VivaChek™ Ino Smart Blood Glucose Monitoring System is comprised of the VivaChek™ Ino Smart Blood Glucose Meter (VGM04) and the VivaChek™ Ino Blood Glucose Test Strips (VGS01). The VivaChek™ Ino Smart Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.
Device Story
System measures glucose in capillary whole blood via amperometric detection; uses glucose oxidase enzyme on test strips. Blood drawn via capillary action into strip; reaction generates electrons proportional to glucose concentration; meter calculates and displays plasma-equivalent results. Used by patients at home for self-monitoring; some models (Smart, Sync, Sound, Plus) include Bluetooth for mobile data transfer; some (Sound, Simple, Bright) include voice output. Provides quantitative glucose readings to aid diabetes management. Requires 0.8µL sample; results displayed after reaction time. Not for multi-patient use; requires specific disinfection protocols.
Clinical Evidence
Four independent user performance studies (n≥350 per study) evaluated lay user accuracy. Results for the most complex model (VGM09) showed 100% of results within ±20% of YSI 2300 comparator for all glucose concentrations. Accuracy at extreme glucose levels (<80 mg/dL and >250 mg/dL) also met acceptance criteria. Usability testing confirmed lay users could correctly perform testing and interpret results.
Indicated for persons with diabetes for quantitative measurement of glucose in fresh capillary whole blood from the fingertip as an aid to monitor diabetes control. Not for neonatal use, diagnosis, or screening. Contraindicated for critically ill patients, hypotensive individuals, patients in shock, severe dehydration, or hyperosmolar states.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
VivaChek™ Ino Blood Glucose Monitoring System (K160179)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k173140
B. Purpose for Submission:
New devices
C. Measurand:
Glucose in capillary whole blood obtained from the fingertip
D. Type of Test:
Quantitative, amperometric method, glucose oxidase
E. Applicant:
VivaChek Laboratories, Inc.
F. Proprietary and Established Names:
VivaChek Ino Smart Blood Glucose Monitoring System
VivaChek Ino Sync Blood Glucose Monitoring System
VivaChek Ino Sound Blood Glucose Monitoring System
VivaChek Ino Plus Blood Glucose Monitoring System
VivaChek Ino Sound Simple Blood Glucose Monitoring System
VivaChek Ino Sound Bright Blood Glucose Monitoring System
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1345, Glucose test system
2. Classification:
Class II
{1}
3. Product code:
NBW, System, Test, Blood Glucose, Over The Counter
4. Panel:
Clinical Chemistry (75)
## H. Intended Use:
1. Intended use(s):
See Indication(s) for use below
2. Indication(s) for use:
### VivaChek Ino Smart Blood Glucose Monitoring System
VivaChek Ino Smart Blood Glucose Monitoring System is comprised of the VivaChek Ino Smart Blood Glucose Meter (VGM04) and the VivaChek Ino Blood Glucose Test Strips (VGS01). The VivaChek Ino Smart Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.
### VivaChek Ino Sync Blood Glucose Monitoring System
VivaChek Ino Sync Blood Glucose Monitoring System is comprised of the VivaChek Ino Sync Blood Glucose Meter (VGM05) and the VivaChek Ino Blood Glucose Test Strips (VGS01). The VivaChek Ino Sync Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.
### VivaChek Ino Sound Blood Glucose Monitoring System
VivaChek Ino Sound Blood Glucose Monitoring System is comprised of the VivaChek Ino Sound Blood Glucose Meter (VGM09) and the VivaChek Ino Blood Glucose Test Strips (VGS01). The VivaChek Ino Sound Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-
{2}
testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.
## VivaChek Ino Plus Blood Glucose Monitoring System
VivaChek Ino Plus Blood Glucose Monitoring System is comprised of the VivaChek Ino Plus Blood Glucose Meter (VGM22) and the VivaChek Ino Blood Glucose Test Strips (VGS01). The VivaChek Ino Plus Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.
## VivaChek Ino Sound Simple Blood Glucose Monitoring System
VivaChek Ino Sound Simple Blood Glucose Monitoring System is comprised of the VivaChek Ino Sound Simple Blood Glucose Meter (VGM26) and the VivaChek Ino Blood Glucose Test Strips (VGS01). The VivaChek Ino Sound Simple Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.
## VivaChek Ino Sound Bright Blood Glucose Monitoring System
VivaChek Ino Sound Bright Blood Glucose Monitoring System is comprised of the VivaChek Ino Sound Bright Blood Glucose Meter (VGM27) and the VivaChek Ino Blood Glucose Test Strips (VGS01). The VivaChek Ino Sound Bright Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.
## 3. Special conditions for use statement(s):
- Very high (above 70%) and very low (below 20%) hematocrit levels can cause false results. Talk to your health care professional to find out your hematocrit level.
- Not for use in hypotensive individuals
- Critically ill patients should not be tested with this system.
- Not for use on patients in shock, or with severe dehydration or from patients in a hyperosmolar state (with or without ketosis).
- Not for neonatal use.
- Not for screening or diagnosis of diabetes mellitus.
{3}
- Do not use at altitudes above 13,123ft (4000 meters) above sea level.
- For single-patient use only
- This device is not intended for use in healthcare or assisted-use settings such as hospitals, physician offices, or long-term care facilities because it has not been cleared by FDA for use in these settings, including for routine assisted testing or as part of glycemic control procedures. Use of this device on multiple patients may lead to transmission of Human Immunodeficiency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or other bloodborne pathogens.
4. Special instrument requirements:
VivaChek Ino Smart Blood Glucose Meter (VGM04)
VivaChek Ino Sync Blood Glucose Meter (VGM05)
VivaChek Ino Sound Blood Glucose Meter (VGM09)
VivaChek Ino Plus Blood Glucose Meter (VGM22)
VivaChek Ino Sound Simple Blood Glucose Meter (VGM26)
VivaChek Ino Sound Bright Blood Glucose Meter (VGM27)
I. Device Description:
The six systems in this submission, the VivaChek Ino Smart, the VivaChek Ino Sync, the VivaChek Ino Sound, the VivaChek Ino Plus, the VivaChek Ino Sound Simple, and the VivaChek Ino Sound Bright Blood Glucose Monitoring Systems are intended to quantitatively measure the glucose concentration in fresh capillary whole blood. The glucose measurement is achieved by using the amperometric detection method. The test is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with glucose enzyme and the mediator. Electrons are generated, producing a current that is positive correlation to the glucose concentration in the sample. After the reaction time, the glucose concentration in the sample is displayed. The meter is calibrated to display plasma-like concentration results.
The same test strips, the VivaChek Ino Blood Glucose Test Strips (VGS01) are used with each of the meters in this submission, the VivaChek Ino Smart (VGM04), VivaChek Ino Sync (VGM05), VivaChek Ino Sound (VGM09), VivaChek Ino Plus (VGM22), VivaChek Ino Sound Simple (VGM26), and VivaChek Ino Sound Bright (VGM27) Blood Glucose Meters to measure the glucose level in fresh capillary whole blood. VivaChek Control Solutions are aqueous solutions containing glucose and are available at three levels (level 1, level 2 and level 3) and are compatible for use with each system.
All of the devices in this 510(k) submission employ the same device technology and have the same intended use. The devices differ in minor design features and certain user functions. For example, the VGM04, VGM05, VGM09, and VGM22 meter models contain Bluetooth functionality for transferring glucose results to a compatible mobile device, and the VGM09, VGM26, and VGM27 meter models contain a talking functionality. Other
{4}
features that differ between meters are minor features such as presence or absence of a backlight, strip ejector port, or rechargeable batteries.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
VivaChek Ino Blood Glucose Monitoring System
2. Predicate 510(k) number(s):
k160179
3. Comparison with predicate:
| Item | Candidate Devices (k173140) | Predicate Device (k160179) |
| --- | --- | --- |
| Indications for use | Quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip as an aid to monitor the effectiveness of diabetes control | Same |
| Detection method | Amperometric | Same |
| Enzyme | Glucose oxidase | Same |
| Measuring range | 20-600 mg/dL | Same |
| HCT range | 20-70% | Same |
| Sample volume | 0.8μL | Same |
| Operating temperature | 5–45°C (41–113°F) | Same |
| Operating relative humidity | 10-90% | Same |
| Sample type | Capillary whole blood from fingertip | Capillary whole blood from fingertip, palm and forearm |
| Altitude | 13,123 feet | 9,777 feet |
{5}
6
# K. Standard/Guidance Document Referenced (if applicable):
Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use, Guidance for Industry and Food and Drug Administration Staff. October 11, 2016.
# L. Test Principle:
The VivaChek Ino Smart Blood Glucose Monitoring System, VivaChek Ino Sync Blood Glucose Monitoring System, VivaChek Ino Sound Blood Glucose Monitoring System, VivaChek Ino Plus Blood Glucose Monitoring System, VivaChek Ino Sound Simple Blood Glucose Monitoring System, and VivaChek Ino Sound Bright Blood Glucose Monitoring System are systems for the quantitative detection of glucose in fresh capillary whole blood from the fingertip. Amperometric technology is used for the detection of glucose from testing the strip (with whole blood sample) on the meter. Reagent consisting of glucose oxidase and mediator is deposited onto the reaction cell section of the test strip with printed electrodes. When a drop of whole blood sample is applied to the reaction cell on the test strip, glucose in the blood sample reacts in the presence of glucose oxidase and mediator, and the reaction yields electrons. Thus, a current signal is produced from the reaction and detected by the meter. The detected current signal is then calculated by the meter, and glucose concentration reading is then displayed on the meter display.
# M. Performance Characteristics (if/when applicable):
Unless otherwise noted, the analytical studies detailed below were performed with the VivaChek Ino Smart Blood Glucose Monitoring System. As the only differences between the six meters in this submission are minor design features, as described above in Device Description (Section I), the analytical studies performed with the VivaChek Ino Smart Blood Glucose Monitoring System are representative of the performance for all devices in this 510(k) submission.
1. Analytical performance:
a. Precision/Reproducibility:
Repeatability (within-run precision):
The sponsor performed repeatability (within-day) precision studies using venous whole blood adjusted to 5 different glucose concentration levels (30 to 50, 51 to 110, 111 to 150, 151 to 250, and 251 to 400 mg/dL). Each glucose concentration level was analyzed in replicates of 10, with 3 test strip lots, and 10 meters, for a total of 300 tests per glucose level for each meter. Results are summarized below:
{6}
7
| Glucose Level (mg/dL) | Strip Lot | n | Average Glucose (mg/dL) | SD (mg/dL) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| 30-50 | 1 | 100 | 49.5 | 2.3 | 4.6 |
| | 2 | 100 | 49.5 | 2.2 | 4.4 |
| | 3 | 100 | 49.5 | 2.2 | 4.5 |
| 51-110 | 1 | 100 | 78.1 | 2.5 | 3.2 |
| | 2 | 100 | 78.1 | 2.7 | 3.4 |
| | 3 | 100 | 78.1 | 2.6 | 3.3 |
| 111-150 | 1 | 100 | 133.5 | 4.0 | 3.0 |
| | 2 | 100 | 133.5 | 3.7 | 2.8 |
| | 3 | 100 | 133.5 | 3.8 | 2.8 |
| 151-250 | 1 | 100 | 200.7 | 5.8 | 2.9 |
| | 2 | 100 | 200.7 | 5.7 | 2.9 |
| | 3 | 100 | 200.7 | 5.4 | 2.7 |
| 251-400 | 1 | 100 | 357.5 | 10.3 | 2.8 |
| | 2 | 100 | 357.5 | 10.5 | 2.8 |
| | 3 | 100 | 357.5 | 10.2 | 2.8 |
## Intermediate Precision:
Intermediate (day-to-day) precision was evaluated using 3 levels of glucose control solutions over 10 days with 3 test strip lots. For each level, on each day, 10 meters were used for testing, with 1 replicate collected per meter for a total of 10 replicates per day for each glucose level. Results are summarized below:
| Glucose Level (mg/dL) | Strip Lot | n | Average Glucose (mg/dL) | SD (mg/dL) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| Level 1 | 1 | 100 | 50.5 | 2.4 | 4.8 |
| | 2 | 100 | 50.6 | 2.4 | 4.8 |
| | 3 | 100 | 50.5 | 2.4 | 4.6 |
| Level 2 | 1 | 100 | 117.5 | 3.2 | 2.7 |
| | 2 | 100 | 119.1 | 2.8 | 2.4 |
| | 3 | 100 | 118.1 | 2.8 | 2.3 |
| Level 3 | 1 | 100 | 360.0 | 8.5 | 2.4 |
| | 2 | 100 | 359.6 | 9.4 | 2.6 |
| | 3 | 100 | 358.2 | 9.1 | 2.5 |
## b. Linearity/assay reportable range:
Linearity testing was performed using venous whole blood samples. Samples with the following glucose concentrations were prepared and confirmed by the laboratory comparator method YSI 2300: 21.7, 77.7, 135.7, 196.3, 250.8, 308.6, 368.1, 425.9,
{7}
481.0, 541.6, 595.0 mg/dL. Each sample was measured in replicates of four using each of three lots of test strips and the mean was used in the regression analysis. The regression analysis results are summarized in the table below:
| Lot | Slope | y-intercept | R² |
| --- | --- | --- | --- |
| 1 | 0.9983 | 0.0150 | 0.9986 |
| 2 | 0.9852 | 2.4035 | 0.9974 |
| 3 | 1.0052 | -2.2806 | 0.9982 |
The results of the study support the sponsor's claimed glucose measuring range of 20 – 600 mg/dL. If a sample is less than 20 mg/dL glucose, the result is flagged by the meter as LO. If a sample result exceeds 600 mg/dL glucose, the result is flagged by the meter as HI. The LO and HI functions were validated and demonstrated to function as intended.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
The system is traceable to NIST SRM #917b glucose reference material. A method comparison was performed using the candidate device and YSI 2300 as the reference method. The meter provides plasma-equivalent results.
Test Strip Stability:
Test strip stability was assessed using accelerated and real time stability studies. Protocols and acceptance criteria were reviewed and found acceptable. The labeling includes claims that the VivaChek Ino Test Strips are stable for 4 months after opening and 24 months unopened when stored between of 36°F to 86°F (2°C-30°C) and 10-90% relative humidity.
d. Detection limit:
The reportable range for the VivaChek Blood Glucose Monitoring System is 20 to 600 mg/dL supported by the linearity assay study above (M.1.b).
e. Analytical specificity:
To assess potential interferences, the sponsor used venous whole blood samples adjusted to 3 glucose concentration ranges: 50-70, 110-130, and 225-270 mg/dL. Each of these samples was divided into a test pool and a control pool, with each of the potential endogenous and exogenous interfering substances added to the test pool. The % difference between the test sample and the control sample was calculated using mean of 10 replicates for each of the 3 strip lots tested. The compounds at the concentrations listed below did not have significant interference (defined by the sponsor as average bias between spiked and control sample glucose results within ±
8
{8}
10% with glucose >100 mg/dL and within 10 mg/dL with glucose <100 mg/dL). The highest concentrations at which no significant interference was observed are presented in the following table:
| Substance | Highest Concentration tested with no observed interference |
| --- | --- |
| Acetaminophen | 20 mg/dL |
| Ascorbic acid | 3 mg/dL |
| Bilirubin | 40 mg/dL |
| Cholesterol | 500 mg/dL |
| Creatinine | 10 mg/dL |
| Dopamine | 20 mg/dL |
| Galactose | 15 mg/dL |
| Gentisic acid | 27.8 mg/dL |
| Glutathione | 92 mg/dL |
| Hemoglobin | 20 g/dL |
| Ibuprofen | 50 mg/dL |
| L-Dopa | 0.5 mg/dL |
| Maltose | 225 mg/dL |
| Methyldopa | 10.5 mg/dL |
| Salicylic acid | 60 mg/dL |
| Sodium | 350 mg/dL |
| Tolbutamide | 64 mg/dL |
| Tolazamide | 10 mg/dL |
| Triglycerides | 1500 mg/dL |
| Uric acid | 24 mg/dL |
| Xylose | 200 mg/dL |
| Mannitol | 0.09 mg/dL |
| Sorbitol | 0.09 mg/dL |
| Lactose | 25 mg/dL |
| Tetracycline | 1.5 mg/dL |
The sponsor lists the following limitations in the User Manual and the Test Strip Insert:
- If you are taking vitamin C (ascorbic acid in your blood >3 mg/dL) then your glucose results using this meter may not be reliable.
- The VivaChek Ino Blood Glucose Monitoring System should not be used following xylose absorption procedures.
f. Assay cut-off:
Not applicable
2. Comparison studies:
{9}
a. Method comparison with predicate device:
Not applicable. See section M.3.c. below for performance in the hands of the intended user.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
User Performance Study:
To assess the performance of the VivaChek Ino Blood Glucose Monitoring Systems in the hands of the intended users, the sponsor performed four independent studies that each included a minimum of 350 lay users with the VivaChek Ino Smart (VGM04), VivaChek Ino Sound (VGM09), VivaChek Ino Plus (VGM22), and VivaChek Ino Sound Simple (VGM26) Blood Glucose Meters (see Section I Device Description for a description of the minor differences between each meter). Each study included lay users that were provided with the device labeling, collected their own fingers:ick sample, and obtained results on a VivaChek Ino meter. After the tests were completed by the lay user, additional blood samples were collected from each subject by a professional for testing with the comparator method, YSI 2300. There were no significant differences demonstrated in performance between the four lay user studies completed. As representative performance, the results from the most complex system (e.g. presence of talking feature, bluetooth function, backlight, strip ejector etc.), the VivaChek Ino Sound Blood Glucose Monitoring System (VGM09), are summarized in the tables below:
VivaChek Ino Sound Blood Glucose Monitoring System (VGM09)
| For glucose concentrations <75 mg/dL | | | |
| --- | --- | --- | --- |
| Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL | Within ± 20 mg/dL |
| 49/65 (75.4%) | 64/65 (98.5%) | 64/65 (98.5%) | 65/65 (100%) |
{10}
11
| For glucose concentrations >75 mg/dL | | | |
| --- | --- | --- | --- |
| Within ± 5 % | Within ± 10 % | Within ± 15 % | Within ± 20 % |
| 209/290 (72.1.0%) | 281/290 (96.9%) | 290/290 (100%) | 290/290 (100%) |
| All glucose concentrations | | | |
| --- | --- | --- | --- |
| Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% |
| 243/355 (68.5%) | 337/355 (94.9%) | 354/355 (99.7%) | 355/355 (100%) |
Results of linear regression analysis:
$$
y = 1.0031x - 0.7024; R^2 = 0.9950
$$
## Accuracy at Extreme Glucose Study:
To assess the performance of the VivaChek Ino Blood Glucose Monitoring Systems at extreme glucose concentrations, a study was conducted using blood from 105 study participants where blood was either aged to lower glucose concentrations or spiked to raise glucose concentration. The results were compared to the YSI 2300 comparator method. Results demonstrate that the meter has acceptable accuracy at extreme glucose values.
For glucose concentrations < 80 mg/dL
| Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| --- | --- | --- |
| 48/53 (90.6%) | 53/53 (100%) | 53/53 (100%) |
For glucose concentrations >250 mg/dL
| Within ± 5 % | Within ± 10 % | Within ± 15 % | Within ± 20 % |
| --- | --- | --- | --- |
| 35/52 (67.3%) | 52/52 (100%) | 52/52 (100%) | 52/52 (100%) |
## Usability/Readability:
The study participants were asked to complete a usability questionnaire regarding ease of understanding of information in the user manual and the ease of use when performing a blood glucose test. The study demonstrated that users of the device were able to understand and follow the instructions provided in the labeling to perform tasks involved in blood glucose testing.
{11}
Flesch-Kincaid readability assessment was conducted on all labeling and demonstrated that the user manuals are written at an 8th grade level or below.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
The sponsor includes the following expected glucose values for people without diabetes in the device labeling.
| Time of day | Glucose Range |
| --- | --- |
| Fasting and before meals | <100 mg/dL |
| 2 hours after meals | <140 mg/dL |
Source: American Diabetes Association (Standards of Medical Care in Diabetes – 2018. Diabetes Care, January 2018, vol. 41, Supplement 1, S13-S27).
N. Instrument Name:
VivaChek Ino Smart Blood Glucose Meter (VGM04)
VivaChek Ino Sync Blood Glucose Meter (VGM05)
VivaChek Ino Sound Blood Glucose Meter (VGM09)
VivaChek Ino Plus Blood Glucose Meter (VGM22)
VivaChek Ino Sound Simple Blood Glucose Meter (VGM26)
VivaChek Ino Sound Bright Blood Glucose Meter (VGM27)
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X ☐ or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ X ☐ or No ☐
The VivaChek Ino Smart, VivaChek Ino Sync, VivaChek Ino Sound, and VivaChek Ino Plus Blood Glucose Monitoring System have the ability to transmit data to a mobile device via Bluetooth. The VivaChek Ino Sound Simple and VivaChek Ino Sound Bright Blood Glucose Monitoring Systems do not have the ability to transmit data to a
12
{12}
computer, webserver, or mobile device.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
The system is intended to be used with capillary whole blood from the finger. The whole blood sample is applied directly to the test strip by capillary action.
5. Calibration:
Calibration is automatic. There is no user input for coding.
6. Quality Control:
VivaChek Control Solutions are aqueous solutions containing glucose and are available at three levels (level 1, level 2 and level 3). Instructions on how to order the control solutions are included in the user manual. The control solution readings are not included in the average of the patient results. An acceptable range for each control level is printed on the test strip vial label. The user is cautioned not to use the meter if the control result falls outside these ranges.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. Hematocrit Study:
To evaluate the effect of hematocrit on the VivaChek Ino Blood Glucose Monitoring System, venous blood samples with hematocrit levels of 20%, 25%, 30%, 35%, 40%, 45%, 50%, 55%, 60%, 65%, and 70% were tested at 5 glucose concentrations (50, 90, 130, 200, and 330 mg/dL). Each sample was tested in replicates of 10 test strips per strip lot, with a total of 3 lots of testes strips tested. Results from the meter were compared to results obtained using a laboratory-based comparator method (YSI 2300). The evaluation of bias relative to the comparator method demonstrated acceptable performance to support the claimed hematocrit range of 20-70%.
13
{13}
14
2. **Altitude Study:**
To evaluate the effect of altitude on the VivaChek Ino Blood Glucose Monitoring System, meters were tested at sea level and 13,123 feet. The evaluation included 3 test strip lots and the glucose range tested was 52 to 538 mg/dL. Results at each altitude were compared to the comparator method, YSI 2300. Results demonstrated that altitudes up to 13,123 feet above sea level have no significant effect on blood glucose measurements.
3. **Sample Volume Study:**
The sponsor performed a study to support the claimed minimum sample volume for the VivaChek Ino Blood Glucose Monitoring System. Venous whole blood samples were tested at 0.6, 0.7, and 0.8 µL at thee glucose levels (60, 110, and 230 mg/dL) as established by the YSI 2300 comparator method. Each sample was tested in duplicate on two meters using 3 test strip lots. Results support the claimed minimum sample volume of 0.8 µL. The sponsor provided validation studies demonstrating that with blood volumes below 0.8 µL, the insufficient sample volume error message functioned as intended.
4. **Operating Conditions Study:**
The sponsor performed temperature and humidity studies using venous blood samples with three glucose levels (60-70, 100-120, and 300-400 mg/dL). Temperatures ranging from 41°F - 113°F (5-45°C) and relative humidity from 10% to 90% were tested. Meter results were compared to the YSI 2300 comparator analyzer. Four temperature and humidity combinations were tested including low temperature/low humidity, low temperature/high humidity, high temperature/low humidity and high temperature/high humidity. The results support the claims in the labeling that the system can be used in conditions of 41°F - 113°F (5-45°C) with relative humidity of 10 to 90%.
5. **Flex Studies:**
Intermittent sampling, sample perturbation, testing with used test strips, drop/shock testing, and vibration testing was completed. The testing performed demonstrated that the VivaChek Ino Blood Glucose Monitoring Systems are robust to intermittent sampling, sample perturbation, drop/shock, and vibration, and that an error message is returned to the user if a used test strip is inserted into the meter.
6. **EMC Testing:**
The sponsor provided appropriate documentation certifying that acceptable electromagnetic testing (EMC) had been performed for each device in this 510(k) submission and that they were compliant.
7. **Infection Control Studies:**
The six VivaChek Ino systems in this submission are intended for single-patient use only. Disinfection efficacy studies were performed on the materials comprising the meters by an outside commercial testing laboratory demonstrating complete inactivation of hepatitis B virus (HBV) with the chosen disinfectant, Clorox Germicidal Wipes, 0.55% sodium hypochlorite (EPA Registration # 67619-12). Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or external materials of the meter after 608 cleanings and disinfection cycles with the Clorox Germicidal
{14}
Wipes. The robustness studies were designed to simulate 5 years of single-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
8. Test Strip Lot Release Protocol:
The test strip lot release protocol and acceptance criteria were reviewed and found to be acceptable.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.