PEEK Biosolution is intended to be used for the fabrication of permanent dental restorations using CAD/CAM techniques. These are: - Fully anatomical crowns and bridges (max. 2 pontics and min. 13 mm2 connector cross-section) - Crown copings and bridge substructures for composite veneering (max. 2 pontics and min. 13 mm2 connector cross-section)) - Telescopic primary and secondary crowns and frameworks - Secondary bar structures on primary bars made of titanium alloy, CoCr alloys, gold alloy, zirconium dioxide.
Device Story
PEEK Biosolution is a polyetheretherketone (PEEK) milling disc used by dental laboratories to fabricate permanent dental restorations via CAD/CAM systems. The device serves as raw material; dental technicians use CAD/CAM software and milling equipment to shape the disc into specific dental prosthetics. The resulting restorations are used for crowns, bridges, telescopic frameworks, and secondary bar structures. The device provides a biocompatible, high-strength alternative to traditional dental materials, benefiting patients by enabling durable, custom-fitted restorations.
Clinical Evidence
Bench testing only. No clinical data. Testing included melting temperature (ISO 11357), water sorption (ISO 62), bond strength (ISO 10477), density (ISO 1183), modulus of elasticity (ISO 20795-1), and flexural strength (ISO 10477). Biocompatibility testing performed per ISO 10993-1 and USP Class VI.
Technological Characteristics
Material: Polyetheretherketone (PEEK) plus pigments. Form: Solid disc with circumferential ridge. Dimensions: 95 mm diameter; 12-30 mm thickness. Fabrication: CAD/CAM milling. Standards: ISO 11357, ISO 62, ISO 10477, ISO 1183, ISO 20795-1, ISO 10993-1, USP Class VI.
Indications for Use
Indicated for patients requiring permanent dental restorations, including fully anatomical crowns and bridges, crown copings, bridge substructures for veneering, telescopic crowns/frameworks, and secondary bar structures.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
December 18, 2018
Merz Dental GmbH % Richard Hunter Principal Washington Regulatory Consultants, LLC 5616 Mariola PI NE Albuquerque, Minnesota 87111
Re: K173124
Trade/Device Name: PEEK Biosolution Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: September 28, 2017 Received: September 29, 2017
Dear Richard Hunter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# 4. Indications for Use
# 510(k) Number (if known): K173124
Device Name: PEEK Biosolution
#### Indications for Use:
PEEK Biosolution is intended to be used for the fabrication of permanent dental restorations using CAD/CAM techniques. These are:
- Fully anatomical crowns and bridges (max. 2 pontics and min. 13 mm2 connector ● cross-section)
- Crown copings and bridge substructures for composite veneering (max. 2 pontics and min. 13 mm2 connector cross-section))
- . Telescopic primary and secondary crowns and frameworks
- . Secondary bar structures on primary bars made of titanium alloy, CoCr alloys, gold alloy, zirconium dioxide.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use
(21 CFR Part 801 Subpart D)
(21 CFR Part 807 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
{3}------------------------------------------------
| 510(k) Summary | K173124 | |
|--------------------|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| PEEK Biosolution | | |
| Date of Summary: | September 28, 2017 | |
| Submitter: | Company Name:<br>Address: | Merz Dental GmbH<br>Eetzweg 20<br>Lutjenburg, Germany D-24321 |
| | Contact Name:<br>Telephone No.<br>Fax No.<br>e-mail: | Dr. Med. Claudia Bobrowski<br>(011) 43 81/4 03-4 11<br>(011) 43 81/4 09-1 07<br>Claudia.bobrowski@merz-dental.de |
| Device Name: | Proprietary name:<br>Common name:<br>Classification reg: | PEEK Biosolution<br>Dental material<br>21CFR 872.3690<br>Classification name: Material, Tooth Shade, Resin |
| Predicate Devices: | breCAM.BioHPP, K152113 | |
{4}------------------------------------------------
# Intended Use:
PEEK Biosolution is intended to be used for the fabrication of permanent dental restorations using CAD/CAM techniques. These are:
- . Fully anatomical crowns and bridges (max. 2 pontics and min. 13 mm² connector cross-section)
- . Crown copings and bridge substructures for composite veneering (max. 2 pontics and min.13 mm² connector cross-section))
- Telescopic primary and secondary crowns and frameworks
- . Secondary bar structures on primary bars made of titanium alloy, CoCr alloys, gold alloy, zirconium dioxide.
### Technological characteristics and substantial equivalence:
PEEK Biosolution is substantially equivalent to breCAM.BioHPP (K152113). See table below for a comparison of the two devices. Both devices are solid discs composed of polyether ether ketone (PEEK) and both are indicated for the fabrication of permanent dental restorations using CAD/CAM techniques and equipment. They only differ in dimensions.
| Device | PEEK Biosolution | breCAM.BioHPP |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Indications | Milling blank for fabrication of permanent dental restorations using CAD/CAM techniques. | Same |
| Intended Use | •Fully anatomical crowns and bridges<br>•Crown copings and bridge substructures for composite veneering<br>•Telescopic primary and secondary crowns and frameworks<br>•Secondary bar structures on primary bars made of titanium alloy, CoCr alloys, gold alloy, zirconium dioxide. | Same |
| Material Class | Polyetheretherketone (PEEK) | Same |
| Material Form | Solid Disc with circumferential ridge | Same |
| Material Comp | PEEK plus pigments | Same |
| Disc Sizes | | |
| Diameter | 95 mm | 84, 95 and 98.5 mm |
| Thickness | 12, 16, 20, 24, 26, 30 mm | 12 to 25 mm |
| Disc Shades | White and A2/B2 | Standard and Dentin |
| Fabrication | Milling | Same |
| Milling<br>Equipment | imes-icor, Wieland Dental, Organical, Zirkonzahn or equivalent | Same |
| CAD/CAM<br>Software | Software and milling equipment operated by independent dental laboratories | Same |
| Clearance | Subject of Application | K152113 |
| Manufacturer | Merz Dental, GmbH | Bredent GmbH & Co.KG |
Comparison of PEEK Biosolution to breCAM.BioHPP
{5}------------------------------------------------
PEEK Biosolution was tested and found to be in compliance with applicable dental materials standards. The results and a comparison to the predicate device, breCAM.BioHPP, are presented below.
| Testing<br>(passing value) | PEEK<br>BioSolution | breCAM BioHPP | PEEK<br>Biosolution<br>Standard |
|----------------------------------------------------------------------|--------------------------------------|-----------------|---------------------------------|
| Material | PEEK | PEEK | NA |
| Shape | Disc | Disc | NA |
| Melting temperature | 339 °C | appr. 340 °C | ISO 11357 |
| Water sorption | < 7.6 µg / mm3 | 6.5 µg / mm3 | ISO 62 |
| Bond strength to veneering resins<br>after thermo cycling, (≥ 15MPa) | 22 Mpa | 21.4 MPa | ISO 10477 |
| Density | 1.49 ± 0.03 | 1.4425 g/cm3 | ISO 1183 |
| Modulus of elasticity | 5100 ± 210 MPa | 4550 - 4620 MPa | ISO 20795-1 |
| Flexural strength, (>150 MPa) | 178 ± 4 MPa | 174 - 182 MPa | ISO 10477 |
| Flexural strength<br>after thermocycling, (>150MPa) | 171 ± 5 MPa | 173 - 174 MPa | ISO 10477 |
| Elongation at fracture, (no fracture) | no fracture | no fracture | ISO 10477 |
| Fracture load 3-unit bridge, > 800 N | > 910 N (limit of<br>testing device) | 1307 N | ISO 10477 |
| Elongation at fracture, load > 7%,<br>no fracture | no fracture | no fracture | ISO 10477 |
Physical Properties of PEEK Biosolution and breCAM.BioHPP
PEEK raw material (granules) used in the production of rod stock was tested by the manufacturer according to ISO 10993-1 and USP Class VI and found to be biocompatible. PEEK Biosolution, which is manufactured from the rod stock, was tested and found to not release extractable inorganic substances, and was not cytotoxic. All tests were performed according to GLP requirements.
# Summary:
Based on the above description of the Merz Dental GmbH PEEK Biosolution and the predicate device. Merz Dental considers PEEK Biosolution to be substantially equivalent to the cited predicate device, and safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.