← Product Code [PND](/productcode/PND) · K173114

# primeMidline Catheters (K173114)

_Pfm Medical, Inc. · PND · Mar 7, 2018 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173114

## Device Facts

- **Applicant:** Pfm Medical, Inc.
- **Product Code:** [PND](/productcode/PND.md)
- **Decision Date:** Mar 7, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The primeMIDLINE™ Catheters are indicated for short term access to the peripheral venous system for selected intravenous therapies, blood sampling and power injection of contrast media. These catheters are indicated for patients older than 28 days and that weigh more than 10kg, with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The primeMIDLINE™ Catheters are suitable for use with power injectors.

## Device Story

Peripherally placed midline catheters; provide short-term (<30 days) vascular access. Radiopaque polyurethane construction; open-ended lumen; 3F, 4F, 5F single lumen and 5F dual lumen configurations. Inserted percutaneously via Modified Seldinger Technique with pre-inserted stiffening stylet. Used in clinical settings for intravenous therapy, blood sampling, and power injection of contrast media. Healthcare providers (physicians/nurses) operate device. Output is vascular access for fluid/contrast delivery or blood withdrawal. Benefits include reliable peripheral access for patients meeting weight/age criteria.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included visual inspection, tensile strength, radiopacity, catheter collapse, leak, priming volume, flow rates (gravity and power injection), and biocompatibility (ISO 10993-1). All tests met predetermined acceptance criteria.

## Technological Characteristics

Radiopaque polyurethane shaft, junction, and extension leg; acetal resin clamp. 3F, 4F, 5F single lumen and 5F dual lumen configurations (20cm length). Open-ended distal tip. Luer connection. Sterilized via 100% Ethylene Oxide (SAL 10^-6). Complies with ISO 10555-1, ASTM F640-07, ISO 594-1/2, and AAMI/ANSI/ISO 11135.

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- PowerMidline Catheter ([K153393](/device/K153393.md))

## Reference Devices

- 4F Single Lumen, 5F Single Lumen and 5F Dual Lumen primePICC Catheters ([K083873](/device/K083873.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 7, 2018

PFM Medical, Inc. Jessica Jho Regulatory Affairs Consultant 1916 Palomar Oaks Way Suite 150 Carlsbad. California 92008

Re: K173114

Trade/Device Name: primeMidline™ Catheters Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: PND Dated: February 1. 2018 Received: February 5, 2018

Dear Jessica Jho:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tina
Kiang -S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K173114

Device Name

primeMidline™ Catheters

#### Indications for Use (Describe)

The primeMIDLINE™ Catheters are indicated for short term access to the peripheral venous system for selected intravenous therapies, blood sampling and power injection of contrast media. These catheters are indicated for patients older than 28 days and that weigh more than 10kg, with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The primeMIDLINE™ Catheters are suitable for use with power injectors.

| Type of Use (Select one or both, as applicable)  |                                             |
|--------------------------------------------------|---------------------------------------------|
| X   Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for pfmmedical. The logo is composed of the letters "pfm" in blue and the word "medical" in black. The letters "pfm" are in a sans-serif font and are connected to each other. The word "medical" is in a sans-serif font and is separated from the letters "pfm".

# 510(k) Summary K173114 As required by section 807.92

| General Information                                                                        |                                                                                                                                     |
|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner:<br>Submitter Address:                                                        | PFM Medical, Inc.<br>1916 Palomar Oaks Way, Suite 150<br>Carlsbad, CA 92008                                                         |
| Contact Person:<br>Telephone Number:<br>Fax Number:<br>Date of Preparation:                | Jessica Jho, RAC<br>760.758.8749<br>760.758.1167<br>March 6, 2018                                                                   |
| Subject Device Information                                                                 |                                                                                                                                     |
| Trade Name:<br>Common Name:<br>Classification Name:<br>Product Code:<br>Regulation Number: | primeMIDLINE™ Catheters<br>Midline Catheter<br>Intravascular Catheter<br>PND<br>21 CFR §880.5200                                    |
| Predicate Device Information                                                               |                                                                                                                                     |
| Trade Name:<br>Premarket Notification<br>Classification<br>Manufacturer                    | PowerMidline Catheter<br>K153393<br>PND, 21 CFR §880.5200<br>Bard Access Systems                                                    |
| Reference Device Information                                                               |                                                                                                                                     |
| Trade Name:<br>Premarket Notification<br>Classification:<br>Manufacturer:                  | 4F Single Lumen, 5F Single Lumen and 5F Dual Lumen<br>primePICC Catheters<br>K083873<br>L JS, 21 CFR §880.5970<br>PFM Medical, Inc. |

{4}------------------------------------------------

Image /page/4/Picture/9 description: The image shows the logo for PFM Medical. The logo consists of the letters "pfm" in blue, followed by the word "medical" in black. The letters "pfm" are in a sans-serif font, and the word "medical" is in a similar font. The logo is simple and modern, and it is likely used to identify PFM Medical as a company.

#### Device Description

The primeMIDLINE™ Catheters are a family of peripherally placed midline catheters designed to provide access to the vascular system. These catheters are offered in 3F, 4F and 5F Single Lumen and 5F Dual Lumen configurations with 20cm lengths. The catheter lumen is an open-ended design comprised of radiopaque polyurethane. Each primeMIDLINE™ Catheter is packaged sterile in a tray with the accessories necessary to establish short-term (less than 30 days) vascular access. The primeMIDLINE™ Catheters are suitable for use with power injectors.

#### Indications for Use

The primeMIDLINE™ Catheters are indicated for short term access to the peripheral venous system for selected intravenous therapies, blood sampling and power injection of contrast media. These catheters are indicated for patients older than 28 days and that weigh more than 10kg, with consideration given to adequacy of vascular anatomy and appropriateness of the primeMIDLINE™ Catheters are suitable for use with power injectors.

#### Technological Characteristics

The proposed primeMIDLINE™ Catheters are substantially equivalent to the Power Midline Catheters, previously reviewed and cleared under K153393. When compared to the proposed primeMIDLINE™ Catheters have equivalent materials, design, components, technological characteristics and Indications for Use statements. The minor differences between the subject primeMIDLINE™ catheters and the predicate device do not change the intended use of the device and they do not raise different questions of safety and effectiveness.

|                                         | Subject Device:<br>primeMIDLINE™ Catheters                           | Predicate Device:<br>PowerMidline Catheters                                                                                                                                              | Reference Device:<br>4F SL, 5F SL, 5F DL primePICC<br>Catheters                                                                                                                     |
|-----------------------------------------|----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                            | PFM Medical, Inc.                                                    | Bard Access Systems                                                                                                                                                                      | PFM Medical, Inc.                                                                                                                                                                   |
| 510(k) Number                           | K173114                                                              | K153393                                                                                                                                                                                  | K083873                                                                                                                                                                             |
| FDA Product<br>Code                     | Identical to Predicate<br>Devices                                    | PND                                                                                                                                                                                      | L JS                                                                                                                                                                                |
| Intended Use                            | Identical to Predicate/<br>Reference Devices                         | The PowerMidline catheter<br>is intended to access the<br>vascular system for<br>administration of fluids<br>intravenously, blood<br>sampling, and power<br>injection of contrast media. | The pimePICC catheters are<br>intended to access the vascular<br>system for administration of<br>fluids intravenously, blood<br>sampling, and power injection<br>of contrast media. |
| Duration of Use                         | Identical to predicate<br>PowerMidline Catheter                      | Short term (<30days)                                                                                                                                                                     | Short term (<30days) or long<br>term (>30 days)                                                                                                                                     |
| Insertion<br>Method                     | Identical to Predicate/<br>Reference Devices                         | Percutaneous, using<br>Modified Seldinger<br>Technique and guidewire                                                                                                                     | Percutaneous, using Modified<br>Seldinger Technique and<br>guidewire                                                                                                                |
|                                         | Subject Device:<br>primeMIDLINE™ Catheters                           | Predicate Device:<br>PowerMidline Catheters                                                                                                                                              | Reference Device:<br>4F SL, 5F SL, 5F DL primePICC<br>Catheters                                                                                                                     |
| Tip Placement<br>Location               | Identical to predicate<br>PowerMidline Catheter                      | Peripheral venous system,<br>with catheter tip<br>terminating prior to the<br>axilla                                                                                                     | Central venous system, lower<br>1/3 of the Superior Vena Cava<br>preferred                                                                                                          |
| Insertion Site                          | Identical to Predicate/<br>Reference Devices                         | Peripheral                                                                                                                                                                               | Peripheral                                                                                                                                                                          |
|                                         |                                                                      | Shaft Tubing<br>Polyurethane                                                                                                                                                             | Shaft Tubing<br>Polyurethane                                                                                                                                                        |
|                                         |                                                                      | Catheter Junction<br>Polyurethane                                                                                                                                                        | Catheter Junction<br>Polyurethane                                                                                                                                                   |
| Catheter Base<br>Materials              | Identical to predicate<br>PowerMidline Catheter                      | Extension Leg<br>Polyurethane                                                                                                                                                            | Extension Leg<br>Polyurethane                                                                                                                                                       |
|                                         |                                                                      | Luer Hub<br>Polyurethane                                                                                                                                                                 | Luer Hub<br>Polycarbonate                                                                                                                                                           |
|                                         |                                                                      | Extension Leg Clamp<br>Acetal Resin                                                                                                                                                      | Extension Leg Clamp<br>Acetal Resin                                                                                                                                                 |
| Catheter<br>Proximal<br>Configuration   | Identical to Predicate/<br>Reference Devices                         | Luer Connection                                                                                                                                                                          | Luer Connection                                                                                                                                                                     |
| Catheter Distal<br>Configuration        | Identical to Predicate/<br>Reference Devices                         | Open Ended                                                                                                                                                                               | Open Ended                                                                                                                                                                          |
|                                         | 3F Single Lumen x 20cm                                               | 3F SL x 20cm usable length                                                                                                                                                               | 4F Single Lumen x 50cm length                                                                                                                                                       |
| Catheter<br>Dimensions                  | length<br>4F Single Lumen x 20cm<br>length                           | 4F SL x 20cm useable length                                                                                                                                                              | 5F Single Lumen x 55cm length                                                                                                                                                       |
|                                         | 5F Single Lumen x 20cm<br>length<br>5F Dual Lumen x<br>20cm length   |                                                                                                                                                                                          | 5F Dual Lumen x 55cm length                                                                                                                                                         |
| Depth Markings                          | Identical to predicate/<br>reference devices                         | Number every 5 cm and<br>depth mark every cm                                                                                                                                             | Number every 5 cm and depth<br>mark every cm                                                                                                                                        |
| Stylet<br>Configuration                 | Identical to predicate/<br>reference devices                         | Pre-Inserted<br>Stiffening Stylet                                                                                                                                                        | Pre-Inserted<br>Stiffening Stylet                                                                                                                                                   |
| Power Injection<br>Maximum Flow<br>Rate | 3F SL: 5mL/sec<br>4F SL: 7mL/sec<br>5F SL: 7mL/sec<br>5F DL: 7mL/sec | 3F SL = 3 mL/sec<br>4F SL = 7 mL/sec                                                                                                                                                     | 5 mL/sec                                                                                                                                                                            |
| Sterility                               | Identical to Predicate/<br>Reference Devices                         | 100% Ethylene Oxide, SAL<br>of 10-6                                                                                                                                                      | 100% Ethylene Oxide, SAL of<br>10-6                                                                                                                                                 |

The following is a summary of the technological characteristics of the primeMIDLINE™ catheters as compared to the predicate and reference devices:

{5}------------------------------------------------

# pfmmedical

{6}------------------------------------------------

### Performance Testing

PFM Medical, Inc. conducted a risk analysis per ISO 14971: Medical devices – Application of risk management of medical devices to assess the risk profile of the primeMIDLINE™ catheter. Control mechanisms, including design verification testing, were defined to mitigate the identified risks and to demonstrate that the devices perform as intended. Below is a listing of non-clinical testing that is included in this submission:

| Visual Inspection           | Tensile Strength          |
|-----------------------------|---------------------------|
| Dimensional Characteristics | Catheter Elongation       |
| Radiopacity                 | Lifecycle Performance     |
| Catheter Collapse           | Static Burst Pressure     |
| Leak                        | Suture Wing Integrity     |
| Priming Volume              | Luer Testing              |
| Flowrate: Gravity           | Flowrate: Power Injection |

The following standards and FDA guidance documents were utilized in the testing listed above:

- ISO 10555-1:2013, Intravascular Catheters Sterile and Single-Use Catheters ■
- ASTM F640-07 – Standard Test Methods for Determining Radiopacity for Medical Use
- ISO 594-1: 1986, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment- Part 1: General Requirements
- . ISO 594-2: 1998, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment- Part 2: Lock Fittings
- FDA Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters
- PFM Medical Internal Standards

The primeMIDLINE™ catheters were subjected to the following biocompatibility testing as required per ISO 10993-1: Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process:

| Cytotoxicity                               | Sensitization                |
|--------------------------------------------|------------------------------|
| Irritation or Intracutaneous<br>Reactivity | Acute Systemic Toxicity      |
| Material Mediated<br>Pyrogenicity          | Subacute/Subchronic Toxicity |
| Genotoxicity                               | Carcinogenicity              |
| Chronic Toxicity                           | Hemocompatibility            |
| Implantation                               |                              |

{7}------------------------------------------------

### Performance Testing

The primeMIDLINE™ Catheters are sterilized by 100% Ethylene Oxide Gas with a Sterility Assurance level of 10 °. The following sterilization is applicable to the primeMIDLINE™ Catheters:

| Sterilization Validation –<br>Overkill Method | EO/ECH Residuals                                   |
|-----------------------------------------------|----------------------------------------------------|
| Direct Product Sterility                      | Bioburden / Organism<br>Characterization           |
| Bacteriostasis Fungistasis<br>Testing         | Bacterial Endotoxins / Limulus<br>Amebocyte Lysate |

The sterilization testing listed above was conducted in accordance with the following standards:

- l AAMI/ANSI/ISO 11135:2014, Sterilization of health-care products - Ethylene oxide -Requirements for the development, validation and routine control of a sterilization process for medical devices
- . ANSI/AAMI/ISO 11737-1: 2006/(R)2011, Sterilization of health care products - Microbiological methods - Part 1: Determination of the population of microorganisms on product
- ANSI/AAMI/ISO 11737-2:2009/(R)2014,Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
- AAMI/ANSI/ISO 10993-7:2008, Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals
- ANSI/AAMI ST72:2011/(R)2016, Bacterial Endotoxin-Test methods, routine monitoring, and alternatives to testing

The subject devices met all predetermined acceptance criteria as defined in the referenced Standards, FDA Guidance Documents or PFM Medical, Inc. Internal Standards.

## Conclusion

Based on detailed comparison of intended use, technological characteristics and performance testing of the subject and predicate device, the primeMidline Catheters are substantially equivalent to the predicate Power Midline Catheters.

---

**Source:** [https://fda-staging.innolitics.com/device/K173114](https://fda-staging.innolitics.com/device/K173114)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
