrated that the proposed ORISE™ Tissue Retractor System is substantially equivalent to the currently marketed LumenR Cannula Retractor System (K153698) and can be safely and effectively used for its proposed indication.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...