The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.
Device Story
Aristotle 14 Guidewire is a 0.014" diameter steerable guidewire with shapeable Nitinol distal tip; stainless steel core; hydrophilic distal coating; PTFE proximal coating; radiopaque marker coil. Supplied sterile for single use. Includes introducer and torque device accessories. Used in neuro and peripheral vasculature to navigate and position catheters/interventional devices. Operated by physicians in clinical settings. Provides tactile feedback and steerability during navigation; radiopaque tip allows fluoroscopic visualization. Benefits include facilitating access to complex vascular anatomy for interventional procedures.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing including biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility, latex content), functional testing (visual inspection, dimensional verification, flexing, tensile strength, tip shape retention, torqueability, torque strength, tip flexibility, fracture, coating lubricity/durability/integrity, particulates, accessory testing), and simulated use/cadaver testing. All tests met acceptance criteria.
Technological Characteristics
0.014" (0.36mm) diameter; 200-300cm length. Stainless steel core; Nitinol distal tip. Hydrophilic distal coating; PTFE proximal coating. Radiopaque marker at distal tip. Sterilized via 100% Ethylene Oxide (EO) per ISO 11135. Accessories: introducer and torque device. MRI Unsafe.
Indications for Use
Indicated for general vascular use in neuro and peripheral vasculatures to introduce and position catheters and interventional devices. Not for use in coronary vasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 22, 2018
Scientia Vascular LLC % Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW Buffalo, Minnesota 55313
Re: K173235
Trade/Device Name: Aristotle 14 Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: December 13, 2017 Received: December 14, 2017
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K173235
Device Name
Aristotle 14 Guidewire
#### Indications for Use (Describe)
The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Scientia Vascular LLC Traditional 510(k) Aristotle 14 Guidewire
Image /page/3/Picture/2 description: The image shows the logo for Scientia. The logo consists of a stylized blue symbol above the word "SCIENTIA" in a futuristic-looking font. The symbol is made up of two curved lines that are parallel to each other.
# 510(K) SUMMARY (Per 21 CFR 807.92)
## SCIENTIA VASCULAR LLC ARISTOTLE 14 GUIDEWIRE
| 510(k) Sponsor: | Scientia Vascular LLC<br>3487 West 2100 South Suite 100<br>West Valley City, UT 84119<br>Tel: (775) 657-6330 | |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------|--|
| Contact Person: | David Sabodski, Director of Quality Assurance<br>Tel: 801.573.5897<br>E-mail: dsabodski@scientiavascular.com | |
| Date Prepared: | December 8, 2017 | |
| Prepared by: | Ryan O'Callaghan, MS, RAC<br>Phil Triolo and Associates LC<br>Tel: 801.699.9846<br>Fax: 801.328.2399<br>E-mail: ryano@philt.com | |
| Trade Name: | Aristotle 14 Guidewire | |
| Common Name: | Guidewire | |
| Classification Name | Catheter Guide Wire per 21 CFR 870.1330 | |
| Product Code: | DQX | |
| Predicate Device: | PVS 1300 Synchro® 0.014" Guidewire (K032146) | |
{4}------------------------------------------------
#### DEVICE DESCRIPTION
The Scientia Vascular Aristotle 14 Guidewire is a 0.014" diameter steerable guidewire with a shapeable tip to aid in accessing vasculature. The guidewire is supplied sterile and is for single use only. It is provided in a range of stiffness profiles, from soft to support, The product is provided in 200cm and 300cm lengths.
The distal portion of the guidewire tip includes a radiopaque platinum wire marker coil to facilitate fluoroscopic visualization. The guidewire has a hydrophilic polymer coating on the distal portion and a polytetrafluoroethylene (PTFE) coating on the proximal portion to reduce friction during manipulation in vessels.
The guidewire is provided with an introducer (to aid with the insertion of the guidewire into a catheter hub and/or a hemostasis valve) and a torque device (to attach to the proximal portion to facilitate gripping and manipulation of the guidewire during use). The introducer and torque accessory devices are included to facilitate use of the guidewire and are not intended to contact the patient's body.
The Aristotle 14 guidewire is substantially equivalent with respect to technological characteristics, design and materials to Boston Scientific - Precision Vascular's currently marketed PVS 1300 Synchro® 0.014" Guidewire cleared under K032146. Both devices are provided with a torque device and introducer.
## INDICATIONS FOR USE
The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.
The indications for use statement is equivalent to that of the predicate device and the differences in wording are not critical to the intended therapeutic, diagnostic, prosthetic, surgical or other use of the device. The differences do not affect the safety and effectiveness of the device when used as labeled, as the same intent is documented for the predicate and new devices.
#### TECHNOLOGICAL CHARACTERISTICS
As shown in the table below, the technological characteristics of the Aristotle 14 Guidewire are equivalent to those of the predicate device, the PVS 1300 Synchro® 0.014" Guidewire.
{5}------------------------------------------------
| Comparison between Subject & Predicate Device Technological Characteristics | | | |
|-----------------------------------------------------------------------------|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|------------|
| Characteristic | Subject Device<br>Aristotle 14 Guidewire | PVS 1300 Synchro®<br>0.014" Guidewire<br>(K032146) | Comparison |
| Anatomical<br>Location | Neuro and peripheral<br>vasculature | Neuro and peripheral<br>vasculature | Same |
| Dimensions | 0.D.: 0.014" (0.36mm)<br>Length: 200cm to 300cm<br>range | 0.D.: 0.014" (0.36mm)<br>Length: 180cm to 300cm<br>range, with 200cm being<br>typical | Equivalent |
| Core Wire | Stainless Steel | Stainless Steel | Same |
| Distal Tip | Shapeable<br>Length: 35cm<br>Material: Nitinol | Shapeable<br>Length: 25cm to 65cm<br>range, with 35 to 45cm<br>being typical<br>Material: Nitinol | Equivalent |
| Stiffness Profiles | Range from support (stiff)<br>to flex (soft) | Range from support (stiff)<br>to flex (soft) | Equivalent |
| Coatings | Distal End: Hydrophilic<br>Proximal End: PTFE | Distal End: Hydrophilic<br>Proximal End: PTFE | Equivalent |
| Radiopaque<br>Marker | 1 radiopaque marker at<br>distal tip | 1 radiopaque marker at<br>distal tip | Same |
| Introducer<br>(Accessory) | Provided with each<br>guidewire | Provided with each<br>guidewire | Equivalent |
| Torque Device<br>(Accessory) | Provided with each<br>guidewire | Provided with each<br>guidewire | Same |
| Sterilization<br>Method | 100% Ethylene Oxide<br>(EO) | Currently radiation<br>sterilization [devices<br>originally cleared were<br>sterilized with100%<br>Ethylene Oxide (EO)] | Same |
{6}------------------------------------------------
Scientia Vascular LLC Traditional 510(k) Aristotle 14 Guidewire
Results of tests performed on the new Aristotle 14 Guidewire demonstrate that the new guidewire performs as well as the predicate device and/or meets requirements of relevant standards. Further, any differences in technological characteristics of the Aristotle Guidewires when compared with predicate device characteristics do not raise different questions of safety and effectiveness.
## NON-CLINICAL PERFORMANCE TESTS
#### Biocompatibilitv
Biocompatibility of the Aristotle 14 Guidewire and accessory materials has been verified in accordance with ISO 10993-1:2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and FDA's Guidance for Industry and Food and Drug Administration Staff Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued June 16, 2016.
The results of the following biological and toxicological safety evaluations verified the biocompatibility of the subject device when tested as an external communicating, blood contact, limited duration (<24 hours) device:
- . Cytotoxicity;
- . Sensitization;
- Irritation/Intracutaneous Reactivity:
- Acute Systemic Toxicity;
- Materials-Mediated Pyrogenicity:
- . Hemocompatibility
- Hemolysis by Direct Contact and Extract: o
- Partial Thromboplastin Time (PTT); о
- Complement Activation of C3a and SC5b-9; O
- Thrombogenicity in a Dog Model; and O
- Latex Antigenic Protein Content
Results of these tests are summarized in the following table.
| Summaries of Biocompatibility Tests Conducted to Support this Premarket<br>Notification | | | |
|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|--|
| Test | Results | Conclusions | |
| Cytotoxicity<br>[L-929 MEM Elution] | Cell culture treated with<br>test sample exhibited no<br>reactivity (Grade 0). | Non-cytotoxic | |
| Summaries of Biocompatibility Tests Conducted to Support this Premarket<br>Notification | | | |
| Test | Results | Conclusions | |
| Sensitization<br>[Maximization<br>(Magnusson-Kligman)] | Challenge sites treated<br>with test sample exhibited<br>no erythema or edema<br>(Grade 0). | Negative for dermal<br>sensitization | |
| Irritation<br>[Intracutaneous Toxicity<br>(ISO)] | The mean test score in<br>0.9% Normal Saline<br>extract was 0, and in<br>Sesame Oil was 0.1. | Non-irritating | |
| Systemic Toxicity<br>(Acute)<br>[Systemic Injection<br>(ISO)] | No study animals were<br>observed with abnormal<br>clinical signs indicative of<br>toxicity during the 72-hour<br>test period. | Non-toxic | |
| Systemic Toxicity<br>(Acute)<br>[Material Mediated<br>Pyrogen in a Rabbit<br>Model] | Temperature increases for<br>the all test animals did not<br>exceed the acceptable test<br>limit for maximum<br>individual temperature<br>rise. | Non-pyrogenic | |
| Hemocompatibility<br>[Hemolysis, direct<br>contact - device/material<br>(human blood)] | The difference between<br>the hemolytic indices of<br>the test article and the<br>negative control was<br>0.00%. | Non-hemolytic | |
| Hemocompatibility<br>[Hemolysis, indirect -<br>device/material (human<br>blood)] | The difference between<br>the hemolytic indices of<br>the test article and the<br>negative control was<br>0.00%. | Non-hemolytic | |
| Summaries of Biocompatibility Tests Conducted to Support this Premarket<br>Notification | | | |
| Test | Results | Conclusions | |
| Hemocompatibility<br>[Partial Thromboplastin<br>Time (PTT test) - Human<br>plasma] | Both the subject device<br>and predicate device fell<br>within the same<br>thrombogenicity category<br>(minimal activator) | Subject and predicate devices<br>showed similar<br>thrombogenicity. | |
| Hemocompatibility<br>[Complement Activation<br>- SC5b-9] | Amounts of SC5b-9<br>generated by the test<br>article and predicate<br>device after exposure<br>times of 30 and 60<br>minutes were not<br>statistically different. The<br>amount of SC5b-9<br>generated by the test<br>article was statistically<br>lower than that released by<br>the predicate after 90<br>minutes exposure. | Subject and predicate devices<br>showed similar complement<br>system activation. | |
| Hemocompatibility<br>[Thrombogenicity in a<br>Dog Model] | Both test animals had a<br>subject device (Aristotle<br>14 Guidewire) and<br>predicate device (Synchro-<br>14 Guidewire) inserted.<br>The average thrombus<br>score was the same for<br>both the subject and<br>predicate device. | Subject and predicate devices<br>showed similar thrombogenic<br>potential. | |
| Latex<br>[LEAP test] | No Latex Antigenic<br>Proteins were detected. | Contains no detectable traces of<br>latex | |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Scientia Vascular LLC Traditional 510(k) Aristotle 14 Guidewire
### Functional Testing
Functional testing was performed in accordance with ISO 11070:2014 Sterile single-use intravascular introducers, dilators and guidewires and the FDA Guidance Document Coronary and Cerebrovascular Guidewire Guidance (January 1995). The following table summarizes the functional tests performed and test results obtained to demonstrate substantial equivalence to the predicate device:
| Summaries of Functional Tests Conducted to Support this Premarket Notification | | | | |
|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Test | Test Method Summary | Results | | |
| Visual<br>Inspection | Tests per ISO 11070:<br>Visual inspection for extraneous<br>matter, process and surface defects<br>or defects that may cause trauma<br>to vessels during use | No extraneous matter, surface<br>defects or visible droplets of coating<br>were present on the Aristotle 14<br>Guidewires. | | |
| Dimensional<br>Verification | Tests per ISO 11070:<br>Dimensional inspection per<br>engineering drawings | All guidewires met dimensional<br>specifications. | | |
| Flexing Test | Tests per ISO 11070:<br>Inspection for defects and damage<br>or flaking of the coating after<br>flexing | No defects or damage / flaking of<br>the coating were observed after<br>flexing. | | |
| Tensile<br>Strength | Tensile testing per ISO 11070 | All guidewires met minimum force<br>breakage requirements specified in<br>ISO 11070. | | |
| Tip Shape,<br>Retention | Guidewires must be shapeable and<br>must retain shaped angle after<br>simulated use | All tips met shaping and shape<br>retention requirements after<br>simulated use. | | |
| Torqueability | Measurement of torque response<br>(average input to output lag) in an<br>anatomical model | All guidewires demonstrated<br>acceptable torque responses. The<br>torque response of the subject<br>device was comparable to that of the<br>predicate device. | | |
| Torque<br>Strength | Torque turns to failure in an<br>anatomical model | All guidewires demonstrated<br>acceptable torque strength. The<br>torque strength of the subject device<br>was comparable to that of the<br>predicate device. | | |
{10}------------------------------------------------
| Summaries of Functional Tests Conducted to Support this Premarket Notification | | |
|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test | Test Method Summary | Results |
| Tip<br>Flexibility | Measure force to deflect guidewire tips to 45 and 90 degrees at 5mm, 10mm, and 20mm test lengths | The forces required to deflect the guidewire tips were acceptable. The flexibility of the tips of all subject devices was comparable to the tip flexibility of the predicate guidewire. |
| Fracture | Tests per ISO 11070:<br>Inspection for fracture, loosening, or failure after wrapping around mandrel | No guidewires showed signs of fracture, loosening, or failure after wrapping them 8 times around a mandrel. |
| Coating<br>Lubricity and<br>Durability | Frictional force of coated guidewires was determined after simulated use in a tortuous path | All guidewires met specified frictional force requirements. |
| Coating<br>Integrity | Coating uniformity and integrity visually examined on dyed samples after simulated use in a tortuous path | All samples showed acceptable coating coverage after simulated use. |
| Particulates | Particulates of various size ranges counted after simulated use in a tortuous path | A comparable number of particulates was recovered from subject and predicate devices following simulated use. |
| Torque<br>Device -<br>Introducer<br>Testing | Various tests on guidewire accessories per ISO 11070: biocompatibility, visual inspection, corrosion resistance, tensile testing, luer taper dimensions, pouch seal strength | Acceptance criteria of all tests were met. |
| Simulated<br>Use Model<br>Testing and<br>Product<br>Compatibility | Anatomical model designed to simulate the tortuous anatomy of the neurovasculature used for simulated use testing | Guidewires and predicate devices were found to perform acceptably in evaluations of: Torqueability in tortuous vasculature, Lubricity, Microcatheter Support & Tracking, Compatibility with Introducer, Compatibility with Torque Device, and Compatibility with Microcatheter |
| Summaries of Functional Tests Conducted to Support this Premarket Notification | | |
| Test | Test Method Summary | Results |
| Cadaver<br>Testing | Physicians evaluated subject and predicate guidewires for various<br>performance characteristics in a human cadaver | Subject and predicate guidewires<br>both exhibited acceptable<br>performance. |
| Radiopacity | Subject and predicate guidewires evaluated by physicians in human<br>cadaver | Both subject and predicate<br>guidewires exhibited acceptable<br>radiopacity. |
| Corrosion<br>Resistance | Test for corrosion resistance per<br>ISO 11070 | There were no signs of corrosion on<br>guidewires after soaking in typical<br>end-use solutions. |
| Chemical<br>Compatibility | Guidewires were exposed to saline<br>and contrast agent/saline solutions<br>and examined for degradation. | All guidewires showed no signs of<br>degradation, corrosion or physical<br>decomposition after exposure. |
| MRI<br>Compatibility | Guidewires are constructed of<br>metallic materials and should not<br>be exposed to MRI procedures. | No testing performed. Aristotle<br>Guidewires are labeled “MRI<br>Unsafe.” |
| Latex | Tested for trace latex proteins per<br>ASTM D6499-07 | No detectable traces of latex were<br>found. |
| Package<br>Integrity | Simulated transportation test per<br>ASTM D 4169:16. Pouch<br>evaluated for seal strength per<br>ASTM F 88-15 and leak tests<br>(bubble test) per ASTM F 2096-11 | Following exposure to typical<br>storage and transportation<br>conditions, all sterile barrier<br>pouches maintained their integrity<br>and labeling remained affixed and<br>legible. |
| Shelf Life | Device performance attributes that<br>can be affected by storage<br>conditions were evaluated after<br>exposure to accelerated aging<br>conditions per ASTM F1980. | After exposure to accelerated aging<br>conditions simulating real-time<br>storage under ambient conditions<br>for 6 months, all device<br>performance acceptance criteria<br>were met, justifying labeling the<br>devices with a 6-month shelf life. |
| Summaries of Functional Tests Conducted to Support this Premarket Notification | | |
| Test | Test Method Summary | Results |
| Sterilization<br>Validation | 100% EO is used to sterilize the<br>device to achieve a SAL of at least<br>10-6. The device was adopted into<br>an EO sterilization processing<br>group in accordance with AAMI<br>TIR 28:2009. Validation of the EO<br>sterilization cycle was performed<br>using the Half-cycle, overkill<br>approach described in Section<br>B.1.2 of ISO11135:2007 Annex B. | Results justified adoption into the<br>EO sterilization processing group:<br>Comparative and Bioburden<br>Resistance study results<br>demonstrated that PCDs are more<br>difficult to sterilize than devices;<br>Bioburden Enumeration and<br>Extraction Efficiency tests were<br>used to enumerate the CFUs present<br>on devices; and Bacteriostasis/<br>Fungistasis test results demonstrated<br>that the product does not inhibit the<br>growth of organisms. |
| Sterilization<br>Validation:<br>EO and ECH<br>Residuals | Measured EO and ECH residuals<br>per AAMI/ANSI/ISO 10993-7 | The residual traces of EO and ECH<br>remaining in the Aristotle 14<br>Guidewire after exposure to the EO<br>sterilization process are well below<br>the limits specified in ISO 10993-7. |
| Sterilization<br>Validation:<br>Bacterial<br>Endotoxin<br>Levels | LAL testing was conducted in<br>accordance with ANSI/AAMI<br>ST72:2011/(R)2016, USP <161>,<br>and USP <85>, using the kinetic<br>chromogenic method. | < 2.15 EU/device |
{11}------------------------------------------------
{12}------------------------------------------------
## CONCLUSION:
Scientia Vascular, LLC has presented information in this premarket notification supporting its contention that the Aristotle 14 Guidewire is substantially equivalent with respect to technological characteristics and indications for use to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.