K173318 · 3M Deutschland GmbH · ELW · Oct 31, 2017 · Dental
Device Facts
Record ID
K173318
Device Name
Rapid HB, Rapid LB, Rapid MB
Applicant
3M Deutschland GmbH
Product Code
ELW · Dental
Decision Date
Oct 31, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3660
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums
Device Story
Polyether impression materials (heavy, medium, light consistency) used by dental professionals to create dental impressions. Rapid HB and MB are mixed via Pentamix device; Rapid LB is applied via Garant dispenser. Materials are placed on preformed trays to capture tooth/gum structure. Curing occurs via cationic ring-opening polymerization initiated by a Sulfonium starter. Accelerated curing mechanism compared to predicate. Used in one-step or mono-phase techniques. Output is a physical dental impression used for restorative planning and fabrication. Benefits include accurate reproduction of dental anatomy for clinical procedures.
Clinical Evidence
Bench testing only. In vitro testing conducted per ISO 4823. Compared consistency, contact angle, working time, intraoral setting time, recovery from deformation, strain in compression, detail reproduction, gypsum compatibility, and linear dimensional change against predicate devices. Results demonstrate substantial equivalence.
Technological Characteristics
Polyether-based impression material (polyalkyleneoxides). Curing via cationic ring-opening polymerization (Sulfonium starter). Heavy/Medium consistency: Penta foil bags, 5:1 mixing ratio (Pentamix). Light consistency: Garant cartridge, 2:1 mixing ratio. Complies with ISO 4823. Biocompatibility per ISO 10993-1, -5, -10, -11, -12, -18 and ISO 7405.
Indications for Use
Indicated for impressions of inlay, onlay, veneer, crown, and bridge preparations, as well as fixation and implant impressions. Suitable for single-unit and double-unit preparations.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
3M Deutschland GmbH % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K173318
Trade/Device Name: Rapid HB, Rapid LB, Rapid MB Regulation Number: 21 CFR 872.3660 Regulation Name: Impression material Regulatory Class: Class II Product Code: ELW Dated: October 18, 2017 Received: October 20, 2017
Dear Mr. Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
October 31, 2017
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows the name "Mary S. Runner -S" in a large, bold font. The text is black and appears to be against a white background. The letters are clearly legible and well-defined.
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Change Control Table, Change History
## Change Control Table
| Version | Document Author | Document Approver | Date Approved |
|---------|---------------------|---------------------|---------------|
| 1.00 | Name, Title, Office | Name, Title, Office | MM/DD/YYYY |
Complete Change Control Table (all versions) retained in SWIFT Docs.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
Device Name Rapid HB Rapid LB Rapid MB
Indications for Use (Describe)
· Impressions of inlay, onlay, veneer, crown, and bridge preparations
· Fixation and implant impressions
RAPID HB, MB and LB are especially suitable for taking impressions of single-unit and double-unit preparations.
| Type of Use (Select one or both, as applicable) | <span style="padding-right:20px;"> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 | EF |
|----------------------|-------------|----------------------------------------|----|
|----------------------|-------------|----------------------------------------|----|
{4}------------------------------------------------
3M Deutschland GmbH
Carl-Schurz-Straße 1 41453 Neuss Germany
Postal Address: ESPE Platz 82229 Seefeld Germany
☎+49 (0)8152 700-0 鳥+49 (0)8152 700-1366 Internet: www.3M.de/OralCare E-Mail: info3MESPE@3M.com WEEE-Reg.-Nr. DE 36963167 VAT-ID: DE 120679179
#### 510(k) Summary K173318
| Submitter: | 3M Deutschland GmbH<br>ESPE Platz<br>82229 Seefeld<br>Germany<br>Establishment Registration Number: 9611385 |
|----------------|-------------------------------------------------------------------------------------------------------------|
| Contact Person | Ruediger Franke<br>Regulatory Affairs Specialist<br>Phone: +49-8152-700 1802<br>Fax: +49-8152-700 1869 |
| Phone: +49-8152-700 1602 |
|--------------------------------|
| Fax: +49-8152-700 1869 |
| e-mail: ruediger.franke@3M.com |
| Date: October 4. 2017 |
| Trade Name: | Rapid HB |
|-------------|----------|
| | Rapid LB |
| | Rapid MB |
Common Name: ........................................ Polyether impression material
Classification Name:.......................................................................................................................................................... ..............................................................................................................................................................................
Device Class: ................................................................................................................................................................
Predicate Devices ....................... Impregum Penta Soft Quick Step HB (K032001) .............................................................................................................................................................................. ..............................................................................................................................................................................
#### Description of Device
Rapid HB, Rapid LB and Rapid MB are a group of polyether impression materials of heavy, light and medium consistency used by dental professionals individually or in combination. Rapid HB and Rapid MB are mixed in the Pentamix mixing device, while Rapid LB can be applied directly using the Garant dispenser.
Rapid HB is a heavy bodied hydrophilic impression material used in one step technique. Thereby Rapid HB is used in combination with the light consistency wash material Rapid LB. Rapid HB is reddish-violet colored.
Sitz: 41453 Neuss · Handelsregister: B 1878 Amtsgericht Neuss Geschäftsführer: Prof. Dr. Joerg Dederichs, Christiane Grün, Nicolas Jennepin, Michael Peters Vorsitzende der Geschäftsführung: Christiane Grün; Vorsitzender des Aufsichtsrates: Günter Gressler
{5}------------------------------------------------
Rapid LB is a light bodied hydrophilic impression material used in one step technique. Thereby Rapid LB is used in combination with the higher consistency tray material Rapid LB is apple green colored.
Rapid MB is a medium bodied hydrophilic impression material used in mono-phase technique (Rapid MB as syringing and as tray material) and in one step technique (Rapid LB as syringing and Rapid MB as tray material). Rapid MB is blueish-violet colored.
#### Applicable Standards for Product Tests
- . ISO 4823: Dentistry - Dentistry - Elastomeric Impression Materials
### Indications for Use for Rapid HB, Rapid LB and Rapid MB
- · Impressions of inlay, onlay, veneer, crown, and bridge preparations
- Fixation and implant impressions
RAPID HB, MB and LB are especially suitable for taking impressions of single-unit and doubleunit preparations.
#### Comparison
Rapid HB, Rapid LB and Rapid MB were compared to Impregum Penta Soft Quick Step MB and Impregum Soft Quick Step LB regarding indications for use, intended use, composition technology and physical and mechanical properties.
The tables below summarize the indications and technology of Rapid HB, Rapid LB and Rapid MB and predicate devices:
| Indications<br>Comparison | Rapid HB<br>Rapid LB<br>Rapid MB | Predicate Devices (K032001) |
|---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use<br>(from the<br>Indications for<br>Use form) | • Impressions of inlay,<br>onlay, veneer, crown, and<br>bridge preparations<br>• Fixation and implant<br>impressions<br>RAPID HB, MB and LB are<br>especially suitable for<br>taking impressions of<br>single-unit and double-unit<br>preparations. | Impregum Penta Soft Quick Step<br>HB<br>Poly Q Penta H:<br>Tray material for dual phase<br>impression techniques.<br>Impregum Soft Quick Step LB<br>Poly Q Garant L:<br>Wash material for dual phase<br>impression techniques.<br>Impregum Penta Soft Quick Step<br>MB<br>Poly Q Penta M:<br>Impression material for monophase<br>technique. |
| Intended Use | Intended to be placed on a<br>preformed impression tray<br>and used to reproduce the<br>structure of a patient's teeth<br>and gums | Intended to be placed on a<br>preformed impression tray and used<br>to reproduce the structure of a<br>patient's teeth and gums |
Table Comparison of indications
{6}------------------------------------------------
The differences between the indications of the Rapid materials and the predicate devices are not critical to the intended clinical use because the Rapid indications described the clinical case whereas the predicate indication described the usable impression technique of each material. All clinical indications (inlay, onlay, veneer, crown, and bridge preparations as well as fixation and implant impressions) are achievable with the described impression techniques (dual phase and monophase technique). This difference, i.e., stating clinical cases rather than impression techniques, does not raise any new questions regarding safety or effectiveness. Rather, it clarifies the appropriate uses within the described impression techniques and does not expand the indications of the new device(s) beyond those of the predicate device(s).
| Technology | Rapid HB<br>Rapid LB<br>Rapid MB | Impregum Penta Soft Quick Step HB<br>Impregum Penta Soft Quick Step MB<br>Impregum Soft Quick Step LB<br>(K032001) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Light bodied material offered in<br>Garant cartridge system with mixing<br>ratio of base and catalyst pastes 2:1<br>(by volume)<br>The pastes are mixed using Garant<br>mixing tips or Intra-oral syringe. | × | × |
| Medium and Heavy bodied material<br>offered in Penta foil bags with mixing<br>ratio of base and catalyst paste 5:1<br>(by volume).<br>The pastes are mixed in automatic<br>mixing machine of the Pentamix<br>family and Penta mixing tip. | × | × |
| Material family | Polyether<br>material | Polyether material |
| Setting characteristic | super quick | quick |
Table Comparison to Predicate Technology
Rapid HB, Rapid LB and Rapid MB, Impregum Penta Soft Quick Step HB, Impregum Penta Soft Quick Step MB and Imprequm Soft Quick Step LB are Polyether based Precision Impression materials. Chemically they are based on polyalkyleneoxides which cure by a cationic ringopening polymerization mechanism which is initiated by a Sulfonium starter. However, Rapid materials have been developed with the goal to accelerate the curing mechanism by adjusting the initiator system. Additives like fillers, softeners, pigments etc. have been adjusted accordingly.
{7}------------------------------------------------
#### Non-clinical testing
In vitro testing was conducted to show that Rapid HB, Rapid LB and Rapid MB fulfils the requirements of FDA recognized standard ISO 4823. Additionally, the performance of Rapid HB, Rapid LB and Rapid MB was compared to the predicate devices Impregum Penta Soft Quick Step HB, Impregum Penta Soft Quick Step MB and Impregum Soft Quick Step LB regarding consistency, contact angle, working time, intraoral setting time, recovery from deformation, strain in compression, detail reproduction, compatibility with gypsum type 3 / type 4 and linear dimensional change. The results of Rapid LB and Rapid MB fulfill the requirements of ISO 4823 and are similar to Impregum Penta Soft Quick Step HB, Impregum Penta Soft Quick Step MB and Impregum Soft Quick Step LB. In summary, 3M Deutschland GmbH concludes that Rapid HB, Rapid LB and Rapid MB is substantially equivalent to the predicate devices regarding performance and physical and mechanical properties.
#### Biocompatibility
The biocompatibility assessment for the products was conducted in accordance with the following guidance:
| Guidance | Edition | Title |
|------------------------------------------------------------------------------|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| US FDA Docket<br>Number FDA-2013-<br>D-0350. CDRH<br>Document Number<br>1811 | June 16,<br>2016 | Use of International Standard ISO 10993-1, Biological<br>evaluation of medical devices - Part 1: Evaluation and testing<br>within a risk management process - Guidance for Industry<br>and Food and Drug Administration Staff |
| ISO 10993-1 | 2009 | Evaluation and testing within a risk management process |
| ISO 10993-5 | 2009 | Tests for <i>in vitro</i> cytotoxicity |
| ISO 10993-10 | 2010 | Tests for irritation and skin sensitization |
| ISO 10993-11 | 2006 | Tests for systemic toxicity |
| ISO 10993-12 | 2012 | Sample preparation and reference materials |
| ISO 10993-18 | 2005 | Chemical characterization of materials |
| ISO 7405 | 2008/Amd 1:<br>2013 | Evaluation of biocompatibility of medical devices used in<br>dentistry |
The biocompatibility of Rapid LB, Rapid LB and Rapid MB has been assessed by a boardcertified toxicologist according to recommendations in FDA quidance and internationally recognized standards for medical and dental devices. The conclusion of the assessment is that Rapid HB, Rapid LB and Rapid MB are safe for their intended use.
### Conclusion
Comparisons of the indications for use/intended use, composition, technology, and physical and mechanical properties showed that the Rapid HB, Rapid LB and Rapid MB group of materials is as safe, as effective, and performs as well as or better than the legally marketed devices Impregum Penta Soft Quick Step HB, Impregum Penta Soft Quick Step MB and Impregum Soft Quick Step LB.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.