The TSV™ BellaTek® Encode® Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.
Device Story
TSV™ BellaTek® Encode® Healing Abutments are dental implant accessories used during the healing phase post-implant placement. They facilitate soft tissue contouring to create an emergence profile for final prostheses. The device features machined markings on the occlusal surface (Encode® Coding Scheme) that identify the mating implant's position and orientation. Clinicians use these markings during abutment-level impressions or intraoral scans to capture digital data for restorative planning. The device is used in a clinical setting by dental professionals. By providing a standardized coding system, the abutment simplifies the transition from the healing phase to the final restoration, potentially improving the accuracy of the final prosthesis fit and reducing chair time.
Clinical Evidence
No clinical data were included in this submission. Bench testing included radiation sterilization validation (ISO 11137-1/2), LAL testing (ANSI/AAMI ST 72), and biocompatibility, shelf-life, and MR evaluation by reference to predicate K170013.
Technological Characteristics
Material: Ti-6Al-4V ELI. Design: Internal connection healing abutment with machined occlusal markings for identification. Sterilization: Radiation (SAL 10^-6). MR conditional. Compatible with Tapered Screw-Vent and Trabecular Metal implant systems.
Indications for Use
Indicated for patients requiring endosseous dental implant healing; used as an accessory to prepare gingival tissue for final abutment and restoration.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Zimmer Dental Trabecular Metal Implant System (K132258)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 13, 2017
Biomet 3i % Linda Schulz Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K173374
Trade/Device Name: TSV™ BellaTek® Encode® Healing Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: October 26, 2017 Received: October 27, 2017
Dear Linda Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
# K173374
Device Name
TSVTM BellaTek® Encode® Healing Abutments
#### Indications for Use (Describe)
The TSV™ BellaTek® Encode® Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments reqarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# 510(k) Summary K173374 Biomet 3i TSV™ BellaTek® Encode® Healing Abutments
October 26, 2017
### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Biomet 3i<br>4555 Riverside Drive<br>Palm Beach Gardens, FL 33410<br>Telephone: +1 (561) 776-6700<br>Fax: +1 (561) 776-1272 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Nichole O'Neil, Regulatory Affairs Specialist |
| Representative/Consultant | Linda K. Schulz, BSDH, RDH<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 (858) 792-1235<br>Fax: +1 (858) 792-1236<br>Email: lschulz@paxmed.com<br>flarson@paxmed.com |
# DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | TSVTM BellaTek® Encode® Healing Abutments |
|----------------------------|-------------------------------------------|
| Common Name | Dental Implant Abutment |
| Classification Name | Endosseous dental implant abutment |
| Classification Regulations | 21 CFR 872.3630, Class II |
| Product Code | NHA |
| Classification Panel: | Dental Products Panel |
| Reviewing Branch: | Dental Devices Branch |
# PREDICATE DEVICE INFORMATION
| Primary Predicate: | | |
|--------------------|--------------------------------------------------------|---------------------|
| K170013 | Eztetic™ BellaTek® Encode® Healing Abutments | Biomet 3i |
| | Reference Predicates for Compatibility: | |
| K061410 | Zimmer Dental Implant System | Zimmer Dental, Inc. |
| K133339 | Zimmer Dental Tapered Screw-Vent® T Implant, HA Coated | Zimmer Dental, Inc. |
| | Zimmer Dental Tapered Screw-Vent® M Implant, HA Coated | |
| K113753 | Tapered Screw-Vent® X Implant | Zimmer Dental, Inc. |
| K132258 | Zimmer Dental Trabecular Metal Implant System | Zimmer Dental, Inc. |
{4}------------------------------------------------
510(k) Summary Page 2 of 4
### INDICATIONS FOR USE
The TSV™ BellaTek® Encode® Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.
# DEVICE DESCRIPTION
The purpose of this submission is to obtain 510(k) premarket clearance for the TSVIM BellaTek® Encode® Healing Abutments compatible with Tapered Screw-Vent® and Trabecular Metal implant systems. Abutments sizing is based on implant platform diameter.
TSV™ BellaTek® Encode® Healing Abutments are designed to aid in soft tissue contouring during the healing period after implant placement, creating an emergence profile for the final prosthesis. They have the added design feature of machined markings for identification when taking an abutment level impression or an intraoral scan/digital impression. The occlusal surface of the device include markings that provide information about the mating implant's position and orientation.
The principal of operation and Encode "Coding Scheme of the same as the primary predicate device. The pattern of the markings for the subject device is specific to the Tapered Screw-Vent", and Trabecular Metal implant lines.
| Compatible Implant Sizes | | | |
|--------------------------|---------------------------|---------------------------|-------------------------|
| Implant System | Implant Diameter<br>(mmD) | Implant Platform<br>(mmD) | Implant Length<br>(mmL) |
| Tapered Screw -Vent® | 3.7 | 3.5 | 8, 10, 11.5, 13,16 |
| Tapered Screw -Vent® | 4.1 | 3.5 | 8, 10, 11.5, 13,16 |
| Tapered Screw -Vent® | 4.7 | 4.5 | 8, 10, 11.5, 13,16 |
| Tapered Screw -Vent® | 6.0 | 5.7 | 8, 10, 11.5, 13,16 |
| Screw-Vent® | 3.3 | 3.5 | 8,10,13,16 |
| Screw-Vent® | 3.7 | 3.5 | 8,10,13,16 |
| Screw-Vent® | 4.7 | 4.5 | 8,10,13,16 |
| Trabecular Metal™ | 3.7 | 3.5 | 10, 11.5, 13 |
| Trabecular Metal™ | 4.1 | 3.5 | 10, 11.5, 13 |
| Trabecular Metal™ | 4.7 | 4.5 | 10, 11.5, 13 |
| Trabecular Metal™ | 6.0 | 5.7 | 10, 11.5, 13 |
Compatible Implant Sizes
# PERFORMANCE DATA
Non-clinical testing data submitted or relied upon to demonstrate substantial equivalence included radiation sterilization validation according to ISO 11137-1 and 11137-2, demonstrating assurance level (SAL) of 10 °, Limulus amebocyte lysate (LAL) testing according to AANSI/AAMI ST 72, biological evaluation according to ISO 10993-1 by reference to K170013, demonstrating acceptable biocompatibility, accelerated and real time aging studies by reference to K170013, demonstrating a shelf life of five years, and MR evaluation by reference to K170013. The subject devices are labeled as MR conditional.
No clinical data were included in this submission.
{5}------------------------------------------------
# EQUIVALENCE TO MARKETED DEVICE
Biomet 3i submits the information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the subject device, TSV™ BellaTek® Encode® Healing Abutments, is substantially equivalent in indications and design principles to the legally marketed primary predicate device.
| Comparison | Subject Device | Primary Predicate Device |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | TSVTM BellaTek® Encode® Healing<br>Abutments | K170013<br>The EzteticTM BellaTek® Encode®<br>Healing Abutments |
| | Biomet 3i | Biomet 3i |
| Indications for Use | The TSVTM BellaTek® Encode®<br>Healing Abutments are intended for use<br>as an accessory to endosseous dental<br>implants during endosseous and<br>gingival healing to prepare gingival<br>tissue for acceptance of a final<br>abutment and restoration. | The EzteticTM BellaTek® Encode®<br>Healing Abutments are intended for use<br>as an accessory to endosseous dental<br>implants during endosseous and gingival<br>healing to prepare gingival tissue for<br>acceptance of a final abutment and<br>restoration. |
| Design | | |
| Scanning Identification | Encode® Coding Scheme<br>(Machined Markings) | Encode® Coding Scheme<br>(Machined Markings) |
| Emergence Profile (mm) | 3.8, 5.0, 5.6, 6.0, 6.8 | 3.8, 5.0 |
| Platform Diameter (mm) | 3.5, 4.5, 5.7 | 2.9 |
| Gingival Height (mm) | 3, 5, 7 | 3, 4, 6, 8 |
| Implant/Abutment<br>Connection | Internal | Internal |
| Material | | |
| Abutment | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Screw | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
The intended use of the subject device as a healing abutment machined markings for identification is the same as the primary predicate K170013. The Indications for the subject device is the same as the primary predicate Eztetic™ BellaTek® Encode® Healing Abutment, K170013, changing only the name.
The principal of operation and basic design of the Encode Coding Scheme of the subject device is the same as the primary predicate device, Eztetic™ BellaTek® Encode® Healing Abutment, K170013. The pattern of the markings for the subject device is specific to the Tapered Screw-Vent, and Trabecular Metal implant lines.
All of the subject device components are manufactured from the same facilities, using the same manufacturing processes as used for the previously cleared primary predicate device in K170013.
{6}------------------------------------------------
# CONCLUSION
The subject device and the primary predicate device have intended use, have same technological characteristics, and are made of the same materials. The subject device and predicate devices encompass the same design principals, including the Encode® Coding Scheme. The subject and predicate devices are packaged in the same materials and are to be sterilized using the same methods.
The data included in this submission demonstrate substantial equivalence to the primary predicate device listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.