F&P NivairoTM RT046 Non-Vented Hospital Full Face Mask, Standard Elbow Version
K173060 · Fisher and Paykel Healthcare Limited · CBK · Jan 25, 2018 · Anesthesiology
Device Facts
Record ID
K173060
Device Name
F&P NivairoTM RT046 Non-Vented Hospital Full Face Mask, Standard Elbow Version
Applicant
Fisher and Paykel Healthcare Limited
Product Code
CBK · Anesthesiology
Decision Date
Jan 25, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fisher & Paykel Healthcare single patient use masks are intended for use as an accessory to ventilators to enable noninvasive positive pressure ventilation (NPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (>30kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained by trained medical practitioners in a hospital/institutional environment with patient monitoring in place.
Device Story
Oronasal full face mask; accessory to ventilators for noninvasive positive pressure ventilation (NPPV). Input: pressurized breathing gases from external flow source/ventilator via dual limb circuit. Output: delivery of gases to patient airway. Used in hospital/institutional environments; operated by trained medical practitioners. Features RollFit rolling bridge and TubeFit zones for facial seal; sits under lip on chin. Provides therapy for respiratory insufficiency/failure; enables CPAP or bi-level ventilation. Benefits patient by facilitating noninvasive respiratory support.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including ISO 5356-1 (connectors), ISO 10993-1 (biocompatibility), and ISO 17510-2 (masks/accessories). Biocompatibility testing included irritation, sensitization, cytotoxicity, genotoxicity, muscle implantation, acute systemic toxicity, particulate matter, chemical characterization, and VOCs.
Technological Characteristics
Oronasal full face mask; 22mm female swivel adaptor for dual limb circuit connection. Pressure range 4-25 cmH2O. Features RollFit rolling bridge and TubeFit zones. Non-sterile, single-use. Complies with ISO 5356-1, ISO 10993-1, and ISO 17510-2.
Indications for Use
Indicated for spontaneously breathing adult patients (>30kg) with respiratory insufficiency or respiratory failure requiring noninvasive positive pressure ventilation (NPPV) therapy (CPAP or bi-level) in a hospital/institutional setting.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
RT041 Hospital Full Face Mask Non-Vented (K083122)
PerformaTrak SE (Image3 SE) Disposable Face Mask (K023135)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the words "U.S. FOOD & DRUG" on the first line and "ADMINISTRATION" on the second line. The FDA logo is in blue.
January 25, 2018
Fisher and Paykel Healthcare Limited Ashley Nguyen Regulatory Affairs Specialist 15 Maurice Paykel Place Auckland, 2013 NZ
Re: K173060
Trade/Device Name: F&P NivairoTM RT046 Non-Vented Full Face Mask, Standard Elbow Version Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: December 20, 2017 Received: December 22, 2017
Dear Ashley Nguyen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Image /page/1/Figure/5 description: This image shows a formal letter closing. The letter is signed by Tara A. Ryan -S, with a digital signature including the date 2018.01.25. The letter is for Tina Kiang, Ph.D., who is the Acting Director of the Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices, Office of Device Evaluation, Center for Devices and Radiological Health.
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMS No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
## 510(k) Number (ifknown)
K173060
Device Name
F&P Nivairo™ RT046 Non-Vented Hospital Full Face Mask, Standard Elbow Version
### Indications for Use (Describe)
The Fisher & Paykel Healthcare single patient use masks are intended for use as an accessory to ventilators to enable noninvasive positive pressure ventilation (NPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (>30kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained by trained medical practitioners in a hospital/institutional environment with patient monitoring in place.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X |Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary 5
| Contact person/submitter | Ashley Nguyen |
|--------------------------|----------------------------------------------------------|
| Date prepared | January 25, 2018 |
| Contact details | Address: 15 Maurice Paykel Place |
| | East Tamaki |
| | Auckland 2013, New Zealand |
| | Telephone: +64 9 574 0100 |
| | Fax: +64 9 574 0158 |
| Trade name | F&P Nivairo™ RT046 Non-Vented Hospital Full Face Mask, |
| | Standard Elbow Version |
| Common name | Full Face Mask |
| Classification name | Non Continuous Ventilator IPPB (accessory to) |
| | Class II, 21 CFR 868.5895 |
| | Product code CBK (Anesthesiology) |
| Predicate device | K083122 RT041 Hospital Full Face Mask Non-Vented |
| Reference devices: | K130328 F&P Simplus Full Face Mask |
| | K170367 F&P Nivairo™ RT045 Full Face Mask |
| | K023135 PerformaTrak SE (Image3 SE) Disposable Face Mask |
#### 5.1 Device Description
The Nivairo™ RT046 Non-Vented Hospital Full Face Mask, Standard Elbow Version (herein referred to as RT046) is a non-vented hospital full face mask with a standard elbow for use with dual limb circuits. The RT046 is a single use device intended for use as an accessory to deliver non-invasive positive pressure ventilation (NPPV) to a patient as part of a passively vented non-invasive ventilation system.
The RT046 is an oronasal full face mask intended to be used in a hospital/institutional environment by trained medical staff with patient monitoring systems in place. It connects to a dual limb breathing circuit via a 22mm female swivel adaptor to receive pressurized breathing gases from an external flow source or ventilator (CPAP or Bi-Level).
The RT046 mask is a prescription only device, provided in a non-sterile state.
#### 5.2 Intended Use
The Fisher & Paykel Healthcare single use masks are intended for use as an accessory to ventilators to enable non-invasive positive pressure ventilation (NPPV) therapy (CPAP or bilevel) to be delivered to spontaneously breathing adult patients (>30kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained by trained medical practitioners in a hospital/institutional environment with patient monitoring in place.
{4}------------------------------------------------
#### Technological Characteristics Comparison 5.3
The RT046 (subject) contains the following key similarities to the previously cleared predicate RT041 Hospital Full Face Mask Non Vented (herein referred to as RT041):
- Substantially equivalent intended use with same patient population and operating . environment.
- · Same mode of operation whereby both masks deliver gases are oronasal.
The key differences between the RT046 are that the RT041:
- · Headgear color is grey and light blue
- · Has a new addition of a crown strap to the head gear
- Uses headgear clips instead of a push button clip
- Uses new headgear materials .
- · Facial Seal includes two new features a rolling bridge (RollFit™) and two TubeFir"™ zones
- . Facial Seal sits under the lip, on the chin
- Does not include a silicone forehead padding ●
- Change in dead space volume ●
- . Has a pressure range of 4-25 cmH2O
- Is available in four sizes extra small, small, medium and large .
The RT046 (subject) contains the following key similarities to the reference devices:
- · F&P Simplus Full Face Mask (K130328)
- Headgear release design ●
- . Face coverage of seal
- Seal (RollFit™) ●
- Materials and components ●
- · F&P Nivairo™ RT045 Full Face Mask (K170367)
- Headgear color is grey and light blue ●
- Has a new addition of a crown strap to the head gear ●
- Uses headgear clips instead of a push button clip ●
- Uses new headgear materials
- Facial Seal includes two new features a rolling bridge (RollFit™) and two ● TubeFit™ zones
- Facial Seal sits under the lip, on the chin ●
- Identical dead space volume
- Has a pressure range of 4-25 cmH2O ●
- Is available in four sizes extra small, small, medium and large ●
- · Image3 SE Disposable Face Mask (K023135)
- Similar dead space ●
{5}------------------------------------------------
#### 5.4 Non-Clinical Performance Data
Performance testing of the RT046 (subject) was conducted and data were compared to the performance data for the RT041 (predicate) for performance. The data demonstrate substantial equivalence of the RT046 to the RT041. The results of the comparative bench testing do not raise any new questions regarding safety or effectiveness for the RT046.
Where applicable the RT046 has been evaluated to:
- . ISO 5356-1:2004 Anaesthetic and respiratory equipment - Conical connectors -Part 1: Cones and sockets.
- ISO 10993-1:2009 Biological Evaluation of medical devices Part 1: Evaluation and ● testing within a risk management process.
- . Clauses 5.3 and 5.5 of ISO 17510-2:2007 Sleep Apnoea Breathing Therapy – Part 2: Masks and Application Accessories
The following biocompatibility testing was performed. The materials passed all parameters:
- Irritation ●
- . Sensitization
- Cytotoxicity .
- Genotoxicity
- . Muscle Implantation
- Acute Systemic Toxicity
- . Particulate Matter
- . Chemical Characterization
- . Volatile Organic Compounds
#### 5.5 Clinical Performance Data
Clinical performance testing was not required to demonstrate substantial equivalence for the RT046.
#### 5.6 Conclusions
The comparison of features, performance, technological characteristics, and intended use demonstrate that the RT046 (subject) is substantially equivalent to the RT041 (predicate).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.