The Pakter Curved Needle Set is intended for discography, diagnostic sampling, aspiration, and injection.
Device Story
Pakter Curved Needle Set is a coaxial needle system comprising a preshaped curved nitinol needle (30° bevel) and a stainless steel introducer needle. Used percutaneously by clinicians to access target anatomy; introducer needle and stylet are inserted first; curved needle is then advanced through the introducer cannula to the target site. Facilitates diagnostic sampling, aspiration, and injection. Device is sterile, single-use, and intended for clinical settings. Provides clinicians with a steerable/curved access tool for precise placement compared to straight needles.
Clinical Evidence
Bench testing only. Testing included tensile strength (BS EN ISO 7864:2016, BS EN ISO 11070:2014), stiffness (BS EN ISO 9626:2016), resistance to breakage/corrosion, liquid leakage, and dimensional compatibility. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogen, hemolysis) confirmed safety.
Technological Characteristics
Coaxial needle set. Curved needle: Nitinol cannula, polycarbonate hub, 30° bevel tip. Introducer: Stainless steel cannula, acrylic hub. Includes stainless steel stylets. Sterilized via ethylene oxide. Standards: BS EN ISO 7864:2016, BS EN ISO 11070:2014, BS EN ISO 9626:2016, ISO 10993-1.
Indications for Use
Indicated for discography, diagnostic sampling, aspiration, and injection in patients requiring these procedures.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
June 7, 2018
Cook Incorporated Johnathan Liu Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47404
Re: K173155
Trade/Device Name: Pakter Curved Needle Set Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: May 7, 2018 Received: May 8, 2018
Dear Johnathan Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173155
Device Name Pakter Curved Needle Set
Indications for Use (Describe)
The Pakter Curved Needle Set is intended for discography, diagnostic sampling, aspiration, and injection.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in large, white, sans-serif font, and the word "MEDICAL" is in a smaller, white, sans-serif font below the word "COOK".
#### 2.0 510(k) SUMMARY
### Pakter Curved Needle Set 21 CFR §807.92(c) Date Prepared: May 2, 2018
### Submitted By:
| Submission: | Traditional 510(k) Premarket Notification | |
|-----------------------|-------------------------------------------|--|
| 510(k) Number: | K173155 | |
| Applicant: | Cook Incorporated | |
| Applicant Address: | Cook Incorporated | |
| | 750 Daniels Way | |
| | Bloomington, IN 47404 | |
| Contact: | Johnathan Liu | |
| Email: | regsubmissions@cookmedical.com | |
| Contact Phone Number: | (812) 335-3575 x104509 | |
| Contact Fax Number: | (812) 332-0281 | |
### Device Information:
| Trade Name: | <b>Pakter Curved Needle Set</b> |
|-----------------------|--------------------------------------------------|
| Common Name: | Needle, Conduction, Anesthetic (W/Wo Introducer) |
| Classification Name: | Anesthesia conduction needle |
| Regulation: | 21 CFR §868.5150 |
| Product Code: | BSP |
| Device Class: | II |
| Classification Panel: | Anesthesiology |
### Predicate Device:
The Pakter Curved Needle Set is considered substantially equivalent to the following device: Atraumatic Standard Sprotte® Needle (K911260, Pajunk GMBH) cleared on October 21, 1991.
{4}------------------------------------------------
#### Device Description:
The Pakter Curved Needle Set is manufactured as a coaxial needle set and includes a preshaped curved needle and an introducer needle. The curved needle consists of a nitinol cannula connected to a polycarbonate hub at the proximal end and is characterized with a distal curve and a 30° bevel distal tip. It also includes a stainless steel stylet that is designed for a transitional fit. The curved needle with stylet is 25 gage in diameter and has a length of 20 or 15 centimeters. The introducer needle consists of a stainless steel cannula connected to an acrylic hub at the proximal end. The introducer needle is also designed with a stainless steel stylet. The introducer needle with stylet is 21 gage in diameter and has a length of 15 or 10 centimeters. When used as intended, the introducer needle and stylet are first inserted percutaneously and advanced adjacent to the target anatomy. After removing the introducer needle stylet, the curved needle is then inserted through the introducer needle cannula and slowly advanced to the target anatomy. The Pakter Curved Needle Set is sterilized by ethylene oxide and intended for one-time use.
#### Indications for Use:
The Pakter Curved Needle Set is intended for discography, diagnostic sampling, aspiration, and injection.
#### Comparison to Predicate Device:
The Pakter Curved Needle Set and the predicate device, Atraumatic Standard Sprotte® Needle (K911260), are considered substantially equivalent in that these devices are identical in principles of operation and fundamental technologies. Additionally, the subject device has similar indications for use, coaxial design, materials, and dimensions as the predicate device.
The differences in technological characteristics between the subject device and the predicate device include the following:
- . Materials - The subject device needle cannula is made of Nitinol, while the predicate device's cannula is made of stainless steel. The subject device's introducer needle cannula hub is acrylic, while the predicate device's hub is made of polycarbonate.
- . Dimensions - The subject device's needle cannula and introducer needle cannula are longer in length than those of the predicate device: the subject device's needle cannula gage size is smaller than that of the predicate device; the needle cannula tip of the subject device is curved with a 30-degree bevel, while that of the predicate
{5}------------------------------------------------
device is straight with a pencil point bevel; and the introducer needle cannula gage size of the subject device is smaller than that of the predicate device.
- . Additionally, the subject device set includes a needle cannula stylet, while the predicate device set does not.
These differences in technological characteristics between the subject and predicate devices do not raise any different questions of safety and/or effectiveness.
| | | PREDICATE DEVICE | SUBJECT DEVICE |
|-------------------------------|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| | | Atraumatic Standard Sprotte®<br>Needle<br>(K911260) | Pakter Curved Needle Set |
| Regulation | | 21 CFR § 868.5150 | Identical |
| Product Code | | BSP | Identical |
| Classification | | II | Identical |
| Indications for Use | | To gain entry into or puncture the<br>spinal cavity permitting<br>injection/withdrawal of fluids for<br>purposes of diagnostic lumbar<br>puncture, myelography/discography<br>and chemonuceolysis procedures. | The Pakter Curved Needle Set is<br>intended for discography, diagnostic<br>sampling, aspiration, and injection. |
| Device is for One-time<br>Use | | Yes | Identical |
| Needle * | Cannula<br>Material | Stainless Steel | Nitinol |
| | Cannula Hub<br>Material | Polycarbonate | Identical |
| | Cannula Length | 7 to 15 cm | 15, 20 cm |
| | Cannula Size | 19.5 to 24 gage | 25 gage |
| | Cannula Tip | Straight, Pencil point | Curved, 30° bevel |
| | Stylet Material | Stainless Steel | Identical |
| Introducer Needle | Cannula<br>Material | Stainless Steel | Identical |
| | Cannula Hub<br>Material | Polycarbonate | Acrylic |
| | Cannula Size | 0.7 to 1.0 mm | 21 gage (0.82 mm) |
| | Cannula Length | 3, 4 cm | 10, 15 cm |
| | Stylet Material | No stylet | Stainless Steel |
| Packaging | | Pouch | Peel Pouch |
| Sterilization Method | | EtO | Identical |
| Sterility Assurance<br>Level | | 10-6 | Identical |
*Subject device needle is curved. Predicate device needle is not curved.
Indications for Use: The indications for use of the subject device is similar to that of the predicate device. Both the subject device and predicate device are indicated for discography. Although the wording is different, the subject device indications for
{6}------------------------------------------------
"diagnostic sampling, aspiration, and injection" are synonymous with the predicate device indications for "injection/withdrawal of fluids." The only difference between the subject device and predicate device intended use statements is that the subject device is not indicated for myelography and chemonucleolysis. The wording differences and exclusion of myelography and chemonucleolysis in the indications for use statement of the subject device do not raise different questions of safety and/or effectiveness.
### Technological Characteristics:
The subject device. Pakter Curved Needle Set, underwent the following applicable testing to ensure reliable design and performance under the specified testing parameters:
### Bench Testing (including time-zero and five-year accelerated aged testing)
- Tensile Strength .
- Curved Needle Cannula Hub-to-Shaft The peak tensile load for the hub-to-. shaft section of the curved needle cannula shall be greater than or equal to 22 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
- . Curved Needle Stylet Hub-to-Shaft – The peak tensile load for the hub-to-shaft section of the curved needle stylet shall be greater than or equal to 11 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
- Introducer Needle Cannula Hub-to-Shaft The peak tensile load for the hub-to-. shaft section of the introducer needle cannula shall be greater than or equal to 44 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
- Introducer Needle Stylet Hub-to-Shaft The peak tensile load for the hub-to-. shaft section of the introducer needle stylet shall be greater than or equal to 22 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
- . Stiffness - Introducer Needle Cannula and Stylet: The introducer needle cannula and stylet shall show a deflection of less than 0.45 mm when a force of 9.6 N is applied in accordance with BS EN ISO 9626:2016. The acceptance criterion was met.
{7}------------------------------------------------
- . Resistance to Breakage
- . Curved Needle Cannula - The curved needle cannula shall show no visible breakage after cyclic testing in accordance with BS EN ISO 9626:2016. The acceptance criterion was met.
- . Introducer Needle Cannula – The introducer needle cannula shall show no visible breakage after cyclic testing in accordance with BS EN ISO 9626:2016. The acceptance criterion was met.
- . Resistance to Corrosion - Following testing in accordance with the method given in Annex B of BS EN ISO 11070. the curved needle and introducer needle will be examined for visual signs of corrosion and the results will be reported. The acceptance criterion was met.
- . Tensile Strength Post-Corrosion
- Curved Needle Cannula Hub-to-Shaft The peak tensile load for the hub-to-. shaft section of the curved needle cannula shall be greater than or equal to 22 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
- . Curved Needle Stylet Hub-to-Shaft – The peak tensile load for the hub-to-shaft section of the curved needle stylet shall be greater than or equal to 11 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
- Introducer Needle Cannula Hub-to-Shaft The peak tensile load for the hub-to-. shaft section of the introducer needle cannula shall be greater than or equal to 44 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
- Introducer Needle Stylet Hub-to-Shaft The peak tensile load for the hub-to-. shaft section of the introducer needle stylet shall be greater than or equal to 22 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
- Liquid Leakage Under Positive Pressure The curved needle cannula shall maintain a . positive pressure of 300 kPa for 30 seconds without leakage in accordance with BS EN ISO 11070:2014. The acceptance criterion was met.
- . Patency of Lumen – A stainless steel stylet with a diameter of 0.009 inches (0.23 mm) shall pass through the lumen of the curved needle cannula. The acceptance criterion was met.
{8}------------------------------------------------
- . Dimensional Analysis and Compatibility - Testing conducted demonstrated that the curved needle and introducer needle dimensions are within the specified tolerances and the components are compatible.
Biocompatibility Testing:
- . Per ISO 10993-1 and FDA guidance, testing for cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity, material mediated pyrogen, and hemolysis study demonstrated the biocompatibility of the subject device.
#### Conclusion:
The results of these tests confirm that the Pakter Curved Needle Set meets the design input requirements based on the intended use and support the conclusion that this device does not raise different questions of safety and/or effectiveness and is substantially equivalent to the predicate device, Atraumatic Standard Sprotte® Needle (K911260).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.