The KJ ZIRCONIA Implant System is intended to use in the treatment of missing teeth to support prosthetic device, such as artificial teeth, in order to restore mastication in partially edentulous patients. It is for delayed loading.
Device Story
KJ ZIRCONIA Implant System is a one-piece endosseous dental implant; manufactured from zirconia; surgically placed into upper or lower jaw bone to support prosthetic devices. Device features threaded body; available in two types (BL and BS) differing by abutment length. Intended for delayed loading in partially edentulous patients. Used by dental clinicians in clinical settings. Provides structural support for artificial teeth; restores mastication function. Benefits patient by replacing missing teeth with biocompatible ceramic material.
Clinical Evidence
No clinical data. Evidence is based on bench testing, including sterilization validation (ISO 11137, EN 552), real-time shelf-life testing (3 years), packaging integrity (ASTM F88, ISO 11607-1), microbial challenge (ISO 11737-2), biocompatibility (ISO 10993-3, ISO 10993-6), and endotoxin testing (USP 39 <85>).
Indicated for partially edentulous patients requiring treatment of missing teeth to support prosthetic devices (e.g., artificial teeth) for restoration of mastication. For delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure. The FDA part of the logo has the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 28, 2018
KJ Meditech Co., Ltd % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 690 Roosevelt Irvine, California 92620
Re: K173116
Trade/Device Name: KJ ZIRCONIA Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: May 24, 2018 Received: May 30, 2018
#### Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) K173116
Device Name KJ ZIRCONIA Implant System
Indications for Use (Describe)
The KJ ZIRCONIA Implant System is intended to use in the treatment of missing teeth to support prosthetic device, such as artificial teeth, in order to restore mastication in partially edentulous patients. It is for delayed loading.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
## 510(k) Summary (K173116)
#### Date: 06/28/2018
#### 1. Submitter
KJ Meditech Co., Ltd. 21, Cheomdan Venture-ro 40 Beon-gil, Buk-gu Gwangju, Jeonranamdo, Republic of Korea 500-470
#### 2. U.S Agent/Contact Person
Priscilla Chung / Regulatory Affairs Consultant LK Consulting Group USA, Inc. 690 Roosevelt, Irvine CA 92620 Phone: 714.202.5789Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
#### 3. Device
- . Trade Name: KJ ZIRCONIA Implant System
- Common Name: Dental Implant System
- Classification Name: Endosseous Dental Implant
- Product Code: DZE ●
- Classification regulation: 21CFR872.3640 ●
#### 4. Predicate Device
- Predicate Device: ● Nobel Biocare (K061971) by Novel Biocare Co., Ltd.
- Reference Device: PURE Ceramic Implant (K151328) by Staumann USA
#### 5. Description
KJ ZIRCONIA Implant System is a dental implant system made of ZIRCONIA intended to be surgically placed in the bone of the upper or lower jaw arches. It is one piece type implant which the implant and the abutment are manufactured as one piece. It offers two types: BL type and BS type, and the difference is only in the length of the abutment part.
{4}------------------------------------------------
## 6. Indication for Use
The KJ ZIRCONIA Implant System is intended to use in the treatment of missing teeth to support prosthetic device, such as artificial teeth, in order to restore mastication in partially edentulous patients. It is for delayed loading.
#### 7. Basis for Substantial Equivalence
The indications for use statement of the subject device does not use exactly the same words as the predicate and the reference devices; however, the statements have the same indications for use which is treatment of missing teeth.
The subject device is substantially equivalent to the predicate and the reference devices in terms of design and material. The size range of the predicate device covers that of the subject device.
| | Subject Device | Predicate Device | Reference Device |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | KJ ZIRCONIA Implant<br>System | Nobel biocare | PURE Ceramic implant |
| 510k # | K173116 | K061971 | K151328 |
| Applicant | KJ Meditech Co., Ltd. | Novel biocare<br>Co., Ltd. | Staumann USA |
| Indications for Use | The KJ ZIRCONIA<br>Implant System is<br>intended to use in the<br>treatment of missing teeth<br>to support prosthetic<br>device, such as artificial<br>teeth, in order to restore<br>mastication in partially<br>edentulous patients.<br>It is for delayed loading. | Nobel Biocare's Zirconia<br>Implants are root-form<br>endosseous dental<br>implants intended to<br>be surgically placed in<br>the bone of the upper or<br>lower jaw arches to<br>provide support for<br>prosthetic devices, such<br>as an artificial tooth, in<br>order to restore patient<br>esthetics and<br>chewing function. Nobel<br>Biocare's Zirconia<br>Implants are indicated for<br>single or multiple<br>unit restorations in<br>splinted or non-splinted<br>applications. Nobel<br>Biocare Zirconia<br>Implants<br>may be placed<br>immediately and put into<br>immediate function<br>providing that the initial<br>stability requirements<br>detailed in the surgical<br>manual are satisfied. | The Straumann® PURE<br>Ceramic Implant<br>(Monotype) is indicated<br>for restoration in single<br>tooth gaps and in an<br>edentulous or partially<br>edentulous jaw. The<br>prosthetic restorations<br>used are single crowns,<br>fixed partial or full<br>dentures, which are<br>connected to the implants<br>through the<br>corresponding<br>components.<br>The ø3.3 mm reduced<br>diameter implants are<br>recommended for central<br>and lateral incisors only.<br>The Straumann® PURE<br>Ceramic Implant<br>Protective Cap is<br>intended to protect the<br>Straumann® PURE<br>Ceramic Implant<br>(Monotype) during the<br>healing phase after<br>implant placement for up<br>to 6 months.<br>Temporary copings are<br>intended to serve as a<br>base for temporary crown<br>or bridge restoration<br>for the Straumann® |
{5}------------------------------------------------
| | | | PURE Ceramic Implant<br>(Monotype) for up to 30 |
|--------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------|
| | | | days. |
| Implant Design | • Implant Type: One<br>Piece Type<br>· Body Design:<br>Threaded body | • Implant Type: One<br>Piece Type<br>· Body Design:<br>Threaded body | • Implant Type: One<br>Piece Type<br>· Body Design:<br>Threaded body |
| | | | |
| Implant/Abutment<br>Connection | None (Monotype) | None (Monotype) | None (Monotype) |
| Allowable Angulation | Angulation not allowed | Angulation not allowed | Angulation not allowed |
| Appearance | | | |
| Material | Zirconia (ISO 13356) | Zirconia | Zirconia |
| Surface Treatment | N/A | N/A | N/A |
| Implant Sterile | Yes | Yes | Yes |
| Sterilization Method | Gamma | Gamma | Gamma |
| Implant Diameters | 3.5, 4.0, 4.5, 5.0 | 3.5, 4.3 , 5.0 | 3.3, 4.1 |
| Implant Lengths | 8, 9, 10, 11, 12, 13, 14,<br>15,16mm | 10, 13, 16mm | 8,10,12,14mm |
| Implant Neck Dimensions | 2.6, 2.8, 3.0 | 3.5, 4.3, 5.0 | 3.5 , 48 |
| Thread Pitch | 1.0mm | | |
| Product Code | DZE | DZE | DZE |
## 8. Non-Clinical Testing
- Sterilization validating testing in accordance with ISO11137 VDmax25, EN552, and ISO 11137
- Real time shelf tests to validate the sterility of the device through the proposed shelf . life have been conducted for three (3) years. The testing includes visual, label, seal inspections (ASTM F88), dye penetration (ISO 11607-1) and microbial challenge (ISO11737-2).
- Biocompatibility tests in accordance with ISO 10993-3 and ISO 10993-6 ●
- Endotoxin Testing in accordance with USP 39 <85> on the endotoxin limit which is ● 0.5 EU/mL
The tests above support that the differences which the subject device might have when comparing to the predicate devices such as in sterilization parameters, packaging configuration, and manufacturing processes do not raise a question in substantial equivalence.
#### 9. Conclusion
Based on the information provided, the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.