K173169 · Sharplight Technologies, Ltd. · GEX · Nov 20, 2017 · General, Plastic Surgery
Device Facts
Record ID
K173169
Device Name
Rapid Family
Applicant
Sharplight Technologies, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 20, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Rapid Family systems and optional Hand-Pieces are intended for use in aesthetic and cosmetic applications and in selective treatments required in the medical specialties of dermatology. Rapid Family systems and optional Hand-pieces with 420 – 1200 nm wavelengths (with and without contact-cooling) are indicated for: * Hair removal and Permanent Hair Reduction in all skin types (I-VI) to the Fitzpatrick scale - Recommended wavelengths in the range of 635-950 nm, 730-950 nm or 580-950 nm * Treatment of Pigmented and Vascular Lesions in all skin types (I-VI) to the Fitzpatrick scale - Recommended wavelength in the range of 535-950 nm * Treatment of Inflammatory Acne (acne vulgaris) in skin types (I-V) to the Fitzpatrick scale – Recommended wavelength in the range of 420-950 nm * Treatment of Benign Pigmented Epidermal and Cutaneous Lesions including warts, scars and striae in all skin types (I-VI) to the Fitzpatrick scale - Recommended wavelength in the range of 580-950 nm * Treatment of Benign Pigmented Epidermal, Cutaneous and Vascular Lesions including warts; scars and striae in skin types (I-V) to the Fitzpatrick scale - Recommended wavelength in the range of 535-680 nm & 860-1200nm Rapid Family systems and optional Infrared (IR) Hand Pieces with 850 - 1750 nm wavelengths (with contact cooling) are indicated for: * Dermatologic Treatment such as, but not limited to: photocoagulation of soft tissue (Scars, Wrinkles, Rhytids and Periorbital Wrinkles) Rapid Family systems and optional Nd: YAG Laser 1064 nm Long Pulse (LP) Hand-Pieces are indicated for: * Removal of unwanted hair, for stable long-term, or permanent, hair reduction through selective targeting of melanin in hair follicles. * Removal or lightening of unwanted hair (with and without adjuvant preparation) * Treatment of pseudofolliculitis barbae (PFB) Rapid Family systems and optional Nd:YAG Laser 1064 nm Long Pulse (LP) and Q-Switched Hand-Pieces are indicated for: * Benign vascular lesions such as, but not limited to treatment of: Port wine stains; Hemangiomas; Warts, Superficial and deep telangiectasias (venulectasias); Reticular veins (0.1-4.0 mm dia.) of the leg; Rosacea; Venus lake; Leg veins; Spider veins; Poikiloderma of Civatte; Angiomas * Benign cutaneous lesions, such as, but not limited to: Warts; Scars; Striae; Psoriasis * Benign pigmented lesions such as, but not limited to: Lentigos (age spots); Solar lentigos (sun spots); Cafe-au-lait macules; Sebortheic; keratoses; Nevi and nevus of Ota, Chloasma; Verrucae; Skin tags; Keratoses; The removal of black, blue or green tattoos (significant reduction in the intensity of blue/black tattoos); Plaques * Pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment and for patients with lesions that have not responded to other laser treatments. * The non-ablative treatment of facial wrinkles, such as, but not limited to: Periocular wrinkles Rapid Family systems and optional Nd:YAG Laser 1064 nm Long Pulse (LP) and Q-Switched Hand-Pieces are indicated for: * Laser skin resurfacing procedures for the treatment of: Acne scars; Wrinkles * Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. * Indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin. Rapid Family systems and optional Nd: YAG Laser 532 nm Frequency Dabbler (FD) Long Pulse (LP) and O-Switched Hand-Pieces are indicated for: * Tattoo removal: Light blue; Yellow; Red; Green * Vascular lesions: Hemangiomas (Port wine stains/birthmarks, cavernous, cherry, spider); hemangiomas; Angiomas (cherry, spider); Telangiectasia; Spider nevi * Benign pigmented lesions, such as, but not limited to: Cafe-au-lait (macules); Lentigines (senile and solar); Freckles (ephelides); Chloasma; Nevi; Nevus spillus, Nevus of Ota; Becker's Nevi; * Other pigmented cutaneous lesions, such as, but not limited to: Verrucae; Skin tags; Plaques Rapid Family systems and optional Er: Y AG Laser 2940 nm Long Pulse (LP) Hand-Pieces, with standard and scanner accessory tips are indicated for: Use in soft tissue (skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs, and glands) such as, but not limited to: * Dermatology and plastic Surgery: Skin resurfacing; Treatment of wrinkles; Epidermal nevi; Telangiectasia; Spider veins: Actinic chelitis: Keloids: Verrucae: Skin tags: Keratoses: Scar revision (including acne scars)
Device Story
Rapid Family is a multi-platform aesthetic laser/light-based system. Device utilizes various hand-pieces with specific wavelengths (420-1200 nm, 850-1750 nm, 1064 nm Nd:YAG, 532 nm FD, 2940 nm Er:YAG) to deliver energy to skin/soft tissue. Principle of operation involves selective photothermolysis or photocoagulation targeting melanin, hemoglobin, or water. Used in clinical settings by trained medical professionals. Output is controlled light/laser energy applied to patient skin. Healthcare providers use device to treat dermatological conditions, vascular/pigmented lesions, and perform skin resurfacing. Benefits include non-invasive or minimally invasive removal of unwanted hair, lesions, tattoos, and improvement of skin texture/appearance.
Technological Characteristics
Multi-wavelength laser/light system. Wavelengths: 420-1200 nm (IPL/light), 850-1750 nm (IR), 1064 nm (Nd:YAG), 532 nm (FD), 2940 nm (Er:YAG). Features contact cooling. Includes standard and scanner accessory tips. Form factor: console with interchangeable hand-pieces.
Indications for Use
Indicated for aesthetic/cosmetic dermatology applications including hair removal/reduction (all skin types I-VI), treatment of pigmented/vascular lesions, inflammatory acne, benign cutaneous lesions (warts, scars, striae), tattoo removal, and skin resurfacing/wrinkle reduction. Applicable to Fitzpatrick skin types I-VI depending on specific hand-piece/wavelength.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Submission Summary (Full Text)
{0}------------------------------------------------
November 20, 2017
SharpLight Technologies Ltd. % Ilan Sharon Ilan Sharon A 109 Neot Golf Caesarea, Il
Re: K173169
Trade/Device Name: Rapid Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 19, 2017 Received: September 29, 2017
Dear Ilan Sharon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820):
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@Ida.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173169
Device Name Rapid Family
Indications for Use (Describe)
Rapid Family systems and optional Hand-Pieces are intended for use in aesthetic and cosmetic applications and in selective treatments required in the medical specialties of dermatology.
Rapid Family systems and optional Hand-pieces with 420 – 1200 nm wavelengths (with and without contact-cooling) are indicated for:
* Hair removal and Permanent Hair Reduction in all skin types (I-VI) to the Fitzpatrick scale - Recommended wavelengths in the range of 635-950 nm, 730-950 nm or 580-950 nm
* Treatment of Pigmented and Vascular Lesions in all skin types (I-VI) to the Fitzpatrick scale - Recommended wavelength in the range of 535-950 nm
* Treatment of Inflammatory Acne (acne vulgaris) in skin types (I-V) to the Fitzpatrick scale – Recommended wavelength in the range of 420-950 nm
* Treatment of Benign Pigmented Epidermal and Cutaneous Lesions including warts, scars and striae in all skin types (I-VI) to the Fitzpatrick scale - Recommended wavelength in the range of 580-950 nm
* Treatment of Benign Pigmented Epidermal, Cutaneous and Vascular Lesions including warts; scars and striae in skin types (I-V) to the Fitzpatrick scale - Recommended wavelength in the range of 535-680 nm & 860-1200nm
Rapid Family systems and optional Infrared (IR) Hand Pieces with 850 - 1750 nm wavelengths (with contact cooling) are indicated for:
* Dermatologic Treatment such as, but not limited to: photocoagulation of soft tissue (Scars, Wrinkles, Rhytids and Periorbital Wrinkles)
Rapid Family systems and optional Nd: YAG Laser 1064 nm Long Pulse (LP) Hand-Pieces are indicated for:
* Removal of unwanted hair, for stable long-term, or permanent, hair reduction through selective targeting of melanin in hair follicles.
* Removal or lightening of unwanted hair (with and without adjuvant preparation)
* Treatment of pseudofolliculitis barbae (PFB)
Rapid Family systems and optional Nd:YAG Laser 1064 nm Long Pulse (LP) and Q-Switched Hand-Pieces are indicated for:
* Benign vascular lesions such as, but not limited to treatment of: Port wine stains; Hemangiomas; Warts, Superficial and deep telangiectasias (venulectasias); Reticular veins (0.1-4.0 mm dia.) of the leg; Rosacea; Venus lake; Leg veins; Spider veins; Poikiloderma of Civatte; Angiomas
* Benign cutaneous lesions, such as, but not limited to: Warts; Scars; Striae; Psoriasis
* Benign pigmented lesions such as, but not limited to: Lentigos (age spots); Solar lentigos (sun spots); Cafe-au-lait macules; Sebortheic; keratoses; Nevi and nevus of Ota, Chloasma; Verrucae; Skin tags; Keratoses; The removal of black, blue or green tattoos (significant reduction in the intensity of blue/black tattoos); Plaques
* Pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment and for patients with lesions that have not responded to other laser treatments.
* The non-ablative treatment of facial wrinkles, such as, but not limited to: Periocular wrinkles
Rapid Family systems and optional Nd:YAG Laser 1064 nm Long Pulse (LP) and Q-Switched Hand-Pieces are indicated for:
* Laser skin resurfacing procedures for the treatment of: Acne scars; Wrinkles
{3}------------------------------------------------
* Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
* Indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin.
Rapid Family systems and optional Nd: YAG Laser 532 nm Frequency Dabbler (FD) Long Pulse (LP) and O-Switched Hand-Pieces are indicated for:
* Tattoo removal: Light blue; Yellow; Red; Green
* Vascular lesions: Hemangiomas (Port wine stains/birthmarks, cavernous, cherry, spider); hemangiomas; Angiomas (cherry, spider); Telangiectasia; Spider nevi
* Benign pigmented lesions, such as, but not limited to: Cafe-au-lait (macules); Lentigines (senile and solar); Freckles (ephelides); Chloasma; Nevi; Nevus spillus, Nevus of Ota; Becker's Nevi;
* Other pigmented cutaneous lesions, such as, but not limited to: Verrucae; Skin tags; Plaques
Rapid Family systems and optional Er: Y AG Laser 2940 nm Long Pulse (LP) Hand-Pieces, with standard and scanner accessory tips are indicated for:
Use in soft tissue (skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs, and glands) such as, but not limited to:
* Dermatology and plastic Surgery: Skin resurfacing; Treatment of wrinkles; Epidermal nevi; Telangiectasia; Spider veins: Actinic chelitis: Keloids: Verrucae: Skin tags: Keratoses: Scar revision (including acne scars)
NOTE Permanent hair reduction is defined as the long-term stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regimen
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
er-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.