The Lumendi DiLumen C2 is an endoscope accessory intended to ensure complete positioning of an endoscope in the large intestine, and assist with optical visualization, diagnosis, and endoscopic treatment.
Device Story
DiLumen C2 is a sterile, single-use sleeve fitting over standard endoscopes; features two inflatable balloons (Fore and Aft) for positioning and stabilization within the large intestine. Clinician manually inflates/deflates balloons using an inflation handle with a squeeze bulb. Once inflated, balloons create an isolated diagnostic/therapeutic zone. Device includes two 6-mm tool channels for flexible endoscopic tools (graspers, scissors, knives). System includes a reusable Tool Mount (stainless steel) that attaches to surgical table rails to hold tool channels in fixed positions. Used by physicians in clinical settings during endoscopic procedures. Benefits include improved endoscope stability and access for endolumenal interventions.
Clinical Evidence
Bench testing only. No clinical or animal data required. Testing included biocompatibility, balloon diameter/inflation/leakage, device slip, fore balloon extension, articulation, colon grip, therapeutic zone creation, Tuohy leak/bond leak, insertion force, packaging/transit, user validation, EO residuals, and Tool Mount cleaning/sterilization/positioning/locking validation.
Technological Characteristics
Sterile, single-use sleeve; materials include low durometer polyurethane (balloons), Pellethane with 2.5% Glycolube (sleeve), Isoplast/Pebax (tubing), Lexan (collet), and FEP/stainless steel braid (tool channels). Manual inflation via ambient air squeeze bulb. Tool Mount uses stainless steel articulating arm. Dimensions: 1680 mm working length, 60 mm outer balloon diameter. Sterilization: Ethylene Oxide (EO).
Indications for Use
Indicated for use as an endoscope accessory in the large intestine to ensure complete positioning, assist with optical visualization, diagnosis, and endoscopic treatment.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two distinct elements: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 17, 2018
Lumendi, LLC % John J. Smith, MD, JD Partner Hogan Lovells U.S. LLP 553 13th Street NW Washington, DC 20004
Re: K173317
Trade/Device Name: DiLumen C2: DiLumen Tool Mount Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDF Dated: March 6, 2018 Received: March 6, 2018
Dear John J. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and is set against a white background. The letters are clearly legible and well-spaced.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173317
Device Name DiLumen C2; DiLumen Tool Mount
Indications for Use (Describe)
The Lumendi DiLumen C2 is an endoscope accessory intended to ensure complete positioning of an endoscope in the large intestine, and assist with optical visualization, diagnosis, and endoscopic treatment.
Type of Use (Select one or both, as applicable)
| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|--------------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY Lumendi, LLC's DiLumen C2 and Tool Mount
### Submitter's Information:
LUMENDI, LLC 253 Post Road West Westport, CT 06880 Phone: (203) 463-2669 Fax: (203) 557-0459
#### Contact Person:
Dennis J. Daniels Senior Director, Regulatory Affairs and Quality Assurance Telephone: (203) 557-6335 Fax: (203) 557-0459 Email: Dennis.Daniels@Lumendi.com
Date Prepared: April 5, 2018
### Device Identification:
Trade Name: DiLumen C2 and Tool Mount Common Name: Endoscope Accessory Classification Regulation: 21 C.F.R. § 876.1500 (Endoscope and accessories) Product Codes: FDF Device Class: II
Predicate Device: DiLumen Endolumenal Interventional Platform (K162428)
Reference Device: FISSO Holding System (K070509)
### Device Description:
The DiLumen C² consists of a sleeve including two inflatable balloons that fits over a standard endoscope to facilitate positioning and stabilization of the endoscope during surgical procedures. The C2 comes into limited (<24 hour) contact with breached or compromised surfaces of the patient. The C2 is provided sterile and intended only for single patient use.
To stabilize the system during clinical use, the C2 is provided with its own designated accessory, known as the Tool Mount, a metal holding system that fastens the tool channels to a surgical table rail. The Tool Mount is re-usable and is provided non-sterile; it must be cleaned and sterilized prior to each use following the instructions in the device labeling.
### Intended Use / Indications for Use:
The Lumendi DiLumen C is an endoscope accessory intended to ensure complete positioning of an endoscope in the large intestine, and assist with optical visualization, diagnosis, and endoscopic treatment.
{4}------------------------------------------------
The identified predicate device for the DiLumen C2 has similar indications for use as the subject device. Like the predicate Lumen (K162428), the DiLumen C² device is intended for use as an accessory to an endoscope to position/stabilize the endoscope and assist in optical visualization, diagnosis, and endoscopic treatment in the large intestine. Accordingly, the differences in indications for use between the DiLumen C2 and its predicate device are not critical to the intended therapeutic/diagnostic use of the device and do not affect its safety or effectiveness when used as labelled.
## Technological Characteristics and Principles of Operation:
The DiLumen C2 Endolumenal Interventional Platform is a sterile, single-use, close-fitting sleeve that fits securely over a standard endoscope.
The DiLumen C2 utilizes two balloons to position and stabilize the endoscope within a patient's large intestine. After the DiLumen C2 is installed over the endoscope, a collet is secured to the endoscope distal tip. The endoscope and DiLumen are navigated to the target zone with the balloons deflated. Once the clinician is at the area of interest, the Aft Balloon, which is attached to the DiLumen C2 sleeve, will be inflated until it contacts the intestinal wall near the proximal end of the articulating section of the endoscope. The second balloon, the Fore Balloon, is also attached to the sleeve via two flexible extension push rods and is deployed at the distal end of the endoscope at a variable distance. Once extended and inflated, the Fore Balloon contacts the patient's intestinal wall, and in combination with the Aft Balloon, creates an isolated diagnostic or therapeutic zone. Both balloons are controlled independently using an Inflation Handle with a squeeze bulb to manually inflate and deflate the two balloons with ambient air. The balloons assist in stabilizing the endoscope and the therapeutic area.
The flexibility, maneuverability and functionalities (such as visualization, suction, insufflations, etc.) of the endoscope are unaffected by the presence of the DiLumen C2. The DiLumen C2 has two 6-mm diameter tool channels. The tool channels are located 180° apart at the distal end of the sleeve. The tool channels provide the clinician access at the therapeutic site so that the clinician can use flexible endoscopic tools such as graspers, scissors, knives, etc. to perform endolumenal interventions.
The DiLumen C-is provided with a designated accessory, the Tool Mount, which holds the tool channels in a fixed position to facilitate their use. The Tool Mount consists of the previously cleared FISSO Holding System (K070509) articulating arm with an added permanently attached tool bar to firmly hold the C-tool channels in the clinician's preferred orientation. The tool bar has flanges that firmly hold the proximal ends of the DiLumen C² tool channels and are adjustable in three angular positions. The Tool Mount and holding system is attached to a standard surgical table rail using a rail clamp. The Tool Mount has similar indications for use and function as the FISSO Holding System reference device, namely to hold endoscopes, equipment, and patient positioning accessories during diagnostic and therapeutic procedures, but because the Tool Mount is designed for use with only the DiLumen C2, its intended use is narrower.
### Performance Data:
Performance testing has demonstrated that the DiLumen C2 meets specifications and is as safe and effective as the predicate. As the DiLumen C2 shares many of the same components as the predicate devices, bench testing primarily targeted verification of the balloon integrity and operation of the DiLumen C2 with an endoscope and with the Tool Mount. The following performance data were provided in support of this Premarket Notification:
{5}------------------------------------------------
- 1. Biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, material mediated pyrogenicity)
- 2. Balloon Diameter, Inflation/Deflation and Leakage Test
- 3. Device Slip Relative to Scope Test
- 4. Fore Balloon Extension Test
- 5. Articulation Test
- 6. Colon Grip Test
- 7. Therapeutic Zone Creation Test
- 8. Tuohy Leak Force / Bond Leak Test
- 9. Insertion Force Test
- 10. Packaging and Transit Test
- 11. User Validation
- 12. EO residuals
- 13. Tool Mount Cleaning and Sterilization Validation
- 14. Tool Mount Positioning and Locking Test
- 15. Tool Mount Validation
In all instances, the device functioned as intended and the results observed were as expected.
No animal or clinical testing was required to demonstrate substantial equivalence to the predicate.
#### Substantial Equivalence:
The subject Dilumen C² device has nearly identical technological characteristics as the DiLumen Endolumenal Interventional Platform cleared under K162428, with the main differences between the devices being:
- 1. The subject device has two tool channels;
- 2. The subject device has a dedicated Tool Mount accessory;
- 3. The subject device is secured to the endoscope distal tip using a collet;
- 4. The internal diameter of the DiLumen C2's balloons is slightly larger; and
- 5. The subject device is EO-sterilized.
Because the DiLumen C2 and the predicate consist of many of the same components to ensure complete positioning of the endoscope and advance the endoscope to the target location in the same manner, and as further demonstrated by performance testing, these minor technological differences do not raise different types of safety or efficacy questions. In addition, the Tool Mount accessory - which is limited to use with the C2 device - is very similar to the FISSO Holding System reference device (K070509).
A table comparing the key features of the subject and predicate devices is provided below.
| | Lumendi DiLumen C2 | Lumendi DiLumen Endolumenal<br>Interventional Platform |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | TBD | K162428 |
| Intended Use /<br>Indications for Use | The Lumendi DiLumen C2 is an<br>endoscope accessory intended to<br>ensure complete positioning of an<br>endoscope in the large intestine,<br>and assist with optical | The Lumendi DiLumen is an accessory to an<br>endoscope. The DiLumen dual balloon<br>accessory is intended for use with any<br>endoscope that has an outer diameter of 12.5 –<br>14.3 mm and a minimum working length of 1680 |
| | Lumendi DiLumen C2 | Lumendi DiLumen Endolumenal<br>Interventional Platform |
| | visualization, diagnosis, and<br>endoscopic treatment. | mm. The device is indicated to ensure complete<br>positioning of an endoscope in the large<br>intestine, and assist with optical visualization,<br>diagnosis, and endoscopic treatment. |
| Sterility | Sterile - EO | Non-sterile |
| Single use/reusable | Single Use | Single Use |
| Balloon | Low durometer Polyurethane | Low durometer Polyurethane |
| Sleeve | Pellethane with 2.5% Glycolube<br>translucent | Pellethane with 2.5% Glycolube translucent |
| Inflation tube/<br>Pushrod Assembly | Isoplast<br>Pebax<br>Pellethane | Isoplast<br>Pebax<br>Pellethane |
| Collet | Lexan | NA |
| Tool Channel<br>Tubing | FEP; Pellethane; stainless steel<br>internal braid; | NA |
| Tool Channel<br>Components | Polyurethane; Pellethane with<br>2.5% Glycolube ;ePTFE Lexan;<br>silicone; | NA |
| Compatible<br>Endoscope Tip<br>Diameter | 11.5 - 14.5 mm | 12.5 - 14.3 mm |
| Fore/Aft Balloons<br>- Inner Diameter | 28.5 mm | 21.5 mm |
| Fore/Aft Balloons<br>- Outer Diameter | 60 mm | 60 mm |
| Balloon Pressure | 45 ± 12 mmHg | 45 ± 12 mmHg |
| Relief Pressure | 55 mmHg | 55 mmHg |
| Working Length | 1680 mm | 1680 mm |
| Inflation Source | Manual inflation bulb | Manual inflation bulb |
| Shelf Life | 12 months | 12 months |
| Accessories | The Lumendi DiLumen C2 has a<br>designated Tool Mount that is<br>intended to mount to a surgical<br>table rail and hold the DiLumen<br>C2 tool channels during<br>endoscopic treatment. The Tool<br>Mount is made of stainless steel | The Lumendi DiLumen can be used with various<br>additional endoscope accessories, but is not<br>supplied with any designated accessories. |
{6}------------------------------------------------
## Conclusion:
The DiLumen C² is as safe and effective as the DiLumenal Interventional Platform cleared under K162428. The device has the same intended use and principles of operation, and similar indications for use and technological characteristics, as the predicate device, and its designated accessory (the Tool Mount) has similar indications and function as the identified reference device (the FISSO Holding System cleared under K070509). The minor differences in indications and technology do not alter the intended therapeutic use of the device and do not raise any new questions of safety or effectiveness when the device is used as labelled. In addition, bench testing data demonstrate that the DiLumen C2 is as safe and effective as the predicate device. Thus, the DiLumen C2 is substantially equivalent to the DiLumen Endolumenal Interventional Platform.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.