THE Graft Bone Substitute is intended for use as a bone grafting material in dental surgery such as: - Filling of extraction sockets to enhance preservation of the alveolar ridge - Elevation of maxillary sinus floor
Device Story
THE Graft Bone Substitute is a resorbable, porcine-derived cancellous bone graft material composed of hydroxyapatite. It functions as a natural, porous mineral matrix with interconnected macro and micro pores that support trabecular architecture. Used in dental surgery, the material is packed into bony sites (extraction sockets or maxillary sinus floor) by a clinician. It acts as a scaffold, gradually resorbing and being replaced by new bone during the healing process. The device is supplied sterile in vials or syringes for single-use.
Clinical Evidence
Evidence includes non-clinical bench testing (ASTM F1581 compliance), comprehensive biocompatibility testing (ISO 10993 series: cytotoxicity, sensitization, reactivity, toxicity, genotoxicity, implantation), and sterilization/viral inactivation validation. In-vivo dog model testing and clinical data were provided to assess performance as an implantable material; all acceptance criteria were met.
Technological Characteristics
Material: Porcine cancellous bone (hydroxyapatite). Structure: Porous, irregular-shaped particles with interconnected macro/micro pores. Dimensions: 0.2–0.355 mm, 0.25–1.0 mm, or 1.0–2.0 mm. Sterilization: Gamma irradiation (ISO 11137, SAL 10^-6). Compliance: ASTM F1581. Packaging: Glass vial or syringe in Tyvek pouch.
Indications for Use
Indicated for dental surgery patients requiring alveolar ridge preservation via extraction socket filling or maxillary sinus floor elevation.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services, which features a stylized human figure. To the right is the logo for the U.S. Food & Drug Administration (FDA). The FDA logo includes the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" written below in a smaller font.
PURGO BIOLOGICS INC. Hana Jung RA Manager E-607, 700, Pangyo-ro, Bundang-gu Seongnam-si, Gyeonggi-do, Korea
Re: K173188
Trade/Device Name: The Graft Bone Substitute Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPM Dated: June 14, 2018 Received: June 21, 2018
Dear Hana Jung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
July 20, 2018
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Mary S. Runner -S For Tina Kiang, Ph.D.
Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173188
Device Name THE Graft Bone Substitute
Indications for Use (Describe)
THE Graft Bone Substitute is intended for use as a bone grafting material in dental surgery such as:
- Filling of extraction sockets to enhance preservation of the alveolar ridge
- Elevation of maxillary sinus floor
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 2em; line-height: 0.5em;">☑</span> |
|----------------------------------------------|------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 2em; line-height: 0.5em;">☐</span> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Purgo Dental Biologics Solution. The logo consists of a gray circular icon with a smaller circle inside, followed by the word "Purgo" in a purple, sans-serif font. To the right of "Purgo" are the words "Dental Biologics Solution" stacked vertically, also in a purple, sans-serif font, but smaller in size.
# 510(k) Summary K173188
#### l. SUBMITTER
PURGO BIOLOGICS INC. E-607, 700, Pangyo-ro, Bundang-gu Seongnam-si, Gyeonggi-do, Korea Postal Code 13516
Contact Person: Hana Jung Regulatory Affairs Manager Phone: +82 70 4827 4863 Fax: +82 2 549 4003 Email: junghn@purgo.co.kr
Date Revised: 7-20-2018
## II. DEVICE
Name of Device: THE Graft Bone Substitute Common or Usual Name: Bone Grafting Material Classification Name: Bone Grafting Material, Animal Source (21 CFR 872.3930) Regulatory Class: Class II (special controls) Product Code: NPM
### III. PREDICATE DEVICE
Collagen Matrix Inc. Porcine Anorganic Bone Mineral, K140714
Reference device: Bio-Oss® , K952617 (Geistlich-Pharma)
### IV. SPECIAL CONTROLS
FDA Guidance "Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices; Guidance for Industry and FDA Staff" was followed during the preparation of this submission.
510(k) Summary
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for "Purgo Dental Biologics Solution". The logo features a gray circular icon with a smaller circle inside and a line extending upwards to a small circle, followed by the word "Purgo" in a bold, purple font. To the right of "Purgo" are the words "Dental Biologics Solution" stacked vertically, also in purple.
### V. DEVICE DESCRIPTION
THE Graft Bone Substitute is a resorbable bone graft material made of porcine cancellous bone consisting of Hydroxyapatite(HA).
THE Graft Bone Substitute is a natural and porous bone mineral matrix available in cancellous granules packaged in a vial or syringe. It is manufactured by removal of most organic components from porcine bone. The composition of THE Graft Bone Substitute meets the requirements of ASTM F1581 Standard Specification for Composition of Anorqanic Bone for Surgical Implants. Due to its natural structure of macro and microscopic structures, the anorqanic bone mineral of THE Graft Bone Substitute is physically and chemically comparable to the mineralized matrix of human bone. When packed into a bony site, THE Graft Bone Substitute is gradually resorbed and replaced with new bone during the healing process. The formation and ingrowth of new bone at the implantation site of THE Graft Bone Substitute is due to its trabecular architecture, interconnecting macro and micro pores and its natural consistency.
THE Graft Bone Substitute is supplied sterile, non-pyrogenic, and for single use only.
### VI. INDICATIONS FOR USE
THE Graft Bone Substitute is intended for use as a bone grafting material in dental surgery such as:
- Filling of extraction sockets to enhance preservation of the alveolar ridge ।
- Grafting of maxillary sinus floor ।
### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
THE Graft Bone Substitute is substantially equivalent to the predicate device: Porcine Anorganic Bone Mineral K140714, in which the basic features, raw materials, material composition, intended uses, technological characteristics and performance characteristics are similar.
THE Graft Bone Substitute and the predicate device have similar Indications for Use. Both devices are intended for the same population and sterilized by gamma irradiation. The devices are made of biocompatible calcium phosphate bone mineral derived from porcine cancellous tissue adequate for bone formation. The devices feature trabecular-like characteristics, including interconnected macro and micro pores.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Purgo Dental Biologics Solution. The logo consists of a gray circular icon with a smaller circle inside and a magnifying glass symbol attached to the top left. To the right of the icon, the word "Purgo" is written in a bold, purple font. Further to the right, the words "Dental Biologics Solution" are stacked vertically in a smaller, purple font.
| Item | THE Graft | ZCORE; K140714 | Bio-Oss; K952617 |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | THE Graft Bone<br>Substitute is intended for<br>use as a bone grafting<br>material in dental surgery<br>such as:<br>- Filling of extraction sockets<br>to enhance preservation of<br>the alveolar ridge<br>- Grafting of maxillary sinus<br>floor | Porcine Anorganic Bone<br>Mineral is intended for use in<br>dental surgery. The products<br>may be used in surgical<br>procedures such as:<br>- Augmentation or<br>reconstructive treatment of<br>alveolar ridge<br>- Filling of infrabony<br>periodontal defects<br>- Filling of defects after root<br>resection, apicoectomy, and<br>cystectomy<br>- Filling of extraction sockets<br>to enhance preservation of<br>the alveolar ridge<br>- Elevation of maxillary sinus<br>floor<br>- Filling of periodontal<br>defects in conjunction with<br>products intended for<br>Guided Tissue Regeneration<br>(GTR) and Guided Bone<br>Regeneration (GBR)<br>- Filling of peri-implant<br>defects in conjunction with<br>products intended for<br>Guided Bone Regeneration | Geistlich Bio-Oss® is<br>intended for the following<br>uses:<br>- Augmentation or<br>reconstructive treatment of<br>the alveolar ridge;<br>- Filling of infrabony<br>periodontal defects;<br>- Filling of defects after root<br>resection, apicoectomy, and<br>cystectomy;<br>- Filling of extraction sockets<br>to enhance preservation of<br>the alveolar ridge;<br>- Elevation of the maxillary<br>sinus floor;<br>- Filling of periodontal<br>defects in conjunction with<br>products intended for<br>Guided Tissue Regeneration<br>(GTR) and Guided Bone<br>Regeneration (GBR); and<br>- Filling of peri-implant<br>defects in conjunction with<br>products intended for<br>Guided Bone Regeneration<br>(GER). |
| Material<br>Source | Porcine | Porcine | Bovine |
| Materials<br>Composition | Calcium phosphate | Calcium phosphate | Calcium phosphate |
| Mineral<br>Structure | Hydroxyapatite<br>containing Carbonate | Carbonate apatite | Hydroxyapatite |
| Form | Granules | Granules | Granules |
| Color | White | White | White |
| Physical<br>Appearance | Porous, irregular-shaped<br>particles | Porous, irregular-shaped<br>particles | Porous, irregular-shaped<br>particles |
| Particle Size<br>Range | 0.2 – 0.355 mm,<br>0.25mm – 1.0mm or<br>1.00mm – 2.0mm | 0.25mm – 1.0mm or<br>1.00mm – 2.0mm | 0.25mm – 1.0mm or<br>1.00mm – 2.0mm |
| Resorption<br>Profile | Gradual resorption | Gradual resorption | Gradual resorption |
| Package | Glass Vial or Glass<br>Syringe in a single Tyvek<br>Pouch | Plastic Jar in a single<br>Blister Tray with Tyvek<br>Lid | Glass Vial in a single<br>Blister Tray with Tyvek Lid |
| Sterility | Sterile, SAL 10-6<br>Gamma irradiation<br>(ISO 11137) | Sterile, SAL 10-6<br>Gamma irradiation<br>(ISO 11137) | Sterile, SAL 10-6<br>Gamma irradiation<br>(ISO 11137) |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Non-pyrogenic |
| Single<br>Use/Reuse | Single use only | Single use only | Single use only |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for "Purgo Dental Biologics Solution". The word "Purgo" is in a large, bold, purple font. To the left of the word is a gray icon that looks like a microscope. To the right of the word "Purgo" is the phrase "Dental Biologics Solution" in a smaller, stacked font.
#### VIII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Mechanical and Physical testing
Non-clinical laboratory performance testing was conducted to confirm that the composition of THE Graft Bone Substitute meets the requirements of ASTM F1581 Standard Specification for Composition of Anorganic Bone for Surgical Implants. This testing was performed in accordance with FDA's Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices, issued on April 28th of 2005.
### Biocompatibility testing
The biocompatibility evaluation for THE Graft Bone Substitute was conducted in accordance with the International Standard ISO 10993-1:2009 "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing Within a Risk Management Process" as recognized by FDA to further ensure substantial equivalence with the predicate device.
Assessment of the candidate device included the following tests:
- Cytotoxicity (ISO 10993-5) .
- 0 Sensitization Test (ISO 10993-10)
- Intracutaneous Reactivity (ISO 10993-10) .
- Acute Systemic Toxicity (ISO 10993-11) ●
- Subchronic Toxicity Test (ISO 10993-11) ●
- Genotoxicity (ISO 10993-3) 0
- Implantation Test (ISO 10993-6) 0
- . Pyrogen Testing
- Shelf-Life 0
#### Sterilization
In accordance with ISO 11137-1. THE Graft Bone Substitute is sterilized with gamma radiation with the use of minimum dose 25 kGy for sterility assurance level of 10 °.
- Sterilization Validation (ISO 11137-2) ●
- Endotoxin Test ●
- Packaging Validation
#### Viral Inactivation
Viral safety validation was performed and the result showed complete inactivation of viruses.
#### Non-clinical and Clinical Evaluation
In-vivo dog model was performed as well as Clinical data was provided to assess THE Graft and the predicate device as implantable materials.
All of the acceptance criteria were met.
510(k) Summary
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Purgo Dental Biologics Solution. The logo features a gray circular icon with a smaller circle inside and a handle extending from the top. To the right of the icon, the word "Purgo" is written in a bold, purple font. Further to the right, the words "Dental Biologics Solution" are stacked vertically, also in purple.
# IX. CONCLUSIONS
Based on the information provided within this 510(k) submission, Purgo Biologics Inc. concludes that THE Graft is substantially equivalent to the legally marketed predicate device listed according to the requirements of the Federal Food, Drug, and Cosmetic Act.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.