The REUSABLE BLOOD PRESSURE CUFF is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in infant, pediatric and adult sizes. The DISPOSABLE BLOOD PRESSURE CUFF is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes.
Device Story
Device consists of blood pressure cuff with bladder and air hose (single or double tube). Operates by wrapping bladder around patient limb; secured via hook and loop closure. Air hose connects to noninvasive blood pressure measurement system. Used in clinical settings; operated by healthcare professionals. Disposable version (non-woven fabrics/PVC) for single-patient use; reusable version (210D Nylon TPU/PVC) for multiple uses. Provides pneumatic interface for blood pressure monitoring; facilitates accurate cuff inflation/deflation for pressure measurement. Benefits include compatibility with standard monitoring systems for patient blood pressure assessment.
Clinical Evidence
No clinical data included. Bench testing only. Compliance demonstrated with ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation and sensitization), and ISO 81060-1 (non-invasive sphygmomanometer requirements).
Technological Characteristics
Materials: Disposable (non-woven fabrics, PVC); Reusable (210D Nylon TPU, PVC). Sensing principle: Pneumatic bladder. Connectivity: Single or double tube air hose interface. Standards: ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 81060-1. Pressure range: 0-300 mmHg; Max pressure: 400 mmHg. Leakage: < 4 mmHg/min.
Indications for Use
Indicated for use as an accessory with noninvasive blood pressure measurement systems for neonatal, infant, pediatric, and adult patient populations.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 20, 2017
Ningbo Albert Novosino Co.,Ltd. % Ms. Doris Dong Manager Shanghai CV Technology Co., Ltd. Room 903 of Dongbao Building No. 19 Dongbao Road, Songjiang Area Shanghai, 201316 Cn
Re: K173044
Trade/Device Name: Reusable Blood Pressure Cuff, Disposable Blood Pressure Cuff Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO Dated: September 20, 2017 Received: September 28, 2017
Dear Ms. Doris Dong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
M.A. Hillebrand
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173044
Device Name REUSABLE BLOOD PRESSURE CUFF; DISPOSABLE BLOOD PRESSURE CUFF
Indications for Use (Describe)
The REUSABLE BLOOD PRESSURE CUFF is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in infant, pediatric and adult sizes. The DISPOSABLE BLOOD PRESSURE CUFF is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5 510(k) Summary
[As required by 21 CFR 807.92]
#### 1. Submission Sponsor
Ningbo Albert Novosino Co.,Ltd. No. 1 Xinheng 3 Road Cicheng Town, Jiangbei District, Ningbo, Zhejiang, P.R. China 315036 Phone: +86 574-83002882 Fax: +86 574-87527880 Contact: Victor Zhou
#### 2. Submission Correspondent
Shanghai CV Technology Co., Ltd. Room 903 of Dongbao Building, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613 China Phone: 86 21-31261348 Fax: 86 21-57712250 Contact: Doris Dong(Consultant) Email: doris_d@126.com
## 3. Date Prepared
Sept 20, 2017
## 4. Device Identification
Trade/Proprietary Name: REUSABLE BLOOD PRESSURE CUFF; DISPOSABLE BLOOD PRESSURE CUFF Common Name/Classification Name: Blood Pressure Cuff Product Code: DXQ Regulation Number: 21 CFR 870.1120 Regulation Class : Class II Review Panel: Cardiovascular
## 5. Predicate Devices
The proposed devices are substantially equivalent to the following predicate devices :
| Applicant | Device name | 510(k) Number | Product code |
|-------------------------|--------------------------------|---------------|--------------|
| Shenzhen Vistar Medical | Disposable Blood Pressure Cuff | K152465 | DXC |
| Supplies Co., Ltd. | | | |
{4}------------------------------------------------
| Shenzhen Vistar Medical<br>Supplies Co., Ltd. | Reusable Blood Pressure Cuff | K152468 | DXQ |
|-----------------------------------------------|------------------------------|---------|-----|
|-----------------------------------------------|------------------------------|---------|-----|
## 6. Device Description
The proposed device is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is available in neonatal, infant, pediatric and adult sizes.
The proposed device includes disposable blood pressure cuff and reusable blood pressure, both disposable and reusable blood pressure cuff has same structure, which contains Cuff with bladder and Air Hose. Air hose has single tube and double tube.
The disposable blood pressure cuff is single use device, and which is made of non-woven fabrics &PVC (Cuff) and PVC (Air Hose), thereinto, the non-woven fabrics & PVC (Cuff) is the material used to contacting with the patient.
The disposable blood pressure cuff has 26 models with different size for different population with different arm size, and the 13 models of 26 are use single tube of air hose and other 13 models are use double tube of air hose, which is used for different noninvasive blood pressure measurement system.
The reusable blood pressure cuff is reusable device, and which is made of 210D Nylon TPU (Cuff) and PVC (Air Hose), thereinto, the 210D Nylon TPU (Cuff) is the material used to contacting with the patient.
The reusable blood pressure cuff has 16 models with different size for different population with different arm size, and the 8 models of 16 are use single tube of air hose and other 8 models are use double tube of air hose, which is used for different noninvasive blood pressure measurement system.
## 7. Indication for Use Statement
## · REUSABLE BLOOD PRESSURE CUFF
Is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in infant, pediatric and adult sizes.
## · DISPOSABLE BLOOD PRESSURE CUFF
Is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes.
## 8. Comparison of Technological Characteristics
The following table compares the proposed devices to the predicate device with respect to K173044
{5}------------------------------------------------
Page 3 of 6
intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| ITEM | Proposed Device | Predicate Device | Remark |
|----------------|--------------------------------------|-----------------------------------|------------|
| Trade name | DISPOSABLE BLOOD | DISPOSABLE BLOOD | |
| | PRESSURE CUFF | PRESSURE CUFF | |
| 510(K) | Ningbo Albert Novosino Co.,Ltd. | Shenzhen Vistar Medical | |
| Submitter | | Supplies Co., Ltd. | |
| 510(K) | K173044 | K152465 | |
| Number | | | |
| Classification | 21CRF 870.1120 | 21CRF 870.1120 | SE |
| Regulation | | | |
| Classification | Class II, | Class II, | SE |
| and Code | DXQ | DXQ | |
| Common | Non-invasive Blood Pressure | Non-invasive Blood Pressure | SE |
| name | Cuff | Cuff | |
| Intended Use | The disposable blood pressure cuff | The disposable blood pressure | SE |
| | is an accessory used in | cuff is an accessory used in | |
| | conjunction with noninvasive | conjunction with noninvasive | |
| | blood pressure measurement | blood pressure measurement | |
| | systems. The cuff is non-sterile | systems. The cuff is non-sterile | |
| | and for single-patient use. It is | and for single-patient use. It is | |
| | available in neonatal, pediatric and | available in neonatal, pediatric | |
| | adult sizes. | and adult sizes. | |
| Patient | Adults/Pediatrics | Adults/Pediatrics | SE |
| Populations | | | |
| Tube | One or two tube | One or two tube | SE |
| Configuration | | | |
| Principles of | Bladder is wrapped around the | Bladder is wrapped around the | SE |
| Operation | patient's limb and secured by hook | patient's limb and secured by | |
| | and loop closure Air hose is | hook and loop closure Air hose is | |
| | connected to the noninvasive | connected to the noninvasive | |
| | blood pres sure measurement | blood pressure measurement | |
| | systems | systems | |
| Size | Conform to AHA bladder sizes | Conform to AHAbladder sizes | Analysis 1 |
| | Recommendations | Recommendations | |
| | Neonatal 1 (3-6 cm) | Neonatal 1 (3-6 cm) | |
| | Neonatal 2 (4-8cm) | Neonatal 2 (4-8 cm) | |
| | Neonatal 3 (6-11 cm) | Neonatal 3 (6-11 cm) | |
| | Neonatal 4 (7-13 cm) | Neonatal 4 (7-13 cm) | |
Table 5A - General Comparison(DISPOSABLE BLOOD PRESSURE CUFFS)
{6}------------------------------------------------
Ningbo Albert Novosino Co.,Ltd.
No.1 Xinheng 3 Road Cicheng Town,Jiangbei District,315036 Ningbo, Zhejiang, P.R. China K173044 Page 4 of 6
| | Neonatal 5 (8-15cm) | Neonatal 5 (8-15 cm) | |
|--------------------|---------------------------------------------------------|---------------------------------------------------------|----|
| | Infant (8-14 cm) | Infant (9-14.8 cm) | |
| | Child (13-20 cm) | Child (13.8-21.5 cm) | |
| | Small Adult (18-26 cm) | Small Adult (20.5-28.5 cm) | |
| | Adult (26-35 cm) | Adult (27.5-36.5 cm) | |
| | Adult Long (29-38 cm) | Adult Long (27.5-36.5 cm) | |
| | Large Adult (32-42 cm) | Large Adult(35.5-46 cm) | |
| | Large Adult Long (35-44 cm) | Large Adult (35.5 – 46 cm) | |
| | Thigh (42 – 50 cm) | Thigh (45 – 56.5 cm) | |
| Single Use | Y | Y | SE |
| Sterile | No | No | SE |
| Pressure<br>Range | 0-300 mmHg | 0-300 mmHg | SE |
| Max.Pressure | 400mmHg | 400mmHg | SE |
| Tubing<br>Size(mm) | OD = 8.0; ID=4.0; L=200; | OD = 8.0; ID=4.0; L=200; | SE |
| Max. Leakage | < 4mm Hg/ min. | < 4mm Hg/ min. | SE |
| Material | non-woven fabrics & PVC (Cuff)<br>and PVC<br>(Air Hose) | non-woven fabrics & PVC (Cuff)<br>and PVC<br>(Air Hose) | SE |
Table 5B - General Comparison(REUSABLE BLOOD PRESSURE CUFFS)
| ITEM | Proposed Device | Predicate Device | Remark |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Trade name | REUSABLE BLOOD<br>PRESSURE CUFF | REUSABLE BLOOD<br>PRESSURE CUFF | |
| 510(K)<br>Submitter | Ningbo Albert Novosino Co.,Ltd. | Shenzhen Vistar Medical<br>Supplies Co., Ltd. | |
| 510(K)<br>Number | K173044 | K152468 | |
| Classification<br>Regulation | 21CRF 870.1120 | 21CRF 870.1120 | SE |
| Classification<br>and Code | Class II,<br>DXQ | Class II,<br>DXQ | SE |
| Common<br>name | Non-invasive Blood Pressure<br>Cuff | Non-invasive Blood Pressure<br>Cuff | SE |
| Intended Use | The reusable blood pressure cuff is<br>an accessory used in conjunction<br>with noninvasive blood pressure<br>measurement<br>systems. The cuff is non-sterile<br>and may be reused. It is available | The reusable blood pressure cuff<br>is an accessory used in<br>conjunction with noninvasive<br>blood pressure measurement<br>systems. The cuff is non-sterile<br>and may be reused. It is available | SE |
{7}------------------------------------------------
Ningbo Albert Novosino Co.,Ltd.
No.1 Xinheng 3 Road Cicheng Town,Jiangbei District,315036 Ningbo, Zhejiang, P.R. China K173044 Page 5 of 6
| | in infant, pediatric and adult sizes. | in infant, pediatric and adult sizes. | |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Patient<br>Populations | Adults/Pediatrics | Adults/Pediatrics | SE |
| Tube<br>Configuration | One or two tube | One or two tube | SE |
| Principles of<br>Operation | Bladder is wrapped around the<br>patient's limb and secured by hook<br>and loop closure Air hose is<br>connected to the noninvasive<br>blood pressure measurement<br>systems | Bladder is wrapped around the<br>patient's limb and secured by<br>hook and loop closure Air hose is<br>connected to the noninvasive<br>blood pressure measurement<br>systems | SE |
| Size | Conform to AHA bladder sizes<br>Recommendations<br>Infant (8-14 cm)<br>Child (13-20 cm)<br>Small Adult (18-26 cm)<br>Adult (26-35 cm)<br>Adult Long (29-38 cm)<br>Large Adult (32-42 cm)<br>Large Adult Long (35-44 cm)<br>Thigh (42 - 50 cm) | Conform to AHA bladder sizes<br>Recommendations<br>Infant (8-13 cm)<br>Child (12-19 cm)<br>Small Adult (17-25 cm)<br>Adult (23-33 cm)<br>Adult Long (23-33 cm)<br>Large Adult(31-40 cm)<br>Large Adult Long (31 - 40 cm)<br>Thigh (38 - 50 cm) | Analysis 1 |
| Single Use | Y | Y | SE |
| Sterile | No | No | SE |
| Pressure<br>Range | 0-300 mmHg | 0-300 mmHg | SE |
| Max.Pressure | 400mmHg | 400mmHg | SE |
| Tubing<br>Size(mm) | OD = 8.0; ID=4.0; L=200; | OD = 8.0; ID=4.0; L=200; | SE |
| Max. Leakage | < 4mm Hg/ min. | < 4mm Hg/ min. | SE |
| Material | 210D Nylon TPU (Cuff) and PVC<br>(Air Hose) | 210D Nylon TPU (Cuff) and PVC<br>(Air Hose) | SE |
Analysis 1:
The proposed device has different cuff sizes compared with the predicate device for different patient arm circumference. The proposed devices are compliant with the ISO 81060-1 for the performance and Biocompatibility test and ISO 10993-1 for safety. We consider that these differences do not affect the substantial equivalence, safety or effectiveness when compared with the predicate device.
| | | | Table 5C - Biocompatibility Comparison |
|--|--|--|-----------------------------------------|
|--|--|--|-----------------------------------------|
| ITEM | Proposed Device | Predicate Device | Remark |
|--------------|-------------------------------------------|----------------------------|--------|
| Cytotoxicity | Under the conditions<br>of the study, not | Comply with ISO<br>10993-5 | SE |
{8}------------------------------------------------
Ningbo Albert Novosino Co.,Ltd.
No.1 Xinheng 3 Road Cicheng Town.Jiangbei District.315036 Ningbo, Zhejiang, P.R. China K173044 Page 6 of 6
| | cyteotoxicity effect | | |
|---------------|----------------------------------------------------|--------------------------|----|
| Irritation | Under the conditions of the study, not an irritant | Comply with ISO 10993-10 | SE |
| Sensitization | Under conditions of the study, not a sensitizer. | | SE |
## Conclusion:
The proposed devices share the same indications for use, device operation, overall technical and functional capabilities, meet the same standards and requirements and therefore, are substantially equivalent to the predicate device(s).
## 9. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
· AAMI / ANSI / ISO 10993-5:2009/(R) 2014, Biological Evaluation Of Medical Devices --Part 5: Tests For In Vitro Cytotoxicity.
· ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
· ISO 81060-1:2007 Non-Invasive Sphygmomanometers - Part 1: Requirements And Test Methods For Non-Automated Measurement Type
## 10. Clinical Test Conclusion
No clinical study is included in this submission.
## 11. Statement of Substantial Equivalence
It has been shown in this 510(k) submission that the difference between the proposed devices and the predicate devices do not raise any questions regarding safety and effectiveness. Performance testing and compliance with voluntary standards, demonstrate that the proposed are substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use. Therefore the proposed devices are determined to be substantially equivalent to the referenced predicate Devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.