← Product Code [HRS](/productcode/HRS) · K173293

# EVOS Small Fragment Upper Extremity Plates (K173293)

_Smith & Nephew, Inc. · HRS · Jan 8, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173293

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Jan 8, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The EVOS Small Fragment Plating System is indicated for adult and pediatric patients, as well as patients with osteopenic bone. It is indicated for fixation of small and long bone fractures, including, but not limited to, those of the tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, metatarsals, and clavicle.

## Device Story

EVOS Small Fragment Upper Extremity Plates are metallic bone fixation implants designed for surgical stabilization of fractures. Plates feature 2.7mm and 3.5mm holes, including threaded, non-threaded, and variable-angle locking configurations. Used by orthopedic surgeons in clinical settings, these plates are implanted to provide structural support during bone healing. The system is an extension of existing Smith & Nephew plating lines, utilizing compatible bone screws and instruments. Clinical benefit includes anatomical fracture fixation, particularly for upper extremity applications. Device performance is validated through finite element analysis (FEA) and mechanical bending/construct fatigue testing to ensure structural integrity comparable to predicate systems.

## Clinical Evidence

Bench testing only. Evidence includes finite element analysis (FEA) to identify worst-case designs, followed by mechanical bending and construct fatigue testing. Results demonstrated performance similar or superior to predicate plates. Packaging verification and bacterial endotoxin testing (per ANSI/AAMI ST72) were performed to confirm sterility maintenance and safety.

## Technological Characteristics

Metallic bone fixation plates with 2.7mm/3.5mm hole diameters. Features include threaded, non-threaded, and variable-angle locking holes. Designed for compatibility with existing EVOS system screws and instruments. Sterilization validated per FDA guidance and ANSI/AAMI ST72 standards.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Smith & Nephew Bone Plate System ([K993106](/device/K993106.md))
- Smith & Nephew Locking Bone Plate System ([K033669](/device/K033669.md))
- PERI-LOC Periarticular Locked Plating System - B Plate Locking Bone Plates and Screws ([K062216](/device/K062216.md))
- PERI-LOC Periarticular Locked Plating System -VLP Locking Bone Plates and Screws ([K071563](/device/K071563.md))
- EVOS Small Fragment Plating System Straight Plates and Screws ([K162078](/device/K162078.md))
- EVOS Small Fragment Lower Extremity Plates ([K170457](/device/K170457.md))
- EVOS Small Fragment Plating System ([K170887](/device/K170887.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 8, 2018

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Smith & Nephew, Inc. Samantha Staubach Regulatory Affairs Specialist II 1450 Brooks Road Memphis, Tennessee 38116

### Re: K173293

Trade/Device Name: EVOS Small Fragment Upper Extremity Plates Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: December 19, 2017 Received: December 20, 2017

Dear Ms. Staubach:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

## Katherine D. Kavlock -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# >< smith&nephew

| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120                        |
|-----------------------------------------|---------------------------------------------------------|
| Food and Drug Administration            | Expiration Date: 06/30/2020<br>See PRA Statement below. |

**Indications for Use**

| 510(k) Number (if known)       | K173293                                                                                                                                                                                                                                                                                                                                             |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                    | EVOS Small Fragment Upper Extremity Plates                                                                                                                                                                                                                                                                                                          |
| Indications for Use (Describe) | The EVOS Small Fragment Plating System is indicated for adult and pediatric patients, as well as patients with osteopenic bone. It is indicated for fixation of small and long bone fractures, including, but not limited to, those of the tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, metatarsals, and clavicle. |

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                  |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="15" width="15"> <rect height="15" style="fill:lightgrey;stroke-width:1;stroke:rgb(0,0,0)" width="15"></rect> <line style="stroke:rgb(0,0,0);stroke-width:2" x1="2" x2="13" y1="2" y2="13"></line> <line style="stroke:rgb(0,0,0);stroke-width:2" x1="13" x2="2" y1="2" y2="13"></line> </svg> </span>                 Prescription Use (Part 21 CFR 801 Subpart D)             </div> | <div> <span> <svg height="15" width="15"> <rect height="15" style="fill:white;stroke-width:1;stroke:rgb(0,0,0)" width="15"></rect> </svg> </span>                 Over-The-Counter Use (21 CFR 801 Subpart C)             </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*

{3}------------------------------------------------

## >< We are smith&nephew

| Submitted by:                                | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116                                       |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary:                             | October 12, 2017                                                                                                                        |
| Name of Device:                              | Samantha Staubach<br>Regulatory Affairs Specialist II<br>T 901-399-6132<br>F 901-566-7596<br>EVOS Small Fragment Upper Extremity Plates |
| Common Name:                                 | Bone Plates                                                                                                                             |
|                                              |                                                                                                                                         |
| Device Classification Name and<br>Reference: | 21 CFR 888.3030 Single/multiple component metallic bone<br>fixation appliances and accessories                                          |
| Device Class:                                | Class II                                                                                                                                |
| Panel Code:                                  | Orthopaedics/87                                                                                                                         |
| Product Code:                                | HRS                                                                                                                                     |

#### Predicates

| Manufacturer         | Description                                                                                                                  | Submission<br>Number | Clearance Date        |
|----------------------|------------------------------------------------------------------------------------------------------------------------------|----------------------|-----------------------|
| Smith & Nephew, Inc. | Smith & Nephew Bone Plate<br>System (primary predicate,<br>now branded as TC-100)                                            | K993106              | December 9,<br>1999   |
| Smith & Nephew, Inc. | Smith & Nephew Locking Bone<br>Plate System (now branded as<br>PERI-LOC)                                                     | K033669              | December 10,<br>2003  |
| Smith & Nephew, Inc. | PERI-LOC Periarticular Locked<br>Plating System - B Plate Locking<br>Bone Plates and Screws (now<br>branded as PERI-LOC VLP) | K062216              | September 15,<br>2006 |
| Smith & Nephew, Inc. | PERI-LOC Periarticular Locked<br>Plating System -VLP Locking<br>Bone Plates and Screws                                       | K071563              | August 8, 2007        |
| Smith & Nephew, Inc. | EVOS Small Fragment Plating<br>System Straight Plates and<br>Screws                                                          | K162078              | November 18,<br>2016  |
| Smith & Nephew, Inc. | EVOS Small Fragment Lower<br>Extremity Plates                                                                                | K170457              | June 14, 2017         |
| Smith & Nephew, Inc. | EVOS Small Fragment Plating<br>System                                                                                        | K170887              | April 24, 2017        |

{4}------------------------------------------------

#### Device Description

Subject of this premarket notification is an extension of the EVOS Small Fragment Plating System, the EVOS Small Fragment Upper Extremity Plates. The subject plates feature similarities to existing Smith & Nephew small fragment plating systems (TC-100 Small Bone Plating System, PERI-LOC/PERI-LOC VLP Plating Systems) and also share some instruments and compatible implants from the existing EVOS MINI Plating System and EVOS Small Fragment Plating System. When compared against EVOS Small Fragment Plates already cleared for market, the subject plates use the same hole features and are designed to work with the same bone screws. The EVOS Small Fragment Upper Extremity Plates are available in a variety of plate designs for specific anatomical areas. These plate designs include plates with 2.7mm holes or 3.5mm holes only, or with a combination of 2.7mm and 3.5mm holes. Plate designs may include threaded holes, non-threaded holes, and variable-angle locking holes.

#### Indications for Use

The EVOS Small Fragment Plating System is indicated for adult and pediatric patients, as well as patients with osteopenic bone. It is indicated for fixation of small and long bone fractures, including, but not limited to, those of the tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, metatarsals, and clavicle.

#### Technological Characteristics

Device comparisons described in this premarket notification demonstrated that the proposed devices are substantially equivalent to legally marketed predicates with respect to intended use, indications, and performance characteristics. The subject devices include plates designs with a variety of hole features, which are available on EVOS Small Fragment Plates previously cleared for market. The subject plates differ from the EVOS Small Fragment Plates previously cleared in that the plate shapes are more anatomically specific for upper extremity fractures.

#### Summary of Pre-Clinical Testing

- . Finite element analysis (FEA) was conducted on the proposed plate designs to determine the worst case plates based upon structural strength for further mechanical testing. Plates were separated into groups for evaluation based upon similar designs or anatomical application.
- . Bending fatigue performance or construct fatigue performance was evaluated for the worst case plate designs identified through FEA. Results of the testing determined that the subject EVOS plates performed similar or superior to the predicate plates tested, when evaluated under the same conditions.
- Engineering rationales were leveraged to justify the bending strength of some EVOS . plates based on previously conducted FEA, previously conducted mechanical testing, and hand calculations.

{5}------------------------------------------------

- . Packaging verification testing was conducted for the proposed packaging configurations and the results of this testing demonstrated that the product will not be damaged during shipment and will adequately maintain sterility post shipment.
- · Bacterial endotoxin testing was completed and met the acceptable endotoxin limits as stated in the FDA Guidance , "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," "Pyrogen and Endotoxins Testing: Questions and Answers," and ANSI/AAMI ST72.

#### Conclusion

This Traditional 510(k) premarket notification is being submitted to request clearance for the EVOS Small Fragment Plating System Upper Extremity Plates. Based on similarities to the predicate plating systems and a review of the mechanical testing performed, the subject devices are substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K173293](https://fda-staging.innolitics.com/device/K173293)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).