← Product Code [ORC](/productcode/ORC) · K173349

# Operio (K173349)

_Toul Meditech AB · ORC · Dec 20, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173349

## Device Facts

- **Applicant:** Toul Meditech AB
- **Product Code:** [ORC](/productcode/ORC.md)
- **Decision Date:** Dec 20, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5070
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K173349 · Dec 20, 2017 | Operio | Toul Meditech AB | Swedish hospital clinical patient records/measurements | Clinical measurements of CFU levels at the surgical site and over instruments were collected from 82 patients across various surgical procedures to demonstrate that the device maintains microbial levels below the threshold of < 5 CFU/m3, supporting substantial equivalence. | CFU measurements; Surgical site contamination; Routine clinical practice; Microbiological monitoring |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective analysis of clinical measurements | Patients undergoing ophthalmic, orthopedic, and neurosurgery in Swedish hospitals; Sample Size: 82 patients; Number of Sites: Multiple Swedish Hospitals | Not applicable for this study | CFU levels at the surgical site and over instruments |

## Indications for Use

Operio is a portable device for use in a surgical operating environment that produces a directed, non-turbulent flow of air to the surgical site during ophthalmic, orthopedic and neuro surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m3 at the surgical site and instruments. The air flow is intended to be directed parallel to the surgical site and/or instruments within: 20" (50 cm) in width, 47" (120 cm) in length and 15" (40 cm) in height. Device effectiveness may not be reliably detectable at a distance of 47 inches (120 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

## Device Story

Mobile air-handling unit; recirculates ambient OR air; HEPA filtration; produces directed, non-turbulent airflow; reduces airborne particulates/microorganisms at surgical site/instruments. Used in OR; operated by surgical staff. Device positioned near OR table; uses sterile shield barrier. Provides localized clean air zone (20"x47"x15"); reduces CFU levels to < 5 CFU/m3. Benefits patient by reducing risk of surgical site contamination.

## Clinical Evidence

Clinical testing included 302 CFU measurements from 82 patients in Swedish hospitals. Results demonstrated mean CFU levels of 0.4 CFU/m3 (ophthalmic), 0.52 CFU/m3 (orthopedic), and 2 CFU/m3 (neurosurgery). All results remained below the < 5 CFU/m3 threshold, demonstrating substantial equivalence to the predicate.

## Technological Characteristics

Mobile unit; materials: aluminum, powder-coated steel, ABS plastic, stainless steel. HEPA filtration (>99.995% efficiency at 0.3 µm) using Nelior membrane media; polyamide mesh pre-filter. Airflow: 235 CFM at 0.4 m/s. Power: 120V AC, 60Hz, 290 VA. Sterile accessories (shields) sterilized via gamma irradiation (10^-6 SAL). Standards: IEC/EN 60601-1, IEC/EN 60601-1-2, IEC 61010-1.

## Regulatory Identification

Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

## Predicate Devices

- Operio ([K153498](/device/K153498.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 20, 2017

Toul Meditech AB Tomas Hansson CEO Tunbytorpsgatan 31 Vasteras, 72137 SE

Re: K173349

Trade/Device Name: Operio Regulation Number: 21 CFR 878.5070 Regulation Name: Air-Handling Apparatus for a Surgical Operating Room Regulatory Class: Class II Product Code: ORC Dated: October 25, 2017 Received: October 25, 2017

Dear Tomas Hansson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

#### Page 2 - Tomas Hansson

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K173349

Device Name Operio

#### Indications for Use (Describe)

Operio is a portable device for use in a surgical operating environment that produces a directed, non-turbulent flow of air to the surgical site during ophthalmic, orthopedic and neuro surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m3 at the surgical site and instruments.

The air flow is intended to be directed parallel to the surgical site and/or instruments within: 20" (50 cm) in width, 47" (120 cm) in length and 15" (40 cm) in height. Device effectiveness may not be reliably detectable at a distance of 47 inches (120 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "toul meditech". The word "toul" is in a large, blue, sans-serif font, with the "o" being partially obscured by a blue square. Below "toul" is the word "meditech" in a smaller, light blue, sans-serif font. The logo is simple and modern, and the colors are calming and professional.

## 510(k) Summary

#### Submission information

| Submission date:  | 12th of October 2017                                                                                                                                           |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) submitter: | Tomas Hansson                                                                                                                                                  |
|                   | Toul Meditech AB<br>Tunbytorpsgatan 31<br>721 37 Västerås, Sweden<br>Phone: + 46 21 13 50 00<br>Fax: + 46 21 13 86 45<br>Email: tomas.hansson@toulmeditech.com |

#### Device information

| Trade name:          | Operio                                               |
|----------------------|------------------------------------------------------|
| Common name:         | Air zone unit                                        |
| Classification name: | Air-handling apparatus for a surgical operating room |
| Regulation number:   | 21 CFR 878.5070                                      |
| Product code         | ORC                                                  |
| Device class         | II                                                   |

### Legally marketed predicate device

Identification of the legally marketed predicate device to which Toul Meditech AB claims SE for.

| Code | Manufacturer     | Device | 510(k) # |
|------|------------------|--------|----------|
| ORC  | Toul Meditech AB | Operio | K153498  |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for "toul meditech". The word "toul" is in large, bold, blue letters. Below "toul" is the word "meditech" in smaller, blue letters. The logo is simple and modern.

## Device description

The air zone unit Operio is used in a surgical operating room for cleaning the air in a given area by re-circulating the ambient air and cleaning it from airborne particles. The air is filtered through a HEPA filter and the air is generated over the areas where the demands of clean air are especially high. The air zone unit Operio is equipped with a control panel for adjustments and for positioning of the unit.

The air zone unit Operio is a mobile unit that with the help of castors easily can be moved around the patient to an optimal position or, when needed, transported inside the hospital between wards. It is also equipped with an optional instrument tray.

By using a unique sterile shield as a protective barrier the air zone unit can be placed close to the OR table and deliver HEPA filtered air to the surgical site and instruments to reduce the presence of airborne particulate and microorganisms.

#### Indications for use

Operio is a portable device for use in a surgical operating environment that produces a directed, non-turbulent flow of air to the surgical site during ophthalmic, orthopedic and neuro surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m3 at the surgical site and instruments.

The air flow is intended to be directed parallel to the surgical site and/or instruments within: 20" (50 cm) in width, 47" (120 cm) in length and 15" (40 cm) in height. Device effectiveness may not be reliably detectable at a distance of 47 inches (120 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

### Summary of technological characteristics compared to predicate devices

The difference for the subject device compared to the predicate device is additional and extended claims in the indications. to include neurosurgery and orthopedic surgery. The subject device and predicate device is the same model and identical in technology and characteristics. The use with additional claims do not introduce any new hazards and test results can confirm that the subject device is safe and effective for use.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for "toul meditech". The word "toul" is in large, blue, sans-serif font, with a gradient blue square behind the "o". Below "toul" is the word "meditech" in a smaller, blue, sans-serif font.

### Comparison table:

| Characteristic                                                        | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                       | Operio                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Operio K153498                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Review Panel                                                          | General Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | General Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Indications for<br>use                                                | Operio is a portable device for use in a<br>surgical operating environment that<br>produces a directed, non-turbulent<br>flow of air to the surgical site during<br>ophthalmic, orthopedic and neuro<br>surgery and to the sterile instruments<br>used during surgery. The air flow from<br>the device is HEPA-filtered to reduce<br>the presence of particulate matter to<br>reduce the microorganisms to a level<br>of < 5 CFU per m³ at the surgical site<br>and instruments.<br>The air flow is intended to be directed<br>parallel to the surgical site and/or<br>instruments within: 20" (50 cm) in<br>width, 47" (120 cm) in length and 15"<br>(40 cm) in height. Device effectiveness<br>may not be reliably detectable at a<br>distance of 47 inches (120 cm) from<br>the air flow outlet, and effectiveness<br>depreciates beyond this specified area. | Operio is a portable device for use in a<br>surgical operating room that produces<br>a directed, non-turbulent flow of air to<br>the surgical site during ophthalmic<br>surgery and to the sterile instruments<br>used during surgery. The air flow<br>from the device is HEPA-filtered to<br>reduce the presence of particulate<br>matter to reduce the microorganisms<br>to a level of < 5 CFU per m³ at the<br>surgical site and instruments.<br>The air flow is intended to be directed<br>parallel to the surgical site and/or<br>instruments within: 20" (50 cm) in<br>width, 47" (120 cm) in length and 15"<br>(40 cm) in height. Device<br>effectiveness may not be reliably<br>detectable at a distance of 47 inches<br>(120 cm) from the air flow outlet, and<br>effectiveness depreciates beyond this<br>specified area. |
| Physical<br>Dimension                                                 | 24 x 18 x 47-67 inches                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 24 x 18 x 47-67 inches                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Weight                                                                | 92 lb                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 92 lb                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Material                                                              | Aluminium, powder coated steel, ABS<br>plastic, stainless steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Aluminium, powder coated steel,<br>ABS plastic, stainless steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Electrical<br>Specifications                                          | 1. Mains power input: 120 V (AC), 60<br>Hz<br>2. Power consumption: 290 VA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 1. Mains power input: 120 V (AC), 60<br>Hz<br>2. Power consumption: 290 VA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Type of device                                                        | Prescription use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Prescription use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Movability                                                            | Operio is a portable device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Operio is a portable device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Air filtration                                                        | HEPA filtered air with at least<br>99.995% efficiency against 0.3 µm<br>particles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | HEPA filtered air with at least<br>99.995% efficiency against 0.3 µm<br>particles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| HEPA filter<br>media                                                  | Nelior membrane media.<br>Folded sheets of nelior.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Nelior membrane media.<br>Folded sheets of nelior.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Air inlet Pre-<br>filter Media                                        | Polyamide mesh                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Polyamide mesh                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Regular<br>maintenance                                                | HEPA filter to be replaced after 2000<br>hours/yearly.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | HEPA filter to be replaced after 2000<br>hours/yearly.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Airflow capacity                                                      | Airflow is 235 CFM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Airflow is 235 CFM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Efficiency                                                            | Bacteria carrying particles at the<br>surgical sites are 0.4, 2 and 0.52                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Bacteria carrying particles at the<br>surgical site is 0.4 CFU/m3.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Airflow speed                                                         | 0.4 m/s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 0.4 m/s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Air volume<br>delivered to<br>incision in CFM                         | Airflow over surgical site is 235 CFM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Airflow over surgical site is 235 CFM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Measured<br>average<br>particulate<br>density at<br>surgical incision | 1.1 particulate of 0.5 $   \mu $ m/f3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 1.1 particulate of 0.5 $   \mu $ m/f3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Air flow position                                                     | Air flow is provided over the surgical<br>site and instruments.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Air flow is provided over the surgical<br>site and instruments.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Sterile accessory                                                     | Single-use sterile shield from Toul<br>Meditech.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Single-use sterile shield from Toul<br>Meditech.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Sterility<br>Assurance (SAL)                                          | 10-6 SAL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 10-6 SAL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Sterilization<br>Method                                               | Gamma irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Gamma irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for "toul meditech". The word "toul" is written in large, blue, sans-serif font, with the "o" appearing to be inside of a blue square. Below "toul" is the word "meditech" in a smaller, blue, sans-serif font.

## Summary of testing

The performance characteristics of the air zone unit Operio have been obtained through both a series of tests, clinical and non-clinical, and performance specifications. Particulate test and CFU test results clearly meet the criteria for achieving reduction of airborne contamination at the intended surgical site and over instruments. The efficiency of HEPA filtration has been tested and is found to have an efficiency > 99.995% efficiency against 0.3 um particles.

In testing, the subject device has demonstrated a level of particulates by 1.1 particles of 0.5 um/f³.

The following non-clinical tests has been conducted: Simulated use test. Particulate counting test, Colony Forming Units counting test, Air velocity test, Air leakage test, Smoke test and Turbulence tests to demonstrate air flow patterns.

The following standards have been used for performance testing:

- IEC/EN 60601-1:2006 Medical electrical equipment Part 1: General requirements ● for basic safety and essential performance
- . IEC/EN 60601-1-2:2007 Medical electrical equipment - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- . IEC 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements

### Summary of clinical testing

Clinical tests have been carried out for surgical procedures such as ophthalmology, orthopedics and neurosurgery. Measurements where made near the surgical site and over the instruments where the results demonstrated a statistically significant reduction in CFU levels.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for "toul meditech". The word "toul" is in large, blue, sans-serif font, with the "o" inside a blue square. Below "toul" is the word "meditech" in a smaller, blue, sans-serif font.

The clinical tests include 302 CFU measurements from 82 patients in Swedish Hospitals. For the predicate device Operio for ophthalmic surgery the mean CFU value is 0.4 CFU/m² and the subject device Operio with additional claims show for orthopaedic surgery a mean CFU value of 0.52 CFU/m3 and for neurosurgery a mean CFU value of 2 CFU/m².

Since CFU level doesn't exceed the predicate device of < 5 CFU it is concluded that substantial equivalence is achieved as clinical validation.

## Conclusion

The conclusions drawn from the nonclinical and clinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device (K153498).

---

**Source:** [https://fda-staging.innolitics.com/device/K173349](https://fda-staging.innolitics.com/device/K173349)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
