← Product Code [OWQ](/productcode/OWQ) · K173262

# Reprocessed ViewFlex Xtra Ice Diagnostic Ultrasound Catheter (K173262)

_Innovative Health, LLC · OWQ · Mar 6, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173262

## Device Facts

- **Applicant:** Innovative Health, LLC
- **Product Code:** [OWQ](/productcode/OWQ.md)
- **Decision Date:** Mar 6, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1200
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Reprocessed ViewFlex Xtra ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.

## Device Story

Reprocessed ViewFlex Xtra ICE is a temporary intracardiac ultrasound catheter; used to visualize cardiac structures, blood flow, and devices (e.g., septal occluders, needles) within the heart. Device connects to compatible intracardiac ultrasound consoles (ViewMate/Phillips CX50) via interface module. Features 9F shaft, 90cm usable length, and four-directional tip deflection (min 120 degrees). Operated by clinicians in clinical settings. Provides real-time visualization to assist procedural guidance; no direct diagnostic use. Reprocessing involves cleaning, decontamination, inspection, and functional testing. Device limited to one reprocessing cycle; tracked and marked to ensure safety.

## Clinical Evidence

Bench testing only. Testing included biocompatibility, cleaning validation, sterilization validation, functional testing (visual inspection, dimensional verification, ultrasound transducer testing, simulated use, mechanical characteristics), electrical safety (dielectric and current leakage), and packaging validation.

## Technological Characteristics

9F diagnostic ultrasound catheter; 90cm usable length; four-directional tip deflection (min 120 degrees). Compatible with St. Jude Medical ViewMate/II/Z and Phillips CX50 consoles. Reprocessed via cleaning and decontamination; sterilized. Single-use reprocessing limit. No changes to original manufacturer specifications.

## Regulatory Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

## Predicate Devices

- ViewFlex Xtra ICE Catheter ([K133853](/device/K133853.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

March 6, 2018

Innovative Health, LLC. Amy Stoklas-Oakes Director. Ouality and Regulatory 1435 North Hayden Road Suite 100 Scottsdale, Arizona 85257

Re: K173262

Trade/Device Name: Reprocessed ViewFlex Xtra Ice Diagnostic Ultrasound Catheter (See enclosed list for model numbers) Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OWO Dated: February 1, 2018 Received: February 5, 2018

Dear Amy Stoklas-Oakes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

M & Hillelemen

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

The item numbers in the scope of this submission are as follows:

| REF Item<br>Number | Catalog<br>ltem<br>Number | Description                                         | Sheath<br>Usable<br>Length<br>(cm) | French<br>Size | System<br>Compatibility                                                                 |
|--------------------|---------------------------|-----------------------------------------------------|------------------------------------|----------------|-----------------------------------------------------------------------------------------|
| D087031            | 100046963                 | ViewFlex Xtra ICE Diagnostic<br>Ultrasound Catheter | 90                                 | 9F             | St. Jude Medical<br>ViewMate.<br>ViewMate II,<br>ViewMate Z<br>Console/Phillips<br>CX50 |

Table 1: Devices in Scope

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K173262

Device Name

Reprocessed ViewFlex Xtra ICE Diagnostic Ultrasound Catheter

Indications for Use (Describe)

The Reprocessed ViewFlex Xtra ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# SECTION 5: 510(k) SUMMARY

As required by 21 CFR 807.92(c)

## Submitter's Name and Address:

Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257

## Contact Name and Information:

Amy Stoklas-Oakes Director. Quality and Requlatory Innovative Health, LLC. (480) 525-5972 (office) (888) 965-7705 (fax) astoklas-oakes@innovative-health.com

### Date prepared:

October 10, 2017

## Device Information:

Trade/Proprietary Name:

Common Name: Classification Name: Classification Number: Product Code:

Reprocessed ViewFlex Xtra ICE Diagnostic Ultrasound Catheters ICE Catheter Reprocessed Intravascular Ultrasound Catheter Class II, 21 CFR 870.1200 owq

## Predicate Device:

| 510(k) Number | 510(k) Title               | Manufacturer                                          |
|---------------|----------------------------|-------------------------------------------------------|
| K133853       | ViewFlex Xtra ICE Catheter | Irvine Biomedical, Inc. a<br>St. Jude Medical Company |

#### Device Description:

The Reprocessed ViewFlex Xtra ICE is a temporary intracardiac ultrasound catheter intended for use in patients to accurately visualize cardiac structures, blood flow and other devices within the heart when connected to a compatible intracardiac ultrasound console via the compatible ViewFlex Catheter Interface Module. Examples of the types of other devices that can be visualized include, and are not limited to, intracardiac catheters, septal occluders, delivery wires, delivery sheaths, sizing balloons and transseptal needles. The use of these images is limited to visualization with no direct or indirect diagnostic use. The ViewFlex Xtra ICE catheter has a useable length of 90 cm, with a 9 French (F) shaft with an ultrasound transducer. A 10F introducer is recommended for use with this catheter for insertion into the femoral or jugular veins. The catheter tip has four-directional deflection allowing for Left-Right and Posterior-Anterior deflection, with an angle of at least 120 degrees in each direction.

{5}------------------------------------------------

| REF Item Number | Catalog/Reorder Item Number | Description                                      | Sheath Usable Length (cm) | French Size | System Compatibility                                                     |
|-----------------|-----------------------------|--------------------------------------------------|---------------------------|-------------|--------------------------------------------------------------------------|
| D087031         | 100046963                   | ViewFlex Xtra ICE Diagnostic Ultrasound Catheter | 90                        | 9F          | St. Jude Medical ViewMate, ViewMate II, ViewMate Z Console/Phillips CX50 |

The item numbers in scope of this submission are as follows:

Table 5.1: Device Scope

# Indications for Use:

The Reprocessed ViewFlex Xtra ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.

# Technological Characteristics:

The purpose, design, materials, function, and intended use of the Reprocessed Diagnostic Ultrasound Catheters are identical to the predicate devices. There are no changes to the claims, clinical applications, patient populations, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of the Diagnostic Ultrasound Catheter includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.

# Functional and Safety Testing:

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed Diagnostic Ultrasound Catheters. This included the followina:

- Biocompatibility .
- Cleaning Validation .
- Sterilization Validation ●
- Functional testing .
	- . Visual Inspection
	- . Dimensional Verification
	- . Ultrasound Transducer Testing
	- . Simulated Use
	- Mechanical Characteristics ■
- Electrical Safety Testing .
	- · Dielectric and Current Leakage
- . Packaging Validation

The Reprocessed Diagnostic Ultrasound Catheters are reprocessed no more than one (1) time. Each device is marked and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.

# Conclusion:

Innovative Health concludes that the Reprocessed Diagnostic Ultrasound Catheter is as safe and effective as the predicate devices described herein.

---

**Source:** [https://fda-staging.innolitics.com/device/K173262](https://fda-staging.innolitics.com/device/K173262)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
