K173194 · Orantech, Inc. · FLL · Oct 10, 2018 · General Hospital
Device Facts
Record ID
K173194
Device Name
Temperature Probe
Applicant
Orantech, Inc.
Product Code
FLL · General Hospital
Decision Date
Oct 10, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Indications for Use
Orantech Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Philips Model MP30 and Nihon Kohden Model BSM-6301A. These devices are indicated for used by qualified medical personnel only.
Device Story
Reusable temperature probes designed for continuous patient temperature monitoring. Device utilizes a thermistor (NTC resistance) at the patient-contact end to measure temperature changes, which are transmitted via a phono plug connector to compatible patient monitors (Philips MP30 or Nihon Kohden BSM-6301A). Probes are used in clinical settings by qualified medical personnel. The output is a temperature reading displayed on the host monitor, assisting clinicians in patient assessment and clinical decision-making. The device is non-sterile and intended for multi-patient use following validated cleaning and disinfection procedures.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC 60601-1), laboratory accuracy (≤ 0.3°C), heating transient time (< 150s), use-life testing, and validation of cleaning/disinfection methods per AAMI TIR 30 and AAMI TIR 12. Biocompatibility testing (cytotoxicity, sensitization, irritation) confirmed safety for surface-contacting use.
Technological Characteristics
Reusable temperature probe; NTC thermistor (2252 Ohms at 25°C); measurement range 25-45°C; accuracy ± 0.1°C. Materials: Epoxy/stainless steel sensor tip, PVC/TPU cable, PA66 plug. Connectivity: Phono plug for direct monitor connection. Sterilization: Non-sterile; validated for multi-patient reuse via cleaning/disinfection. Standards: ISO 10993 (biocompatibility), IEC 60601-1 (electrical safety).
Indications for Use
Indicated for adult patient temperature monitoring using skin or rectal contact. For use by qualified medical personnel only.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Unimed Medical Supplies Inc. Temperature Probe (K121427)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 10, 2018
Orantech Inc. % Kevin Wang Chonconn Medical Device Consulting Co., Ltd. 22A, Haijing Square, No. 18, Taizi Road Nanshan District, Shenzhen, 518067 China
Re: K173194
Trade/Device Name: Temperature Probe (Models: TS-Y400-AG30. TS-Y400-AS30. TS-PH-AG30 and TS-PH-AS30) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: September 13, 2018 Received: September 13, 2018
Dear Mr. Kevin Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173194
### Device Name
Temperature Probe (Model: TS-Y400-AG30, TS-Y400-AS30, TS-PH-AG30 and TS-PH-AS30)
Indications for Use (Describe)
Orantech Temperature Probes are intended to be used for monitoring temperature probes are reusable and designed for use with monitors of Philips Model MP30 and Nihon Kohden Model BSM-6301A. These devices are indicated for used by qualified medical personnel only.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| <span> <span style="font-family: DejaVu Sans, sans-serif">✔</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
# 510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
## Prepared Date: 2018/02/05
### 1. Submission sponsor
Name: Orantech Inc.
Address: Zone#A,4F,1st Bld,7th Industrial Zone, Yulv Community, GongMing, Guangming New District, Shenzhen, P.R. China 518106 Contact person: Hsin Xiong Title: General manager E-mail: Yunxi.xiong@orantech.cn Tel: +86-755 23699939
### 2. Submission correspondent
Name: Chonconn Medical Device Consulting Co., Ltd. Address: 22A, Haijing Square, No. 18, Taizi Road, Nanshan District, Shenzhen, P.R. China 518067 Contact person: Kevin Wang E-mail: kevin(@chonconn.com Tel: +86-755 33941160
| Trade/Device Name | Temperature Probe |
|-------------------|--------------------------------------------------------|
| Model | TS-Y400-AG30, TS-Y400-AS30, TS-PH-AG30 and TS-PH-AS30 |
| Common Name | Temperature Probe |
| Regulatory Class | Class II |
| Classification | 21CFR 880.2910 / Clinical electronic thermometer / FLL |
| Submission type | Traditional 510(K) |
#### 3. Subject Device Information
### 4. Predicate Device
- 1. Unimed Medical Supplies Inc. Temperature Probe under K121427.
### 5. Device Description
Reusable temperature probes are used during patient temperature measurement for multi-patient use. These probes consist of a phone plug connector on the monitor end and a thermistor on the patient end. Temperature probes measure temperature by a resistor that is sensitive to temperature changes. These probes are connected to the patient monitor either directly by using a phono plug or by an interconnect cable. These probes have a skin or core contact with a patient.
These temperature probes are typically used with legacy Philips monitor model MP30 and Nihon Kohden monitor model BSM-6301A.
{4}------------------------------------------------
Products are packed individually into a plastic bag in non-sterile condition. Package label describes product REF codes, manufacturing date, CE-mark, legal entity information and a caution "Rx Only ".
### Product Specification
| Model | Application<br>site | length | Compatible<br>monitors | Measurement<br>range | Accuracy |
|--------------|---------------------|--------|-----------------------------------|----------------------|-------------|
| TS-PH-AG30 | Rectum | 3.0m | Philips<br>Model: MP30 | 25-45°C | $\pm$ 0.1°C |
| TS-PH-AS30 | Skin | 3.0m | Philips<br>Model: MP30 | 25-45°C | $\pm$ 0.1°C |
| TS-Y400-AG30 | Rectum | 3.0m | Nihon Kohden<br>Model: BSM-6301A | 25-45°C | $\pm$ 0.1°C |
| TS-Y400-AS30 | Skin | 3.0m | Nihon Kohden<br>Model: BSM-6301 A | 25-45°C | $\pm$ 0.1°C |
### Indication for use 6.
Orantech Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Philips Model MP30 and Nihon Kohden Model BSM-6301A.
These devices are indicated for used by qualified medical personnel only.
### Comparison to the Predicate Device 7.
| Orantech Model | Predicate Model | Description |
|----------------|-----------------|------------------------------------------------|
| TS-PH-AG30 | THP-AG | Reusable Temperature Probe, adult rectum, 3.0M |
| TS-PH-AS30 | THP-AS | Reusable Temperature Probe, adult skin, 3.0M |
| TS-Y400-AG30 | T2252-AG | Reusable Temperature Probe, adult rectum, 3.0M |
| TS-Y400-AS30 | T2252-AS | Reusable Temperature Probe, adult skin, 3.0M |
| Features | Subject Device<br>Orantech Temperature<br>Probe | Predicate Device<br>Unimed Temperature<br>Probe<br>K121427 | Comparison |
|------------------------|-------------------------------------------------|------------------------------------------------------------|---------------|
| Classification<br>Name | Temperature Probe | Temperature Probe | Same |
| Product Code | FLL | FLL | Same |
| Regulation Number | 880.2910 | 880.2910 | Same |
| | | | |
| Panel | General Hospital | General Hospital | Same |
| Class | II | II | Same |
| Thermistor | NTC resistance | NTC resistance | Same |
| | (2252 Ohms in 25°C) | (2252 Ohms in 25°C) | |
| Accuracy range | 25-45°C | 25-45°C | Same |
| Accuracy | +0.1℃C | +0.1℃C | Same |
| Measure site | Skin & Rectum | Skin & Rectum | Same |
| Patient population | Adult | Adult | Same |
| Material contact to | Thermistor Sensor Tip: | Thermistor Sensor Tip: | Different (1) |
| body | Epoxy & Stainless steel | Stainless steel | |
| Length | 3m | 1.5m & 3m | Same |
| Cable material | PVC & TPU | PVC & TPU | Same |
| Plug material | PA66 | ABS | Different (2) |
| Biocompatibility | Cytotoxicity | Cytotoxicity | Same |
| | Sensitization | Sensitization | |
| | Irritation | Irritation | |
| Operational | Continual | Continual | Same |
| Principles | | | |
| Indication for Use | Orantech Temperature | Unimed Temperature Probes | Different (3) |
| | Probes are intended to be | are intended to be used for | |
| | used for monitoring | monitoring temperature. The | |
| | temperature. The | temperature probes are | |
| | temperature probes are | designed for use with | |
| | reusable and designed for | monitors of Philips, | |
| | use with monitors of | Marquette, Mindray, | |
| | Philips Model MP30 and | Spacelabs, Siemens, | |
| | Nihon Kohden Model | Artema/S&W and other | |
| | BSM-6301A. These | monitors compatible with | |
| | devices are indicated for | YSI 400 series temperature | |
| | used by qualified medical | probes. | |
| | personnel only. | These devices are indicated | |
| | | for used by qualified | |
| | | medical personnel only. | |
{5}------------------------------------------------
Justifications for differences between Orantech Temperature Probe and the predicate device are shown as below:
Different (1): It's different with the patient contacted material of predicate device. However, it's conformed with ISO 10993 series standard. According to analysis results, all the Patient Contacting Materials shows no cytotoxicity, irritation response was negligible and no sensitization. Thus, different material will not raise any safety or biocompatibility issues.
Different (2): It's different with the plug material of predicate device. However, it's conformed with ISO 10993 series standard. According to analysis results, all the materials shows no cytotoxicity, irritation
{6}------------------------------------------------
response was negligible and no sensitization. Thus, different material of plug will not raise any safety or biocompatibility issues.
Different (3): The compatible monitors of proposed devices are different from the predicate device. The proposed devices are intended to use with Philips and Nihon Kohden monitors and the predicate devices are intended to use with Philips, Marquette, Mindray, Spacelabs, Siemens, Artema/S&W and other monitors. The core component of temperature probe is Negative Coefficient (NTC) which is identical to the NTC used in predicate device. The NTC determines the accuracy and range of temperature measurement. Therefore, the indication for use of proposed devices is essentially the same as predicate device.
### Performance Data 8.
## Biocompatibility testing
The biocompatibility evaluation for the Orantech Temperature Probe was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- Sensitization
- Irritation
The Orantech Temperature Probe is considered surface contacting for a duration of exceed 24 hours but not 30 days.
## Non-clinical data
The Orantech Temperature Probe has been tested according to the following standards:
| Test | Description | Result |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Safety | Ensures the temperature probes meet the requirements of IEC 60601-1 electrical safety | Pass |
| Laboratory accuracy | Not greater than 0.3 °C for a continuous clinical thermometer that is not an adjusted mode clinical thermometer | Pass |
| Time response | Heating transient time < 150s | Pass |
| Use life testing | Ensures the temperature probes meet the requirements of the expected service life of the probe. | Pass |
| Cleaning and Disinfection | Validate the cleaning and disinfection method is effective. The study was conducted based on following documents:<br>1) AAMI TIR 30: 2011 A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices;<br>2) AAMI TIR 12: 2010 Designing, testing and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers; | Pass |
{7}------------------------------------------------
# Oral ntech
| 3) FDA Guidance, Reprocessing Medical Devices in Health Care<br>Settings: Validation Methods and Labelling document issued on:<br>March 17, 2015. | |
|---------------------------------------------------------------------------------------------------------------------------------------------------|--|
|---------------------------------------------------------------------------------------------------------------------------------------------------|--|
# 9. Conclusion
Performance testing and compliance with voluntary standards demonstrate that the proposed are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.