The Visual-ICE MRI Cryoablation System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, proctology, and urology. This system is designed to destroy tissue (including prostate and kidney tissue, liver metastases, tumors, and skin lesions) by the application of extremely cold temperatures. The Visual-ICE MRI Cryoablation System has the following specific indications: Urology Ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia (BPH) . Oncology Ablation of cancerous or malignant tissue and benign tumors, and palliative intervention Dermatology Ablation or freezing of skin cancers and other cutaneous disorders: Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratosis, cavernous hemangiomas, peri-anal condylomata, and palliation of tumors of the skin Gynecology Ablation of malignant neoplasia or benign dysplasia of the female genitalia . General surgery Palliation of tumors of the rectum, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenomas ENT Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth Thoracic surgery (with the exception of cardiac tissue) • Proctology Ablation of benign or malignant growths of the anus or rectum
Device Story
Mobile cryoablation system for minimally invasive tissue destruction under MRI guidance. System comprises console (control room), mobile connection panel (magnet room), and junction boxes. Uses gas-based cryoablation needles to freeze tissue. Console features touch screen interface for procedure control and monitoring. Fiberoptic lines connect console and mobile connection panel to minimize MRI interference. Physician operates system to ablate prostate, kidney, liver, or skin lesions. Output is controlled freezing/thawing cycles; enables precise tumor destruction. Benefits include targeted tissue ablation with visual guidance.
Clinical Evidence
Bench testing included system, electrical, mechanical, packaging, and labeling verification. Software testing covered individual units and full package functionality. Validation included design, usability, and animal testing. Results confirmed system meets specifications and performs equivalently to predicates.
Technological Characteristics
Mobile console with touch screen interface; non-ferromagnetic mobile connection panel; fiberoptic signal lines. Gas-based cryoablation principle. Compatible with IceRod/IceSeed MRI needles. Compliant with IEC 60601-1-2 (4th ed) for EMC/EMI. Computer-controlled.
Indications for Use
Indicated for patients requiring cryosurgical tissue destruction in general surgery, dermatology, neurology, thoracic surgery (excluding cardiac), ENT, gynecology, proctology, and urology. Specific applications include prostate cancer, BPH, malignant/benign tumors, skin lesions, and various gynecological/proctological conditions.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's seal, which includes the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern. Inside the circle is a stylized image of three human profiles facing right, with a triple-line design above them, resembling a stylized human form.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 21, 2017
Galil Medical Incorporated Ms. Lynne Davies Senior Manager, Regulatory Affairs 4364 Round Lake Road Arden Hills, Minnesota 55112
Re: K173037
Trade/Device Name: Visual-ICE MRI Cryoablation System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH Dated: September 27, 2017 Received: September 28, 2017
Dear Ms. Davies:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
{3}------------------------------------------------
# 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Galil Medical Inc.<br>4364 Round Lake Road<br>Arden Hills, MN 55112 |
|-----------------------------|---------------------------------------------------------------------------------------|
| Company Contact Person: | Ms. Lynne A Davies, BS<br>Sr. Manager, Regulatory Affairs<br>Galil Medical Inc. |
| Phone: | 651-287-5098 |
| Fax: | 651-846-4516 |
| Email: | lynne.davies@btgplc.com |
| Alternate Contact Person: | Rebeka Stoltman, BS, MBA<br>Director, Regulatory Affairs<br>Galil Medical Inc. |
| Phone: | 651-287-5020 |
| Fax: | 877-510-7757 |
| Email: | Rebeka.stoltman@btgplc.com |
| Device Name: | Visual-ICE™ MRI Cryoablation System |
| Device Classification Name: | Cryosurgical unit and accessories |
| Regulation Number: | 21 CFR 878.4350 |
| Product Code: | GEH |
| Predicate Devices 510(k): | Visual-ICE Cryoablation System (K113860)<br>SeedNet MRI Cryoablation System (K021261) |
| Date of Preparation: | September 27, 2017 |
### Device Description:
The Visual-ICE™ MRI Cryoablation System is a mobile system intended for cryoablative tissue destruction using a minimally invasive procedure. The Visual-ICE MRI Cryoablation System has been designed to permit the delivery of cryoablation treatments under visual guidance in an MR environment. The Visual-ICE MRI System is comprised of a Console, Mobile Connection Panel (MCP) and a Junction Box Assembly. The system is computercontrolled with a touch screen user interface that allows the user to control and monitor the procedure. The console is located in the control room, while the Mobile Connection Panel is located in the magnet room. Junction boxes mounted in the two rooms, and connection of the gas and signal lines between the junction boxes occurs via a penetration panel. The Mobile Connection Panel is constructed of non-ferromagnetic materials to allow for placement within the magnet room close to the scanner. The Visual-ICE MRI Cryoablation System was designed for compatibility with Galil Medical's previously cleared IceRod MRI and IceSeed MRI cryoablation needles (K051052).
The table below provides a summary comparison of the submitted device compared to the predicate devices.
{4}------------------------------------------------
# K173037
| Description of Submitted<br>Device:<br>Visual-ICE MRI Cryoablation<br>System | Comments related to Predicates:<br>SeedNet MRI and Visual-ICE Cryoablation Systems<br>(K021261,K113860) |
|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design and Construction | |
| Console, MCP and Junction<br>Boxes | Similar to predicate; Visual-ICE MRI is comprised of<br>updated components to address more stringent ESD<br>requirements in the 4th edition of Medical Electrical<br>Equipment EMC standard IEC 60601-1-2 |
| Needle ports | Same as predicate |
| Channel Lock | Same as predicate |
| User interface | Same as predicate |
| Performance and Function | |
| Principles of Operation:<br>Freezing / Thawing | Same as predicate |
| Needle Compatibility | Same as predicate |
| Signals between Console and<br>MCP | Similar to predicate; Fiberoptic lines were selected for<br>Visual-ICE MRI to minimize interference as compared<br>to the electrical conductor lines for the legacy SeedNet<br>MRI System |
In summary, the submitted Visual-ICE MRI Cryoablation System has the same technology and principle of operation as the predicate devices.
#### Indications for Use / Intended Use:
The Visual-ICE MRI Cryoablation System is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, proctology, and urology. This system is designed to destroy tissue (including prostate and kidney tissue, liver metastases, tumors, and skin lesions) by the application of extremely cold temperatures.
The Visual-ICE MRI Cryoablation System has the following specific indications:
- Urology Ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia (BPH)
- . Oncology Ablation of cancerous or malignant tissue and benign tumors, and palliative intervention
- Dermatology Ablation or freezing of skin cancers and other cutaneous disorders: Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratosis, cavernous hemangiomas, peri-anal condylomata, and palliation of tumors of the skin
- Gynecology Ablation of malignant neoplasia or benign dysplasia of the female genitalia
- . General surgery Palliation of tumors of the rectum, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenomas
- ENT Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth
- Thoracic surgery (with the exception of cardiac tissue) •
- Proctology Ablation of benign or malignant growths of the anus or rectum
{5}------------------------------------------------
# K173037
The intended use and indications for use are the predicates with the exception that warts, hemorrhoids and ablation of cardiac tissue have been excluded for international indication alignment and marketing reasons. Galil Medical contends that there are no differences to the intended therapeutic, diagnostic, prosthetic, or surgical use of the device, and that the device is safe and effective when used as labelled.
## Summary of Performance Data and Substantial Equivalence:
A full battery of verification and validation testing was conducted on the Visual-ICE MRI Cyroablation System to ensure that the design, functionality, and performance met all the specified requirements and that the features of the system satisfy its intended use. Testing was conducted according to protocols based on international standards and in-house requirements. Verification testing included system testing, electrical testing, mechanical testing, packaging and labeling testing, software testing. System testing assessed whether the functional requirements of the system as a whole were satisfied by the design. Mechanical testing evaluated the mechanical robustness of the system and sub-systems. functional testing of the gas system, and tests of mechanical safety requirements. Electrical testing assessed the functional aspects of the circuit assemblies, tests of electromagnetic compatibility and immunity (EMC/EMI), and tests of electrical safety requirements. Software testing exercised individual units of software as well as tests of the functionality of the entire software package. Validation testing included design and usability testing and animal testing. Test results demonstrated that the Visual-ICE MRI Cryoablation System meets defined specifications, is substantially equivalent to the predicate devices, and does not raise any new issues of safety and effectiveness for its intended use.
# Conclusion:
The information and data provided in this Traditional 510(k) Premarket Notification establish that the Visual-ICE MRI Cryoablation System is substantially equivalent to the legally marketed predicate devices with regard to performance, safety, and effectiveness for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.