K173273 · Imaging Dynamics Company , Ltd. · MQB · Nov 9, 2017 · Radiology
Device Facts
Record ID
K173273
Device Name
Aquarius 8600 1417WCI; Aquarius 8600 1717WCI
Applicant
Imaging Dynamics Company , Ltd.
Product Code
MQB · Radiology
Decision Date
Nov 9, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not intended for mammography.
Device Story
Medical x-ray image acquisition system; utilizes x-ray generator/tube to penetrate patient body; flat panel detector converts x-rays to digital files. Detector interfaces via tethered Gigabit Ethernet or wireless connection. Digital files transmitted to PC running Magellan software (cleared in K170202) for display on monitor. Used in clinical settings by physicians or technologists; patient positioned sitting, standing, or lying. Output consists of digital radiographic images; supports DICOM 3.0 protocol for transfer. System facilitates diagnostic imaging of skeletal and soft tissue structures; aids clinical decision-making by providing high-resolution radiographic visualization.
Clinical Evidence
Bench testing only. Evaluated electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and imaging performance (MTF, DQE) per FDA guidance for solid-state x-ray imaging devices. Clinical images were provided as supplemental evidence to demonstrate system functionality, though not required for substantial equivalence determination.
Technological Characteristics
Stationary digital x-ray system. Components: Amorphous Silicon TFT flat panel detector with CsI:Tl scintillator. Connectivity: Tethered Gigabit Ethernet or wireless. Power: 100-250V AC or rechargeable battery. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-54 (x-ray equipment), IEC 62133 (batteries), NEMA PS 3.1-3.18 (DICOM). Software: Magellan software (previously cleared).
Indications for Use
Indicated for adult and pediatric patients requiring diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, and extremities. Contraindicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
November 9, 2017
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Imaging Dynamics Company Ltd. % Daniel Kamm, P.E. Principal Engineer Kamm & Associates 8870 Ravello Ct. NAPLES FL 34114
Re: K173273
Trade/Device Name: Aquarius 8600 1417WCI; Aquarius 8600 1717WCI Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: October 4, 2017 Received: October 12, 2017
Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert Oels
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173273
Device Name
Aquarius 8600 1417WCI; Aquarius 8600 1717WCI
Indications for Use (Describe)
Intended for use by a qualified/trained doctor or technologist on both and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts on both adult and pediatic patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not intended for mammography
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
- 1. Submitter: Name -Imaging Dynamics Company Suite 130, 3510-29th Street NE Calgary, Alberta, Canada T1Y 7E5 Tel: 403.251.9939 Fax: 403.251.1771 www.imagingdynamics.com Contact - Nicole Wherry Date prepared: September 7, 2017
- 2. Identification of the Device: Device: Aquarius 8600 1417WCl; Aquarius 8600 1717WCI Classification Name: Stationary x-ray system Product Code: MQB Common/Usual Name: Stationary x-ray system (digital) Device Class/Regulation Number: Class II per 21 CFR 892.1680
- 3. Predicate Device: Manufacturer: Imaging Dynamics Company Device: Aquarius 8600 1417TC/1717TC Number: K170202 Classification Name: Stationary x-ray system Product Code: MQB Common/Usual Name: Stationary x-ray system (digital) Device Class/Regulation Number: Class II per 21 CFR 892.1680
- 4. Reference Device: K160810
| Trade name: | ViZion DR + Wireless |
|--------------------|--------------------------|
| Regulation name: | Stationary x-ray system. |
| Regulation number: | 21 CFR 892.1680 |
| Product code: | MQB |
| Device class: | Class II. |
The flat panel detector integrated into the Aquarius 8600 1417WCI is identical to the flat panel detector cleared in K160810
- A description of the device: This device is a medical x-ray image acquisition device. X-rays generated by ട. X-ray generator/ tube that penetrate patient's body are converted to a digital file by the detector. After that the detector sends this digital file to a pc where the companion software has been installed. This software was cleared in our previous submission, K170202. A monitor displays this image. Images can then be transferred via the DICOM protocol. This device is nearly identical to the predicate but now the digital panels have Wi-Fi in addition to Ethernet interfaces. The Aquarius 8600 will be marketed in two possible configurations:
Aquarius 8600 1717WCl (tethered or wireless), 17 x 17 inch flat panel as a retrofit package with Magellan software).
Aquarius 8600 1417WCl (tethered or wireless), 14 x 17 inch flat panel as a retrofit package with Magellan software).
{4}------------------------------------------------
## IDC Recommended Generator Specifications.
Exposure is controlled manually by the user through a hand held trigger. Recommended generator specifications are provided to the user and identified below. If additional generator compatibility information is required customers will be notified to contact the Imaging Dynamics Help Desk.
| Generator | Property | Power Requirement |
|-------------------------------------|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 40 kW CMP 200 DR<br>X-ray generator | Line Voltage | 208 VAC – 5% to 230 VAC + 10%, 1 phase<br>208 VAC – 5% to 230 VAC + 10%, 3 phase<br>400 VAC ± 10%, 3 phase<br>480 VAC ± 10%, 3 phase |
| | Line Frequency | 50/60 Hz |
| | Momentary Current | 275 Amps at 208 VAC (1 phase)<br>154 Amps/phase at 208 VAC (3 phase)<br>250 Amps at 230 VAC (1 phase)<br>139 Amps/phase at 230 VAC (3 phase)<br>80 Amps/phase at 400 VAC<br>65 Amps/phase at 480 VAC |
| | Nominal Current | ≤ 5 Amps |
| | Momentary Power Consumption | 55 kVA |
| | | |
| 50 kW CMP 200 DR<br>X-ray generator | Line Voltage | 208 VAC – 5% to 230 VAC + 10%, 3 phase<br>400 VAC ± 10%, 3 phase<br>480 VAC ± 10%, 3 phase |
| | Line Frequency | 50/60 Hz |
| | Momentary Current | 192 Amps/phase at 208 VAC<br>174 Amps/phase at 230 VAC<br>100 Amps/phase at 400 VAC<br>80 Amps/phase at 480 VAC |
| | Nominal Current | ≤ 5 Amps |
| | Momentary Power Consumption | 65 kVA |
| 65 kW CMP 200 DR<br>X-ray generator | Line Voltage | 400 VAC ± 10%, 3 phase<br>480 VAC ± 10%, 3 phase |
| | Line Frequency | 50/60 Hz |
| | Momentary Current | 125 Amps/phase at 400 VAC<br>105 Amps/phase at 480 VAC |
| | Nominal Current | ≤ 5 Amps |
| | Momentary Power Consumption | 85 kVA |
Indications for Use: Intended for use by a qualified/trained doctor or technologist on both adult and 6. pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not intended for mammography. (Rx Only)
{5}------------------------------------------------
- 7. The Magellan 1417WCl/Magellan 1717WCl panels have essentially the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate device. See the comparison table below.
| Characteristic | Aquarius 8600 Aquarius 8600 1417TC/1717TC<br>K170202<br>Imaging Dynamics Company | Aquarius 8600 1417WCI;<br>Aquarius 8600 1717WCI<br>Imaging Dynamics Company |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications | Intended for use by a qualified/trained doctor<br>or technologist on both adult and pediatric<br>patients for taking diagnostic radiographic<br>exposures of the skull, spinal column, chest,<br>abdomen, extremities, and other body parts<br>on both adult and pediatric patients.<br>Applications can be performed with patient<br>sitting, standing or lying in the prone or supine<br>positions. Not intended for mammography.<br>(Rx Only) | Intended for use by a qualified/trained doctor or<br>technologist on both adult and pediatric patients<br>for taking diagnostic radiographic exposures of<br>the skull, spinal column, chest, abdomen,<br>extremities, and other body parts on both adult<br>and pediatric patients. Applications can be<br>performed with patient sitting, standing or lying<br>in the prone or supine positions. Not intended<br>for mammography. (Rx Only)<br>(EXACTLY THE SAME) |
| Panel Communi-<br>cation | Tethered Gigabit Ethernet | Tethered Gigabit Ethernet or<br>Wireless via previously cleared panel (K160810) |
| Sensor Type | Amorphous Silicon, TFT | Amorphous Silicon, TFT |
| Scintillator | Csl: <span style="text-transform:uppercase">TI</span> | Csl: <span style="text-transform:uppercase">TI</span> |
| Panel Resolution | 2816 x 3328 or 3328 x 3328 | 2304 x 2800 or 3072 x 3072 |
| Panel Size | 14x17 inches or 17 X 17 inches | 14x17 inches or 17 X 17 inches SAME |
| Pixel Size | 127 μm | 150 μm. (14x17) or 139 μm (17x17)<br>NOT A MEANINGFUL DIFFERENCE |
| Resolution | 3.9 lp/mm | 3.59 lp/mm or 3.33 lp/mm<br>NOT A MEANINGFUL DIFFERENCE |
| Image depth | 14 bits | 14 bits SAME |
| Preview Image | 2 seconds | Less than 3 seconds EQUIVALENT |
| Data Output | RAW *the RAW files are convertible into<br>DICOM 3.0 by console software | RAW *the RAW files are convertible into DICOM<br>3.0 by console software |
| DICOM | Yes | SAME |
| Safety/EMC | EN/IEC 60601-1, Safety<br>EN/IEC 60601-1-2 EMC | SAME |
| Operating<br>Environment | Storage and transportation<br>Conditions: Temperature -10 to +50°C<br>Humidity 10 to 80% | SAME |
| MTF @1 lp/mm | 14 x 17: 0.608<br>17 x 17: 0.517 | 14 x 17: 0.667<br>17 x 17: 0.706<br>The new panels have slightly better MTF |
| DQE (0) | 14 x 17: 0.74<br>17 x 17: 0.68 | 14 x 17: 0.40<br>17 x 17: 0.53<br>The new panels have slightly lower DQE |
| Trigger | Auto sense or manual | SAME |
| Characteristic | Aquarius 8600 Aquarius 8600 1417TC/1717TC<br>K170202<br>Imaging Dynamics Company | Aquarius 8600 1417WCI;<br>Aquarius 8600 1717WCI<br>Imaging Dynamics Company |
| Power Source | 100 -250V ~ 50/60Hz 250V ~ | 100 -250V ~ 50/60Hz 250V OR Rechargeable<br>Battery |
| Photo | Image: Aquarius 8600 | Image: Aquarius 8600 |
{6}------------------------------------------------
Comparison Conclusion:
The proposed new device has the following similarities to the predicate:
- Sensor Type a)
- c) Active Area
- d) Sensor Pixel
- e) Dimensions
- Performance characteristics. f)
- IDENTICAL Indications for Use g)
The main difference is the new proposed panels may operate wirelessly or tethered, whereas the predicate panels could only work tethered.
- 8. Description of non-clinical tests. The unit has undergone electrical safety and electromagnetic compatibility testing, as well as integration validation and risk analysis. Bench testing for imaging characteristics such as MTF and DQE was performed in accordance with the FDA Guidance Document on Solid State Imaging Devices: "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices." Labeling complies with FDA guidelines.
Testing was performed in accordance with the following standards:
- a. IEC 60601-1:2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- b. IEC 60601-1-2:2007 Medical electrical equipment - Part 1-2: General requirements for safety; Collateral standard: Electromagnetic compatibility; Requirements and tests
- NEMA PS 3.1~PS 3.18 Digital Imaging and Communications in Medicine (DICOM)][7/31/2008] ﻥ
- d. IEC 60601-2-54 Particular requirements for basic safety and essential performance of X-ray equipment for radiography and radioscopy
{7}------------------------------------------------
- IEC 62133 Secondary cells and batteries containing alkaline or other non-acid electrolytes Safety e. requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications.
- f. Our device software and labeling has addressed cybersecurity via reference to the FDA's cybersecurity guidance: "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices." It was used in the development of the device and precautions were added to the labeling.
- 9. Description of clinical tests. Clinical images were provided; these images were not necessary to establish substantial equivalence based on the modifications to the device (note x-ray system that uses previously cleared detectors) but they provide further evidence in addition to bench testing data to show that the complete system works as intended.
- 10. Conclusions drawn: The nonclinical and clinical tests that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph 3, above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.