← Product Code [FRO](/productcode/FRO) · K173276

# Medline Sterile Water USP and Medline Sterile 0.9% Normal Saline USP (K173276)

_Medline Industries, Inc. · FRO · Jul 2, 2018 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173276

## Device Facts

- **Applicant:** Medline Industries, Inc.
- **Product Code:** [FRO](/productcode/FRO.md)
- **Decision Date:** Jul 2, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

For Over-the-Counter Use: For moistening absorbent wound dressings and cleaning minor cuts, minor burns, superficial abrasions and minor irritations of the skin. For Prescription Use: For moistening absorbent wound dressings and for moistening, debriding and cleaning acute and chronic dermal lesions, such as Stage I-IV pressure ulcers, stasis ulcers, diabetic ulcers, foot ulcers, post-surgical wounds, first and second degree burns, cuts, abrasions and minor skin irritations and for device irrigation.

## Device Story

Device consists of sterile, preservative-free, aqueous solutions (0.9% Sodium Chloride or Sterile Water USP) in HDPE bottles. Used for wound irrigation, debridement, and moistening of dressings. Operates via mechanical action of fluid flow across wound or device surfaces to remove debris and moisturize. Intended for use in clinical settings or home environments by clinicians or patients. Output is the physical irrigation fluid; healthcare providers use this to facilitate wound healing and maintenance. Benefits include effective cleaning and hydration of compromised skin surfaces.

## Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-1:2009 (cytotoxicity, irritation, sensitization, systemic toxicity) and USP <151> pyrogen testing. Sterilization validated per AAMI TIR33 (SAL 10^-6). Shelf life verified per ASTM F1980-16.

## Technological Characteristics

Single-use, sterile, preservative-free aqueous solutions (0.9% NaCl or Sterile Water USP). Container: HDPE bottles with induction-heat seal and liner. Sterilization: Gamma radiation (AAMI TIR33). Biocompatibility: ISO 10993-1 compliant. Meets USP monograph criteria for Water for Irrigation (USP <1231>).

## Predicate Devices

- Nurse Assist Wound Flush, Sterile Water and Sterile Normal Saline ([K083042](/device/K083042.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, which features a stylized depiction of human figures. To the right of this symbol, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The word "FDA" is in a larger, bolder font, while "U.S. FOOD & DRUG" is in a slightly smaller font, and "ADMINISTRATION" is in an even smaller font below.

Medline Industries, Inc. Claire Pigman Senior Regulatory Affairs Specialist Three Lakes Drive Northfield, Illinois 60093

Re: K173276

Trade/Device Name: Medline Sterile Water USP and Medline Sterile 0.9% Normal Saline USP Regulatory Class: Unclassified Product Code: FRO Dated: June 1, 2018 Received: June 5, 2018

Dear Claire Pigman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

K173276

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely, Neil R Oaden -5 For Binita S. Ashar, M.D., M.B.A., F.A.C Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K173276

Device Name

Medline Sterile Water USP and Medline Sterile 0.9% Normal Saline USP

Indications for Use (Describe)

For Over-the-Counter Use:

For moistening absorbent wound dressings and cleaning minor burns, superficial abrasions and minor irritations of the skin.

For Prescription Use:

For moistening absorbent wound dressings and for mostening acute and chronic dernal lesions, such as Stage I-IV pressure ulcers, stasis ulcers, diabetic ulcers, post-surgical wounds, first and second degree burns, cuts, abrasions and minor skin irritations and for device irrigation.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Medline logo. The logo features the word "MEDLINE" in white, bold, sans-serif font, positioned to the left of a white, stylized starburst symbol. The background is a solid dark blue color. The logo is simple, clean, and professional, reflecting the company's focus on medical supplies and services.

**Medline Industries, Inc.**
Three Lakes Drive
Northfield, IL 60093

K173276

# SECTION 5 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]

# Submitter / 510(k) Sponsor

Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093 Registration Number: 1417592

# Contact Person

Claire Pigman Sr. Regulatory Affairs Specialist Phone: 847-643-4071 Email: cpigman@medline.com

Summary Preparation Date October 9, 2017

Type of 510(k) Submission Traditional

# Device Name / Classification

Name of Device: Medline Sterile Water USP and Medline Sterile 0.9% Normal Saline USP Device Common Name: Sterile Water and Sterile Saline Classification Name: Dressing, Wound, Drug Product Code: FRO Classification Panel: General and Plastic Surgery Devices Device Classification: Unclassified Regulation Number: Pre-Amendment

# Predicate Device

Predicate Device Name: Wound Flush, Sterile Water and Sterile Normal Saline Predicate Device Sponsor: Nurse Assist, Inc. Predicate Device K Number: K083042

# Device Description

The subject devices are 100 mL, 250 mL or 500 mL sterile normal saline (0.9% Sodium Chloride). The design of these single-use devices incorporates clear, high density polyethylene (HDPE)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is written in white, with a white star-like symbol above it. The star symbol has four points, and it is positioned so that the word "MEDLINE" is centered below it.

edline Industries. hree Lakes Drive Northfield II 60093

bottles filled and sealed with a sterile, preservative-free, clear, colorless, aqueous solution, as labeled for device irrigation and wound debridement (not for injection). The subject device will function by means of mechanical action; specifically, by the mechanical mechanism of action associated with applying and moving an aqueous solution across a wound or device surface, which facilitates the moisturizing, debridement and irrigation of these surfaces.

The chemical composition of its aqueous solution is either sterile saline (0.9% Sodium Chloride in sterile water) for irrigation, which have been substantiated to each meet their respective USP monograph criteria, and its formulation contains no additives. The components of the subject device's container and closure system, made from polyethylene (PE) resins and an induction-heat seal and liner, meet the requirements of the Agency's food additive regulations. Medline Sterile Water USP and Sterile 0.9% Normal Saline USP are available in the following configurations:

| Medline Item Number | Description                                      |
|---------------------|--------------------------------------------------|
| RDI30295            | Medline Sterile Water, USP - 100 mL              |
| PCS1550             | Medline Sterile Water, USP - 250 mL              |
| DYND500MLW          | Medline Sterile Water, USP - 500 mL              |
| RDI30296            | Medline Sterile 0.9% Normal Saline, USP - 100 mL |
| PCS1650             | Medline Sterile 0.9% Normal Saline, USP - 250 mL |
| DYND500MLS          | Medline Sterile 0.9% Normal Saline, USP - 500 mL |

Please note that all six of these configurations are designed to consist of either sterile water USP or sterile normal saline (0.9% Sodium Chloride) USP, packaged and sealed in HDPE bottles and terminally sterilized using gamma irradiation. Please refer to Section 11 of this Premarket Notification for a more comprehensive description of this device.

#### Indications for Use

For Over-the-Counter Use: For moistening absorbent wound dressings and cleaning minor cuts, minor burns, superficial abrasions and minor irritations of the skin.

For Prescription Use: For moistening absorbent wound dressings and for moistening, debriding and cleaning acute and chronic dermal lesions, such as Stage I-IV pressure ulcers, diabetic ulcers, foot ulcers, post-surgical wounds, first and second degree burns, cuts, abrasions and minor skin irritations and for device irrigation.

# Summary of Technological Characteristics

Medline Sterile Water USP and Medline Sterile 0.9% Normal Saline USP are single-use devices and their design incorporates clear, high density polyethylene (HDPE) bottles filled and sealed with a sterile, preservative-free, clear, colorless, aqueous solution, as labeled for device irrigation and wound debridement (not for injection). The chemical composition of its aqueous solution is either sterile water

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is written in white, bold, sans-serif font. Above the word is a white, stylized star shape with four points. The star is oriented so that one point is facing upwards.

ledline Industries, Inc. Three Lakes Drive Northfield. IL 60093

or sterile saline (0.9% Sodium Chloride) for irrigation, which have been substantiated to each meet their respective USP monograph criteria, and its formulation contains no additives.

In addition, the subject device is similar in design, intended use and function to the predicate device, Nurse Assist Wound Flush, Sterile Water and Sterile Normal Saline, cleared under K083042. The table below provides a side-by-side comparison of the proposed and predicate devices. Please refer to Section 12 for a further discussion of the similarities and differences between the two devices.

{6}------------------------------------------------

# Table 1: Comparison of Proposed and Predicate Devices

| Device<br>Characteristic          | Proposed Device                                                                                                                                                                                                                                                                                                                                                        | Predicate Device                                                                                                                                                                                                                                                                                                                                                       | Comparison<br>Analysis                                                                                                                                                                                                                                                     |                                                                                                                                    |      |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------|
| Product Name                      | Medline Sterile Water USP and Medline Sterile 0.9% Normal Saline<br>USP                                                                                                                                                                                                                                                                                                | Nurse Assist Wound Flush, Sterile Water and Sterile Normal<br>Saline                                                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                                                                        |                                                                                                                                    |      |
| 510(k) Reference                  | K173276                                                                                                                                                                                                                                                                                                                                                                | K083042                                                                                                                                                                                                                                                                                                                                                                | N/A                                                                                                                                                                                                                                                                        |                                                                                                                                    |      |
| Product Owner                     | Medline Industries, Inc.                                                                                                                                                                                                                                                                                                                                               | Nurse Assist Incorporated                                                                                                                                                                                                                                                                                                                                              | N/A                                                                                                                                                                                                                                                                        |                                                                                                                                    |      |
| Product Code                      | FRO                                                                                                                                                                                                                                                                                                                                                                    | FRO                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                       |                                                                                                                                    |      |
| Indication for Use                | Over-the-Counter Use: For moistening absorbent wound dressings<br>and cleaning minor cuts, minor burns, superficial abrasions and<br>minor irritations of the skin                                                                                                                                                                                                     | Over-the-Counter Use: For moistening absorbent wound dressings<br>and cleaning minor cuts, minor burns, superficial abrasions and<br>minor irritations of the skin.                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                       |                                                                                                                                    |      |
|                                   | For Prescription Use: For moistening absorbent wound dressings and<br>for moistening, debriding and cleaning acute and chronic dermal<br>lesions, such as Stage I-IV pressure ulcers, stasis ulcers, diabetic<br>ulcers, foot ulcers, post-surgical wounds, first and second degree<br>burns, cuts, abrasions and minor skin irritations and for device<br>irrigation. | For Prescription Use: For moistening absorbent wound dressings<br>and for moistening, debriding and cleaning acute and chronic<br>dermal lesions, such as Stage I-IV pressure ulcers, stasis ulcers,<br>diabetic ulcers, foot ulcers, post-surgical wounds, first and second<br>degree burns, cuts, abrasions and minor skin irritations and for<br>device irrigation. | Same                                                                                                                                                                                                                                                                       |                                                                                                                                    |      |
| Regulation Number                 | Pre-Amendment                                                                                                                                                                                                                                                                                                                                                          | Pre-Amendment                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                       |                                                                                                                                    |      |
| Technological<br>Characteristics: | Principle mechanism of action is through the mechanical action<br>associated with the application and movement of fluid (either Sterile<br>Water or Saline) across a wound or device surface to facilitate the<br>moisturizing, debridement and irrigation of these surfaces.                                                                                          | Mechanical action of fluid moving across the wound or device<br>provides for the mechanism of action and aids in the removal of<br>foreign objects such as dirt and debris.                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                       |                                                                                                                                    |      |
| Chemical<br>Composition           | Sterile Saline (0.9% Sodium Chloride) or Sterile Water for irrigation<br>substantiated to meet USP; no antimicrobial or other substance<br>added.                                                                                                                                                                                                                      | Sterile Saline (0.9% Sodium Chloride) or Sterile Water for<br>irrigation substantiated to meet USP; no antimicrobial or other<br>substance added.                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                       |                                                                                                                                    |      |
| Design<br>Configurations          | <b>CONTENTS</b><br>100 mL Sterile Water, USP<br>250 mL Sterile Water, USP<br>100 mL Normal Saline, USP<br>250 mL Normal Saline, USP<br>500 mL Sterile Water, USP<br>500 mL Normal Saline, USP                                                                                                                                                                          | <b>CONTAINER</b><br>HDPE Bottles made from<br>polyethylene (PE) resin.<br>Include an induction-heat<br>seal and liner with a<br>polypropylene (PP) screw<br>cap closure                                                                                                                                                                                                | <b>CONTENTS</b><br>100 mL USP Sterile Water<br>250 mL USP Sterile Water<br>100 mL USP Normal Saline<br>250 mL USP Normal Saline                                                                                                                                            | Similar                                                                                                                            |      |
|                                   |                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                        | <b>CONTAINER</b><br>HDPE Bottles made from<br>polyethylene (PE) resin.<br>Include an induction sealed<br>inner pull tab with a<br>polypropylene (PP) screw cap<br>closure                                                                                                  |                                                                                                                                    |      |
|                                   |                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                        | 120 mL USP Sterile Water<br>120 mL USP Normal Saline<br><br>210 mL USP Normal Saline                                                                                                                                                                                       | White polypropylene (PP) cup<br>with heat sealed foil lid<br>Nitrogen pressurized<br>aluminum bottle with latex-<br>free inner bag |      |
| Performance<br>Specifications     | The sterile water and sterile saline is manufactured from Water for<br>Irrigation that is designed, tested and operated in accordance with<br>the USP monograph for Water for Irrigation, as specified in USP<br>General Chapter: Water for Pharmaceutical Purposes <1231>                                                                                             |                                                                                                                                                                                                                                                                                                                                                                        | The sterile water and sterile saline is manufactured from Water for<br>Irrigation that is designed, tested and operated in accordance with<br>the USP monograph for Water for Irrigation, as specified in USP<br>General Chapter: Water for Pharmaceutical Purposes <1231> |                                                                                                                                    | Same |

{7}------------------------------------------------

# Table 1: Comparison of Proposed and Predicate Devices

| Device<br>Characteristic          | Proposed Device            | Predicate Device           | Comparison<br>Analysis |
|-----------------------------------|----------------------------|----------------------------|------------------------|
| Prescription vs. OTC              | Prescription Use & OTC Use | Prescription Use & OTC Use | Similar                |
| Contact Durations                 | <24 hours                  | <24 hours                  | Same                   |
| Sterile vs. Non-<br>Sterile       | Sterile                    | Sterile                    | Same                   |
| Sterilization Method              | Gamma Radiation            | Gamma Radiation            | Same                   |
| Disposable vs. Non-<br>Disposable | Disposable                 | Disposable                 | Same                   |
| Single Use vs.<br>Reusable        | Single Use                 | Single Use                 | Same                   |
| Shelf Life                        | í year                     | 2 years                    | Different              |

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image features the Medline logo set against a blue background. The logo consists of the word "MEDLINE" in a sans-serif font, positioned to the left of a stylized white starburst symbol. The starburst has four points, with the top and bottom points being longer than the side points. A small gray rectangle is visible in the upper right corner of the image.

Medline Industries. Inc Three I akes Drive

### Shelf Life and Sterilization

Medline Sterile Water USP and Medline Sterile 0.9% Normal Saline USP are sterilized by gamma radiation and its sterilization has been conducted in accordance with AAMI TIR33, Sterilization of healthcare products - Radiation - Substantiation of a selected sterilization dose - Method VDmax.to ensure that the determined minimum irradiation dose achieves a Sterility Assurance Level (SAL) of 10-0.

Additionally, in accordance with ASTM F1980-16: Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices, aging studies have been conducted to verify the shelf life of the subject device and ensure its sterility is successfully maintained throughout its shelf life.

# Summary of Non-Clinical Testing

Non-clinical verification of Medline Sterile Water USP and Medline Sterile 0.9% Normal Saline USP has been conducted to evaluate its safety, performance and functionality. The results of these tests have demonstrated the overall safety of the proposed device and ultimately support a substantial equivalence determination.

The biocompatibility evaluation for the Medline Sterile Water USP and Medline Sterile 0.9% Normal Saline USP was conducted in accordance with ANSI/AAMI/ISO 10993-1:2009 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process, as recognized by FDA. The proposed device is classified as a surface device with a limited duration of contact (<24 hours) on potentially breached or compromised surfaces. Based on this classification, the following tests were performed to evaluate the biocompatibility of the Medline Sterile Water USP and Sterile 0.9% Normal Saline USP:

- ISO 10993-5: 2009 Biological evaluation of medical devices- Part 5: Tests for in vitro ● cytotoxicity
- ISO 10993-10 Biological evaluation of medical devices- Part 10: Tests for irritation and ● delayed-type hypersensitivity
- ISO 10993-10 Biological evaluation of medical devices- Part 10: Tests for irritation and ● skin sensitization
- ISO 10993-11 Biological evaluation of medical devices- Part 11: Tests for systemic ● toxicity
- USP <151> Pyrogen Test Procedure, Material-Mediated Rabbit Pyrogen Testing ●

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the Medline logo. The logo is set against a blue square background. A white star-like symbol is in the center of the square, and the word "MEDLINE" is written in white letters across the top of the star. There is a gray bar on the right side of the square.

**Medline Industries, Inc.**
Three Lakes Drive
Northfield, IL 60093

## Performance Testing (Bench)

Non-clinical verification of Medline Sterile Water USP and Medline Sterile 0.9% Normal Saline USP has been conducted to evaluate the safety, performance and functionality of the proposed device. The results of these tests have demonstrated the overall safety of the proposed device and ultimately support a substantial equivalence determination.

#### Summary of Clinical Testing

Not applicable. No clinical testing was done on the proposed device.

#### Conclusion

In accordance with 21 CFR Part 807. and based on the information provided in this premarket notification, Medline Industries, Inc. concludes that the Medline Sterile Water USP and Medline Sterile 0.9% Normal Saline USP are as safe and as effective for their intended use as the predicate device Nurse Assist Wound Flush, Sterile Water and Sterile Normal Saline (K083042).

---

**Source:** [https://fda-staging.innolitics.com/device/K173276](https://fda-staging.innolitics.com/device/K173276)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com