The device is for use on adult patients for the treatment of Obstructive Sleep Apnoea (OSA). The device is for use in the home or sleep laboratory.
Device Story
F&P SleepStyle is a non-invasive, auto-adjusting CPAP flow generator with integrated heated humidifier; treats OSA by delivering physician-prescribed positive airway pressure to splint airways open. Device pressurizes room air, transports it to patient interface; humidifier adds water vapor to gas to reduce nasal dryness. Features include SensAwake, expiratory relief, ramp, auto-altitude adjustment, and leak compensation. Operated by patients in home or sleep lab. Integrated Bluetooth pairs with mobile apps for patient self-assessment; integrated cellular modem and USB stick upload therapy data to InfoSmart software for healthcare provider review. Output allows providers to monitor compliance/efficacy and adjust therapy settings. Benefits include improved patient comfort and effective airway management.
Clinical Evidence
Clinical validation study conducted at Fisher & Paykel Sleep Laboratory (New Zealand) with 50 patients. Study validated device performance for OSA treatment using APAP and CPAP modes, including pressure relief technologies (expiratory relief and SensAwake). Results demonstrated substantial equivalence to the predicate device.
Technological Characteristics
Non-invasive CPAP flow generator; 4–20 cmH2O pressure range. Integrated heated humidifier (ISO 8185). Connectivity: Bluetooth, cellular modem, USB. Materials tested per ISO 10993 (permanent tissue contact). Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, ISO 80601-2-70. Software: Moderate level of concern, developed per IEC 62304 and ISO 14971.
Indications for Use
Indicated for adult patients with Obstructive Sleep Apnea (OSA) for use in home or sleep laboratory settings.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below that.
August 8, 2018
Fisher & Paykel Healthcare Ltd Jayanti Karandikar Regulatory Affairs Specialist 15 Maurice Paykel Place East Tamaki, Auckland 2013 NZ
Re: K173193
Trade/Device Name: F&P SleepStyle TM Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: July 5, 2018 Received: July 9, 2018
Dear Jayanti Karandikar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone
(1-800-638-2041 or 301-796-7100).
Sincerely,
# James J. Le
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K173193
Device Name
F&P SleepStyle™
Indications for Use (Describe)
The device is for use on adult patients for the treatment of Obstructive Sleep Apnoea (OSA). The device is for use in the home or sleep laboratory.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
F&P SleepStyle- Traditional 510(k)
Image /page/3/Picture/1 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are in a larger, bold font on the top line. Below that, there is a horizontal line, and then the word "HEALTHCARE" is in a smaller, bold font.
# 510(k) Summary
| Contact person/submitter | Jayanti Karandikar |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Date prepared | 04 Aug 2018 |
| Contact details | Address: 15 Maurice Paykel Place<br>East Tamaki<br>Auckland 2013, New Zealand<br>Telephone: +64 9 574 0100<br>Fax: +64 9 574 0158 |
| Trade name | F&P SleepStyle™ |
| Classification name | Non Continuous Ventilator (IPPB)<br>Class II (21 CFR §868.5905)<br>Product code BZD (Anaesthesiology) |
| Predicate device | F&P ICON (K094040) |
#### Intended Use and Indications for Use
The device is for use on adult patients for the treatment of Obstructive Sleep Apnea (OSA). The device is for use in the home or sleep laboratory.
#### Device Description
The F&P SleepStyle™ is a non-invasive, auto adjusting Continuous Positive Airway Pressure (CPAP) flow generator, incorporating a heated respiratory humidifier. The device is intended to treat Obstructive Sleep Apnea (OSA) by delivering a flow of positive airway pressure at a level prescribed by a physician to splint open the airway and prevent airway collapse.
The F&P SleepStyle™ system is comprised of the following devices/accessories:
- F&P SleepStyle™ models SPSAAN and SPSCAN: .
- . Breathing tubes:
- ThermoSmart™ heated breathing tube; o
- Standard breathing tube, i.e. non-heated, and elbow connector; O
- Water chamber and water chamber seal; .
- USB stick (F&P InfoUSB) .
The F&P SleepStyle™ system is available in two modes, fixed and auto-adjusting CPAP.
Similar to the predicate device, the SleepStyle™ works by pressurizing room air which is then transported to a patient interface. The delivery of positive air pressure to the patient holds the soft tissue in the throat open, preventing airway collapse and thus disruptions to the patient's breathing and sleep.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are in a large, bold, sans-serif font, with the "&" symbol connecting the two words. Below the company name is the word "HEALTHCARE" in a smaller, bold, sans-serif font. A horizontal line separates the company name from the word "HEALTHCARE".
Humidified CPAP devices include a water chamber which holds water that is heating element in the CPAP device, thus adding water vapor to the gas prior to delivery to the patient. This is integrated into the device. The humidifier reduces nasal dryness and improves patient comfort.
An optional heated breathing tube (ThermoSmart) reduces condensation and improves humidity delivery to the patient.
The proposed device has integrated Bluetooth which allows therapy data to be sent to general wellness applications which can be used by patients for self-assessment and monitoring of therapy progress.
The proposed device also has an integrated modem which automatically uploads therapy data to the software reporting tool InfoSmart (K161686) managed by the healthcare provider. Alternatively data from the device can be stored onto an USB stick and uploaded to the software reporting tool InfoSmart (K161686).
The following table outlines the features of the F&P SleepStyle™ models.
| Performance features | SleepStyle Auto CPAP<br>(SPSAAN) | SleepStyle CPAP<br>(SPSCAN) |
|-------------------------------|----------------------------------|-----------------------------|
| Fully integrated humidifier | ● | ● |
| Auto-adjusting pressure | ● | ● |
| ThermoSmart technology | ● | ● |
| SensAwake | ● | ● |
| Expiratory relief | ● | ● |
| Ramp | ● | ● |
| Auto-altitude adjustment | ● | ● |
| Leak compensation | ● | ● |
| Efficacy reporting | ● | ● |
| Compliance reporting | ● | ● |
| F&P InfoUSB | ● | ● |
| Bluetooth wireless technology | ● | ● |
{5}------------------------------------------------
## F&P SleepStyle— Traditional 510(k)
Image /page/5/Picture/1 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are in a bold, dark blue font on the top line. Below that is a horizontal line, and below that is the word "HEALTHCARE" in a smaller, bold, dark blue font.
| Cellular modem | • | • |
|----------------|---|---|
|----------------|---|---|
## Technological Characteristics Comparison
| Design /<br>technological<br>characteristic<br>for<br>comparison | Subject device<br>(F&P SleepStyleTM<br>CPAP) | Predicate device<br>(F&P ICON<br>SERIES CPAP,<br>K094040) | Comments | |
|------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Intended Use and Indications for Use | Intended Use/<br>Indications<br>for use | The device is for use<br>on adult patients for<br>the treatment of<br>Obstructive Sleep<br>Apnea (OSA).<br>The device is for use<br>in the home or sleep<br>laboratory. | The F&P ICON<br>Series CPAP is for<br>use on adult<br>patients for the<br>treatment of<br>Obstructive Sleep<br>Apnea (OSA).<br>The device is for<br>use in the home or<br>sleep laboratory. | Identical |
| Classification | Device<br>classification | II (868.5905) | II (868.5905) | Identical |
| | Product code | BZD | BZD | Identical |
| | Classification<br>panel | Anesthesiology | Anesthesiology | Identical |
| Pressure delivery | | | | |
| | Fixed CPAP | Yes | Yes | Identical feature offered |
| | Automatic<br>adjusting<br>pressure | Yes (Auto CPAP<br>model) | Yes (Auto CPAP<br>model) | Identical feature offered |
| Algorithm | | | | |
| | Automatic<br>adjusting<br>СРАР<br>algorithm | Yes | Yes | Modified algorithm - Refer<br>to §5.3.1 a) |
| | AHI<br>measurement<br>algorithm | Yes | Yes | Modified algorithm - Refer<br>to §5.3.1 b) |
| Design /<br>technological<br>characteristic<br>for<br>comparison | Subject device<br>(F&P SleepStyle™<br>CPAP) | Predicate device<br>(F&P ICON<br>SERIES CPAP,<br>K094040) | Comments | |
| Performance specifications | | | | |
| Pressure<br>range | 4 – 20 cmH2O | 4 – 20 cmH2O | Identical | |
| Humidity<br>output | ≥ 10 mg/L, as<br>required by ISO<br>8185 | ≥ 10 mg/L, as<br>required by ISO<br>8185 | Identical | |
| Features | | | | |
| Altitude<br>adjustment | 0 to 3000m (0 to<br>9000ft) | 0 to 3000m (0 to<br>9000ft) | Identical | |
| Leak<br>compensation | Yes | Yes | Identical feature offered | |
| Comfort features | | | | |
| Pressure<br>ramp | Up to 20 minutes | Up to 20 minutes | Identical feature offered | |
| Humidification | Yes | Yes | Similar feature with option of<br>7 comfort settings for both,<br>subject and predicate<br>device. | |
| SensAwake™ | Yes | Yes | Similar feature offered –<br>Refer to §5.3.1 c) | |
| Expiratory<br>Relief | Yes | No | New comfort feature – Refer<br>to §5.3.1 c) | |
| User interface | | | | |
| User controls | Buttons | Dial | Updated user interface –<br>Refer to §5.3.1 d) | |
| Menu | Text | Symbols | | |
| Reuse / sterilization / shelf life | | | | |
| Reuse | Multi-patient, multi-<br>use | Multi-patient, multi-<br>use | Similar.<br>The SleepStyle heated<br>breathing tube is not<br>intended for multi-patient<br>use | |
| Design /<br>technological<br>characteristic<br>for<br>comparison | Subject device<br>(F&P SleepStyle™<br>CPAP) | Predicate device<br>(F&P ICON<br>SERIES CPAP,<br>K094040) | Comments | |
| Sterility | Components<br>(including<br>accessories) are not<br>provided sterile or<br>intended to be<br>sterilized | Components<br>(including<br>accessories) are<br>not provided sterile<br>or intended to be<br>sterilized | Identical | |
| Cleaning | Single patient,<br>multi-use:<br>cleaning Multi-patient,<br>multi-use:<br>disinfection of<br>components in<br>the humidified air<br>path | Single patient,<br>multi-use:<br>cleaning Multi-patient,<br>multi-use:<br>disinfection of<br>components in<br>the humidified<br>air path | Similar types of cleaning<br>identified for single-patient,<br>multi-use and multi-patient,<br>multi-use - Refer to §5.3.1<br>e) | |
| Shelf Life | 2 years | Not defined | Shelf life defined for<br>SleepStyle accessories:<br>heated breathing tube,<br>standard breathing tube,<br>water chamber, inlet filter | |
| Data reporting | | | | |
| Data reporting<br>tool | InfoSmart<br>(K161686) | InfoSmart<br>(K094040) | Identical - Refer to §5.3.1 f) | |
| USB data stick | Yes | Yes | Similar function of data<br>storage and data transfer -<br>Refer to §5.3.1 g) | |
| Cellular<br>modem | Yes (built-in) | Yes (external -<br>InfoGSM,<br>K110316) | Similar function of data<br>transfer - Refer to §5.3.1 h) | |
| Bluetooth® | Yes | No | New feature - Refer to<br>§5.3.1 i) | |
| Accessories | | | | |
| Heated<br>breathing<br>tube | Available | Available | Identical accessory offered | |
| Non-heated<br>breathing<br>tube | Available | Available | Identical accessory offered | |
| Design /<br>technological<br>characteristic<br>for<br>comparison | Subject device<br>(F&P SleepStyle™<br>CPAP) | Predicate device<br>(F&P ICON<br>SERIES CPAP,<br>K094040) | Comments | |
| Water<br>chamber | Yes | Yes | Identical accessory type<br>offered; detachable lid seal<br>added - Refer to §5.3.1 j) | |
| Operating conditions | | | | |
| Temperature | 12 to 35° C<br>54 to 95° F | 5 to 35° C<br>41 to 95° F | Similar | |
| Humidity | 15 to 90% RH | 5 to 95%RH | | |
| Altitude | 0 to 3000m<br>0 to 9,000 ft | 0 to 3000m<br>0 to 9,000 ft | Identical | |
| Storage and Transport conditions | | | | |
| Temperature | -10 °C to 60 °C<br>14 to 140 °F | -10 °C to 60 °C<br>14 to 140 °F | Identical | |
| Humidity | 15 to 90% RH | Not specified | Humidity has been specified<br>for the SleepStyle device<br>storage conditions. | |
| Electrical ratings | | | | |
| Supply<br>voltage | 100-115 V | 100-115 V | Identical | |
| Current input | 1.2 A (2.5 A max.) | 1.27 A (1.43 A<br>max.) | Similar | |
| Supply<br>Frequency | 50-60 Hz | 50-60 Hz | Identical | |
{6}------------------------------------------------
F&P SleepStyle— Traditional 510(k)
Image /page/6/Picture/1 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are in a large, bold, blue font. Below that, the word "HEALTHCARE" is in a smaller, bold, blue font. There is a blue line separating the two lines of text.
{7}------------------------------------------------
F&P SleepStyle— Traditional 510(k)
Image /page/7/Picture/1 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are in a bold, dark blue font on the top line. Below that is a horizontal line, and below that is the word "HEALTHCARE" in a bold, dark blue font.
{8}------------------------------------------------
F&P SleepStyle= Traditional 510(k)
Image /page/8/Picture/1 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are in a large, bold, blue font. Below that, the word "HEALTHCARE" is in a smaller, bold, blue font. There is a blue line separating the two lines of text.
#### Key Differences
The key differences between the subject device and predicate device, are that F&P SleepStyle™:
- a) Has a revised auto adjusting CPAP algorithm. The algorithm from the predicate device has been modified.
- b) Has a revised AHI measurement algorithm. Changes were made to the algorithm used in the predicate device.
- c) Has a revised SensAwake algorithm and new comfort feature expiratory relief. Changes were made to the SensAwake algorithm used in the predicate device. Expiratory relief reduces pressure delivered to the patient each time the patient exhales.
{9}------------------------------------------------
F&P SleepStyle= Traditional 510(k)
Image /page/9/Picture/1 description: The image contains the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are in a bold, dark blue font on the top line. Below that is a horizontal line, and below that is the word "HEALTHCARE" in a bold, dark blue font.
- d) Uses a modified user interface The SleepStyle™ device uses dedicated buttons for key features i.e. therapy on/off and a graphical display to display text.
- e) Has similar reprocessing instructions to the predicate device, with additional disinfection options. Performance and verification testing was conducted as per the recommendations of FDA quidance "Reprocessing Medical Devices in Health Care Settings: Validations Methods and Labelling." Similar to the predicate device, chemical disinfection for SleepStyle™ parts (excluding the heated breathing tube) can be performed using Cidex OPA.
The main differences are described below:
- The heated breathing tube used with the SleepStyle™ is for single-patient use only and . therefore is not intended for high level disinfection.
- Thermal disinfection of some SleepStyle™ parts can be carried out at 2 different . temperatures: 75°C for 30 min and 90°C for 1 min. Thermal disinfection of the predicate device was validated at 70°C for 30 min.
- f) Is compatible with InfoSmart, a software reporting tool. The client version of InfoSmart was cleared with the predicate device through 510K K094040. InfoSmart was recently cleared through K161686 to be able to work with compatible flow generators and incorporated the client and web versions. InfoSmart simply displays data as recorded by the compatible CPAP device. Introducing a new compatible device does not change the basic functionality of InfoSmart.
- g) Includes an USB stick which is used to store sleep data that is captured by the device during therapy. The data is transferred from the memory of the CPAP device when the USB stick is inserted into the device's USB port. The USB stick can then be sent to the healthcare provider to view/report on the data, and/or adjust therapy settings. Alternatively, the application in the stick can enable data to be uploaded via the internet to a remote server which can be accessed by the healthcare provider via a web based software called InfoSmart Web (K161686). The stick, in the predicate device, functions in a similar fashion to the stick above but does not have the software application to support web based data transfer.
- h) Includes an integrated modem as opposed to an external modem InfoGSM (K110316) which is compatible with the predicate F&P ICON (K094040). The purpose of the integrated modem within the proposed device is to enable data transfer via cellular network to a software reporting tool InfoSmart (K161686).
The external modem used with the predicate device performs a similar function.
- i) Includes integrated Bluetooth which enables data transfer by pairing the device with the mobile phone or tablet. This allows the user to view their data on the SleepStyle™ mobile application or the SleepStyle™ website.
- j) Includes a water chamber that has a detachable seal to allow easier cleaning, and filling of the water chamber. The predicate device has the chamber seal permanently attached to the chamber.
{10}------------------------------------------------
F&P SleepStyle= Traditional 510(k)
Image /page/10/Picture/1 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are in a bold, dark blue font on the top line. Below that is a horizontal blue line, and below that is the word "HEALTHCARE" in a bold, dark blue font.
#### Non-Clinical Performance Data
Non-clinical performance testing of the SleepStyle™ device was carried out covering mechanical. electrical and thermal safety, environmental conditions, electromagnetic compatibility and functional verification.
The SleepStyle™ device has been tested according to the following standards:
- ANSI/AAMI ES60601-1:2005 (R) 2012, A1:2012, C1:2009 (R) 2012 and A2:2010 (R) . 2012. Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance (FDA recognition number: 19-4)
- IEC 60601-1-2:2014. Medical electrical equipment Part 1-2: General requirements for . basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests. (FDA recognition number:19-8)
- . IEC 60601-1-6:2010 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability
- . IEC 60601-1-11:2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- ISO 8185:2008 Respiratory tract humidifiers for medical use – Particular requirements for respiratory humidification systems. (FDA recognition number:1-86)
- . 10993-1:2009 Biological Evaluation of Medical Devices -- Part 1: Evaluation and Testing within a Risk Management Process. (FDA recognition number: 2-220)
- . ISO 10993-3:2014. Biological Evaluation of Medical Devices -- Part 3: Tests for Genotoxicity, Carcinogenicity and Reproductive Toxicity. (FDA recognition number: 2-228)
- . 10993-5:2009 Biological Evaluation of Medical Devices -- Part 5: Tests for in Vitro Cytotoxicity. (FDA recognition number: 2-245)
- . ISO 10993-10:2010. Biological Evaluation of Medical Devices -- Part 10: Tests for Sensitization and Irritation. (FDA recognition number: 2-174)
- ISO 10993-11 Biological evaluation of medical devices Part 11: Tests for systemic . toxicity (FDA recognition number:2-176)
- ISO 80601-2-70:2015 Medical Electrical Equipment - Part 2-70: Particular Requirements for Basic Safety and Essential Performance of Sleep Apnea Breathing Therapy Equipment. (FDA recognition number: 1-115)
- . ISO 5356-1 Anaesthetic and Respiratory Equipment - Conical Connectors: Part 1: Cones and Sockets. (FDA recognition number: 1-62)
- . ISO 5367:2014 Anaesthetic and respiratory equipment - Breathing sets and connector
#### Functional Bench Testing:
The following functional testing was conducted on the SleepStyle device:
{11}------------------------------------------------
F&P SleepStyle- Traditional 510(k)
Image /page/11/Picture/1 description: The image contains the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are in a bold, dark blue font on the top line. Below that is a horizontal line, and below that is the word "HEALTHCARE" in a bold, dark blue font.
- Static and Dynamic Pressure Accuracy ●
- . Flowrate at Maximum Pressure
- Max Pressure Single Fault
- Noise Verification
- . Water Chamber Spillage Test
- Transport, Storage, Drop testing ●
- . Altitude adjustment
- Pull test .
#### Biocompatibility:
The F&P SleepStyle™ is classified as an externally communicating device, tissue contact, permanent duration (>30 days). The humidified gas pathway components of the F&P SleepStyle™ device system were tested together as a full system. The testing of cytotoxicity, sensitization, intracutaneous reactivity, and acute systemic toxicity were conducted based on applicable ISO 10993 test standards. In addition, the chemical characterization study plus a toxicological risk assessment for the identified chemical extractables were conducted.
The testing of particulate matters and volatile organic compounds (VOCs) for both the humidified and dry gas path, and testing of carbon monoxide, carbon dioxide and ozone for the dry gas pathway were also performed.
#### Software:
The device software level of concern is moderate and software testing was done as per IEC 62304:2006 (Medical device software - Software life-cycle processes) and ISO 14971:2007 (Medical devices - Application of risk management to medical devices).
#### Clinical Performance Data
The clinical validation study, involving 50 patients, was conducted at the Fisher & Paykel Sleep Laboratory in New Zealand following GCP guidelines.
Clinical studies were carried out:
- . To validate the device, for the treatment of OSA using APAP and CPAP modes with and without pressure relief technologies (expiratory relief and SensAwake).
#### Conclusion
Results obtained from non-clinical and clinical testing demonstrate that, SleepStyle™ is substantially equivalent to the predicate device F&P ICON (K094040).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.