← Product Code [ODC](/productcode/ODC) · K173105

# Endoscopic Cap, Check-Flo Adapter, Side-arm Adapter, Tuohy-Borst Adapter (K173105)

_Cook Incorporated · ODC · Mar 22, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173105

## Device Facts

- **Applicant:** Cook Incorporated
- **Product Code:** [ODC](/productcode/ODC.md)
- **Decision Date:** Mar 22, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

This device is intended to resist the backflow of fluid around an instrument inserted through the working channel of a ureteroscope, cystoscope, or other endoscope.

## Device Story

Endoscopic accessories (Cap, Check-Flo Adapter, Side-Arm Adapter, Tuohy-Borst Adapter) attach to endoscope working channels or female Luer locks. Devices function as mechanical seals to resist fluid backflow during instrument insertion. Operated by clinicians in endoscopic settings. Adapters feature various configurations (e.g., Y-shape, side-arm, stopcocks) to accommodate specific instrument sizes (up to 12.0 Fr) and facilitate fluid management. Output is a sealed interface allowing instrument passage while maintaining pressure/fluid control. Benefits include reduced fluid leakage during procedures, improving visualization and procedural efficiency.

## Clinical Evidence

Bench testing only. Testing included biocompatibility, sterilization validation, leakage testing, device compatibility/integrity, flow rate analysis, and accelerated aging studies to verify design requirements.

## Technological Characteristics

Materials: silicone, acetal, polycarbonate, polyethylene, vinyl. Design: linear/cylindrical or Y-shaped adapters with Luer locks, valves, o-rings, and stopcocks. Connectivity: mechanical attachment to endoscope ports. Sterilization: not specified. Software: none.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- OBP Self Sealing Endoscopic Seal ([K091838](/device/K091838.md))
- Gateway™ Advantage Y-Adapter ([K130909](/device/K130909.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is in blue and consists of the letters "FDA" and the words "U.S. Food & Drug Administration".

March 22, 2018

Cook Incorporated Samuel Engelman Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47404

Re: K173105

> Trade/Device Name: Endoscopic Cap, Check-Flo® Adapter, Side-Arm Adapter, Tuohy-Borst Adapter Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: ODC Dated: February 20, 2018 Received: February 21, 2018

Dear Samuel Engelman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joyce M. Whang -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K173105

Device Name

Endoscopic Cap, Check-Flo® Adapter, Side-Arm Adapter, Tuohy-Borst Adapter

Indications for Use (Describe)

This device is intended to resist the backflow of fluid around an instrument inserted through the working channel of a ureteroscope, cystoscope, or other endoscope.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Cook Medical logo. The logo is white text on a red background. The word "COOK" is in a large, bold font, and the word "MEDICAL" is in a smaller font below it.

6
COOK INCORPORATE
750 DANIELS WA
BLOOMINGTON, IN 47404 US
PHONE: 812.339.2235 TOLL FREE: 800.457.450
WWW.COOKMEDICAL.CO

K173105 Page 1 of 3

#### 2.0 510(k) Summary

# Endoscopic Cap, Side-Arm Adapter, Check-Flo Adapter, and Tuohy-Borst Adapter 21 CFR §807.92 Date Prepared: September 28, 2017

| Submitted By:                      |                                           |
|------------------------------------|-------------------------------------------|
| Submission:                        | Traditional 510(k) Premarket Notification |
| Applicant:                         | Cook Incorporated                         |
| Contact:                           | Samuel Engelman                           |
| Applicant Address:                 | Cook Incorporated                         |
|                                    | 750 Daniels Way                           |
|                                    | Bloomington, IN 47404                     |
| Contact Phone:                     | (812) 339-2235 x104340                    |
| Contact Fax:                       | (812) 332-0281                            |
| Device Information:                |                                           |
| Trade Name:                        | Endoscopic Cap                            |
|                                    | Check-Flo® Adapter                        |
|                                    | Side-Arm Adapter                          |
|                                    | Tuohy-Borst Adapter                       |
| Common Name:                       | Endoscope Channel Accessory               |
| Classification Name:               | Endoscope and accessories                 |
| Classification Regulation:         | 21 CFR §876.1500, Product Code ODC        |
| Device Class/Classification Panel: | Class II, Gastroenterology/Urology        |

# Predicate Device:

- Primary predicate device: OBP Self Sealing Endoscopic Seal (K091838). ■
- Secondary predicate device: Gateway™ Advantage Y-Adapter (K130909) ■

# Device Description:

The Endoscopic Cap covers female Luer locks and bulb-type endoscopic fittings. The Endoscopic Cap is constructed from silicone. The proximal end is manufactured in three ways: without a hole, with a small hole, and with a large hole. The small hole is compatible with instruments that are 0.025 - 0.038 inches in diameter. The large hole is compatible with devices that are 7.0-12.0 Fr.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a dark red background. The dark red background is shaped like a banner.

OK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA NE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM

#### K173105 Page 2 of 3

The Check-Flo Adapters have a linear, cylindrical shape and attach to female Luer locks and bulb-type endoscopic fittings. The Check-Flo Adapter is constructed from four components: male Luer lock, body, valve, and cap. The distal side of the Check-Flo Adapter is an acetal male Luer lock. The body is constructed from acetal and connects all the components. The valve is a silicone disc and is seated in the cap. The cap is constructed from acetal and is located at the distal end of the Check-Flo Adapter. The Check-Flo Adapters are compatible with devices that are less than or equal to 9.0 Fr.

The Check-Flo Adapter design has two variations: a side-arm assembly and universal tubing. The side-arm assembly is a 10 cm length of vinyl tubing with a female Luer lock made from polycarbonate. The side-arm assembly comes with an additional one-way or three-way stopcock constructed from polycarbonate and polyethylene. The Check-Flo Adapter with universal tubing includes a segment of white silicone tubing. It protrudes 2.5 cm from the distal end of the male Luer lock and is 4.0 French in diameter.

The Side-Arm Adapter is shaped like the letter "y" and attaches to female Luer locks. It is constructed from four components: the body, male Luer lock, o-ring, and cap. The body is constructed from acetal. The protruding side-arm of the body has a female Luer lock. A one-way stopcock is provided and is constructed from polycarbonate and polyethylene. The male Luer lock is constructed from acetal and is located on the distal end of the body. The o-ring is seated inside the cap located on the proximal end of the body. The Side-Arm Adapter is compatible with instruments that are 3.0 - 6.0 Fr.

The Tuohy-Borst Adapter has a linear, cylindrical shape and attaches to female Luer locks. The Tuohy-Borst Adapter is constructed from four components: tightening knob, body, sealing ring, and rotator. The tightening knob is located on the proximal side of the Tuohy-Borst Adapter. The body is constructed from polycarbonate and connects all the components. The sealing ring is seated inside the rotator located on the distal end of the body. The Touhy-Borst is compatible with instruments that are 6.0 Fr and smaller.

The Tuohy-Borst has an alternate design where the body has an additional side-arm with female Luer lock. This design has the same four basic components.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in white, bold, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a red background with a diagonal cut on the lower left side.

8
COOK INCORPORATE
750 DANIELS WA
BLOOMINGTON, IN 47404 US
PHONE: 812.339.2235 TOLL FREE: 800.457.450
WWW.COOKMEDICAL.COM

Page 3 of 3

#### Indications for Use:

This device is intended to resist the backflow of fluid around an instrument inserted through the working channel of a ureteroscope, cystoscope, or other endoscope.

# Comparison to Predicate Device:

The subject devices have similar indications for use, methods of operation, and fundamental technological characteristics as the predicate devices. Differences between the subject devices and the predicate devices include a difference in product code, indications for use, components, design, and material. Characteristics of the subject devices that differ from the predicate devices are supported by testing and analysis.

#### Performance Data:

The following testing was performed in order to demonstrate that the proposed Endoscopic Cap, Side-Arm Adapter, Check-Flo Adapter, and Tuohy-Borst Adapter met applicable design and performance requirements.

- Biocompatibility
- 트 Sterilization
- Leakage
- Device Compatibility and Integrity
- Flow Rate
- Accelerated Age

# Conclusion:

The results of these tests support a conclusion that the Endoscopic Cap, Side-Arm Adapter, Check-Flo Adapter, and Tuohy-Borst Adapter will perform as intended. The subject devices do not raise new questions of safety or effectiveness as compared to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K173105](https://fda-staging.innolitics.com/device/K173105)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
