DSM Biomedical Calcium Phosphate Cement is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities and pelvis). These defects may be surgically created or osseous defects created from traumatic injury to the bone. The DSM Biomedical Calcium Phosphate Cement is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The DSM Biomedical Calcium Phosphate Cement cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K-Wires, plates, screws) to help support bone fragments during the surgical procedure. The cured cement acts only as a temporary support media and is not intended to provide structural support during the healing process. The Calcium Phosphate Cement resorbs and is replaced by bone during the healing process.
Device Story
Injectable, fast-setting bone void filler; composed of calcium phosphate powder and bovine collagen. Mixed with saline, patient blood, or bone marrow in a provided syringe; converts to hydroxyapatite in vivo. Used by surgeons in clinical settings to fill non-structural bony voids or gaps in extremities and pelvis; augments provisional hardware (plates, screws, K-wires) to support bone fragments during procedures. Acts as temporary support media; resorbs and is replaced by bone during healing. Provided sterile via gamma irradiation for single use.
Clinical Evidence
Bench testing, biocompatibility, viral inactivation, and animal study. Biocompatibility per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, hemocompatibility, subacute toxicity, implantation). Ovine bilateral femoral defect animal study compared bone healing to predicate. Results indicate acceptable biocompatibility and performance profile.
Technological Characteristics
Calcium phosphate powder with bovine collagen; converts to hydroxyapatite. Injectable, sculptable, fast-setting. Sterile (SAL 10^-6) via gamma irradiation. Single-use. Complies with ASTM F1185 for hydroxylapatite composition. Packaged in mixing/delivery syringe.
Indications for Use
Indicated for filling bony voids or gaps in the skeletal system (extremities and pelvis) that are not intrinsic to structural stability. Suitable for surgically created or traumatic osseous defects. Can augment provisional hardware (K-wires, plates, screws) for temporary support during surgery. Not for permanent structural support.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The word "Administration" is on the second line.
Kensey Nash Corporation dba DSM Biomedical Susan Pileggi Senior Regulatory Specialist 735 Pennsylvania Drive Exton, Pennsylvania 19431
April 3, 2018
Re: K173362
Trade/Device Name: DSM Biomedical Calcium Phosphate Cement Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV, OIS Dated: February 26, 2018 Received: February 28, 2018
Dear Ms. Pileggi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K173362
Device Name DSM Biomedical Calcium Phosphate Cement
#### Indications for Use (Describe)
DSM Biomedical Calcium Phosphate Cement is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities and pelvis). These defects may be surgically created or osseous defects created from traumatic injury to the bone. The DSM Biomedical Calcium Phosphate Cement is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The DSM Biomedical Calcium Phosphate Cement cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K-Wires, plates, screws) to help support bone fragments during the surgical procedure. The cured cement acts only as a temporary support media and is not intended to provide structural support during the healing process. The Calcium Phosphate Cement resorbs and is replaced by bone during the healing process.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DSM, a global science-based company active in health, nutrition, and materials. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold blue font on the right. The graphic is composed of overlapping, translucent shapes in various colors, including blue, green, orange, and purple, creating a sense of dynamism and innovation.
# 510(k) Summary
| Submitted By: | DSM Biomedical<br>735 Pennsylvania Drive<br>Exton, PA 19341 |
|-----------------|--------------------------------------------------------------------------------|
| Contact Person: | Susan Pileggi<br>susan.pileggi@dsm.com<br>(P) 484-713-2100<br>(F) 484-713-2903 |
| Date Prepared: | February 23, 2018 |
Device:
| Trade Name: | DSM Biomedical Calcium Phosphate Cement |
|----------------------------|-----------------------------------------------------------------------|
| Common/Usual Name: | Bone Void Filler |
| Classification Name: | Resorbable Calcium Salt Bone Void Filler Device |
| Classification Regulation: | 21 CFR 880.3045 |
| Device Class: | II |
| Device Code: | MQV: Filler, Bone Void, Calcium Compound |
| | OIS: Calcium Salt Bone Void Filler, Drillable, Non-screw Augmentation |
| Advisory Panel: | Orthopedic |
Predicate Device: Stryker® Injectable Cement (K060061) [Stryker® CMF]
#### Device Description
DSM Biomedical Calcium Phosphate Cement as an injectable, fast self-setting bone substitute. DSM Biomedical Calcium Phosphate Cement is composed of calcium phosphate which converts to hydroxyapatite in vivo, and bovine collagen powder. The device can also be used to augment provisional hardware to help support bone fragments during the surgical procedure. The cement is provided in a powder form packaged in a mixing syringe. The mixing syringe allows for the combination of saline, the patient's blood, or the patient's bone marrow at the required powder-to-liquid ratio. DSM Biomedical Calcium Phosphate Cement is supplied sterile by gamma irradiation and is non-pyrogenic.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DSM, a global science-based company. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold blue font on the right. The graphic is composed of overlapping curved shapes in various colors, including blue, green, purple, and orange. The overall design is modern and corporate.
#### Indications For Use:
DSM Biomedical Calcium Phosphate Cement is indicated to fill bony voids or gaps of the sketal system (i.e. extremities and pelvis). These defects may be surgically created or osseous defects created from traumatic injury to the bone. The Calcium Phosphate Cement is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The Calcium Phosphate Cement cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K-Wires, plates, screws) to help support bone fragments during the surgical procedure. The cured cement acts only as a temporary support media and is not intended to provide structural support during the healing process. The Calcium Phosphate Cement resorbs and is replaced by bone during the healing process.
## Basis for Substantial Equivalence:
The DSM Biomedical Calcium Phosphate Cement is substantially equivalent in terms of indications for use, material composition, technological characteristics and performance characteristics to that of the predicate device, Stryker® Injectable Cement (K060061).
DSM Biomedical Calcium Phosphate Cement and the predicate device have comparable indications for use. Both are calcium phosphate powders that when mixed with a mixing liquid form a fast setting cement. The addition of the collagen to the subject device does not raise questions of substantial equivalence as the safety of the collagen has been shown through biocompatibility, an animal study, and a viral inactivation The subject and predicate devices are both injectable, and fast setting calcium study. phosphate cement that converts to hydroxyapatite over time. Both the subject and predicate devices are packaged with accessories to aid in the mixing and delivery of the subject and predicate devices are provided sterile and are for single use only. The subject and predicate devices are biocompatible and non-pyrogenic.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the DSM company logo. The logo consists of a colorful abstract design on the left and the letters "DSM" in blue on the right. The abstract design is made up of overlapping shapes in various colors, including blue, green, yellow, and purple. The letters "DSM" are in a bold, sans-serif font.
| Charactestic | DSM Biomedical Calcium Phosphate<br>Cement (Proposed) | Stryker® Injectable Cement<br>(K060061) |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications For<br>Use | DSM Biomedical Calcium Phosphate<br>Cement is indicated to fill bony voids or<br>gaps of the skeletal system (i.e.<br>extremities and pelvis). These defects<br>may be surgically created or osseous<br>defects created from traumatic injury<br>to the bone. The Calcium Phosphate<br>Cement is indicated only for bony voids<br>or gaps that are not intrinsic to the<br>stability of the bony structure. The<br>Calcium Phosphate Cement cured <i>in situ</i><br>provides an open void/gap filler that<br>can augment provisional hardware (e.g.<br>K-Wires, plates, screws) to help support<br>bone fragments during the surgical<br>procedure. The cured cement acts only<br>as a temporary support media and is not<br>intended to provide structural support<br>during the healing process. The<br>Calcium Phosphate Cement resorbs and<br>is replaced by bone during the healing<br>process. | Stryker Injectable Cement is a self-setting<br>calcium phosphate cement indicated to fill<br>bony voids or gaps of the skeletal system<br>(i.e. extremities, craniofacial, spine, and<br>pelvis.) These defects may be surgically<br>created or osseous defects created from<br>traumatic injury to the bone. The Stryker<br>Injectable Cement is indicated only for<br>bony voids or gaps that are not intrinsic to<br>the stability of the bony structure. Stryker<br>Injectable Cement cured <i>in situ</i> provides an open<br>void/gap filler that can augment<br>provisional hardware (e.g. K-Wires,plates,<br>screws) to help support bone fragments<br>during the surgical procedure. The cured<br>cement acts only as a temporary support<br>media and is not intended to provide<br>structural support during the healing<br>process. |
| Material | Calcium phosphate powder with bovine<br>collagen mixed with saline, patient's<br>blood, or patient's bone marrow to form<br>hydroxyapatite | Calcium phosphate powder mixed with<br>mixing liquid to form hydroxyapatite |
| Form | Injectable, sculptable, and fast setting<br>calcium phosphate cement that<br>converts to hydroxyapatite | Injectable, sculptable, and fast setting<br>calcium phosphate cement that converts to<br>hydroxyapatite |
| Packaging | Powder prepackaged in a mixing and<br>delivery syringe, packaged with<br>accessories to aid in the mixing and<br>delivery of cement | Calcium phosphate powder in a bowl,<br>packaged with accessories to aid in the<br>mixing and delivery of the cement |
| Sterilization | Sterile, SAL 10-6 | Sterile, SAL 10-6 |
| Reusable | Single Use Device | Single Use Device |
| Biocompatible | Yes | Yes |
# Performance Data:
Material characterization, including chemical and material physical characterization, bench testing, biocompatibility testing, viral inactivation, and an animal study have been conducted to evaluate the performance characteristics and biological safety of DSM Biomedical Calcium Phosphate Cement.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DSM, a global science-based company active in health, nutrition, and materials. The logo consists of a colorful abstract graphic on the left and the letters "DSM" in bold blue font on the right. The graphic is composed of overlapping shapes in various colors, including blue, green, purple, and orange, forming a circular design with a hexagonal shape in the center.
Material characterization and performance testing of the DSM Biomedical Calcium Phosphate Cement was completed in accordance with the FDA Guidance Document, Resorbable Calcium Salt Bone Void Filler Device and ASTM F1185 Standard Specification for Composition of Hydroxylapatite for Surgical Implants.
An ovine bilateral femoral defect animal study was performed to evaluate bone healing following treatment with the DSM Biomedical Calcium Phosphate Cement compared to the predicate device.
Biocompatibility testing was completed in accordance with the requirements of ISO 10993-1: 2009, Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing within a Risk Management Process for a permanent implant device with tissue/bone contact. Testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, genotoxicity, hemocompatibility, subacute toxicity, and implantation. Other safety testing included a viral inactivation study and residual chemical assessment. Results indicate that the device's biocompatibility profile is acceptable.
## Substantial Equivalence:
DSM Biomedical Calcium Phosphate Cement has the same intended use and similar indications, technological characteristics, and principles of operation as the predicate device. The minor technological differences that exist between the DSM Biomedical Calcium Phosphate Cement do not raise any new issues of safety and effectiveness. The bench, animal, and biocompatibility testing demonstrates that DSM Biomedical Calcium Phosphate Cement is substantially equivalent to the predicate device.
#### Conclusion:
The DSM Biomedical Calcium Phosphate Cement is substantially equivalent in terms of indications for use, material composition, technological characteristics, and performance characteristics to the predicate device, Stryker® Injectable Cement (K060061) [Stryker® CMF].
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.