The ToxCO® Breath Carbon Monoxide Monitor and accessories are used by healthcare professionals to determine levels of Carbon Monoxide (CO) poisoning.
Device Story
Handheld, battery-powered exhaled breath CO monitor; uses electrochemical sensor to detect CO; input is exhaled breath via disposable SteriBreath mouthpiece (standard) or face mask sampling system (for unconscious/injured patients). Microcontroller amplifies and processes sensor signal; displays peak CO concentration (ppm) and calculated %COHb on LCD touch screen. Used by healthcare professionals in medical institutions/healthcare environments. Provides real-time CO levels to assist in diagnosing CO poisoning. Benefits include rapid, non-invasive assessment of CO exposure in diverse patient populations, including those unable to perform standard breath-hold maneuvers.
Clinical Evidence
Bench testing only. Validation included accuracy and repeatability testing using calibrated CO gases (0-605 ppm) across operating (0-45°C) and storage (0-50°C) temperatures. Face mask sampling performance verified using a volume/flow simulator to mimic human respiratory rates (12-20 bpm) and tidal volumes, confirming readings remain within specified tolerances.
Technological Characteristics
Handheld, battery-powered (3x AA). Electrochemical sensor. Enclosure: Polycarbonate/ABS blend with SteriTouch antimicrobial additive. Mouthpiece: Polypropylene. Face mask system: PVC/Polyethylene, Styrene Butadiene, Polypropylene, Silicone. Connectivity: Standalone. Software: Moderate level of concern. Complies with IEC 60601-1 (Electrical Safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for healthcare professionals to determine levels of Carbon Monoxide (CO) poisoning in patients, including those who are unconscious or injured, where standard breath sampling is not possible.
Regulatory Classification
Identification
A carbon monoxide gas analyzer is a device intended to measure the concentration of carbon monoxide in a gas mixture to aid in determining the patient's ventilatory status. The device may use techniques such as infrared absorption or gas chromatography.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Bedfont Scientific Ltd Louise Bateman OA & RA Manager Station Road, Harrietsham Maidstone, ME17 1JA UNITED KINGDOM
Re: K173238
Trade/Device Name: ToxCO Regulation Number: 21 CFR 868.1430 Regulation Name: Carbon Monoxide Gas Analyzer Regulatory Class: Class II Product Code: CCJ Dated: March 5, 2019 Received: March 8, 2019
Dear Louise Bateman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Todd D. Courtney -5 for Tina Kiang, Ph.D. Acting Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173238
Device Name ToxCO®
Indications for Use (Describe)
The ToxCO® breath Carbon Monoxide monitor and accessories are used by healthcare professionals to determine levels of Carbon Monoxide (CO) poisoning.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Bedfont Scientific Ltd. The logo features a series of colorful circles in yellow, pink, orange, and purple above a larger blue circle. Below the circles, the word "bedfont" is written in a bold, sans-serif font, with the registered trademark symbol next to it. Underneath "bedfont", the text "est. 1976" is displayed.
ToxCO®
# 510K Number _K173238
### 510(k) Summary
This summary of 510(k) information is being submitted in accordance with the requirement of 21 CFR 807.92.
- l. SUBMITTER
Identification:
Bedfont Scientific Ltd Station Road, Harrietsham Maidstone, Kent. ME17 1JA United Kingdom
- Telephone: +44 1622 851122 Fax: +44 1622 854860
Contact Person: Louise Bateman Email louise@bedfont.com Date of Summary: 04 April 2019
### II. DEVICE
| Device Name: | ToxCO® |
|-----------------------------|-----------------------------------------------|
| Classification Name: | Analyzer, Gas, Carbon Monoxide, Gaseous-Phase |
| Product Regulatory Class: | |
| Product Regulatory Code: | CCJ |
| Panel: | Anesthesiology |
| Code of Federal Regulation: | 21 CFR 868.1430 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Bedfont Scientific Ltd. The logo features the word "bedfont" in a bold, lowercase font, with the registered trademark symbol next to it. Above the word, there are five colorful circles in yellow, pink, orange, purple, and blue. Below the word "bedfont" is the text "est. 1976".
#### III. PREDIČATE AND REFERENCE DEVIČE
Predicate Device Manufacturer: Device Trade Name: Device 510(k):
Bedfont Scientific Ltd Micro*™ Smokerlyzer K082315
Reference only Predicate Device Manufacturer: Bedfont Scientific Ltd Device Trade Name: EC ToxCO® + Device 510(k): K070259
EC ToxCO® +
K070259
#### IV. DEVICE DESCRIPTION
The ToxCO® is a hand held exhaled breath monitor for the detection of Carbon Monoxide on the breath. The battery powered monitor uses an electrochemical sensor, designed to react specifically to carbon monoxide producing an electrical output. The output is then amplified and a microcontroller processes the signal and converts it to a meaningful displayed result on an LCD touch screen.
The sample system for the ToxCO® mimics the predicate device channeling the breath sample directly over the sensor during test ensuring the sensor is exposed to the gas sample for the required length of time to give an accurate reading.
The ToxCO® Monitor uses a non-patient contacting D-piece™ sampling system, with integrated bacterial and viral filter and a one-way valve, attached directly to the monitor. A patient contacting, single patient use SteriBreath™ mouthpiece is connected to the D-piece™ sampling system to perform a breath sample. The D-piece™ is designed to have minimal dead space and therefore initial dilution of the sample is reduced.
The patient is required to hold their breath for a 15 second countdown. This is displayed on the LCD screen of the device. At the end of the breath hold, the patient shall blow gently but fully into the ToxCO®, exhaling as much of the breath in their lungs as possible. The reading on the display shall rise until the peak reading is held on the display.
Use of the breath sampling D-piece™ and disposable SteriBreath™ mouthpiece may be impossible if the patient is unconscious or injured. In this event, a modified sampling technique can be used, which consists of a specially constructed face mask sampling system, which allows exhaled breath to be directed to the instrument's sensor for analysis.
The Face mask sampling system is a single-use, pre-assembled adapter to enable a breath sample to be taken with a single use face mask, connected to the ToxCO® breath Carbon Monoxide monitor by means of a breath sampling D-piece™.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Bedfont Scientific. The logo features five colorful circles in yellow, pink, orange, purple, and blue. Below the circles is the word "bedfont" in a sans-serif font, with the registered trademark symbol next to it.
The face mask sampling procedure does not require a 15 second breath hold to be performed before the test can begin. Sampling will last 60 seconds as the %COHb/ppm levels rise and then hold at the peak level. The result will be shown on the LCD monitor screen.
The ToxCO® breath Carbon Monoxide monitor and accessories are used by healthcare professionals in medical institutions and healthcare environments where Carbon Monoxide exposure is suspected.
#### V. INTENDED USE / INDICATIONS FOR USE
The ToxCO® Breath Carbon Monoxide Monitor and accessories are used by healthcare professionals to determine levels of Carbon Monoxide (CO) poisoning.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The ToxCO® Breath Carbon Monoxide Monitor and its accessories are used as indication of Carbon Monoxide (CO) poisoning and Carboxyhaemoglobin (COHb) when blood testing is not available. The predicate Micro+® Smokerlyzer and the ToxCO® monitor are substantially equivalent as they are both breath carbon monoxide monitors that can be used to provide a reading to determine levels of Carbon Monoxide (CO) poisoning and calculation of the carboxyhaemoglobin (COHb) level.
The predicate Micro+® Smokerlyzer and ToxCO® monitor are both for multi-patient use by health professionals; both used in health environments and used to determine levels of carbon monoxide (CO) poisoning.
The predicate and the ToxCO® monitor are manufactured using the same processes. They are both hand held, battery operated breath Carbon Monoxide Monitors manufactured with identical materials, components, technology and use the same electrochemical sensor.
The Reference Device EC ToxCO® + (K070259) has been included to determine substantial equivalence as the EC ToxCO+ and the subject ToxCO® monitor both utilize a facemask sampling system, using commercially available facemasks.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Bedfont Scientific Ltd. The logo features the word "bedfont" in a lowercase, sans-serif font, with the "o" in "bedfont" colored in a bright blue that matches the color of a large circle above the word. Above the word "bedfont" are five smaller circles in a variety of colors: yellow, pink, orange, purple, and blue.
The ToxCO® Breath Carbon Monoxide Monitor is substantially equivalent to the predicate Micro+ Smokerlyzer Device (K082315) in the following areas:
Equivalence, Similarities and Differences: ToxCO® and Predicate Micro+™ Smokerlyzer®
| Substantial<br>Equivalent<br>comparison<br>Criteria | ToxCO® Monitor | Predicate Device<br>Micro+™ Smokerlyzer® | Equivalence, Similarities &<br>Differences |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Bedfont Scientific Ltd | Bedfont Scientific Ltd | EQUIVALENT |
| Trade name | ToxCO® | Micro+™ Smokerlyzer® | n/a |
| Product code | CCJ | CCJ | EQUIVALENT |
| Regulation<br>number | 868.1430 | 868.1430 | EQUIVALENT |
| Intended use /<br>Indications for<br>use | The ToxCO® Breath<br>Carbon Monoxide<br>Monitor and accessories<br>are used by healthcare<br>professionals to<br>determine levels of<br>Carbon Monoxide (CO)<br>poisoning. | The Micro+™ Smokerlyzer®<br>is a Breath Carbon<br>Monoxide Monitor<br>intended for multi-<br>patient use by healthcare<br>professionals in smoking<br>cessation programs and<br>as an indicator of Carbon<br>Monoxide poisoning in<br>healthcare environments. | EQUIVALENT<br>Both devices are breath<br>carbon monoxide monitors<br>that can be used to provide<br>a reading to<br>determine/indicate levels of<br>Carbon Monoxide (CO)<br>poisoning. |
| Design | The ToxCO® was designed<br>to be hand held, easy to<br>use with different<br>sampling techniques. | The Micro+™ Smokerlyzer®<br>was designed to be hand<br>held, easy to use and has<br>one sampling technique. | EQUIVALENT<br>Both devices are handheld,<br>easy to use Breath Carbon<br>Monoxide (CO) Monitors.<br>Both devices offer a Breath<br>Sampling technique that<br>uses the same disposable<br>single use SteriBreath<br>mouthpiece and D-Piece<br>mouthpiece filter.<br>DIFFERENCES<br>The ToxCO® device offers a<br>modified sampling<br>technique if the patient is<br>unconscious or injured,<br>using a disposable single<br>use Facemask sampling<br>system and single use<br>Facemask. |
| CO measurement<br>range - Breath<br>Test | 0-500 ppm | 0-500 ppm | EQUIVALENT |
| CO measurement<br>range - modified<br>Breath Test | 0-200 ppm | n/a | DIFFERENT<br>The predicate Micro+ does<br>not include the option to<br>perform a breath test with a<br>modified sampling<br>technique, consisting of a<br>specially constructed face<br>mask sampling system. This<br>was available on the EC<br>ToxCO+ reference device.<br>This method for use has<br>been validated as an<br>effective method to indicate<br>levels of CO poisoning up to<br>up to 200ppm. |
| Detection<br>principle | Electrochemical sensor | Electrochemical sensor | EQUIVALENT |
| Sensor sensitivity | 1 ppm | 1 ppm | EQUIVALENT |
| Accuracy | ≤ ±3ppm/10% -<br>whichever is greater | ≤±2ppm/5% whichever is<br>greater | SIMILAR<br>The operating range on the<br>ToxCO, is the same as the<br>Micro+ & the reference<br>ToxCO+, which is 0-500ppm<br>(0 – 50%COHb -<br>carboxyhaemoglobin).<br>The accuracy of the ToxCO<br>is ≤±3ppm or 10%<br>(whichever is greater),<br>which is due to the extend<br>temperature range being 0<br>to 45°C. |
| Power | 3 x AA batteries | 3 x AA batteries | EQUIVALENT |
| Monitor outer<br>enclosure<br>material | Polycarbonate/ABS blend<br>with SteriTouch® Anti-<br>microbial Additive | Polycarbonate/ABS blend<br>with SteriTouch® Anti-<br>microbial Additive | EQUIVALENT |
| Mouthpiece<br>Sample System | Single Use SteriBreath<br>Mouthpiece -<br>Polypropylene | Single Use SteriBreath<br>Mouthpiece -<br>Polypropylene | EQUIVALENT |
| Modified Sample<br>System | Single Use Facemask -<br>PVC/Polyethylene<br>Facemask sampling<br>system -<br>Styrene Butadiene<br>Polypropylene<br>Silicone | Not Applicable | DIFFERENT<br>Use of the breath sampling<br>D-piece™ and disposable<br>SteriBreath™ mouthpiece<br>may be impossible if the<br>patient is unconscious or<br>injured. In this event, a<br>modified sampling<br>technique can be used,<br>which consists of a specially<br>constructed face mask<br>sampling system, which<br>allows exhaled breath to be<br>directed to the instrument's<br>sensor for analysis. Refer to<br>reference EC ToxCO+ below<br>for equivalence. |
| Medical class | II | II | EQUIVALENT |
| Electrical Safety<br>Testing<br>IEC 60601-1 | Yes | Yes | EQUIVALENT |
| EMC Testing<br>IEC 60601-1-2 | Yes | Yes | EQUIVALENT |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Bedfont logo. The logo consists of five colored circles in yellow, pink, orange, purple, and blue. Below the circles is the word "bedfont" in blue, and below that is the text "est. 1976".
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Bedfont Scientific Ltd. The logo features the word "bedfont" in a bold, dark blue font. Above the word, there are five colorful circles in yellow, pink, orange, purple, and blue. The text "est. 1976" is located below the word "bedfont".
### Equivalence, Similarities and Differences: ToxCO® and Predicate Reference Device EC ToxCO+
| Substantial<br>Equivalent<br>comparison<br>Criteria | ToxCO® Monitor | Reference Device<br>EC ToxCO+ | Equivalence & Differences |
|-----------------------------------------------------|------------------------|-------------------------------|---------------------------|
| Manufacturer | Bedfont Scientific Ltd | Bedfont Scientific Ltd | EQUIVALENT |
| Trade name | ToxCO® | EC ToxCO+ | n/a |
| Product code | CC1 | CCJ | EQUIVALENT |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Bedfont Scientific Ltd. The logo features five colorful circles in yellow, pink, orange, purple, and blue. Below the circles is the word "bedfont" in a bold, sans-serif font, with the registered trademark symbol next to it. Underneath "bedfont" is the text "est. 1976".
| Regulation<br>number | 868.1430 | 868.1430 | EQUIVALENT |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CO measurement<br>range - Breath<br>Test | 0-500 ppm | 0-500 ppm | EQUIVALENT |
| CO measurement<br>range - modified<br>Breath Test | 0-200 ppm | 0-500 ppm | SIMILAR<br>The ToxCO and reference<br>predicate EC ToxCO+<br>includes the option to<br>perform a breath test with a<br>modified sampling<br>technique, consisting of a<br>specially constructed face<br>mask sampling system. This<br>method for use has been<br>validated as an effective<br>method to indicate levels of<br>CO poisoning up to up to<br>200ppm. |
| Modified Sample<br>System | Single Use Facemask -<br>PVC/Polyethylene<br>Facemask sampling<br>system –<br>Styrene Butadiene<br>Polypropylene<br>Silicone | Single Use Facemask -<br>PVC/Polyethylene<br>Facemask sampling<br>system –<br>Styrene Butadiene<br>Polypropylene<br>Silicone | EQUIVALENT |
### Differences & Similarities discussion
The ToxCO® and its predicate Micro+™ Smokerlyzer® are breath carbon monoxide monitors that can be used to provide a reading to determine levels of Carbon Monoxide (CO) poisoning. Both Devices are for Multi-patient use by Health Professionals and both can be used in Healthcare environments.
Both the ToxCO® and its predicate Micro+™ Smokerlyzer® calculate the carboxyhaemoglobin (COHb) level and can display the % carboxyhaemoglobin (COHb) level on their LCD touch screen, using the same sensor to enable calculation via the same internal processor.
Although the ToxCO® is not specifically indicated for smoking cessation programs as the predicate Micro+™ Smokerlyzer®, this does not raise different questions of safety and effectiveness of the ToxCO® monitor as both devices determine levels of Carbon Monoxide (CO) poisoning, which is monitored regularly during smoking cessation programs. Both devices use the same sensor and internal processor to determine levels of Carbon Monoxide (CO) poisoning.
The ToxCO® breath Carbon Monoxide monitor's predicate device, the Micro+™ Smokerlyzer® is not for use on unconscious patients, the ToxCO® breath Carbon Monoxide Monitor has been developed with an alternative sampling system via a Facemask and sampling adapter for use in these circumstances. The Facemask sampling system was previously used in the Bedfont Scientific Ltd EC50
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "ToxCO" in bold, black font. A circled capital R symbol is to the right of the word. The text is simple and clear, with a focus on the company name.
Image /page/10/Picture/1 description: The image shows the logo for Bedfont Scientific Ltd. The logo features five colorful circles in yellow, pink, orange, purple, and blue. Below the circles is the word "bedfont" in a bold, sans-serif font, with the registered trademark symbol. Underneath "bedfont" is the text "est. 1976".
ToxCO® + breath Carbon Monoxide Monitor cleared for market under K070259. This device also used a Facemask sampling system, utilizing commercially available facemasks. Use of the breath sampling Dpiece™ and disposable SteriBreath™ mouthpiece may be impossible if the patient is unconscious or injured. In this event, the modified sampling technique can be used, which consists of a specially constructed face mask sampling system, which allows exhaled breath to the instrument's sensor for analysis. The performance of the ToxCO® using the alternative sampling system has been comprehensively tested to confirm its effectiveness.
These minor differences in the intended use and breath sampling method do not raise different questions of safety and effectiveness of the ToxCO® breath Carbon Monoxide monitor, therefore it can be considered substantially equivalent to its predicate device.
#### VII. PERFORMANCE DATA
The following performance data was provided in support of the substantial equivalence determination:
### Biocompatibility Evaluation
The evaluation of biocompatibility requirements for the ToxCO® breath Carbon Monoxide monitor and accessories was performed in accordance with FDA recognized Consensus Standard ISO 10993-1, Part 1: Evaluation and testing within a risk management process.
### Electrical Safety and Electromagnetic Compatibility (EMC)
The ToxCO® breath carbon monoxide monitor was subject to Electrical Safety Testing and EMC testing. The monitor complies with IEC 60601-1 Electrical Safety Standard and IEC 60601-1-2 EMC standards.
### Software (Firmware) Verification and Validation Testing
The ToxCO® breath carbon monoxide monitor includes Firmware that was subject to Verification and Validation testing performed as recommended by FDA's Guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software was determined to be a Moderate level of concern.
### Performance testing
Bench testing was performed during the Validation and Verification cycle to confirm the device performs as intended across its operational range. Validation tests were carried out with calibrated CO gases (0, 20, 50, 158, 605 ppm) to confirm Accuracy of readings and repeatability within specification .Further testing was executed to confirm the device performs across the specified Operating (0-45°c) and Storage (0-50°c) temperatures, carried out with calibrated CO gases (0, 20, 50, 158, 500, 605 ppm) and to also to ensure the Device responds to CO gas, calculating and displaying the correct PPM and %COHB, carried out with calibrated CO gases (0, 20, 50, 158, 605 ppm).
Comprehensive testing was carried out to verify the performance of the ToxCO® device when used in facemask mode. A calibrated Volume/Flow simulator was used to adjust the relevant respiratory rates and tidal volume parameters enabling simulation of human lungs. A Facemask sampling adapter was connected to the ToxCO® device and the simulator, with a CO Monitor to measure the sampled air and to measure the CO concentration when the calibrated test CO gas was applied (0 (clean air), 50, 158, 605 ppm). It was concluded the ToxCO® is able to provide readings within the operating tolerances specified when tested across the detection range.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Bedfont Scientific Ltd. The logo features five colorful circles in yellow, pink, orange, purple, and blue. Below the circles is the word "bedfont" in a lowercase, sans-serif font, with the registered trademark symbol. Underneath "bedfont" is the text "est. 1976".
#### CONCLUSION VIII.
After analyzing Performance data, bench and non-clinical data it is the conclusion of Bedfont Scientific Ltd that the ToxCO® breath Carbon Monoxide monitor is as safe and effective as the predicate device, with equivalence in design, materials, technology, classification and Intended Use as a breath Carbon Monoxide monitor. The subject and predicate device both use identical moldings made from the same material, size and weight. The same mouthpiece is used for the ToxCO® and predicate device again using the same material, classification of applied part, touch times and internal breath sampling system.
Equivalence is also established between the predicate and subject Device as the ToxCO® is a Multipatient use device, used by health professionals in healthcare environments, providing a reading to determine levels of Carbon Monoxide (CO) poisoning. These are all identical characteristics to the Predicate device.
The few minor differences present when compared to the Micro+™ Smokerlyzer® predicate device include the alternative breath sampling method and the face mask breath test concentration range. These do not raise different questions of safety and effectiveness of the ToxCO® breath Carbon Monoxide monitor as comprehensive testing of the Facemask sampling method was performed to declare the device is able to provide readings within the operating tolerances specified. The face mask testing protocol has been developed for use on patients with a respiration rate of between 12-20 breaths per minute. The manual advises if face mask testing is used outside of these specifications, this may result in decreased accuracy of readings.
Both devices are Product Regulatory Class II and Product Regulatory Code CCJ and no additional risks are generated to achieve the desired intended use, therefore the ToxCO® breath carbon monoxide monitor can be considered substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.