The Medline Disposable Electronic Thermometer Probe Covers are intended for use as barriers between any SureTemp and SureTemp Plus digital thermometer probe and users' oral, rectal, or axillary measuring sites to avoid possible contamination and infection during temperature measuring. The probe covers are non-sterile and intended for single use only.
Device Story
Disposable, single-use, non-sterile shell-like barrier; designed to fit over SureTemp and SureTemp Plus digital thermometer probes. Used in clinical settings by healthcare professionals during oral, rectal, or axillary temperature measurements. Acts as a physical barrier to prevent microbiological cross-contamination between patients and the reusable thermometer probe. Does not interfere with heat transfer from patient to the thermistor. Benefits include infection control and patient safety.
Clinical Evidence
Bench testing only. Performance testing conducted in accordance with ASTM E1104-98 (2016). Biocompatibility testing performed per ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Irritation and Delayed-Type Hypersensitivity).
Technological Characteristics
Disposable, single-use, shell-like probe cover. Conforms to ASTM E1104-98 (2016) standards. Non-sterile. Material properties verified via ISO 10993 biocompatibility testing. No electronic components, software, or energy sources.
Indications for Use
Indicated for use as a barrier between SureTemp/SureTemp Plus digital thermometer probes and oral, rectal, or axillary sites to prevent contamination/infection during temperature measurement in patients requiring clinical temperature monitoring.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right is written in a clear, sans-serif font. The FDA logo is a recognizable symbol of the agency's mission to protect and promote public health.
January 17, 2018
Medline Industries, Inc. Dinah Rincones Regulatory Specialist Three Lake Drive Northfield. Illinois 60093
Re: K173113
Trade/Device Name: Medline Disposable Electronic Thermometer Probe Cover Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: December 15, 2017 Received: December 18, 2017
Dear Dinah Rincones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tina Kiang
S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K173113
Device Name
Medline Disposable Electronic Thermometer Probe Cover
Indications for Use (Describe)
The Medline Disposable Electronic Thermometer Probe Covers are intended for use as barriers between any SureTemp and SureTemp Plus digital thermometer probe and users' oral, rectal, or axillary measuring sites to avoid possible contamination and infection during temperature measuring. The probe covers are non-sterile and intended for single use only.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4-001
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, set against a blue square background. Above the word, there is a white star-like symbol with four points. To the right of the blue square, there is a gray vertical bar.
Medline Industries, Inc. Three Lakes Drive Northfield. IL 60093
#### K173113
# 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]
## Submitter / 510(k) Sponsor
Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093 Registration Number: 1417592
## Contact Person
Dinah Rincones Regulatory Specialist Phone: 847-949-2687 Fax: 224-931-1271 Email: DRincones@medline.com
Summary Preparation Date September 28, 2017
Type of 510(k) Submission Traditional
#### Device Name / Classification
Name of Device: Clinical Electronic Thermometer Probe Cover Proprietary Name: Medline Disposable Electronic Thermometer Probe Cover Common Name: Clinical Electronic Thermometer Probe Cover Product Code: FLL Classification Panel: General Hospital Device Class: Class II Regulation #: 21 CFR 880.2910
#### Predicate Device
The Welch Allyn SureTemp Plus Thermometer (K030580) is identified herein as the predicate device of the Medline Disposable Electronic Thermometer Probe Cover.
Please note that the primary predicate 510(k) to support this submission. K030580. included both the thermometer probe cover, as well as the SureTemp Plus thermometer itself. However, only the thermometer probe cover is included within the scope of this premarket notification; Medline does not
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image features the Medline logo, which consists of the word "MEDLINE" in bold, white letters positioned to the left of a stylized white star. The star is composed of four triangular shapes that converge at a central point, creating a dynamic and modern design. The background is a solid, dark blue color, providing a strong contrast that makes the logo stand out. A small gray rectangle is visible in the upper right corner.
edline Industries, Inc hree Lakes Drive
intend to market any other component or accessory of the SureTemp Plus thermometer system, nor the thermometer itself.
#### Device Description
The proposed device is a disposable, single use, non-sterile shell-like device used to cover electronic thermometer probes of temperature taking devices. The Medline Disposable Electronic Thermometer Probe Cover is compatible with the intended use of all SureTemp and SureTemp Plus reusable digital thermometers probes, which require the use of probe covers for oral, axillary, or rectal temperature measurements.
#### Indications for Use
The Medline Disposable Electronic Thermometer Probe Covers are intended for use as barriers between any SureTemp and SureTemp Plus digital thermometer probe and users' oral, rectal, or axillary measuring sites to avoid possible contamination and infection during temperature measuring. The probe covers are non-sterile and intended for single use only.
#### Summary of Technological Characteristics
The Medline Disposable Electronic Thermometer Probe Cover is similar in design, intended use and function to the thermometer probe cover cleared under K030580.
| Device Characteristic | Proposed Device | Predicate Device | Comparison<br>Analysis |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | The Medline Disposable<br>Electronic Thermometer<br>Probe Covers | The Welch Allyn SureTemp<br>Plus Thermometer | N/A |
| 510(k) Reference | TBD | K030580 | N/A |
| Product Owner | Medline Industries, Inc. | Welch Allyn Inc. | N/A |
| Product Code | FLL | FLL | Same |
| Intended Use | The Medline Disposable<br>Electronic Thermometer<br>Probe Covers are intended<br>for use as barriers between<br>any SureTemp and<br>SureTemp Plus digital<br>thermometer probe and users'<br>oral, rectal, or axillary<br>measuring sites to avoid the<br>possible contamination and<br>infection during temperature<br>measuring. The probe covers<br>are non-sterile and intended | Electronic Thermometer:<br>The Welch Allyn SureTemp<br>Plus thermometer enables the<br>health care professional to make<br>an accurate prediction of oral,<br>rectal or axillary temperature.<br>Electronic Thermometer Probe<br>Cover:<br>Probe covers are intended to<br>cover the thermometer probe<br>prior to its use. This cover must<br>not be a significant barrier to<br>the transfer of heat from the | Same*<br>*The predicate's<br>thermometer probe<br>covers are<br>accessories to the<br>thermometer itself,<br>and were included in<br>its 510(k) clearance<br>under K030580.<br>Only the probe<br>covers are included<br>in this 510(k), as<br>Medline does not |
#### TABLE 1: COMPARISON OF PROPOSED AND PREDICATE DEVICES
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image features the logo of Medline, a healthcare company. The logo consists of the word "MEDLINE" in a bold, sans-serif font, positioned to the left of a stylized starburst or cross-like symbol. The background is a solid dark blue color, providing a strong contrast to the white text and symbol.
Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093
| | for single use only. | patient to the probe body (and<br>thermistor). In addition, the<br>disposable nature of the probe<br>cover prevents microbiological<br>cross-contamination among<br>patients which might occur with<br>a reusable probe. | intend to market any<br>other component or<br>accessory of the<br>SureTemp Plus<br>thermometer system,<br>nor the thermometer<br>itself. The proposed<br>device's Indications<br>for Use are,<br>therefore, more<br>narrow than the<br>Indications for Use<br>stated in K030580,<br>which applied to<br>both the<br>thermometer and the<br>probe covers, as the<br>proposed device's<br>indications are only<br>applicable to the<br>probe covers. |
|-----------------------------------|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number | 21 CFR 880.2910 | 21 CFR 880.2910 | Same |
| Design Features | Conforms to ASTM Standard<br>E1104 | Conforms to ASTM Standard<br>E1104 | Same |
| Design Configurations | One size | One size | Same |
| Performance Specifications | Conforms to ASTM Standard<br>E1104 | Conforms to ASTM Standard<br>E1104 | Same |
| Prescription vs. OTC | Prescription Only | Prescription Only | Same |
| Sterile vs. Non-Sterile | Non-sterile | Non-sterile | Same |
| Disposable vs. Non-<br>Disposable | Disposable | Disposable | Same |
| Single Use vs. Reusable | Single Use Only | Single Use Only | Same |
#### Summary of Non-Clinical Testing
Non-clinical verification of the Medline Disposable Electronic Thermometer Probe Cover has been conducted to evaluate its performance and functionality. The results of these tests have demonstrated the overall performance of the proposed device and ultimately support a substantial equivalence determination.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image features the logo of Medline, a healthcare company. The logo is set against a blue square background. The word "MEDLINE" is written in bold, white letters, stacked on top of a white, stylized starburst symbol. A gray bar is visible on the right side of the image.
ledline Industries. Three Lakes Drive Northfield II 60093
Specifically, the evaluation of the proposed device includes:
- . Performance Testing (Bench)
- o In accordance with ASTM E1104-98 (2016)
- . Biocompatibility Testing
- ISO 10993-5: Cytotoxicity MEM Elution; o
- ISO 10993-10: Irritation Intracutaneous reactivity; and, o
- o ISO 10993-10: Delayed-Type Hypersensitivity (Sensitization) Guinea Pig Maximization Test.
## Conclusion
In accordance with 21 CFR Part 807, and based on a comparison of 'Indications for Use,' technological characteristics and performance data, Medline Industries, Inc. concludes that the proposed Medline Thermometer Probe Cover is substantially equivalent to the predicate device cleared under K030580.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.