K173093 · Tama Research Corporation · NFO · Jun 5, 2018 · Neurology
Device Facts
Record ID
K173093
Device Name
TAMA BEMS device
Applicant
Tama Research Corporation
Product Code
NFO · Neurology
Decision Date
Jun 5, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TAMA BEMS Device uses microcurrent electrical to stimulate facial tissues for aesthetic purposes
Device Story
TAMA BEMS is a battery-operated microcurrent stimulator for aesthetic facial tissue stimulation. Device delivers electrical currents up to 800 microamps via probes. User adjusts output current (0-800 micro-amps) and modality (frequency) via pushbuttons. LED indicators display intensity, battery level, and output mode; 'Flow' light confirms skin contact and current delivery. Device is disabled during battery charging. Intended for aesthetic use. Healthcare provider or user operates device to stimulate facial tissues. Output affects tissue stimulation; benefits include aesthetic facial toning. Device features automatic overload, no-load, and shut-off safety mechanisms.
Clinical Evidence
No clinical data required. Substantial equivalence supported by benchtop performance testing, including biocompatibility (cytotoxicity), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), nerve/muscle stimulator safety (IEC 60601-2-10), and human factors engineering (IEC 60601-1-6, ISO 62366).
Technological Characteristics
Microcurrent stimulator; output up to 800 microamps; biphasic rectangular waveform. Housing: Anodized aluminum 6061. Power: Internal 4.3V rechargeable Lithium Polymer battery. Connectivity: Standalone. Safety: Automatic overload, no-load, and shut-off trips. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 14971, IEC 60601-1-6, ISO 62366.
Indications for Use
Indicated for aesthetic stimulation of facial tissues using microcurrent electrical energy.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
Beautiful Image Model 900 Facial Toning Device (K130065)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The first logo on the left is the Department of Health & Human Services-USA logo. The second logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
June 5, 2018
TAMA Research Corporation % Robert Packard President Medical Devices Academy, Inc. 345 Lincoln Hill Road Shrewsbury, Vermont 05728
Re: K173093
Trade/Device Name: TAMA BEMS device Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: May 7, 2018 Received: May 7, 2018
Dear Robert Packard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# VivekJ. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173093
Device Name TAMA BEMS Device
Indications for Use (Describe)
The TAMA BEMS Device uses microcurrent electrical to stimulate facial tissues for aesthetic purposes
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
### 1. SUBMITTER TAMA Research Corporation 7600 North 16th Street. Suite 205 Phoenix, AZ, 85020 USA Tel: +1 (602) 354-8185
| Contact Person: | Ali Shambayati |
|-----------------|----------------|
| Date Prepared: | June 4, 2018 |
| Name of Device: | TAMA BEMS Device |
|------------------------------|---------------------------------|
| Classification Name: | Neurological Therapeutic Device |
| Regulation: | 21 CFR §882.5890 |
| Regulatory Class: | Class II |
| Product Classification Code: | NFO |
#### PREDICATE DEVICE 111
| Predicate Manufacturer: | Biosonic Technologies, LLC |
|-------------------------|------------------------------------------------|
| Predicate Trade Name: | Beautiful Image Model 900 Facial Toning Device |
| Predicate 510(k): | K130065 |
No reference devices were used in this submission.
#### DEVICE DESCRIPTION IV.
The TAMA BEMS Device is a microcurrent stimulator producing electrical currents up to 800 microamps. It is intended for aesthetic purposes. It has built-in pushbuttons to allow the user to adjust the output current from 0 to 800 micro-amps, as well as changing the modality (output frequency). The BEMS device is a battery-operated device that uses a rechargeable battery. When the device is plugged into the AC outlet for charging, it becomes disabled and no treatments can be performed. The BEMS device also has LED indicators to show output intensity, battery level, and output mode. A Flow light shows when probes contact the skin, and current is flowing through the patient. If the measured current through the probes does not match the set point selected by the user, the Flow LED turns off to indicate insufficient current flow.
#### V. INDICATIONS FOR USE
The TAMA BEMS Device uses microcurrent electrical to stimulate facial tissues for aesthetic purposes
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVCE
The following characteristics were compared between the subject device and the predicate device in order to demonstrate substantial equivalence:
{4}------------------------------------------------
- Indications for Use
- Materials of Construction ●
- Design Features
- . Energy Source
### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility Testing
A biocompatibility certification statement was provided indicating that the components with patient contact are fabricated from biocompatible metal that is also used for other TENS devices legally marketed in the USA. In addition, cytotoxicity testing was performed to demonstrate that there are no residuals present that negatively impact biocompatibility.
### Electrical safety and electromagnetic compatibility (EMC)
The following electrical safety and EMC tests have been performed:
- . IEC 60601-1 Electrical Safety Testing
- IEC 60601-1-2 EMC Testing ●
- . IEC 60601-2-10 Medical Electrical Equipment Safety Standard for Nerve and Muscle Stimulators
### Software Verification and Validation Testing
Verification and validation testing of the software was conducted in accordance with IEC 62304.
### Mechanical and acoustic Testing
A risk analysis was completed and risk controls were implemented in accordance with ISO 14971. Human factors testing was conducted in accordance with IEC 60601-1-6. ISO 62366 and the FDA guidance on human factors engineering to demonstrate that the ergonomics of patient and user interfaces for the subject device are substantially equivalent to the predicate device.
### Animal Study
Animal performance testing was not required to demonstrate safety and effectiveness of the device.
### Clinical Studies
Clinical testing was not required to demonstrate the safety and effectiveness of the TAMA BEMS Device. Instead, substantial equivalence is based upon benchtop performance testing.
{5}------------------------------------------------
| | TAMA BEMS Device –<br>Subject Device | Beautiful Image Model<br>900 Facial Toning Device – K130065 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The TAMA BEMS<br>Device uses microcurrent<br>electrical to stimulate<br>facial tissues for aesthetic<br>purposes | The Beautiful Image<br>Model 900 Facial Toning<br>Device uses microcurrent<br>electrical energy to<br>stimulate facial tissues for<br>aesthetic<br>purposes. |
| Design | | |
| Power Source | Internal 4.3V<br>rechargeable Lithium<br>Polymer battery | One 6V Battery |
| Method of Line Current Isolation | N/A | N/A |
| Patient Leakage Current | N/A | N/A |
| Normal Condition (μΑ) | N/A | N/A |
| Single Fault Condition (µA) | N/A | N/A |
| Average DC current through electrodes when<br>device is on but no pulses are being applied (μΑ) | None | None |
| Number of Output Modes | Four | One |
| Number of Output Channels | One | One |
| Synchronous or Alternating? | N/A | N/A |
| Method of Channel Isolation | N/A | N/A |
| Manufacturer | TAMA Research<br>Corporation | Biosonic Technologies,<br>LLC |
| Regulated Current or Regulated Voltage? | Both | Both |
| Software/Firmware/Microprocessor Control? | Yes | Yes |
| Automatic Overload Trip? | Yes | Yes |
| Automatic No-Load Trip? | Yes | Yes |
| Automatic Shut Off? | Yes | Yes |
| User Override Control? | Yes | Yes |
| Indicator Display Status | | |
| On/Off Display Status? | Yes | Yes |
| Low Battery? | Yes | Yes |
| Voltage/Current Level? | Yes | Yes |
| Timer Range (minutes) | 0-30 minutes | None |
| Compliance with Voluntary Standards? | IEC60601-1 | IEC 60601-1 |
| Compliance with 21 CFR 989? | Yes | Yes |
| Weight (lbs., oz.) | 9.5 ounces | 10 lbs |
| Dimensions (in.) [WxHxD] | 3x5x0.7 | 5.5x15.3x11.3 |
| Housing Materials and Construction | Anodized aluminum 6061 | Thermoplastic |
| Waveform (e.g., pulsed monophasic, biphasic) | Biphasic | Biphasic |
| Shape (e.g., rectangular, spike, rectified sinusoidal) | Rectangular | Rectangular |
| Maximum Output Voltage (volts) (+/-5%) | ±0.400 @500Ω<br>±1.600 @2 k Ω<br>±8.00 @10k Ω | 0.347 @500Ω<br>1.242 @2 k Ω<br>5.780 @10k Ω |
| Maximum Output Current (mA) (+/-5%) | ±0.800 @500 Ω<br>±0.800 @2 k Ω<br>±0.800 @10k Ω | 0.647 @500 Ω<br>0.625 @2 k Ω<br>0.584 @10k Ω |
| Duration of primary (depolarizing) phase (msec) | 26.3 – 1200 | 0.648 – 322 |
| Pulse Duration (msec) | 52.6 – 2400 | 3.24 – 1610 |
| Frequency (Hz) [or Rate (pps)] | 0.045 – 2560 | 0.621 – 308.6 |
| For multiphasic waveforms only:<br>Symmetrical phases?<br>Phases Duration (msec), (state range, if applicable), (both phases, if asymmetrical) | Yes<br>26.3 – 1200 | Yes<br>0.324-161 |
| Net Charge (micro coulombs (µC) per pulse) (If zero, state method of achieving zero net charge.) | 0µC @500 Ω | 0µC @500 Ω |
| Maximum Phase Charge, (µC) | 400 µC positive phase,<br>400 µC positive phase,<br>(50% duty cycle), all loads | 190 @500 Ω |
| Maximum Current Density (mA/cm², r.m.s.) | 1.591 @500 Ω (1) | 1.486 @500 Ω |
| Maximum Average Current (average absolute value), mA | 0.800 @500 Ω | 0.493 @500 Ω |
| Maximum Average Power Density, (W/cm²), (using smallest electrode conductive surface area) | 318E-6 @500Ω (2)<br>0.012 @18.75 KΩ (3) | 366E-6 @500 Ω |
| Burst Mode (i.e., pulse trains)<br>(a) Pulses per burst<br>(b) Bursts per second<br>(c) Burst duration (seconds)<br>(d) Duty Cycle: Line (b) x Line (c) | 5 - 20<br>10 - 40<br>0.00195 - 0.0078<br>0.0195 – 0.1719 (4) | N/A<br>N/A<br>N/A<br>N/A |
| ON Time (seconds) | Constant | 10-30 |
| OFF Time (seconds) | None | 1-6 |
| Additional Features (specify if applicable) | None | None |
The following table provides a Substantial Equivalence Comparison of TAMA BEMS with Beautiful Image Model 900 Facial Toning Device (K130065):
{6}------------------------------------------------
## VIII. CONCLUSIONS
Based on a comparison of indications for use, technological characteristics and performance data; it can be concluded that the TAMA BEMS Device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.