The Cocoon device is intended for use in dental radiographic examination and diagnosis of the teeth, jaw, and oral structure for both pediatric and adult patients.
Device Story
Cocoon is a handheld, portable dental X-ray system; utilizes internal X-ray tube to expose intraoral receptors (analog film, digital phosphorous plates, or digital sensors) to ionizing radiation. Operated by dental professionals in clinical settings; handheld or tripod-mounted. Device transforms electrical energy from a 16.8V DC Li-Polymer battery into controlled X-ray pulses (70kVp; 1.7mA or 2.0mA). Exposure parameters (0.05–1.00s) set by operator; output used for diagnostic imaging of teeth, jaw, and oral structures. Benefits include portability for dental radiographic examinations. Clinical decision-making supported by resulting images.
Clinical Evidence
Bench testing only. Performance validated through electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), radiation protection (IEC 60601-1-3), usability (IEC 60601-1-6, IEC 62366-1), and dental-specific safety (IEC 60601-2-65). No clinical data provided.
Technological Characteristics
Handheld X-ray system; 16.8V DC Li-Polymer battery powered. Tube voltage 70kVp (fixed); tube current 1.7mA (DX-7017) or 2.0mA (DX-7020). Dimensions: 7.6"x5.2"x10.6"; weight 5.0 lbs. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-65, ISO 10993-1, ISO 14971. Software lifecycle per IEC 62304.
Indications for Use
Indicated for taking diagnostic dental x-rays for both pediatric and adult patients using intraoral film or digital sensors.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are connected by a flowing line that resembles a ribbon or wave.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 26, 2017
Dexcowin Co., Ltd. % Ms. Joyce St. Germain Regulatory Department Manager 100 East Granada Blvd., Suite 219 ORMOND BEACH FL 32176
Re: K173046
Trade/Device Name: Cocoon Portable Handheld X-ray System Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: September 26, 2017 Received: September 28, 2017
Dear Ms. St. Germain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173046
## Device Name
Cocoon Portable Handheld X-ray system
Indications for Use (Describe)
Cocoon is a portable X-ray System indicated for taking diagnostic dental x-rays for both pediatric and adult patients using intraoral film or digital sensors.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a stylized drawing of an eagle's head and wing. The eagle's head is facing to the right, and its beak is slightly open. The wing is spread out behind the head, and it has five feathers. The drawing is done in black and white, and the lines are clean and simple.
510k FDA Consulting
Medical Device Clearance
100 East Granada Blvd., Suite 219
Ormond Beach, FL 32176
386-506-8711
# Special 510(k) Summary / K173046
#### Submitter/Applicant
Dexcowin Co., Ltd. #905, 2, Gasan digital 1-ro, GeumCheon-Gu Seoul, Korea 08591
| Phone: | 82-2-2027-2880 |
|----------|-------------------------------------------------|
| Fax: | 82-2-2027-2884 |
| Contact: | Ryu Seung-Bum (sbryu@dexcowin.com)Date Prepared |
September 26, 2017
## Preparer/Consultant
Denterprise International, Inc. 100 East Granada Blvd., Suite 219 Ormond Beach, FL 32176
Phone: 386-506-8711 Fax: 386-672-4402 Primary Contact: Joyce St. Germain, Regulatory Dept. Mgr. (joyce@510kfda.com) Secondary Contacts: Claude Berthoin, President (claude@510kfda.com).
#### Device Classification
Trade Name: Common Name: Regulation Name: Regulation Number: Primary Product Code Regulatory Class: 510k Review Panel:
Cocoon Portable Handheld X-ray system Portable x-ray System Extraoral Source X-ray System 21 CFR 872.1800 MUH II Dental
{4}------------------------------------------------
### Predicate Device
The following predicate was cleared as a Special 510k to the original (K083532) reference predicate.
The following predicate is a legally marketed, post-amendment device:
| 510(k) Number: | K133007 |
|-----------------------|-------------------------------|
| Date Cleared: | April 24, 2014 |
| Device Name: | DX-3000 Portable X-ray System |
| Common Name: | Portable X-ray System |
| Regulation Name: | Extraoral Source X-ray System |
| Regulation Number: | 21 CFR 872.1800 |
| Primary Product Code: | MUH |
| Regulatory Class: | II |
| 510k Review Panel: | Dental |
#### Reference Predicate
| 510(k) Number: | K083532 |
|-----------------------|-------------------------------|
| Clearance Date: | March 10, 2009 |
| Device Name: | ADX4000-L, DX3000-L |
| Common Name: | Portable X-ray System |
| Regulation Name: | Extraoral source X-ray System |
| Regulation Number: | 21 CFR 872.1800 |
| Regulatory Class: | II |
| Primary Product Code: | MUH |
#### Indications for Use
Cocoon is a portable X-ray System indicated for taking diagnostic dental x-rays for both pediatric and adult patients using intraoral film or digital sensors.
#### Intended Use
The Cocoon device is intended for use in dental radiographic examination and diagnosis of the teeth, jaw, and oral structure for both pediatric and adult patients.
#### Device Description
Cocoon Portable X-ray System from Dexcowin Co., Ltd., of Seoul, Korea is a handheld x-ray device. The technology of this device was originally developed in Korea more than a decade ago and global production is still concentrated in that country.
The subject device is designed for dental radiographic examination and diagnosis
{5}------------------------------------------------
of diseases of the teeth, jaw, and oral structure by exposing an x-ray image receptor to ionizing radiation. The x-ray source, a tube, is located inside the handheld device. All three conventional types of intraoral receptors can be used with this device-analog x-ray film, digital phosphorous plates, and digital x-ray sensors.
## Comparison of Technological Characteristics with Predicate
The following table compares technological and other characteristics of the subject and predicate device.
# Table 5 -- Technological Comparison Subject Models: DX-7017 and DX-7020
| Feature | Cocoon<br>(Subject Device)<br>DX-7017 / DX-7020<br>K173046 | Dexcowin DX3000<br>(Predicate K133007) | Dexcowin DX3000-L<br>(Predicate K083532) |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use | Indicated for taking<br>diagnostic dental x-rays<br>for both pediatric and<br>adult patients using<br>intraoral film or digital<br>sensors. | Indicated for taking<br>diagnostic dental x-rays<br>for both pediatric and<br>adult patients using<br>intraoral film or digital<br>sensors. | Indicated for taking<br>diagnostic dental x-rays<br>for both pediatric and<br>adult patients using<br>intraoral film or digital<br>sensors. |
| Mechanical | | | |
| Size : Body | 7.6"(H)×5.2"(W)×10.6"(D) | 5.5"(H)×6.5"(W)×3.2"(D) | 5.5"(H)×6.5"(W)×3.2"(D) |
| Weight | 5.0 lbs. (without battery) | 4.4 lbs. (without<br>battery ) | 4.4 lbs. (without<br>battery) |
| Source to<br>skin<br>distance | 180mm | 200mm | 200mm |
| Cone<br>diameter | 58mm | 55mm | 55mm |
{6}------------------------------------------------
| Exposure<br>switch | On tube-head assembly,<br>or at control panel | On tube-head assembly,<br>or at control panel | On tube-head assembly,<br>or at control panel |
|-----------------------|-----------------------------------------------------|-----------------------------------------------------|-----------------------------------------------------|
| Tube head<br>mounting | Handheld, or on a tripod | Handheld, or on a tripod | Handheld, or on a tripod |
| Electrical | | | |
| Energy<br>Source | Rechargeable 16.8V<br>DC Li-Polymer<br>battery pack | Rechargeable 16.8V<br>DC Li-Polymer battery<br>pack | Rechargeable 16.8V<br>DC Li-Polymer battery<br>pack |
| Exposure<br>Time | 0.05~1.00 seconds in<br>0.001 increments | 0.05~1.35 seconds in<br>0.005 increments | 0.05~1.35 seconds in<br>0.005 increments |
| Timer<br>Accuracy | ±(10% + 1ms) | ±(10% + 1ms) | ±(10% + 1ms) |
| Tube<br>Voltage | 70kVp fixed for<br>DX-7017 and DX-7020 | 60kVp fixed | 60kVp fixed |
| Tube<br>Current | 1.7mA = DX-7017<br>2.0mA = DX-7020 | 2mA fixed | 1mA fixed |
The above comparison shows the subject and predicate devices have substantially the same technology characteristics.
The differences of the subject device are shown in blue highlights as compared to the predicate devices. The differences of the subject are due to design changes and the differences have all been tested according to ISO standards and have proven to be as safe and effective as the previous devices. The indications for use and the technological characteristics are the same in these devices and the subject device does not cause any new issues of concern.
All Standard Test Performed for the Subject Device Models: DX-7017 and DX-7020 are listed below:
{7}------------------------------------------------
# Declaration of Conformity
- · Council Directive 93/42/EEC as amended by 2007/47/EC of 14 June 1993 concerning medical devices
- MEDDEV 2.12-1 Rev.8 Guidelines on a medical device vigilance system
| No | Standard | Application |
|----|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | EN ISO 13485:2012 | Medical devices - Quality management systems -<br>Requirements regulatory purposes(ISO 13485:2003)<br>EN ISO13485:2003/AC:2007 |
| 2 | EN ISO 14971:2012 | Medical devices - Application of risk management to<br>medical devices(ISO14971:2012) |
| 3 | IEC 60601-1:2005<br>/AMD1:2012 | Medical electrical equipment - part1: General<br>requirements for basic safety and essential<br>performance |
| 4 | IEC 60601-1-2:2007 | Medical electrical equipment - Part1-2:General<br>requirements for basic safety and essential<br>performance - Collateral standard: Electromagnetic<br>compatibility - Requirements and tests |
| 5 | IEC 60601-1-3:2008 | Medical electrical equipment - Part1-3: General<br>requirements for basic safety essential performance -<br>Collateral standard: Radiation protection in diagnostic<br>X-ray equipment |
| 6 | IEC 60601-1-6:2010<br>/AMD1:2013 | Medical electrical equipment-part1-6:General<br>requirements for basic safety and essential<br>performance-Collateral standard: Usability |
| 7 | IEC 60601-2-28:2010 | Medical electrical equipment-part2-28: Particular<br>requirements for the basic safety and essential<br>performance of X-ray tube assemblies for medical<br>diagnosis |
| 8 | IEC 60601-2-65:2012 | Medical electrical equipment-part2-65:Particular<br>requirements for the basic safety and essential<br>performance of dental intra-oral X-ray equipment |
| 9 | IEC 62366-1:2015 | Medical devices-part1: Application of usability |
| | | engineering to medical devices |
| 10 | IEC62304:2015 | Medical device software - software life cycle<br>processes. |
| 11 | ISO 14155-1:2010 | Clinical investigation of medical devices for human<br>subjects |
| 12 | ISO 15223-1:2010 | Medical devices: Symbols to be used with medical<br>devices labels, labeling and information to be supplied |
| 13 | EN1041:2008 | Information supplied by the manufacturer with medical<br>devices |
| 14 | ISO 10993-1:2009 | Biological evaluation of medical devices-Part1:<br>Evaluation and testing within a risk management<br>process |
| 15 | ISO 17050-1:2010 | Conformity assessment-supplier`s declaration of<br>conformity - part 1:General requirements |
{8}------------------------------------------------
## Non-Clinical Performance Data
The following performance was completed on the subject device in support of the substantial equivalence determination.
- Electrical Safety and EMC 0
- Specifications of Raw Material 0
- Software Validation 0
- Biocompatibility 0
- Usability 0
- Clinical Comparison O
- Risk Assessment O
- All tests were performed in accordance with ISO standards and recognized by 0 FDA.
- None of the standards was adapted for application to the device under review. 0
- There were no requirements of any standard that were not applicable to the 0 device.
- No deviations from the standards were applied. 0
{9}------------------------------------------------
- No differences exist between the tested device and the device to be marketed. 0
- Conformity with all standards was determined by the device manufacturer, O Dexcowin Co., Ltd., Korea.
- Electrical test performed by Met Laboratories, Inc. O
#### Specific Guidance Document
There is a FDA Specific Guidance Document associated with the device: Radiation Safety Consideration for X-ray Equipment Designed for Hand-Held Use. This submission utilized this guidance to develop this device to ensure the safety of this device for both the operators and the patients.
#### Labels
The labels on the device show that this device conforms to the following:
21 CFR 1020 Subchapter J: Performance Standards for Ionizing Radiation Emitting Products,
21 CFR 1020.30: Diagnostic x-ray systems and their major components,
21 CFR 1020.31: Radiographic Equipment
#### Conclusion
The Cocoon Portable X-ray System shows substantial equivalence to its original predicate and modified predicate devices. All differences were evaluated according to the tests and deemed not to raise new issues of safety or effectiveness. The modifications did not affect the device's intended use or alter the device's fundamental scientific technology. This modified subject device is at least as safe and effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.