MasterLoc Stem: Lateralized Plus

K173267 · Medacta International S.A. · LZO · Dec 7, 2017 · Orthopedic

Device Facts

Record IDK173267
Device NameMasterLoc Stem: Lateralized Plus
ApplicantMedacta International S.A.
Product CodeLZO · Orthopedic
Decision DateDec 7, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The hip prosthesis MasterLoc™ is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. Hip replacement is indicated in the following cases: · Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia; · Avascular necrosis of the femoral head; · Acute traumatic fracture of the femoral head or neck; · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

Device Story

MasterLoc™ Stem: Lateralized Plus is a line extension of the MasterLoc™ cementless femoral stem system. Device provides additional caput-collum-diaphyseal (CCD) angle of 122° and varied neck offsets to accommodate diverse patient anatomies. Implants are used by orthopedic surgeons in hospital settings during total or partial hip arthroplasty. The stem is implanted into the femur to restore joint function and alleviate pain associated with degenerative or traumatic hip conditions. The device is a passive orthopedic implant; it does not involve electronic inputs, software, or active processing. Clinical benefit is derived from the restoration of hip joint biomechanics and stability.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including fatigue testing of the stem and head/neck region per ISO 7206-4 and ISO 7206-6, and range of motion testing per EN ISO 21535.

Technological Characteristics

Cementless, flat, dual-tapered femoral stem. Material: Ti6Al7Nb titanium alloy (ISO 5832-11). Surface: 700 µm Ti coating on proximal 2/3; distal portion glass bead blasted. Taper: Eurocone (12/14, 5° 42' 30"). Sizes: 4-12. Lengths: 126.5-147.5 mm.

Indications for Use

Indicated for patients requiring total or partial hip arthroplasty (primary or revision) due to arthritis, traumatic arthritis, rheumatoid polyarthritis, congenital hip dysplasia, avascular necrosis of the femoral head, acute traumatic fracture of the femoral head/neck, or failure of previous hip surgery.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ December 7, 2017 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Medacta International SA % Elizabeth Rose Manager, Regulatory Affairs Mapi USA, Inc. 2343 Alexandria Drive, Suite 100 Lexington, Kentucky 40504 ## Re: K173267 Trade/Device Name: MasterLoc™ Stem: Lateralized Plus Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, MEH, KWY, LPH, LZY Dated: October 10, 2017 Received: October 11, 2017 Dear Ms. Rose: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Katherine D. Kavlock -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K173267 Device Name MasterLoc™ Stem: Lateralized Plus ### Indications for Use (Describe) The hip prosthesis MasterLoc™ is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. Hip replacement is indicated in the following cases: - · Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia; - · Avascular necrosis of the femoral head; - · Acute traumatic fracture of the femoral head or neck; - · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 3.0 510(k) Summary #### I. Submitter Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66 Contact Person: Stefano Baj, Regulatory Affairs Manager, Medacta International SA Consultant: Elizabeth Rose, Regulatory Affairs Manager, Medical Devices, Mapi USA, Inc. Date Prepared: October 10, 2017 Revised: November 29, 2017 #### II. Device | Device Proprietary Name: | MasterLoc™ Stem: Lateralized Plus | |--------------------------|-------------------------------------------------------| | Common or Usual Name: | Total Hip Prosthesis | | Classification Name: | Hip joint metal/polymer semi-constrained cemented or | | | nonporous uncemented prosthesis | | FDA Product Code: | LZO, MEH, KWY, LPH, LZY | | Regulation Number: | 21 CFR 888.3353, 21 CFR 888.3390, 21 CFR 888.3358, 21 | | | CFR 888.3360 | | Device Classification | 2 | #### III. Predicate Device Substantial equivalence is claimed to the following devices: Primary Predicate: - MasterLoc™ Stem, K151531, Medacta International SA . - MasterLoc™ Stem, K160289, Medacta International SA ● Additional Predicates: - Anthology Hip Stem, K052792, Smith & Nephew, Inc. ● #### Device Description IV. The MasterLoc™ Stem: Lateralized Plus implants are a line extension to Medacta's MasterLoc Stem product line (cleared under K151531 and K160289). The MasterLoc "" Stem: Lateralized Plus implants offer an additional caput-collum-diaphyseal (CCD) angle of 122° and neck offsets for various {4}------------------------------------------------ patient anatomies. The MasterLoc™ Stem: Lateralized Plus implants include cementless, flat, dual tapered design stems intended for total hip arthroplasty in primary or revision surgery. The MasterLoc™ Stem: Lateralized Plus implants are made with a titanium alloy substrate (Ti6Al7Nb) according to ISO 5832-11 Second Edition 2014-09-15: Implants for Surgery -Metallic Materials - Part 11: Wrought Titanium 6-Aluminium 7-Niobium Alloy. The surface treatment consists of Ti coating with a thickness of 700 um in the proximal 2/3 of the shaft to improve proximal fixation. The distal portion of the stems are uncoated with a satin finish obtained from glass bead blasting. The MasterLoc " Stem: Lateralized Plus implants are available in sizes 4 to 12, with a range of stem lengths from 126.5 mm to 147.5 mm. The stems have a Eurocone (12/14 taper with an angle of 5° 42' 30'') and the necks are polished. #### V. Indications for Use The hip prosthesis MasterLoc™ is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. Hip replacement is indicated in the following cases: - · Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia; - · Avascular necrosis of the femoral head; - · Acute traumatic fracture of the femoral head or neck; - · Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement. #### VI. Comparison of Technological Characteristics The MasterLoc™ Stem: Lateralized Plus and the predicate devices share the following characteristics: - . sizes: - lengths; - taper; - substrate material; - coating; - device usage; - sterility; - shelf life; and - packaging. {5}------------------------------------------------ The MasterLoc" Stem: Lateralized Plus implants are technologically different from the predicate devices as follows: - CCD angle, and - neck offset. #### VII. Performance Data Testing was conducted according to written protocols with acceptance criteria that were based on standards. The following mechanical studies were performed in support of a substantial equivalence determination: Non-Clinical Studies - Performance Tests - range of motion (ROM): EN ISO 21535:2009 Non-Active Surgical Implants o Joint Replacement Implants - Specific Requirements for Hip-Joint Replacement Implants: - fatigue testing: ISO 7206-4 Third Edition 2010-06-15 Implants for Surgery o Partial and Total Hip Joint Prostheses - Part 4: Determination of Endurance Properties and Performance of Stemmed Femoral Components [Including AMENDMENT 1 (2016)]; and - fatigue testing: ISO 7206-6 Second Edition 2013-11-15 Implants for Surgery o Partial and Total Hip Joint Prostheses - Part 6: Determination of Endurance Properties of Head and Neck Region of Stemmed Femoral Components. - Pyrogenicity - O the Bacterial Endotoxin Test (LAL test) was conducted according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>) and the pyrogen test according to USP chapter <151> for pyrogenicity determination; and - the subject devices are not labeled as non-pyrogenic or pyrogen free. o Clinical Studies - no clinical studies were conducted. # VIII. Conclusion Based on the above information, the MasterLoc ""Stem: Lateralized Plus can be considered substantially equivalent to the identified predicate devices. Substantial equivalence has been demonstrated through a comparison of intended use, design, and technological characteristics, as well as performance evaluations. The MasterLoc "Stem: Lateralized Plus implants are as safe and effective as the predicate devices, Medacta's MasterLoc " Stem (K151531 and K160289) and Smith & Nephew's Anthology Hip Stem (K052792).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...